K Number
K181415

Validate with FDA (Live)

Date Cleared
2018-09-10

(103 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Angio Workstation (XIDF-AWS801) is used in combination with an interventional angiography system (Alphenix series systems, Infinix-i series systems and INFX series systems) to provide 2D and 3D imaging of selective catheter angiography procedures for the whole body (includes heart, chest, abdomen, brain and extremity).

When XIDF-AWS801 is combined with Dose Tracking System (DTS), DTS is used with selective catheter angiography procedures for the heart, chest abdomen, pelvis and brain.

Device Description

The XIDF-AWS801, Angio Workstation (Alphenix Workstation), V8.0 is used for images input from Diagnostic Imaging System and Workstation, image processing and display. The processed images can be outputted to Diagnostic Imaging System and Workstation.

AI/ML Overview

The provided text describes a 510(k) submission for a medical device called XIDF-AWS801, Angio Workstation (Alphenix Workstation), V8.0. However, it does not contain a specific study that directly proves the device meets acceptance criteria in the way a clinical trial or performance study would typically be presented.

Instead, the submission outlines changes made to an existing cleared device (XIDF-AWS801, Angio Workstation, V7.0 - K172646) and asserts substantial equivalence to that predicate device. The "acceptance criteria" are implicitly tied to the safety and performance standards of the predicate device and the new modifications meeting those same standards, or better.

Here's an analysis based on the information provided:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly define "acceptance criteria" in a quantitative table with corresponding performance metrics as would be typical for a new device's efficacy study. Instead, the "acceptance" is based on demonstrating that the modified device's performance is equal to or better than the predicate system and that all requirements for the modifications have been met.

Acceptance Criteria (Implied)Reported Device Performance
Conformance to IEC60601-1-2:2014 & IEC62304:2006 + A1:2015Testing of modified system conducted in accordance with applicable standards.
Compliance with IEC62366:2007 + A1:2014 (Usability)Not explicitly detailed for modifications, but assumed as part of overall system design and safety.
Compliance with ISO 14971:2007 (Risk Management)Risk analysis and verification/validation testing conducted; all known risks mitigated to an acceptable level.
Software Documentation (Moderate Level of Concern)Included in submission, software modules subject to verification/validation.
Image Quality of AlphaCT (3D acquisition) Artifact ReductionPhantom image evaluations employed in bench testing, demonstrating requirements met. Modifications implemented to reduce artifacts.
Stent visualization in AlphaCT Device acquisitionApplication of a spatial filter (11x11 pixels) to reconstructed slice images in the axial direction to improve stent visualization.
Functionality of new features (BP Auto Angle Support, Needle Guidance Function Oblique View, CAA Vessel Track)Verification/validation testing conducted, demonstrates requirements met.
Overall safety and effectiveness compared to predicatePerforms in a manner similar to and intended for the same use as the predicate device. Substantially equivalent in safety and effectiveness.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not explicitly stated. The testing appears to involve bench testing using phantom images and software verification/validation. This suggests the "test set" would primarily comprise simulated scenarios, phantom data, and potentially internal system testing data rather than a large clinical dataset.
  • Data Provenance: The testing describes "bench testing" and "phantom image evaluations." This implies prospective data generated specifically for the verification of the modifications. There is no mention of country of origin for this specific test data, but the manufacturer is Canon Medical Systems Corporation in Japan, implying testing would likely occur there or at associated facilities.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified. Given the nature of the device (angiography workstation software), "experts" would likely be clinical users (e.g., interventional radiologists) or biomedical engineers with expertise in image processing and medical imaging, but this is not mentioned.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not specified. Since the testing primarily involved bench testing and phantom evaluations, formal adjudication by a panel of experts for clinical endpoints is unlikely to have occurred in the context described. Adjudication might have been integrated into the internal verification and validation processes by design engineers and subject matter experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Was an MRMC study done? No. The document does not mention an MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The device is an angiography workstation with advanced image processing and viewing features, not explicitly an "AI" device as typically evaluated in MRMC studies for diagnostic accuracy improvement.
  • Effect Size: Not applicable, as no such study was reported.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Was a standalone study done? Yes, implicitly. The "bench testing" and "phantom image evaluations" are forms of standalone performance assessment for the modifications (e.g., artifact reduction, stent visualization improvement, functionality of new features like angle support). The software modules were also subjected to verification and validation testing to ensure proper integration and performance.

