K Number
K181220

Validate with FDA (Live)

Date Cleared
2018-10-12

(158 days)

Product Code
Regulation Number
872.3760
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Juell 3D Volo Base is a light-cured resin indicated for the fabrication of denture bases fabricated in dental laboratories, including full and partial removable dentures. The material is an alternative to traditional heat cured and auto polymerization resins. Juell 3D Volo Base is intended exclusively for professional dental work. Fabrication of denture bases with Juell 3D Volo Base requires a computer-aided and manufacturing (CAD/CAM) system that includes the following: scanner, design software, additive printer and post-cure unit.

Device Description

Juell 3D Volo Base 3D-printing material is a light-cured resin indicated for the manufacturing of denture bases. The material is used in a 3D printer, which prints the shape determined by a 3D stereolithographic drawing. After printed product is placed in a UV-light curing box for final polymerization.

3D printer is not included with the device.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a dental device, Juell 3D Volo Base. The device is a light-cured resin intended for the fabrication of denture bases using a CAD/CAM system. The submission argues for substantial equivalence to a predicate device, NextDent Denture/E-Denture.

However, the provided document does not contain information about a study that describes acceptance criteria, device performance against those criteria, or details regarding expert involvement, ground truth establishment, or sample sizes related to AI/ML device testing.

The document discusses "Bench Testing" in the context of material properties (Flexural Strength, Flexural Modulus, Water Solubility, and Residual Monomer) and states that "All test were passed" in compliance with ISO 20795-1 Type 4 Material. It also mentions biocompatibility. This is standard for material-based devices, not for AI/ML performance.

Therefore, I cannot provide the specific information requested about acceptance criteria and a study proving an AI/ML device meets them based on the provided text. The device described (Juell 3D Volo Base) is a material, not an AI/ML system.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 12, 2018

Park Dental Research Corporation % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030

Re: K181220

Trade/Device Name: Juell 3D Volo Base Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: July 3, 2018 Received: July 16, 2018

Dear Angela Blackwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mary S. Runner -S3

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181220

Device Name Juell 3D Volo Base

Indications for Use (Describe)

Juell 3D Volo Base is a light-cured resin indicated for the fabrication of denture bases fabricated in dental laboratories, including full and partial removable dentures. The material is an alternative to traditional heat cured and auto polymerization resins. Juell 3D Volo Base is intended exclusively for professional dental work. Fabrication of denture bases with Juell 3D Volo Base requires a computer-aided and manufacturing (CAD/CAM) system that includes the following: scanner, design software, additive printer and post-cure unit.

Juell 3D Volo Base is compatible with the following CAD/CAM systems components:

DesignBrandType
ScannerDental WingsI series
Design SoftwareDental WingsDWOS
PrintingBrandTypeSoftware
PrinterJuell 3DDLPFlash OC
Post-CuringBrandType
Post-Cure UnitJuell 3DCure Cube 360

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K181220

October 3, 2018

Park Dental Research Corporation

2401 N. Commerce St. Suite E Ardmore, OK 73401

800-243-7372

Contact Person: Dr. Ronald A Bulard

Trade Name: Juell 3D Volo Base

Common Name: Denture Base

Product Code: EBI

Regulation No.: 872.3760 Denture Relining, Repairing or Rebasing Resin

Classification: Class II

Predicate 510k: K162572 NextDent Denture/E-Denture from Vertex-Dental BV

Device Description:

Juell 3D Volo Base 3D-printing material is a light-cured resin indicated for the manufacturing of denture bases. The material is used in a 3D printer, which prints the shape determined by a 3D stereolithographic drawing. After printed product is placed in a UV-light curing box for final polymerization.

3D printer is not included with the device.

Indications for Use:

Juell 3D Volo Base is a light-cured resin indicated for the fabrication of denture bases fabricated in dental laboratories, including full and partial removable dentures. The material is an alternative to traditional heat cured and auto polymerization resins. Juell 3D Volo Base is intended exclusively for professional dental work. Fabrication of denture bases with Juell 3D Volo Base requires a computer-aided and manufacturing (CAD/CAM) system that includes the following: scanner, design software, additive printer and post-cure unit.

