K Number
K181434

Validate with FDA (Live)

Date Cleared
2018-11-14

(166 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 270°F (132°C) for 4 minutes exposure time.

Device Description

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 270°F (132°C) for 4 minutes exposure time.

The test pack consists of multiple layers of paper cards with a SporView Rapid Read Biological Indicator, chemical process indicators and load record card. The load record card is used to record the detailed information from the sterilization cycle.

The SporView Rapid Read Biological Indicator, cleared under 510(k) K172432, is composed of a polypropylene vial containing a spore carrier and media ampoule enclosed with a vented cap.

AI/ML Overview

The provided text describes the regulatory clearance for the "SporView Rapid Read Biological Indicator SteamPlus Test Pack." The document includes a summary of non-clinical performance data, which outlines the acceptance criteria and the results of the study conducted to demonstrate the device's performance.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Test Pack PerformanceAll test packs have complete kill in a 270°F (132°C) pre vacuum steam sterilization cycle for an exposure time of 4 minutes.
Resistance PerformanceThe test pack has resistance greater than or equal to the AAMI Towel Pack.
Resistance (vs. self-contained BI)The test pack has resistance greater than the self-contained biological indicator itself.
Chemical Indicator PerformanceTesting to confirm process indicators contained in the test pack perform as intended per the FDA Guidance for Industry and FDA Staff - Premarket Notification [510(k)] Submissions for Chemical Indicators.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample size used for the test set or the country of origin of the data. It also does not specify if the data was retrospective or prospective. The "Summary of Non-Clinical Performance Data" implies prospective testing as it describes performance testing conducted to demonstrate compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This information is not applicable to a biological indicator device. Biological indicators are evaluated based on their ability to show the presence or absence of viable microorganisms after a sterilization process, not on expert interpretation of images or other subjective assessments. The "ground truth" for a biological indicator is the objective determination of microbial growth or no growth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation by multiple human readers (e.g., in medical image analysis). For a biological indicator, the outcome (growth or no growth) is an objective biological response.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This device is a biological indicator for sterilization monitoring, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study or assessment of AI assistance is irrelevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical biological indicator, not a software algorithm. Its performance is inherent in its design and biological response, not in an algorithm's output.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for a biological indicator is the presence or absence of viable Geobacillus stearothermophilus spores after exposure to sterilization conditions. This is determined by observing microbial growth (or lack thereof) in a controlled incubation process. For the "complete kill" criteria, the ground truth is the absence of microbial growth, indicating a successful sterilization cycle. For resistance testing, the ground truth relates to the time/temperature exposure required to inactivate the spores.

8. The sample size for the training set

This information is not applicable. The device is a physical biological indicator, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable. As stated above, the device does not involve a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

November 14, 2018

Crosstex/SPSmedical, A Division of Cantel Medical Megan Skaar Regulatory Affairs Specialist 6789 West Henrietta Road Rush, New York 14543

Re: K181434

Trade/Device Name: SporView Rapid Read Biological Indicator SteamPlus Test Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: October 11, 2018 Received: October 15, 2018

Dear Megan Skaar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K181434

Device Name

SporView Rapid Read Biological Indicator SteamPlus Test Pack

Indications for Use (Describe)

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 270°F (132°C) for 4 minutes exposure time.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "CANTEL" in blue font next to a blue circular logo. The logo is a circle with several horizontal lines running through it. The word "CANTEL" is in all capital letters and appears to be a sans-serif font. The image is likely a company logo.

510(k) Summary - K181434

Manufacturer:Crosstex/SPSmedical, a Cantel Medical Company
Address:6789 W. Henrietta RoadRush, NY 14543(800) 722-1529
Official Contact:Megan SkaarRegulatory Affairs Specialist, Cantel Medical
Date:9 November 2018
Trade Name:SporView Rapid Read Biological Indicator SteamPlus Test Pack
Common Name:Sterilization Biological Test Pack
Classification Name:Sterilization Process Indicator
Product Code:FRC
Device Class:II
Regulation No:Subject Device – SporView Rapid Read Biological Indicator Test Pack,880.2800Predicate Device – 3M Attest 3M Attest 41382 Rapid Readout Steam-PlusProcess Challenge Device, 880.2800

1. Device Description

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 270°F (132°C) for 4 minutes exposure time.

The test pack consists of multiple layers of paper cards with a SporView Rapid Read Biological Indicator, chemical process indicators and load record card. The load record card is used to record the detailed information from the sterilization cycle.

The SporView Rapid Read Biological Indicator, cleared under 510(k) K172432, is composed of a polypropylene vial containing a spore carrier and media ampoule enclosed with a vented cap.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Cantel Medical. The logo consists of a blue circle with several horizontal lines running through it on the left, and the word "CANTEL" in blue on the right. The font is sans-serif and the overall design is simple and modern.

