AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cepheid Xpert® Xpress Flu Assay, performed on the GeneXpert® Instrument Systems, is an automated, multiplex real-time, reverse transcriptase polymerase chain reaction (RT-PCR) assay intended for the in vitro qualitative detection and differentiation of influenza A and influenza B viral RNA. The Xpert Xpress Flu Assay uses nasopharyngeal (NP) swab and nasal swab (NS) specimens collected from patients with signs and symptoms of respiratory infection. The Xpert Xpress Flu Assay is intended as an aid in the diagnosis of influenza infections in conjunction with clinical and epidemiological risk factors.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.

Performance characteristics for influenza A were established during the 2015-2016 influenza season for NP swab specimens and the 2016-2017 influenza season for NS specimens. When other novel influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The Xpert Xpress Flu Assay is a rapid, automated in vitro diagnostic test for qualitative detection and differentiation of influenza A (Flu A) and influenza B (Flu B) viral RNA directly from nasopharyngeal (NP) swab and nasal swab (NS) specimens. The assay is performed on the Cepheid GeneXpert® Instrument Systems.

The Xpert Xpress Flu Assay includes reagents for the detection and differentiation of influenza A and influenza B viral RNA directly from NP swab and NS specimens from patients with signs and symptoms of respiratory tract infection. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are internal controls utilized by the GeneXpert Instrument System platform. The SPC is present to control for an adequate extraction and processing of the target sequences and to monitor for the presence of inhibitor in the PCR reaction. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

The specimens are collected in viral transport medium and transported to the GeneXpert area. The specimen is prepared according to package insert instructions and transferred to the sample chamber (large opening) of the Xpert Xpress Flu Assay cartridge. The GeneXpert cartridge is loaded onto the GeneXpert Xpress System platform, which performs hands-off automated sample processing and real-time RT-PCR for detection of Flu viral RNA. Summary and detailed test results are obtained in approximately 30 minutes or less. The results are automatically generated at the end of the process in a report that can be viewed and printed.

AI/ML Overview

Here's a detailed breakdown of the acceptance criteria and study findings for the Cepheid Xpert® Xpress Flu Assay:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally established by regulatory bodies through guidance documents and are demonstrated through the clinical performance relative to a comparator device (often an FDA-cleared molecular assay). The reported performance is presented as Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). While explicit "acceptance criteria" are not listed in a numerical form in this document, the clinical study results are implicitly compared against regulatory expectations for substantial equivalence.

MetricTarget (Implicit Acceptance Criteria - high agreement with comparator)Reported Device Performance (Xpert Xpress Flu Assay)
NP Swab Specimens (Clinical Comparison)
Influenza A PPA (Combined)High (e.g., >90%)98.1% (93.3 - 99.5% CI)
Influenza A NPA (Combined)High (e.g., >95%)98.8% (98.2 - 99.2% CI)
Influenza B PPA (Combined)High (e.g., >90%)100.0% (95.3 - 100.0% CI)
Influenza B NPA (Combined)High (e.g., >95%)99.1% (98.6 - 99.5% CI)
NS Specimens (Clinical Comparison)
Influenza A PPAHigh (e.g., >90%)98.9% (96.2 - 99.7% CI)
Influenza A NPAHigh (e.g., >95%)97.6% (96.6 - 98.3% CI)
Influenza B PPAHigh (e.g., >90%)98.4% (91.7 - 99.7% CI)
Influenza B NPAHigh (e.g., >95%)99.3% (98.7 - 99.6% CI)
Analytical Sensitivity (LoD)Reproducibly distinguishable from negative with 95% confidence (19/20 positive)Values range from 0.006 TCID50/mL to 0.750 TCID50/mL depending on strain and specimen type. All met LoD definition.
Analytical Specificity100% (No false positives with tested organisms)100%
Analytical ReactivityAll Flu strains tested positive in 3 replicates (or 2/3 at 0.1 TCID50/mL for one strain)Met criteria (one strain tested positive in 2/3 replicates at 0.1 TCID50/mL, others 3/3)
Interfering SubstancesNo interference observedNone of the tested substances caused interference.
Carry-Over ContaminationNo carry-over contamination with negative samples preceded by high positive samplesAll 20 positive samples and 21 negative samples correctly reported.
Fresh vs. Frozen Sample EquivalencyNo statistically significant effectNo statistically significant effect. All positive/negative replicates correctly identified.
Competitive InterferenceExpect no/minimal inhibitory effects (with caveats)Observed competitive inhibitory effects on targets when high concentrations of competing strains were present (addressed in limitations). Otherwise no effects.
Assay Success RateHigh (implicitly >95%)99.3%
Indeterminate RateLow (implicitly <5%)0.7%
Reproducibility (Agreement by Sample)High (e.g., >90% for low positive, 100% for moderate positive & negative)Range from 93.6% (Flu A Low Pos) to 100% (Negative, Flu A Mod Pos, Flu B Mod Pos). Flu B Low Pos: 95.1%.
Reproducibility (Ct value variability)Low coefficient of variation (CV) for SD componentsAll CVs for between-site, lot, day, operator, and within-assay were low (max 4.2%).

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Comparison Study (Test Set):
    • NP Swab Specimens: 2051 total specimens.
      • 1139 fresh, prospectively collected.
      • 912 consecutively collected, frozen specimens.
    • NS Specimens: 1598 total specimens.
    • Provenance: Collected at eleven institutions in the U.S. (2015-2016 influenza season for NP swabs) and fourteen institutions in the U.S. (2016-2017 influenza season for NS specimens). This indicates a multi-center, prospective (for fresh samples) and retrospective (for frozen samples) study within the U.S.
  • Analytical Sensitivity (LoD): Not specified for specific numbers of specimens, but "replicates of 20 per concentration of virus in each matrix."
  • Analytical Specificity: 44 cultures (16 viral, 26 bacterial, 2 yeast strains), each tested in triplicate.
  • Analytical Reactivity: 48 strains (35 influenza A, 13 Influenza B), each tested in triplicate.
  • Interfering Substances: 8 negative samples tested per substance, 8 positive samples tested per substance (with 6 influenza strains).
  • Carry-Over Contamination: 20 positive followed by 20 negative samples.
  • Fresh vs. Frozen Sample Equivalency Study: Not explicitly stated, but replicates of 20 were tested for each specimen type and concentration (low, moderate, high positive, and negative) across fresh, one freeze-thaw, and two freeze-thaw cycles for two influenza strains.
  • Competitive Interference Study: Replicates of 20 for each target strain and each competitive strain combination.
  • Reproducibility Study: 5-member specimen panel tested across 3 sites, 2 operators per site, 6 days, 2 testing days per lot, duplicate testing twice per day. This means each panel member was tested approximately 144 times (3 sites * 2 operators * ~24 tests/operator).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for the clinical comparison study was established primarily using an FDA-cleared molecular comparator assay. For discrepant results between the Xpert Xpress Flu Assay and the comparator assay, bi-directional sequencing was performed. The document does not specify the number of experts or their qualifications for interpreting the sequencing results or for judging the "clinical and epidemiological risk factors" mentioned in the intended use. However, the use of an FDA-cleared molecular assay and sequencing provides a strong, objective ground truth.