Juell 3D Volo Base is compatible with the following CAD/CAM systems components:

{4}------------------------------------------------

Design
BrandType
ScannerDental Wingsl series
Design SoftwareDental WingsDWOS
Printing
BrandTypeSoftware
PrinterJuell 3DDLPFlash OC
Post-Curing
BrandType
Post-Cure UnitJuell 3DCure Cube 360

Technological Characteristics:

Both Volo Base and NextDent Denture are mixtures of dimethacrylate resins with photoinitiator and pigments. The way they are used to make dentures is identical, using 3 D printers to form the dentures and UV light boxes for curing.

The results of this comparison demonstrate that the design, technology, materials, and composition of Juell 3D Volo Base are substantially equivalent to the predicate devices.

Mechanism of Action:

The mechanism of action is the same as the predicate devices and supports a determination of substantial equivalence. Dentures are used the same way by the patient no matter how they are formed. All types of denture are checked the same way for fit and the concerns regarding non-cured resin are the same no matter the type of curing.

Bench Testing:

Juell 3D Volo Base was tested for compliance with ISO 20795-1 Type 4 Material by testing the following: Flexural Strength, Flexural Modulus, Water Solubility, and Residual Monomer. All test were passed. These same tests were passed by the predicate device except for water sorption which exceeded the value for Type 4 but passed for Type 2.

Biocompatibility Testing:

The constituents of Juell 3D Volo Base are very similar to the predicate device, NextDent Denture/E-Denture. This fact along with the very low level of residual monomer in the cured product meant no biocompatibility testing was done.

{5}------------------------------------------------

Predicate Device Comparison Table

Juell 3D Volo BaseNextDent Denture/E-Denture
RegulationNumber21 CFR 872.376021 CFR 872.3760
DeviceClassificationName/Device CommonNameDenture Relining, Repairing or RebasingResinDenture Relining, Repairing orRebasing Resin
ProductCodeEBIEBI
ClassificationIIII
Indicationsfor useJuell 3D Volo Base is a light-cured resinindicated for the fabrication of denturebases fabricated in dental laboratories,including full and partial removabledentures. The material is an alternativeto traditional heat cured and autopolymerization resins. Juell 3D VoloBase is intended exclusively forprofessional dental work. Fabrication ofdenture bases with Juell 3D Volo Baserequires a computer-aided andmanufacturing (CAD/CAM) system thatincludes the following: scanner, designsoftware, additive printer and post-cureunit.NextDent Denture/E-Denture is a light-cured resin indicated for thefabrication of denture bases fabricatedin dental laboratories, including fulland partial removable dentures. Thematerial is an alternative to traditionalheat cured and auto polymerizationresins. NextDent Denture/E-Denture isintended exclusively for professionaldental work. Fabrication of denturebases with NextDent Denture/E-Denture requires a computer-aidedand manufacturing (CAD/CAM) systemthat includes the following: scanner,design software, additive printer andpost-cure unit.
DeviceDescriptionJuell 3D Volo Base 3D-printing materialis a light-cured resin indicated for themanufacturing of denture bases. Thematerial is used in a 3D printer, whichprints the shape determined by a 3Dstereolithographic drawing. Afterprinting, the printed product is placed ina UV-light curing box for finalpolymerization.3D printer is not included with thedevice.NextDent Denture/E-Denture 3D-printing material is a light-cured resinindicated for the manufacturing ofdenture bases. The material is used ina 3D printer, which prints the shapedetermined by a 3D stereolithographicdrawing. After printing, the printedproduct is placed in a UV-light curingbox for final polymerization.3D printer is not included with thedevice.
MaterialDimethacrylate resinsDimethacrylate resins
MaterialPropertiesMeets ISO 20795-1 for a Type 4Material.Meets ISO 20795-1 for a Type 4material except for water solubilitywhich meets the Type 2 criteria.

{6}------------------------------------------------

Conclusion:

Juell 3D Volo Base has the same Indications for Use and Device Description as the identified predicate device. Volo Base is as safe, as effective, and is substantially equivalent to the predicate device in regards to design, technological characteristics, mechanism of action, materials, performance test and biocompatibility.

§ 872.3760 Denture relining, repairing, or rebasing resin.

(a)
Identification. A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.(b)
Classification. Class II.