7. Type of Ground Truth Used:

  • Ground Truth Type: For the image quality evaluations (e.g., artifact reduction, stent visualization), the ground truth would likely be based on physical phantom measurements, known image characteristics, and engineering specifications. For the functional aspects of the software modifications, the ground truth would be adherence to the design requirements and expected output. There is no mention of pathology, expert consensus on clinical cases, or outcomes data being used as ground truth for this submission's testing.

8. Sample Size for the Training Set:

  • Training Set Sample Size: Not specified. The document does not explicitly state that the device uses machine learning or AI that requires a "training set" in the conventional sense. The "AlphaCT (3D acquisition) Artifact Reduction" and "application of the spatial filter" are described as modifications to existing algorithms, rather than newly trained AI models. If there were any learning components, the training data details are not provided.

9. How the Ground Truth for the Training Set Was Established:

  • Ground Truth for Training Set: Not applicable, as the document does not describe the device as having a "training set" for AI or machine learning models. The modifications appear to be software updates/enhancements based on algorithmic refinement and engineering design.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' in blue, and 'ADMINISTRATION' in a smaller font size below.

Canon Medical Systems Corporation % Janine Reyes Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive TUSTIN, CA 92780

Re: K181415

Trade/Device Name: XIDF - AWS801, Angio Workstation (Alphenix Workstation), V8.0 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA Dated: August 22, 2018 Received: August 23, 2018

September 10, 2018

Dear Janine Reyes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jeff Baffges

for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181415

Device Name

XIDF-AWS801, Angio Workstation (Alphenix Workstation), V8.0

Indications for Use (Describe)

The Angio Workstation (XIDF-AWS801) is used in combination with an interventional angiography system (Alphenix series systems, Infinix-i series systems and INFX series systems) to provide 2D and 3D imaging of selective catheter angiography procedures for the whole body (includes heart, chest, abdomen, brain and extremity).

When XIDF-AWS801 is combined with Dose Tracking System (DTS), DTS is used with selective catheter angiography procedures for the heart, chest abdomen, pelvis and brain.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92

1. CLASSIFICATION and DEVICE NAME

Classification NameSolid State X-ray System, Interventional
Regulation Number21 CFR 892.1650 (Class II)
Product CodeOWB, JAA
Trade Proprietary NameXIDF-AWS801, Angio Workstation (Alphenix Workstation)
Model NumberXIDF-AWS801, V8.0

2. SUBMITTER'S NAME

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550

3. OFFICIAL CORRESPONDENT

Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance

4. CONTACT PERSON, U.S. AGENT and ADDRESS

Contact Person

Janine F. Reyes Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7853 Fax: (714) 730-1310 jfreyes@us.medical.canon

5. MANUFACTURING SITE

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan

6. ESTABLISHMENT REGISTRATION 9614698

    1. DATE PREPARED August 20, 2018

8. TRADE NAME(S)

XIDF-AWS801, Angio Workstation (Alphenix Workstation), V8.0

Official Correspondent/U.S. Agent

Paul Biggins Sr. Director, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7808 Fax: (714) 730-1310 pbiggins@us.medical.canon

{4}------------------------------------------------

9. CLASSIFICATION PANEL

Radiology

10. DEVICE CLASSIFICATION

Class II (per 21 CFR 892.1650)

11. PRODUCT CODE / DESCRIPTION

Product Code: OWB, JAA

12. PERFORMANCE STANDARD

This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard].

13. PREDICATE DEVICE

XIDF-AWS801, Angio Workstation, V7.0 (K172646)

ProductXIDF-AWS801, Angio Workstation,V7.0XIDF-AWS801, Angio Workstation(Alphenix Workstation), V8.0
Marketed byCanon Medical Systems USACanon Medical Systems USA
Indications ForUseThe Angio Workstation (XIDF-AWS801)is used in combination with aninterventional angiography system(Infinix-i series systems and INFXseries systems) to provide 2D and 3Dimaging of selective catheterangiography procedures for the wholebody (includes heart, chest, abdomen,brain and extremity).When XIDF-AWS801 is combined withDose Tracking System (DTS), DTS isused with selective catheterangiography procedures for the heart,chest abdomen, pelvis and brain.The Angio Workstation (XIDF-AWS801)is used in combination with aninterventional angiography system(Alphenix series systems, Infinix-iseries systems and INFX series systems)to provide 2D and 3D imaging ofselective catheter angiographyprocedures for the whole body(includes heart, chest, abdomen, brainand extremity).When XIDF-AWS801 is combined withDose Tracking System (DTS), DTS isused with selective catheterangiography procedures for the heart,chest abdomen, pelvis and brain.
510(k) NumberK172646
Clearance DateOctober 30, 2017