4. Adjudication Method for the Test Set

The primary comparison was against an FDA-cleared molecular comparator assay.
For samples where the Xpert Xpress Flu Assay and the comparator assay produced discrepant results, bi-directional sequencing was used as an adjudication method. The sequencing results were "provided for informational purposes only," suggesting that while they informed the understanding of discrepancies, the primary performance metrics (PPA/NPA) were against the comparator.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

This device, the Xpert Xpress Flu Assay, is an automated, multiplex real-time RT-PCR assay. It is a standalone diagnostic test for qualitative detection and differentiation of viral RNA. It does not involve human interpretation of images or other subjective data that would necessitate a multi-reader multi-case (MRMC) study or human assistance with AI. Therefore, this section is not applicable. The results are automatically generated as "POSITIVE" or "NEGATIVE" for Influenza A and Influenza B.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

Yes, a standalone performance evaluation was done. The entire clinical comparison study (for NP and NS specimens) and all non-clinical analytical studies (sensitivity, specificity, reactivity, interference, carry-over, fresh vs. frozen, competitive interference) as well as the reproducibility study evaluate the performance of the Xpert Xpress Flu Assay directly, without human interpretation of the assay's primary output (the detection of viral RNA). The GeneXpert Instrument Systems automatically perform sample processing and real-time RT-PCR, generating quantitative results (Ct values) which are then interpreted by the system's definition file to provide a qualitative POSITIVE/NEGATIVE result.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

  • Clinical Comparison Study: The primary ground truth was an FDA-cleared molecular comparator assay. For discrepant results, bi-directional sequencing was used.
  • Analytical Studies (LoD, specificity, reactivity, interference, carry-over, fresh vs. frozen, competitive interference): Ground truth was established by known concentrations of purified or cultured viruses/organisms and/or negative controls.

8. The Sample Size for the Training Set

This document does not explicitly mention a separate "training set" for the Xpert Xpress Flu Assay itself in the context of machine learning. As a molecular diagnostic assay, its "training" or development would typically involve optimization of primers, probes, and reaction conditions based on known viral sequences and biological principles, rather than a data-driven machine learning training set in the conventional sense. The "clinical study" described serves as the validation/test set to demonstrate performance.

9. How the Ground Truth for the Training Set Was Established

As noted above, a traditional "training set" with ground truth in the machine learning context is not directly applicable here. The development of such molecular assays relies on extensive scientific and laboratory work to design and optimize the assay components (primers, probes) to specifically target the desired viral RNA sequences and to establish robust reaction conditions. This involves:

  • Known viral strains: Using characterized strains of influenza A and B.
  • Nucleic acid sequencing: Detailed knowledge of viral genomic sequences to design specific and sensitive PCR targets.
  • Controlled experiments: Iterative testing and optimization of assay parameters using spiked samples with known concentrations.

{0}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

Indications for Use (Describe)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{1}------------------------------------------------

8.0 510(k) Summary

As required by 21 CFR Section 807.92(c).

Submitted by:Cepheid904 Caribbean DriveSunnyvale, CA 90489Phone number: (408) 400-6838
Contact:Yi-Ping Lin, PhD
Date of Preparation:May 14, 2018
Device:
Trade name:Xpert® Xpress Flu
Common name:Xpert Xpress Flu Assay
Type of Test:Automated, multiplex real-time reverse transcription-polymerase chain reaction (RT-PCR) assay intended for the in vitro qualitative detection and differentiation of influenza Aand influenza B virus.
Regulation number/Classification name/Product code:866.3980/Respiratory viral panel multiplex nucleic acidassay/OCC866.2570/Instrumentation for clinical multiplex testsystems/OOI
ClassificationClass II
Advisory PanelMicrobiology (83)
Prescription UseYes
Predicate DevicesAssay:1) For the detection and differentiation of influenza A,influenza B, and RSV A/B viral RNA in nasopharyngealswab specimens:Xpert® Flu/RSV XC Assay [510(k) #K142045]2) For the Sample Collection Kits:Cepheid Xpert® Nasopharyngeal Sample Collection Kitfor Viruses[510(k) # K171552]Cepheid Xpert® Nasal Sample Collection Kit forViruses[510(k) # K171552]

{2}------------------------------------------------

Device Description:

The Xpert Xpress Flu Assay is a rapid, automated in vitro diagnostic test for qualitative detection and differentiation of influenza A (Flu A) and influenza B (Flu B) viral RNA directly from nasopharyngeal (NP) swab and nasal swab (NS) specimens. The assay is performed on the Cepheid GeneXpert® Instrument Systems.

The Xpert Xpress Flu Assay includes reagents for the detection and differentiation of influenza A and influenza B viral RNA directly from NP swab and NS specimens from patients with signs and symptoms of respiratory tract infection. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are internal controls utilized by the GeneXpert Instrument System platform. The SPC is present to control for an adequate extraction and processing of the target sequences and to monitor for the presence of inhibitor in the PCR reaction. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

The specimens are collected in viral transport medium and transported to the GeneXpert area. The specimen is prepared according to package insert instructions and transferred to the sample chamber (large opening) of the Xpert Xpress Flu Assay cartridge. The GeneXpert cartridge is loaded onto the GeneXpert Xpress System platform, which performs hands-off automated sample processing and real-time RT-PCR for detection of Flu viral RNA. Summary and detailed test results are obtained in approximately 30 minutes or less. The results are automatically generated at the end of the process in a report that can be viewed and printed.

{3}------------------------------------------------

Device Intended Use:

The Cepheid Xpert® Xpress Flu Assay, performed on the GeneXpert® Instrument Systems, is an automated, multiplex real-time, reverse transcriptase polymerase chain reaction (RT-PCR) assay intended for the in vitro qualitative detection and differentiation of influenza A and influenza B viral RNA. The Xpert Xpress Flu Assay uses nasopharyngeal (NP) swab and nasal swab (NS) specimens collected from patients with signs and symptoms of respiratory infection. The Xpert Xpress Flu Assay is intended as an aid in the diagnosis of influenza infections in conjunction with clinical and epidemiological risk factors.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.

Performance characteristics for influenza A were established during the 2015-2016 influenza season for NP swab specimens and the 2016-2017 influenza season for NS specimens. When other novel influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Ancillary Nasopharyngeal Swab Specimen Collection Kit for Viruses

The Xpert® Nasopharyngeal Sample Collection Kit is designed to collect, preserve, and transport nasopharyngeal swab specimens and to preserve and transport nasal aspirate/wash specimens containing viruses from patients with signs and symptoms of respiratory infection prior to analysis with the Xpert Flu Assay or the Xpert Flu/RSV XC Assay. The Xpert® Nasopharyngeal Sample Collection Kit is designed to collect, preserve, and transport nasopharyngeal swab specimens containing viruses from patients

{4}------------------------------------------------

with signs and symptoms of respiratory infection prior to analysis with the Xpert Flu+RSV Xpress Assay, Xpert Xpress Flu/RSV Assay or the Xpert Xpress Flu Assay.

