K Number
K181457

Validate with FDA (Live)

Device Name
EndoStream(TM)
Date Cleared
2019-07-11

(402 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EndoStream™ tubing is intended to provide irrigation via sterile water supply during gastrointestinal (GI) endoscopic procedures when used in conjunction with an irrigation pump.

Device Description

The EndoStream™ is available in two (2) configurations: PFE130 and PFE230. The EndoStream™ is used to convey sterile water from an external water source to a flexible endoscope to aid its function. This sterile water is used to clean the lens of the flexible endoscope during operation or to assist in improving the flexible endoscopes visual field. The device consists of PVC tubing and an endoscope connector with a medical grade plastic check valve. It functions by using a single lumen in conjunction with a peristaltic pump to provide water to the flexible endoscope. EndoStream™ connects a sterile water supply into the proximal end of an endoscope. Once connected, EndoStream™ allows sterile water to flow into the endoscope which carries it into the Gastrointestinal (GI) tract to flush out the GI tract during endoscopic procedures. It is a tubing with a cap which directly screws onto a sterile water bottle. The tubing extremity connects to the endoscope via the pump head on a peristaltic irrigation pump. A back flow valve is integrated in the extremity of each EndoStream™ tubing ensuring the unidirectional flow from the sterile water bottle to the endoscope, thus guaranteeing the prevention of risks of cross contamination.

AI/ML Overview

The EndoStream™ tubing is intended to provide irrigation via sterile water supply during gastrointestinal (GI) endoscopic procedures when used in conjunction with an irrigation pump. The performance data section indicates that the device underwent several tests to demonstrate functionality and performance integrity.

Here's the breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Sterilization Validation (ISO 11135)Not explicitly stated what was measured as performance, but the test was completed "per respective standards." The implication is that the sterilization process successfully rendered the device sterile and met the requirements of ISO 11135.
Shelf-Life / Aging Testing (ASTM 1980)Not explicitly stated what was measured as performance, but the test was completed "per respective standards." This included: - Dye Penetration - Seal Strength Testing - Visual Seal The implication is that the device maintained its integrity and functionality over its claimed shelf-life.
Biocompatibility (ISO 10993)Tested for: - Cytotoxicity - Intracutaneous Irritation Test - Sensitization The implication is that the materials used are biocompatible and met the requirements of ISO 10993.
Functionality & Performance IntegrityThe EndoStream™ product was also tested to demonstrate functionality and performance integrity. Specific quantitative metrics are not provided, but the conclusion states "Any differences do not raise any new issues of safety or effectiveness," suggesting successful performance.

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes used for the sterilization validation, shelf-life/aging, biocompatibility, or functionality tests. No information is provided regarding the country of origin of the data or whether the studies were retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device is a medical accessory (tubing for irrigation), and the described tests are primarily engineering and material science evaluations, not clinical performance studies requiring expert interpretation of results for ground truth establishment.

4. Adjudication method for the test set

Not applicable. The tests mentioned (sterilization, shelf-life, biocompatibility, functionality) are objective measurements conducted according to established standards. They do not involve subjective interpretation or a need for adjudication methods like 2+1 or 3+1.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The EndoStream™ is an irrigation tubing, not an AI-powered diagnostic or assistive device. Therefore, an MRMC study comparing human readers with or without AI assistance is irrelevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. The EndoStream™ is a physical medical device (tubing), not an algorithm or AI system.

7. The type of ground truth used

The "ground truth" for the tests performed would be defined by the specifications and acceptance criteria of the respective standards (ISO 11135, ASTM 1980, ISO 10993) and the device's own functional requirements. For example:

  • Sterilization: Ground truth is the absence of viable microorganisms after the sterilization process, confirmed by biological indicator tests or sterility assurance levels defined by ISO 11135.
  • Shelf-Life/Aging: Ground truth is the device maintaining its physical and functional properties (e.g., no dye penetration, specified seal strength, no visual degradation) over the tested period, according to ASTM 1980.
  • Biocompatibility: Ground truth is the absence of cytotoxic effects, irritation, or sensitization as per the criteria of ISO 10993.
  • Functionality: Ground truth would be the device's ability to consistently and reliably convey sterile water from the source to the endoscope without leaks, blockages, or failure of the check valve, as designed.