14. REASON FOR SUBMISSION

Modification of a cleared device

15. SUBMISSION TYPE

Traditional 510(k)

16. DEVICE DESCRIPTION

The XIDF-AWS801, Angio Workstation (Alphenix Workstation), V8.0 is used for images input from Diagnostic Imaging System and Workstation, image processing and display. The processed images can be outputted to Diagnostic Imaging System and Workstation.

{5}------------------------------------------------

17. INDICATIONS FOR USE

The Angio Workstation (XIDF-AWS801) is used in combination with an interventional angiography system (Alphenix series systems, Infinix-i series systems and INFX series systems) to provide 2D and 3D imaging of selective catheter angiography procedures for the whole body (includes heart, chest, abdomen, brain and extremity).

When XIDF-AWS801 is combined with Dose Tracking System (DTS), DTS is used with selective catheter angiography procedures for the heart, chest abdomen, pelvis and brain.

18. SUMMARY OF CHANGE(S)

This submission is to report the following items have been changed:

  • . BP Auto Angle Support: supports the planning of bi-plane arm angle without any collision by using 3D volume on 3D viewer at AWS. Once each angle is found, INFX system performs automatic positioning by these angles.
  • . Needle Guidance Function Oblique View: addition of oblique cross-section view within the Needle Guide 3D viewer.
  • . CAA (Cerebral Aneurysm Analysis) Vessel Track: modification to existing CAA (Cerebral Aneurysm Analysis) tool provides the vessel extraction and the virtual stent display feature.
  • . 3D Viewer Kit: the existing 3D Viewer software function has been made optional by way of the 3D Viewer kit.
  • Indications For Use: Indications for use statement was modified to include the Alphenix series systems marketing name.
  • AlphaCT (3D acquisition) Artifact Reduction: the following modifications have been . implemented to reduce artifacts within AlphaCT (3D acquisition) reconstructed images: oNew RFC parameters
    • o New scattered radiation correction parameters for abdominal images
  • Application of the spatial filter to reconstructed images in AlphaCT Device acquisition: in order to improve stent visualization in AlphaCT Device acquisition, a spatial filter of 11 × 11 pixels is applied to reconstructed slice images in the axial direction.

19. SAFETY

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report.

LIST OF APPLICABLE STANDARDS

  • IEC60601-1-2:2014
  • . IEC62304:2006 + A1:2015
  • IEC62366:2007 + A1:2014
  • . IEC60950-1:2005 + A1:2009, A2:2013
  • ISO 14971:2007

{6}------------------------------------------------

20. TESTING

This submission contains test data that demonstrates that the system modifications result in performance that is equal to or better than the predicate system. Testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices and XR Systems.

Risk analysis and verification/validation testing, conducted through bench testing, are included in this submission which demonstrates that the requirements for the modifications made to the system have been met. Additionally, Image Quality metrics utilizing phantom image evaluations were employed in bench testing which demonstrates that the requirements for the modifications made to the system have been met.

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission. Software modules were subject to verification and/or validation testing to ensure that they were properly integrated into the existing software platform. Additionally, the design controls used for this device included risk management and all known risks were mitigated to an acceptable level.

21. SUBSTANTIAL EQUIVALENCE

The XIDF-AWS801, Angio Workstation (Alphenix Workstation), V8.0 is substantially equivalent to the XIDF-AWS801, Angio Workstation, V7.0, which received premarket clearance under K172646, marketed by Canon Medical Systems. XIDF-AWS801, Angio Workstation (Alphenix Workstation), V8.0, includes modifications to the cleared device consisting of software change from V7.0 to V8.0, BP Auto Angle Support, Needle Guide Oblique View, CAA (Cerebral Aneurysm Analysis) Vessel Track, 3D Viewer Kit, AlphaCT (3D acquisition) artifact reduction, application of the spatial filter to reconstructed images in AlphaCT Device acquisition, addition of the Alphenix marketing name to the Indications For Use statement.

The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device.

22. CONCLUSION

The XIDF-AWS801, Angio Workstation (Alphenix Workstation), V8.0 performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.