Ancillary Nasal Swab Specimen Collection Kit for Viruses

The Xpert® Nasal Sample Collection Kit is designed to collect, preserve, and transport nasal swab specimens containing viruses from patients with signs and symptoms of respiratory infection prior to analysis with the Xpert Xpress Flu Assay.

Substantial Equivalence:

The Xpert Xpress Flu Assay is substantially equivalent to the current Xpert® Flu/RSV XC Assay [510(k) #K142045]. The Xpert Xpress Flu Assay detects influenza A and influenza B from NP swab and NS specimens and the Xpert® Flu/RSV XC Assay detects influenza A, influenza B, and RSV from NP swab specimens and nasal aspirate/wash (NA/W) specimens. Both assays utilize the same technology by determining the presence of the target organisms through real-time RT-PCR amplification and fluorogenic target-specific hybridization detection. A multi-center clinical study was conducted and obtained data using the Xpert Xpress Flu/RSV Assay which was then reanalyzed with the Xpert Xpress Flu Assay Definition File (ADF). The reanalyzed data was used to determine the performance characteristics of the Xpert Xpress Flu Assay relative to the comparator Flu tests. Discordant results between the Xpert Xpress Flu Assay and the comparator methods Prodesse ProFlu+ Assay [510(k) #K132129] for NP swabs (FDA cleared for NP swabs) and Lyra Influenza A+B Assay [510(k) #K131728] for NS (FDA cleared for NP swabs and NS) were analyzed by bi-directional sequencing using primers different from those used in the Xpert Xpress Flu Assay. The study results showed that the Xpert Xpress Flu Assay is acceptable for its intended use and is substantially equivalent to the predicate device.

Table 8-1 shows the similarities and differences between the Xpert Xpress Flu Assay and the predicate device.

{5}------------------------------------------------

Similarities
DevicePredicate
ItemCepheid Xpert® Xpress FluCepheid Xpert® Flu/RSV XC510(k)# K142045
Regulation866.3980Same
Product CodeOCC, OOISame
Device ClassIISame
TechnologyPrinciple ofOperationMultiplex real time RT-PCRSame
Intended UseThe Cepheid Xpert® XpressFlu Assay, performed on theGeneXpert® InstrumentSystems, is an automated,multiplex real-time, reversetranscriptase polymerasechain reaction (RT-PCR)assay intended for the in vitroqualitative detection anddifferentiation of influenza Aand influenza B viral RNA.The Xpert Xpress Flu Assayuses nasopharyngeal (NP)swab and nasal swab (NS)specimens collected frompatients with signs andsymptoms of respiratoryinfection. The Xpert XpressFlu Assay is intended as anaid in the diagnosis ofinfluenza infections inconjunction with clinical andepidemiological risk factors.The Cepheid Xpert® Flu/RSVXC Assay is an automated,multiplex real-time, reversetranscriptase polymerase chainreaction (RT-PCR) assayintended for the in vitroqualitative detection anddifferentiation ofinfluenza A, influenza B, andrespiratory syncytial virus(RSV) viral RNA. The XpertFlu/RSV XC Assay usesnasopharyngeal swab andnasal aspirate/wash specimenscollected from patients withsigns and symptoms ofrespiratory infection. TheXpert Flu/RSV XC Assay isintended as an aid in thediagnosis of influenza andrespiratory syncytial virusinfections in conjunction withclinical and epidemiologicalrisk factors.
Similarities
ItemDevicePredicate
Cepheid Xpert® Xpress FluCepheid Xpert® Flu/RSV XC510(k)# K142045
Negative results do notpreclude influenza virusinfection and should not beused as the sole basis fortreatment or other patientmanagement decisions.Negative results do notpreclude influenza virus orrespiratory syncytial virusinfection and should not beused as the sole basis fortreatment or other patientmanagement decisions.
Performance characteristicsfor influenza A wereestablished during the 2015-2016 influenza season for NPswab specimens and the 2016-2017 influenza season for NSspecimens. When other novelinfluenza A viruses areemerging, performancecharacteristics may vary.Performance characteristics forinfluenza A were establishedduring the 2013-2014influenza season. When othernovel influenza A viruses areemerging, performancecharacteristics may vary.
If infection with a novelinfluenza A virus is suspectedbased on current clinical andepidemiological screeningcriteria recommended bypublic health authorities,specimens should be collectedwith appropriate infectioncontrol precautions for novelvirulent influenza viruses andsent to state or local healthdepartments for testing. Viralculture should not beattempted in these casesunless a BSL 3+ facility isavailable to receive andculture specimens.If infection with a novelinfluenza A virus is suspectedbased on current clinical andepidemiological screeningcriteria recommended bypublic health authorities,specimens should be collectedwith appropriate infectioncontrol precautions for novelvirulent influenza viruses andsent to state or local healthdepartments for testing. Viralculture should not beattempted in these cases unlessa BSL 3+ facility is availableto receive and culturespecimens.

Table 8-1: Comparison of Similarities and Differences of the Xpert Xpress Flu Assay with the Predicate Devices

{6}------------------------------------------------

{7}------------------------------------------------

Similarities
DevicePredicate
ItemCepheid Xpert® Xpress FluCepheid Xpert® Flu/RSV XC510(k)# K142045
Indication forUsePatients with signs andsymptoms of respiratoryinfection in conjunction withclinical and epidemiologicalrisk factorsSame
Nucleic AcidExtractionYesSame
ExtractionMethodsSample preparation integratedin GeneXpert Cartridge andGeneXpert InstrumentationSystemSame
Assay ResultsQualitativeSame
InstrumentSystemCepheid GeneXpertInstrument Systems; sameCepheid I-core technologySame
Primers andprobesPrimers and probes to detectthe presence of nucleic acidsequences of influenza A,influenza B, and RSV. Onlyresults for influenza A andinfluenza B are reported.Primers and probes to detectthe presence of influenza A,influenza A subtype H7N9,influenza B and RSV. Resultsfor influenza A, influenza Band RSV analytes are reported.
LaboratoryUsersLaboratory users in moderateand high complexitylaboratory settings.Same
SamplePreparationSelf-contained and automatedafter mixed specimen is addedto cartridge. All otherreagents are contained in thecartridge.Same