2. Indications for Use

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 132°C for 4 minutes exposure time.

3. Technological Characteristics Comparison Table

Table 05.1 below provides a detailed comparison between the subject, predicate and reference devices –

TechnologicalCharacteristicsSubject Device -SporView RapidRead BiologicalIndicator TestPackPredicate Device- 3M Attest 41382Rapid ReadoutSteam-PlusProcessChallenge Device(K101910)Reference Device- SporView PlusBI Test PackProcessChallenge Device(K140620)ComparisonSummary
Intended UseMonitor 270°F(132°C)prevacuum steamsterilization cyclesfor 4 minutesexposure time.Monitor 270°F(132°C)prevacuum steamsterilization cycleswith exposure timeof 4 minutes.Monitor 250°F(121°C) gravitydisplacementcycles and 270°F(132°C)prevacuum steamsterilization cyclesfor 4 minutesexposure time.Same
Product CodeFRCFRCFRCSame

Table 05.1 - Predicate Device Comparison Table

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for CANTEL. The logo consists of a blue circular graphic on the left and the word "CANTEL" in blue on the right. The circular graphic has several horizontal lines running through it, giving it a stylized appearance.

FDARegulation21 CFR 880.280021 CFR 880.280021 CFR 880.2800Same
Device ClassIIIIIISame
Organism> 90% geneticsimilarity to G.stearothermophilusATCC™ strain7953> 90% geneticsimilarity to G.stearothermophilusATCC™ strain7953> 90% geneticsimilarity to G.stearothermophilusATCC™ strain7953Same
Viable SporePopulation$1.0 x 10^5$$1.0 x 10^5$$1.0 x 10^5$Same
D-Value ofBiologicalIndicatorSteam 132°C:10.0 SecondsSteam 132°C:10.0 SecondsSteam 132°C:10.0 SecondsSame
Includes 510(k)Cleared ChemicalIndicators?YesYesYesSame
Paper of TestPackPaper Index CardsPaper Index CardsPaper Index CardsSame
Shelf life12 Months24 Months18 MonthsSimilar
StorageControlled roomtemperatureControlled roomtemperatureControlled roomtemperatureSame
IncubationTemperature60±2°C60±2°C55-60°CSame asPredicateDevice
BI IncubationDuration3 hours3 hours10 hoursSame asPredicateDevice
BiocompatibilityNo direct patientcontactNo direct patientcontactNo direct patientcontactSame

The subject device – the SporView Rapid Read Biological Indicator SteamPlus Test Pack and its predicate device - 3M Attest 41382 Rapid Readout Steam-Plus Process Challenge Device are similar in that they both have the same intended use, fundamental technology and general performance. Both the subject device and the 3M Attest 41382 Rapid Readout Steam-Plus Process Challenge Device have 510(k) cleared self-contained biological indicators with a spore carrier inoculated with the same organism, G. stearothermophilus. The subject and predicate test packs both provide a significant challenge to the steam sterilization process.

The subject device is very similar to its reference device - the SporView Plus Bl Test Pack cleared under K140620. The 510(k) cleared chemical indicators in the subject device, including the locations, of the subject test pack are identical to the reference device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "CANTEL" in blue font. To the left of the word is a blue circular logo with several horizontal lines running through it. The logo and the word "CANTEL" are the only elements in the image. The background is white.

4. Summary of Non-Clinical Performance Data

The following testing has been conducted in accordance with the FDA Guidance on Biological Indicators and the FDA Guidance for Industry and FDA Staff – Premarket Notification [510(k)] Submissions for Chemical Indicators and the FDA Guidance for Industry and FDA Staff -Premarket Notification [510(k)] Submissions for Chemical Indicators. Please refer to the table below for a brief description of the subject device testing

Results

Pass

Pass

Pass

Pass

Performance TestingDetails and Acceptance Criteria
Test Pack PerformanceAll test packs have complete kill in a 270°F (132°C)pre vacuum steam sterilization cycle for anexposure time of 4 minutes.
ResistancePerformanceThe test pack has resistance greater than or equal tothe AAMI Towel Pack.

contained biological indicator itself.

  • . Non-Clinical Performance Testing of the Subject:
    Performance testing demonstrates that the subject device has more resistance than the self-contained biological indicator itself and the subject device has resistance greater than or equal to the AAMI Towel Pack. There was no growth of the SCBI contained in the subject device after a full sterilization cycle.

The test pack has resistance greater than the self-

Testing to confirm process indicators contained in the

test pack perform as intended per the FDA Guidance

for Industry and FDA Staff - Premarket Notification [510(k)] Submissions for Chemical Indicators.

5. Conclusion

Chemical Indicator

Performance

Based on the intended uses, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as the legally marketed predicate cleared under 510(k) K101910.

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).