8. The sample size for the training set

Not applicable. This document describes performance testing for a physical medical accessory, not the development or training of an AI algorithm.

9. How the ground truth for the training set was established

Not applicable. No training set was used as this is not an AI/algorithm-based device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 11, 2019

Partnership Medical Limited Silbiano Gonzales Consultant QRC Consulting, LLC 10422 Huebner Road, Apt# 508 San Antonio, TX 78240

Re: K181457

Trade/Device Name: EndoStream™ Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCX Dated: June 3, 2019 Received: June 11, 2019

Dear Silbiano Gonzales:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181457

Device Name EndoStream™

Indications for Use (Describe)

The EndoStream™ tubing is intended to provide irrigation via sterile water supply during gastrointestinal (GI) endoscopic procedures when used in conjunction with an irrigation pump.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K181457

510(k) Summary – Page 1 of 6

l. SUBMITTER

Applicant Name:Address:Partnership Medical LimitedHartley House, Galveston GroveFenton, Stoke on Trent, ST4 3PE
Contact Person:Rob Hartley, Director
US Contact / Correspondent:QRC Consulting, LLC.Silbiano Gonzales, Consultant
Address:5623 Hamilton Wolfe Rd.,Apt. #921 San Antonio, TX78240
Phone:Email:(210) 287-9558silbianog@sbcglobal.net

II. DEVICE

Trade Name: Catalog Number:

EndoStream™

CATALOGUE (PART) NUMBERS
CataloguenumberDescriptionCompatibility
PFE130EndoStream™ Irrigation TubingOlympus®, Pentax®,EndoStratus™, Scope Assist®
PFE230EndoStream™ Irrigation TubingEGP-100E®, Olympus®,ScopeAssist® and ERBE™
Common or Usual Name:
Regulation Number:
Regulatory Name:
Regulatory Class:
Product Code:

Endoscopic Irrigation/Suction System 21 CFR 876.1500 Endoscope and accessories Class II ОСХ

The Gastroenterology/Urology devices panel has classified Endoscopic Irrigation/Suction System as Class II under 21 CFR §870.1500. OCX is the product code that has been assigned for these types of devices.

PART 876 -- GASTROENTEROLOGY-UROLOGY-UROLOGY DEVICES

Subpart B--Diagnostic Devices

{4}------------------------------------------------

K181457 510(k) Summary – Page 2 of 6

Sec. 876.1500 Endoscope and accessories

(a) Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

(b) Classification. (1) Class II (performance standards).

(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube. rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b) (2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in 876.9.

Product Code: OCX

Device: Endoscopic irrigation/suction system

Definition: To supply sterile water, other solutions and/or suction to endoscopes during endoscopic procedures.

Premarket Review: Office of Product Evaluation and Quality

OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices

DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices

THT3A2: Gastroenterology and Endoscopy Devices Team

{5}------------------------------------------------

K181457 510(k) Summary – Page 3 of 6

III. DEVICE DESCRIPTION

The EndoStream™ is available in two (2) configurations:

CATALOGUE (PART) NUMBERS
Catalogue numberDescriptionCompatibility
PFE130EndoStream™ Irrigation TubingOlympus®, Pentax®,EndoStratus™, Scope Assist®
PFE230EndoStream™ Irrigation TubingEGP-100E®, Olympus®,ScopeAssist® and ERBE™

Partnership Medical Ltd introduced the EndoStream™ (product code PFE230) in 2014 which fits the EGP-100, Olympus and ERBE pumps and this product has been supplied to over 50 Hospitals in the UK since that time. The Scope Assist tubing version (product code PFE130) was introduced in 2016 to fit our own Scope Assist pump, but also fits the EndoStratus and Olympus OPF-2 pumps.

The EndoStream™ is used to convey sterile water from an external water source to a flexible endoscope to aid its function.

The EndoStream™ consists of the following:

  1. PVC tubing is used to provide the method of transferring the sterile water from the external source to the flexible endoscope. This sterile water is used to clean the lens of the flexible endoscope during operation or to assist in improving the flexible endoscopes visual field.

  2. An endoscope connector is located at the distal end of the device. The EndoStream™ uses a medical grade plastic check valve, which connects to the flexible endoscope via an adaptor.

The EndoStream™ functions by using a single lumen in conjunction with a peristaltic pump to provide water to the flexible endoscope.