{8}------------------------------------------------

Similarities
ItemDevicePredicate
Primers andprobes forinfluenza A,influenza BPrimers and probes to detectthe presence of nucleic acidsequences of influenza A,influenza B, and RSV A/B.The Xpert Xpress Flu Assaycontains primers and probes todetect additional RNAsegments in order to protectthe assay sensitivity andspecificity from mutations inthe influenza genome due toantigenic drifts and shifts.Only results for influenza Aand influenza B are reported.Primers and probes to detectthe presence of nucleic acidsequences of influenza A,influenza B, and RSV A/B.The Xpert Flu/RSV XC Assaycontains primers and probes todetect additional RNAsegments in order to protectthe assay sensitivity andspecificity from mutations inthe influenza genome due toantigenic drifts and shifts.
TargetSequencesInfluenza A: Matrix protein(MP),basic polymerase (PB2)and acidic protein (PA)Influenza B: Matrix protein(MP) and Non-structuralproteins (NS 1 and NS 2)RSV A and RSV B:Nucleocapsid proteinOnly results for influenza Aand influenza B are reported.Influenza A: Matrix protein(MP),basic polymerase (PB2)and acidic protein (PA)Influenza B: Matrix protein(MP) and Non-structuralproteins (NS 1 and NS 2)RSV A and RSV B:Nucleocapsid protein
InternalControlsSample processing control(SPC) and probe checkcontrol (PCC).Same
Early assayterminationfunctionYesYes
Cepheid Xpert® Xpress FluCepheid Xpert® Flu/RSV XC510(k)# K142045
Differences
DevicePredicate
ItemCepheid Xpert® Xpress FluCepheid Xpert® Flu/RSV XC510(k)# K142045
Assay TargetsInfluenza A Virus andInfluenza B VirusInfluenza A Virus, Influenza BVirus, and RSV viral RNA
SpecimenTypesNasopharyngeal (NP) swaband nasal swab (NS)specimensNasal aspirate/wash (NA/W)specimens andNasopharyngeal (NP) swabspecimens
Assay ControlsEncapsulated (armored) RNApseudovirus as a sampleprocessing control.Available but not provided areinactivated virus controls forinfluenza A/B as externalpositive controls, andCoxsackie virus as an externalnegative control.Encapsulated (armored) RNApseudovirus as a sampleprocessing control.Available but not provided areinactivated virus controls forinfluenza A/B and RSV asexternal positive controls, andCoxsackie virus as an externalnegative control.
Time to obtaintest resultsApproximately 30 minutes orless for sample preparationand RT-PCRApproximately 60 minutes orless for sample preparation andRT-PCR
CombinatorialAssaySelectionsNot applicableYes, user may select combinedassay with all targets or a Fluonly assay or a RSV onlyassay.

{9}------------------------------------------------

{10}------------------------------------------------

The Xpert Xpress Flu Assay and the predicate device have the same general intended use and technological characteristics, and both detect influenza Aand influenza B viral RNA from NP swab specimens. The clinical study demonstrates that the Xpert Xpress Flu Assay is acceptable for its intended use and is substantially equivalent to the predicate device.

The predicate device for the ancillary specimen collection kit, the Xpert Nasopharyngeal Sample Collection Kit for Viruses is the Cepheid Nasopharyngeal Sample Collection Kit for Viruses, [510(k) # K171552]. The similarities are shown in Table 8-2. There is no difference between the Nasopharyngeal Sample Collection Kit for Viruses cleared in 510(k) # K171552 and this 510(k).

The predicate device for the ancillary specimen collection kit, the Xpert Nasal Sample Collection Kit for Viruses is the Xpert Nasal Sample Collection Kit for Viruses, [510(k) # K171552]. The similarities are shown in Table 8-3. There is no difference between the Nasal Sample Collection Kit for Viruses cleared in 510(k) # K171552 and this 510(k).

{11}------------------------------------------------

Similarities
DevicePredicate
ItemXpert® NasopharyngealSample Collection Kit forVirusesXpert® NasopharyngealSample Collection Kit forViruses510(k)# K171552
Intended UseThe Xpert® NasopharyngealSample Collection Kit isdesigned to collect, preserve,and transport nasopharyngealswab specimens and to preserveand transport nasalaspirate/wash specimenscontaining viruses from patientswith signs and symptoms ofrespiratory infection prior toanalysis with the Xpert FluAssay or the Xpert Flu/RSV XCAssay. The XpertNasopharyngeal SampleCollection Kit is designed tocollect, preserve, and transportnasopharyngeal swab specimenscontaining viruses from patientswith signs and symptoms ofrespiratory infection prior toanalysis with the XpertFlu+RSV Xpress Assay, XpertXpress Flu/RSV Assay or theXpert Xpress Flu Assay.Same
Single-use DeviceYesSame

Table 8-2: Comparison of Similarities of the Xpert Nasopharyngeal Sample Collection Kit with the Predicate Device

{12}------------------------------------------------

Similarities
ItemDevicePredicate
Xpert® NasopharyngealSample Collection Kit forVirusesXpert® NasopharyngealSample Collection Kit forViruses510(k)# K171552
Transport MediumFormulationHank's Balanced SaltSolution Bovine SerumAlbuminL-cysteineGelatinSucroseL-glutamic acidHEPES bufferVancomycinAmphotericin BColistinPhenol redSame
pH$7.3 \pm 0.2$Same
Storage Temperature2 - 25°C (refrigeratedand room temperature)Same
Volume3 mlSame
Glass Beads3 x 3 mmSame
ContainerPlastic (medical-gradepolypropylene)Same
Product ConfigurationMedium Tube in Kit withindividually-wrapped sterileswab.Same

{13}------------------------------------------------

Similarities
DevicePredicate
ItemXpert® Nasal SampleCollection Kit for VirusesXpert® Nasal SampleCollection Kit for Viruses510(k)# K171552
Intended UseThe Xpert® Nasal SampleCollection Kit is designed tocollect, preserve, and transportnasal swab specimenscontaining viruses from patientswith signs and symptoms ofrespiratory infection prior toanalysis with the Xpert XpressFlu Assay.Same
Single-use DeviceYesSame
Transport MediumFormulationHank's Balanced SaltSolution Bovine SerumAlbuminL-cysteineGelatinSucroseL-glutamic acidHEPES bufferVancomycinAmphotericin BColistinPhenol redSame
pH$7.3 \pm 0.2$Same
Storage Temperature2 - 25°C (refrigeratedand room temperature)Same
Volume3 mlSame
Glass Beads3 x 3 mmSame
ContainerPlastic (medical-gradepolypropylene)Same
Similarities
ItemDevicePredicate
Product ConfigurationXpert® Nasal Sample Collection Kit for VirusesMedium Tube in Kit withindividually-wrapped sterile swab.Xpert® Nasal Sample Collection Kit for Viruses510(k)# K171552
SwabNylon flockedSame

Table 8-3: Comparison of Similarities and Differences of the Xpert Nasal Sample Collection Kit for Viruses with the Predicate Device

{14}------------------------------------------------

The proposed collection kits and predicate collection kits have the same general intended use and the same technology to collect, store and transport clinical specimens, including viruses, to the laboratory for further testing. The prospective component of the multicenter clinical study of the Xpert Xpress Flu Assay was conducted using Xpert Nasopharyngeal Sample Collection Kit for Viruses [510(k) # K171552] and Xpert Nasal Sample Collection Kit for Viruses [510(k) # K171552] demonstrating that the Xpert Nasopharyngeal Sample Collection Kit for Viruses and Xpert Nasal Sample Collection Kit for Viruses are acceptable for their intended use and substantially equivalent to the predicate devices.

Non-Clinical Studies:

Analytical Sensitivity (Limit of Detection)

Studies were performed to determine the analytical limit of detection (LoD) of the Xpert Xpress Flu Assay with two lots of reagents across three testing days. The higher LoD observed per strain and per lot was selected for verification. Verification of the estimated LoD claim was performed on one reagent lot across a minimum of three testing days. LoD was established using two influenza A H3N2 strains, two influenza A 2009 H1N1 strains and two influenza B strains. Viruses were diluted into negative pooled NP swab and NS clinical matrices for testing. The LoD is defined as the lowest concentration (tissue culture infective dose. TCID50/mL) per sample that can be

{15}------------------------------------------------

reproducibly distinguished from negative samples with 95% confidence or the lowest concentration at which 19 of 20 replicates were positive. Each strain was tested in replicates of 20 per concentration of virus in each matrix in NP swab and NS clinical matrix. The LoD point values for each strain tested are summarized in Table 8-4 through Table 8-6.

Virus StrainConfirmed LoD Probit(TCID50/mL)
NP SwabNS
Influenza A/California/7/20090.0200.018
Influenza A/Florida/27/20110.0400.04

Table 8-4: Confirmed LoD (TCID50/mL): Influenza A 2009 H1N1

Table 8-5: Confirmed LoD (TCID50/mL): Influenza A H3N2

Virus StrainConfirmed LoDProbit(TCID50/mL)
NP SwabNS
Influenza A/Perth/16/20090.0130.006
Influenza A/Victoria/361/20110.7500.21
Table 8-6: Confirmed LoD (TCID50/mL): Influenza B
-----------------------------------------------------
Virus StrainConfirmed LoDProbit(TCID50/mL)
NP SwabNS
Influenza B/Mass/2/20120.4000.07
Influenza B/Wisconsin/01/20110.1900.17

{16}------------------------------------------------

Analytical Specificity (Exclusivity)

The analytical specificity of the Xpert Xpress Flu Assay was evaluated by testing a panel of 44 cultures consisting of 16 viral, 26 bacterial, and two yeast strains representing common respiratory pathogens or those potentially encountered in the nasal passage and nasopharynx. Three replicates of all bacterial and yeast strains were tested at concentrations of ≥ 1 x 106 CFU/mL with the exception of one strain that was tested at 1 x 105 CFU/mL (Chlamydia pneumoniae). Three replicates of all viruses were tested at concentrations of ≥ 1 x 105 TCID50/mL. The analytical specificity was 100%. Results are shown in Table 8-7.

{17}------------------------------------------------

Result
OrganismConcentration(per cartridge)InfluenzaAInfluenzaB
No Template ControlN/ANEGNEG
Adenovirus Type 11.12E+06 TCID50/mLNEGNEG
Adenovirus Type 71.87E+05 TCID50/mLNEGNEG
Human coronavirus OC432.85E+05 TCID50/mLNEGNEG
Human coronavirus 229E1.00E+05 TCID50/mLNEGNEG
Cytomegalovirus1.00E+05 TCID50/mLNEGNEG
Echovirus3.31E+07 TCID50/mLNEGNEG
Enterovirus3.55E+05 TCID50/mLNEGNEG
Epstein Barr Virus7.16E+07 TCID50/mLNEGNEG
HSV8.90E+05 TCID50/mLNEGNEG
Measles6.31E+05 TCID50/mLNEGNEG
Human metapneumovirus1.00E+05 TCID50/mLNEGNEG
Mumps virus6.31E+06 TCID50/mLNEGNEG
Human parainfluenza Type 11.15E+06 TCID50/mLNEGNEG
Human parainfluenza Type 26.31E+05 TCID50/mLNEGNEG
Human parainfluenza Type 33.55E+06 TCID50/mLNEGNEG
Rhinovirus Type 1A1.26E+05 TCID50/mLNEGNEG
Acinetobacter baumannii1.00E+06 CFU/mLNEGNEG
Burkholderia cepacia3.30E+06 CFU/mLNEGNEG
Candida albicans3.20E+06 CFU/mLNEGNEG
Candida parapsilosis3.00E+06 CFU/mLNEGNEG
Bordetella pertussis3.30E+06 CFU/mLNEGNEG
Chlamydia pneumoniae1.00E+05 CFU/mLNEGNEG
Citrobacter freundii3.30E+06 CFU/mLNEGNEG
Corynebacterium sp.3.30E+06 CFU/mLNEGNEG
Escherichia coli1.00E+07 CFU/mLNEGNEG
Enterococcus faecalis1.30E+06 CFU/mLNEGNEG
Hemophilus influenzae1.00E+06 CFU/mLNEGNEG
Lactobacillus reuteri1.00E+06 CFU/mLNEGNEG
Legionella spp.1.00E+06 CFU/mLNEGNEG
Moraxella catarrhalis1.00E+07 CFU/mLNEGNEG
OrganismConcentration(per cartridge)Result
InfluenzaAInfluenzaB
Mycobacterium tuberculosis(avirulent)1.00E+06 CFU/mLNEGNEG
Mycoplasma pneumoniae1.00E+06 CFU/mLNEGNEG
Neisseria meningitides2.15E+06 CFU/mLNEGNEG
Neisseria mucosa1.00E+07 CFU/mLNEGNEG
Propionibacterium acnes2.40E+07 CFU/mLNEGNEG
Pseudomonas aeruginosa3.70E+06 CFU/mLNEGNEG
Staphylococcus aureus (protein Aproducer)2.20E+06 CFU/mLNEGNEG
Staphylococcus epidermidis3.40E+06 CFU/mLNEGNEG
Staphyloccus haemolyticus4.00E+06 CFU/mLNEGNEG
Streptococcus agalactiae3.50E+06 CFU/mLNEGNEG
Streptococcus pneumoniae1.00E+06 CFU/mLNEGNEG
Streptococcus pyogenes1.00E+07 CFU/mLNEGNEG
Streptococcus salivarius1.00E+07 CFU/mLNEGNEG
Streptococcus sanguinis3.10E+06 CFU/mLNEGNEG

Table 8-7: Analytical Specificity of the Xpert Xpress Flu/RSV Assay

{18}------------------------------------------------

{19}------------------------------------------------

Analytical Reactivity (Inclusivity)

The analytical reactivity of the Xpert Xpress Flu Assay was evaluated against multiple strains of influenza A H1N1 (seasonal pre-2009), influenza A H1N1 (pandemic 2009), influenza A H3N2 (seasonal), avian influenza A (H5N1, H5N2, H6N2, H7N2, H7N3, H2N2, H7N9, and H9N2) and influenza B (representing strains from both Victoria and Yamagata lineages) at levels near the analytical LoD. A total of 48 strains comprised of 35 influenza A and 13 Influenza B strains were tested in this study with the Xpert Xpress Flu Assay. Three replicates were tested for each strain. All Flu strains tested positive in all three replicates, except for one Flu A H1N1 strain (A/New Jersey/8/76), which tested positive in 2 of 3 replicates at 0.1 TCID50/mL. Results are shown in Table 8-8.

Predicted cross reactivity from in silico analyses showed 100% sequence homology for additional pandemic H1N1 strains.