EndoStream™ connects a sterile water supply into the proximal end of an endoscope. Once connected, EndoStream™ allows sterile water to flow into the endoscope which carries it into the Gastrointestinal (GI) tract to flush out the GI tract during endoscopic procedures. There are 2 configurations for use with different pumps.

CATALOGUE (PART) NUMBERS
CataloguenumberDescriptionCompatibility
PFE130EndoStream™ Irrigation TubingOlympus®, Pentax®,EndoStratus™, Scope Assist®
PFE230EndoStream™ Irrigation TubingEGP-100E®, Olympus®,ScopeAssist® and ERBE™

{6}------------------------------------------------

K181457 510(k) Summary – Page 4 of 6

The EndoStream™ tubing is designed to be used in conjunction with a peristaltic pump which attaches to a flexible endoscope to supply sterile water for irrigation during endoscopic procedures. The EndoStream™ tubing is compatible with all commercial sterile water bottles.

EndoStream™ is a tubing with a cap which directly screws onto a sterile water bottle. The tubing extremity connects to the endoscope via the pump head on a peristaltic irrigation pump.

A back flow valve is integrated in the extremity of each EndoStream™ tubing ensuring the unidirectional flow from the sterile water bottle to the endoscope, thus quaranteeing the prevention of risks of cross contamination.

IV. INDICATIONS FOR USE

The EndoStream™ tubing is intended to provide irrigation via sterile water supply during gastrointestinal (GI) endoscopic procedures when used in conjunction with an irrigation pump.

V. PREDICATE DEVICES

The EndoStream™ is designed as a disposable water irrigation tubing for irrigation pumps (direct connection between the endoscope and sterile water bottle via an irrigation pump.)

The EndoStream™ Iegally marketed predicates, to which Partnership Medical Limited is claiming equivalence are:

Predicate Device for the EndoStream™
CompanyPredicate Device NameFDA 510(k) Number
Byrne Medical, Inc.EndoGatorK092429
EndoChoice, Inc.EndoChoice Water Bottle CapIrrigation SystemK16482

{7}------------------------------------------------

K181457

510(k) Summary – Page 5 of 6

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICES

The EndoStream™ is substantially equivalent to the Byrne Medical, Inc. (EndoGator, K092429), and the EndoChoice, Inc. (EndoChoice Water Bottle Cap Irrigation System, K161482)

CharacteristicComparedOur DevicePredicate 1Predicate 2
ManufacturerPartnershipMedical, LtdByrne Medical,Inc.EndoChoice,Inc.
DeviceEndoStream™EndoGatorEndoChoiceWater BottleCapIrrigationSystem
510(k) Number-K092429K161482
IndicationTheEndoStream™ tubing isintended toprovideirrigation viasterile watersupply duringgastrointestinal (GI)endoscopicprocedureswhen used inconjunctionwith anirrigationpump.SameSame
ProductCode/ClassOCXOCXOCX
CharacteristicComparedOur DevicePredicate 1Predicate 2
Principle ofOperationConnects to aperistalticpump to eithertheendoscopesauxiliarychannel toallow flushingof waterIdenticalIdentical
DifferenceNoneNoneNone
Materials ofConstructionPer section 16IdenticalIdentical
SterilizationEthyleneOxideEthylene OxideEthyleneOxide

{8}------------------------------------------------

K181457 510(k) Summary - Page 6 of 6

VII. PERFORMANCE DATA

The following performance testing was completed (per respective standards).

  • Sterilization Validation (ISO 11135)
  • Shelf-Life / Aging Testing (ASTM 1980) .
    • Dye Penetration o
    • Seal Strength Testing o
    • Visual Seal O

Biocompatibility:

The materials used were tested for biocompatibility per ISO 10993. The tests performed were for cytotoxicity, Intracutaneous Irritation Test and Sensitization.

The EndoStream™ product was also test to demonstrate functionality and performance integrity.

VIII. CONCLUSIONS

Partnership Medical Limited considers the EndoStream™ to be substantially equivalent to the legally marketed predicate device listed above, and safe and effective for the intended use. The conclusion is based on similarities in indications for use, materials, performance testing, technological characteristics, principle of operation and design features. Any differences do not raise any new issues of safety or effectiveness.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.