{20}------------------------------------------------

VirusStrainTargetConcentrationResult
No Template Controln/aNEGNEG
InfluenzaA H1N1(pre-2009)A/swine/Iowa/15/300.1 TCID50/mLPOSNEG
A/WS/330.1 TCID50/mLPOSNEG
A/PR/8/340.1 TCID50/mLPOSNEG
A/Mal/302/540.1 TCID50/mLPOSNEG
A/Denver/1/570.1 TCID50/mLPOSNEG
A/New Jersey/8/760.1 TCID50/mLPOSNEG
A/New Caledonia/20/19990.1 TCID50/mLPOSNEG
A/New York/55/20040.1 TCID50/mLPOSNEG
A/Soloman Island/3/20060.1 TCID50/mLPOSNEG
A/Taiwan/42/060.1 TCID50/mLPOSNEG
A/Brisbane/59/20070.1 TCID50/mLPOSNEG
InfluenzaA H1N1(pdm2009)A/swine/NY/02/20090.1 TCID50/mLPOSNEG
A/Colorado/14/20120.1 TCID50/mLPOSNEG
A/Washington/24/20120.1 TCID50/mLPOSNEG
InfluenzaA H3N2(Seasonal)A/Aichi/2/682.0 TCID50/mLPOSNEG
A/HongKong/8/682.0 TCID50/mLPOSNEG
A/Port Chalmers/1/732.0 TCID50/mLPOSNEG
A/Hawaii/15/20012.0 TCID50/mLPOSNEG
A/Wisconsin/67/052.0 TCID50/mLPOSNEG
A/Brisbane/10/20072.0 TCID50/mLPOSNEG
A/Minnesota/11/2010 (H3N2)v2.0 TCID50/mLPOSNEG
A/Indiana/08/2011 (H3N2)v2.0 TCID50/mLPOSNEG
A/Texas/50/20122.0 TCID50/mLPOSNEG
Avianinfluenza AA/duck/Hunan/795/2002 (H5N1)≤ 1pg/μLaPOS
A/chicken/Hubei/327/2004(H5N1)≤ 1pg/μLaPOSNEG
A/Anhui/01/2005 (H5N1)≤ 1pg/μLaPOSNEG
TargetResult
VirusStrainConcentrationFlu AFlu B
A/Japanese white eye/ HongKong/ 1038/2006 (H5N1)≤ 1pg/μLaPOSNEG
A/mallard/WI/34/75 (H5N2)≤ 1pg/μLaPOSNEG
A/chicken/CA431/00 (H6N2)≤ 1pg/μLaPOSNEG
A/duck/LTC-10-82743/1943(H7N2)≤ 1pg/μLaPOSNEG
A/chicken/NJ/15086-3/94(H7N3)≤ 1pg/μLaPOSNEG
A/Anhui/1/2013 (H7N9)N/AbPOSNEG
A/Shanghai/1/2013 (H7N9)N/AbPOSNEG
A/chicken/Korea/38349-p96323/1996 (H9N2)≤ 1pg/μLaPOSNEG
A/Mallard/NY/6750/78 (H2N2)≤ 1pg/μLaPOSNEG
Influenza BB/Lee/401.0 TCID50/mLNEGPOS
B/Allen/451.0 TCID50/mLNEGPOS
B/GL/1739/541.0 TCID50/mLNEGPOS
B/Maryland/1/591.0 TCID50/mLNEGPOS
B/Panama/45/90C1.0 TCID50/mLNEGPOS
B/Florida/07/2004d1.0 TCID50/mLNEGPOS
B/Florida/02/06C1.0 TCID50/mLNEGPOS
B/Florida/04/06d1.0 TCID50/mLNEGPOS
B/Hong Kong/5/721.0 TCID50/mLNEGPOS
B/Wisconsin/01/2010d1.0 TCID50/mLNEGPOS
B/Malaysia/2506/04c1.0 TCID50/mLNEGPOS
B/Taiwan/2/621.0 TCID50/mLNEGPOS
B/Brisbane/60/2008c1.0 TCID50/mLNEGPOS

Table 8-8: Analytical Reactivity (Inclusivity) of the Xpert Xpress Flu/RSV Assay

{21}------------------------------------------------

a. Purified viral RNA in simulated background matrix was used for avian influenza A viruses due to biosafety regulations.

b. Inactivated avian influenza A (H7N9) viruses without viral titer was diluted 100,000 fold in simulated background matrix and tested due to biosafety regulations.

  • c. Known Victoria lineage.
    d. Known Yamagata lineage.

{22}------------------------------------------------

Potentially Interfering Substances

In a non-clinical study, potentially interfering substances that may be present in the nasal passage and nasopharynx were evaluated directly relative to the performance of the Xpert Xpress Flu Assay. Potentially interfering substances in the nasal passage and nasopharynx may include, but are not limited to: blood, nasal secretions or mucus, and nasal and throat medications used to relieve congestion, nasal dryness, irritation, or asthma and allergy symptoms, as well as antibiotics and antivirals. Negative samples (n = 8) were tested per each substance to determine the effect on the performance of the sample processing control (SPC). Positive samples (n = 8) were tested per substance with six influenza (four influenza A and two influenza B) strains spiked at 3X the analytical LoD determined for each strain. All results were compared to positive and negative simulated background matrix controls. The simulated background matrix consisted of 2.5% (w/v) porcine mucin, 1% (v/v) human whole blood in 0.85% sodium chloride (NaCl) formulated in 1x PBS solution with 15% glycerol, which was then diluted 1:5 in UTM.

The evaluated substances are listed in Table 8-9 with active ingredients and concentrations tested shown. None of the substances caused interference of the assay at the concentrations tested in this study. All positive and negative replicates were identified correctly using the Xpert Xpress Flu Assay.

Substance/ClassDescription/ActiveIngredientConcentrationTested
ControlSimulated backgroundmatrix100% (v/v)
Beta-adrenergic bronchodilatorAlbuterol Sulfate0.83 mg/mL(equivalent to 1dose per day)
BloodBlood (Human)2% (v/v)
BDTM Universal ViralTransport SystemTransport Media100% (v/v)
Remel M4®Transport Media100% (v/v)
Remel M4RT®Transport Media100% (v/v)
Remel M5®Transport Media100% (v/v)

Table 8-9: Potentially Interfering Substances in the Xpert Xpress Flu Assay

{23}------------------------------------------------

Substance/ClassDescription/Active IngredientConcentration Tested
Remel M6®Transport Media100% (v/v)
Throat lozenges, oralanesthetic and analgesicBenzocaine, Menthol1.7 mg/mL
MucinPurified Mucin protein(Bovine or porcinesubmaxillary gland)2.5% (w/v)
Antibiotic, nasal ointmentMupirocin10 mg/mL
Saline Nasal SpraySodium Chloride (0.65%)15% (v/v)
Anefrin Nasal SprayOxymetazoline, 0.05%15% (v/v)
PHNY Nasal DropsPhenylephrine, 0.5%15% (v/v)
Tamiflu Anti-viral drugZanamivir7.5 mg/mL
Antibacterial, systemicTobramycin4 µg/mL
Zicam Nasal GelLuffa opperculata,Galphimia glauca,Histaminumhydrochloricum Sulfur15% (w/v)
Nasal corticosteroidFluticasone Propionate5 µg/mL

Carry-Over Contamination

A study was conducted to demonstrate that single-use, self-contained GeneXpert cartridges prevent carry-over contamination of negative samples if preceded by very high positive samples in the same GeneXpert module. The study consisted of a negative sample processed in the same GeneXpert module immediately following a very high influenza A sample (A/Victoria/361/2011, 2x107 TCID50/mL) spiked into a simulated background matrix. This testing scheme was repeated 20 times on two GeneXpert modules for a total of 41 runs resulting in 20 positive and 21 negative specimens for each virus type. All 20 positive samples were correctly reported as Flu A POSITIVE; Flu B NEGATIVE. All 21 negative samples were correctly reported as Flu A NEGATIVE; Flu B NEGATIVE.

Fresh vs. Frozen Sample Equivalency Study

Fresh and frozen specimen equivalency in the Xpert Xpress Flu Assay was evaluated by testing individual influenza strains at three different concentrations representing low positives (2X LoD), moderate positives (5X LoD), and high positives (10X LoD) in

{24}------------------------------------------------

pooled negative NP swab clinical matrix. Negative samples consisted of pooled negative NP swab clinical matrix only. Fresh and frozen specimen equivalency was determined using one seasonal Flu A H3N2 strain (A/Victoria/361/2011) and one Flu B strain (B/Mass/2/2012). Replicates of 20 were tested for each specimen type and concentration. All positive and negative specimens were tested fresh, after one freezethaw cycle, and after two freeze-thaw cycles. There was no statistically significant effect in the performance of the Xpert Xpress Flu Assay between fresh virus dilutions and two sequential freeze thaw cycles for positive and negative samples. All positive and negative replicates were correctly identified using the Xpert Xpress Flu Assay.

Competitive Interference Study

Competitive interference of the assay caused by the presence of two targets in the Xpert Xpress Flu Assay was evaluated by testing individual influenza strains near the LoD in the presence of different influenza strains at a higher concentration in a simulated background matrix. Analytical competitive interference was assessed using one (1) seasonal Flu A H3 strain (H3/Victoria/361/2011) at 0.8 TCID50/mL and one (1) Flu B strain (B/Mass/2/2012) at 0.45 TCID50/mL; the strains were tested at in the presence of competing strains at either 1 x 102 TCID50/mL or 1 x 103 TCID50/mL. Replicates of 20 were tested for each target strain and each competitive strain combination. The normal binomial distribution with 20 replicate samples at LoD is between 17 and 20 positive results based on the binomial distribution with N=20, p=0.95 (X~Bin(20,0.95)). Therefore, sets of 20 with 16 or less positives would be rare and an indication of a competitive inhibitory effect due to high levels of a competing analyte.

  • . With Flu A/Victoria/361/2011 at a concentration of 0.8 TCID50/mL no competitive inhibitory effects were observed in the presence of 1x103 TCID50/mL of Flu B/Mass/2/2012.
  • With Flu B/Mass/2/2012 at a concentration of 0.45 TCID50/mL competitive . inhibitory effects were observed in the presence of 1x103 TCID50/mL of Flu

{25}------------------------------------------------

A/Victoria/361/2011. No competitive inhibitory effects were observed in the presence of 1x102 TCID50/mL of Flu A/Victoria/361/2011.

Under the conditions of this study, internal competitive inhibitory effects were observed on the targets (Flu A and Flu B) in the presence of two targets for the Xpert Xpress Flu Assay. The competitive inhibitory effect on the Xpert Xpress Flu targets is addressed in the Limitations section of this package insert.

Clinical Studies

Clinical Comparison Study

Performance characteristics of the Xpert Xpress Flu Assay were evaluated at eleven institutions in the U.S. during the 2015-2016 influenza season for NP swab specimens and at fourteen institutions in the U.S. during the 2016-2017 influenza season for NS specimens. Due to the low prevalence of influenza viruses and the difficulty in obtaining fresh influenza specimens, the specimen population for this study was supplemented with consecutively collected, frozen specimens.

Specimens were collected from the following:

  • . Individuals exhibiting signs and symptoms of respiratory infection who provided informed consent for the collection of a NP swab or NS specimen.
  • . Individuals with signs and symptoms of respiratory infection and whose routine care called for collection of NP swab specimens for influenza testing. For eligible subjects, aliquots of leftover specimens were obtained for testing with the Xpert Xpress Flu Assay and reference testing, and patient management continued at the site per their standard practice.

The Xpert Xpress Flu Assay performance was compared to an FDA-cleared molecular comparator assay. Bi- directional sequencing was performed on specimens where the Xpert Xpress Flu Assay and the comparator assay were discrepant, and is provided for informational purposes only.

{26}------------------------------------------------

Overall Results – NP Swab Specimens

A total of 2051 NP swab specimens were tested for influenza A and influenza B by the Xpert Xpress Flu Assay and the comparator assay. Of the 2051 NP swab specimens. 1139 were fresh, prospectively collected and 912 were consecutively collected, frozen specimens.

For the fresh, prospectively collected NP swab specimens, the Xpert Xpress Flu Assay demonstrated a PPA and NPA of 94.6% and 99.4% for the detection of influenza A and 100% and 99.3% for influenza B, respectively (Table 8-10).

For the consecutively collected, frozen NP swab specimens, the Xpert Xpress Flu Assay demonstrated a PPA and NPA of 100% and 98.0% for the detection of influenza A, respectively; 100% and 99.0% for influenza B, respectively (Table 8-10).

For the combined dataset, the Xpert Xpress Flu Assay demonstrated a PPA and NPA of 98.1% and 98.8% for the detection of influenza A and 100% and 99.1% for influenza B, respectively (Table 8-10).

{27}------------------------------------------------

SpecimenTypeTargetnTPFNTNFPPPA(95% CI)NPA(95% CI)
FreshFlu A1139352a10957b94.6%(82.3 - 98.5)99.4%(98.7 - 99.7)
FreshFlu B113942010898c100.0%(91.6 - 100.0)99.3%(98.6 - 99.6)
FrozenConsecutivelyCollectedFlu A91268082717d100.0%(94.7 - 100.0)98.0%(96.8 - 98.7)
FrozenConsecutivelyCollectedFlu B9123608679e100.0%(90.4 - 100.0)99.0%(98.1 - 99.5)
CombinedFlu A20511032a192224f98.1%(93.3 - 99.5)98.8%(98.2 - 99.2)
CombinedFlu B2051780195617g100.0%(95.3 - 100.0)99.1%(98.6 - 99.5)

Table 8-10: Xpert Xpress Flu Assay Performance on NP Swab Specimens

a. Testing results by sequencing: 2 of 2 were Flu A Negative.

b. Testing results by sequencing: 3 of 7were Flu A Positive; 3 of 7 were Flu A Negative; 1 of 7 insufficient specimen for sequencing.

c. Testing results by sequencing: 6 of 8 were Flu B Positive: 1 of 8 were Flu B Negative; 1 of 8 insufficient specimen for sequencing.

d. Testing results by sequencing: 7 of 17 were Flu A Positive; 7 of 17 were Flu A Negative; 3 of 17 insufficient specimen for sequencing.

e. Testing results by sequencing: 7 of 9 were Flu B Positive; 0 of 9 were Flu B Negative; 2 of 9 insufficient specimen for sequencing.

f. Testing results by sequencing: 10 of 24 were Flu A Positive; 10 of 24 were Flu A Negative; 4 of 24 insufficient specimen for sequencing.

g. Testing results by sequencing: 13 of 17 were Flu B Positive; 1 of 17 were Flu B Negative; 3 of 17 insufficient specimen for sequencing.

For the 98 pre-selected frozen NP swab specimens that were excluded from the analysis above, the Xpert Xpress Flu Assay demonstrated a PPA and NPA of 100% and 97.8%, for influenza A, respectively; and 100% and 96.6% for influenza B, respectively.

Overall Results - NS Specimens

A total of 1598 NS specimens were tested for influenza A and influenza B by the Xpert Xpress Flu Assay and the comparator assay.

The Xpert Xpress Flu Assay demonstrated a PPA and NPA of 98.9% and 97.6% for the detection of influenza A, and 98.4% and 99.3% for influenza B, respectively, relative to the comparator assay (Table 8-11).

{28}------------------------------------------------

TargetaNTPFNTNFPPPA(95% CI)NPA(95% CI)
Flu A15981862a137634b98.9%(96.2-99.7)97.6%(96.6-98.3)
Flu B1598631c152311d98.4%(91.7-99.7)99.3%(98.7-99.6)
Table 8-11: Xpert Xpress Flu Assav Performance on NS Specimens
--------------------------------------------------------------------------------

aDiscrepant Testing: 1 of 2 Flu A NEG; 1 of 2 Flu A POS.

bDiscrepant Testing: 16 of 34 Flu A NEG; 11 of 34 Flu A POS; 7 of 34 inconclusive.

°Discrepant Testing: 1 of 1 inconclusive.

dDiscrepant Testing: 5 of 11 Flu B POS; 6 of 11 inconclusive.

Of the Xpert Xpress Flu Assay runs performed with eligible NP swab and NS specimens, 97.8% (3594/3674) of these specimens were successful on the first attempt. The remaining 80 gave indeterminate results on the first attempt (39 ERROR, 32 INVALID, and 9 NO RESULT). Sixty of the 80 indeterminate cases were retested, of which 54 yielded valid results upon repeat testing; 20 specimens were not retested. The overall rate of assay success was 99.3% (3648/3674). The overall indeterminate rate was 0.7%.

Reproducibility Study

Reproducibility was established in a multi-center, blinded study using a 5-member specimen panel. Testing was performed at three sites (one internal, two external) using the GeneXpert Dx system, the Infinity-48 system, and the Infinity-80 system. Testing was conducted for 6 (not necessarily consecutive) days, with three lots of cartridges and consisted of two testing days per lot. Each site had two operators, one experienced and one inexperienced, who tested each panel in duplicate twice each day. Results are summarized in Table 8-12.

{29}------------------------------------------------

Sample IDTiter ofVirus(TCID50/mLSite 1/Infinity-80Site 2/DXSite 3/Infinity-48% Total
Op 1Op 2SiteOp 1Op 2SiteOp 1Op 2SiteAgreement bySample®
Negative0100%(24/24)100%(24/24)100%(48/48)100%(24/24)100%(24/24)100%(48/48)100%(24/24)100%(24/24)100%(48/48)100%(144/144)
Flu A-0.7587.0%95.8%91.5%95.7%91.7%93.6%100%91.3%95.7%93.6%
Low Pos(20/23)(23/24)(43/47)(22/23)(22/24)(44/47)(23/23)(21/23)(44/46)(131/140)b
Flu A-1.5100%100%100%100%100%100%100%100%100%100%
Mod Pos(24/24)(24/24)(48/48)(23/23)(23/23)(46/46)(24/24)(24/24)(48/48)(142/142)b
Flu B-0.295.8%95.8%95.8%95.8%95.8%95.8%95.8%91.7%93.8%95.1%
Low Pos(23/24)(23/24)(46/48)(23/24)(23/24)(46/48)(23/24)(22/24)(45/48)(137/144)
Flu B-0.4100%100%100%100%100%100%100%100%100%100%
Mod Pos(23/23)(24/24)(47/47)(24/24)(24/24)(48/48)(24/24)(23/23)(47/47)(142/142)b

Table 8-12: Summary of Reproducibility Results

Agreement calculated based on expected result: Negative (targeted positivity: 0%); Positive for Low Pos (targeted positivity: 95%) and a. Mod Pos (targeted positivity: 100%) samples.

Eight samples indeterminate [Flu A Low Pos (4); Flu A Mod Pos (2); Flu B Mod Pos (2)] b.

The reproducibility of the Xpert Xpress Flu Assay was also evaluated in terms of the fluorescence signal expressed in Ct values for each target detected. The mean, standard deviation (SD), and coefficient of variation (CV) between-sites, between- days, between-lots and between-operators for each panel member are presented in Table 8-13.

{30}------------------------------------------------

AssayChannel(Analyte)NaMeanCtBetween-SiteBetween-LotBetween-DayBetween-OperatorWithin-AssayTotal
SampleSDCV(%)SDCV(%)SDCV(%)SDCV(%)SDCV(%)SDCV(%)
NegativeSPC14432.3000.72.10.10.4000.61.90.92.8
Flu A-Low PosFluA113135.3000.61.600001.13.01.23.4
Flu A-Mod PosFluA114233.10000.10.20.6000.61.80.61.9
Flu B-Low PosFluB13734.600000.51.30.41.21.33.91.54.2
Flu B-Mod PosFluB14232.30.10.30.30.8000.30.80.82.40.92.7

Table 8-13: Summary of Reproducibility Data

Results with non-zero Ct values of 144. a.

{31}------------------------------------------------

Conclusions

The results of the nonclinical analytical and clinical performance studies summarized above demonstrate that the Xpert Xpress Flu Assay is substantially equivalent to the predicate device.

{32}------------------------------------------------

Image /page/32/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 15, 2018

Cepheid Jim Kelly, Ph.D. Executive Director, Regulatory Affairs 904 Caribbean Drive Sunnyvale, CA 94089

Re: K181289

Trade/Device Name: Xpert Xpress Flu, Xpert Nasopharyngeal Sample Collection Kit, Xpert Nasal Sample Collection Kit, GeneXpert Dx Systems (GX-I, GX-II, GX-IV, GX-XVI), GeneXpert Infinity-48S System and GeneXpert Infinity-80 System Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OCC, OOI, JSM Dated: May 14, 2018 Received: May 17, 2018

Dear Jim Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{33}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.