Synermed ISE Reagents
K181201 · Infrared Laboratory Systems, LLC (Dba Synermed) · JGS · Jul 10, 2018 · Clinical Chemistry
Device Facts
| Record ID | K181201 |
| Device Name | Synermed ISE Reagents |
| Applicant | Infrared Laboratory Systems, LLC (Dba Synermed) |
| Product Code | JGS · Clinical Chemistry |
| Decision Date | Jul 10, 2018 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1665 |
| Device Class | Class 2 |
Intended Use
The Synermed ISE reagents are intended for the in vitro quantitative measurement of sodium, potassium, and chloride in serum. This device is for use in clinical laboratories only. Sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus and other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Device Story
Synermed ISE Reagents are chemical solutions used for in-vitro diagnostic quantification of sodium, potassium, and chloride in serum samples; intended for clinical laboratory use. Reagents include sodium chloride, potassium chloride, sodium bicarbonate, potassium phosphate, and specific buffers/preservatives. Device functions via Ion Selective Electrode (ISE) technology on the Synermed IR-1200 analyzer. Laboratory technicians operate the system; reagents are loaded into the analyzer to process serum samples. Output consists of numerical electrolyte concentration values (mEq/L or mmol/L). Healthcare providers use these results to diagnose and monitor electrolyte disorders, metabolic conditions, and specific diseases like cystic fibrosis or diabetic acidosis. The device provides essential diagnostic data to guide clinical decision-making regarding patient electrolyte balance.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated per CLSI EP05-A3 and NCCLS EP15-A2; all results met Westgard requirements. Linearity assessed per NCCLS EP6-A; R2 values > 0.999 for all analytes. Interference testing per CLSI EP07-A2 showed no significant bias (>10%) from hemoglobin, bilirubin, or triglycerides. Method comparison (IR-1200 vs. Hitachi 717) per CLSI EP09-A3 showed high correlation (r > 0.99) for all analytes.
Technological Characteristics
Ion Selective Electrode (ISE) technology. Reagents: Sodium chloride, potassium chloride, sodium bicarbonate, potassium phosphate, triethanolamine, phosphoric acid, and non-reactive preservatives. Connectivity: Designed for use with Synermed IR-1200 analyzer. Measuring ranges: Sodium 80-180 mmol/L, Potassium 1.5-10 mmol/L, Chloride 60-140 mmol/L.
Indications for Use
Indicated for in-vitro diagnostic quantitative measurement of sodium, potassium, and chloride in serum for clinical laboratory use. Used in diagnosis/treatment of electrolyte imbalances, aldosteronism, diabetes insipidus, cystic fibrosis, and diabetic acidosis.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Predicate Devices
- Synermed ISE Reagents (K952179)
Related Devices
- K991870 — ELECTROLYTES (NA+, K+, CL-), PRODUCT NO'S 114-01, 114-02, 114-03 · A.P. Total Care, Inc. · Jul 30, 1999
- K051646 — ISE DILUENT, 1N KCL, INTERNAL REFERENCE AND ISE STANDARDS LOW AND HIGH · Pointe Scientific, Inc., · Oct 7, 2005
- K070531 — ISE MODULE AND BAR CODE READER ADDITION FOR THE DATAPRO CLINICAL CHEMISTRY ANALYZER · Thermo Fisher Scientific · Sep 4, 2007
- K052724 — TECO ISE REAGENT SET FOR CX SYSTEM · Teco Diagnostics · Dec 14, 2005
- K040971 — GENCHEM ISE ELECTROLYTE REFERENCE · Genchem, Inc. · Dec 22, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 10, 2018
Infrared Laboratory Systems, LLC (dba Synermed) Julie Paschal Regulatory Affairs Specialist 17408 Tiller Court Suite 1900 Westfield. IN 40674
Re: K181201
Trade/Device Name: Synermed ISE Reagents Regulation Number: 21 CFR 862.1665 Regulation Name: Sodium test system Regulatory Class: Class II Product Code: JGS, CGZ, CEM Dated: February 27, 2018 Received: May 11, 2018
Dear Julie Paschal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181201
Device Name Synermed ISE Reagents
#### Indications for Use (Describe)
The Synermed ISE Reagents are used for in-vitro diagnostic use to quantitate levels of Sodium, Potassium and Chloride in serum. This device is for use in clinical laboratories only. Sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus and other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K181201
## 1. Company Information
Infrared Laboratory Systems, LLC 17408 Tiller Court Suite 1900 Westfield, Indiana 46074 Telephone; (317) 896-1565 FAX: (317) 896-1566
# 2. Contact Information
Julie Paschal Regulatory Affairs Specialist Telephone: 336-235-3057 Email: jpaschal@slplabs.com
# 3. Date Prepared: July 2, 2018
# 4. Device Trade Name: Synermed ISE Reagents
#### 5. Reagent Device Classification:
| Pro<br>Code | Classification<br>Regulation | Classification Name | Device Class | Panel |
|-------------|------------------------------|-----------------------|--------------|-------|
| JGS | 862.1665 | Sodium test system | Class II | 75 |
| CEM | 862.1600 | Potassium test system | Class II | 75 |
| CGZ | 862.1170 | Chloride test system | Class II | 75 |
### 6. Identification of Predicates:
| K-number | Manufacturer | Trade Name | Method |
|----------|--------------|-----------------------------|----------------------------|
| K952179 | Synermed | Synermed<br>ISE<br>Reagents | Ion Selective<br>Electrode |
### 7. Device Description
The Synermed ISE reagents contain the following ingredients: Sodium Chloride, Potassium Chloride, Sodium Bicarbonate, Potassium Phosphate. The ISE buffer contains trlethanolamine preservatives 0.1M, phosphoric acid 0.3% and nonreactive preservatives. The ISE Mid-Standard contains sodium chloride 2.96mM, potassium chloride 0.12M, buffer and non-reactive preservatives. The ISE Reference contains 1M potassium chloride.
{4}------------------------------------------------
# 8. Substantial Equivalence
| Items | Candidate Device: | Predicate Device: |
|--------------|---------------------------------------------------------------------------|----------------------------------|
| | Synermed ISE Reagents<br>K181201 | Synermed ISE Reagents<br>K952179 |
| Similarities | | |
| Intended Use | Quantitative Measurement<br>of Sodium, Potassium and<br>Chloride in serum | Same |
| Environment | Clinical Laboratory Use<br>Only | Same |
| Specimen | Serum | Same |
| Differences | | |
| For use with | Synermed IR-1200 | Hitachi 717 |
# 9. Intended Use
### ISE Reagents
The Synermed ISE reagents are intended for the in vitro quantitative measurement of sodium, potassium, and chloride in serum. This device is for use in clinical laboratories only. Sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus and other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
### 10. Summary of Performance Testing
### a. Precision/Reproducibility
Testing for Precision was accomplished by following the procedures and protocols outlined in the CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition, and in NCCLS EP15-A2 User Verification of Performance for Precision and Trueness; Approved Guideline - Second Edition
Results: Data Examination - NCCLS document EP15-A2; Section 8 pp. 6-10.
Precision Data Analysis Section Figure 1.1 - Mean, Standard Deviation & %CV for all 80 Results for Sodium
{5}------------------------------------------------
| Target<br>Concentration<br>of Sodium in<br>mEq/L | Actual<br>Mean<br>in<br>mEq/L | Westgard<br>%CV Within<br>Run<br>Requirement | Within<br>Run<br>%CV | Within<br>Run<br>SD | %CV<br>Pass<br>(Y/N) | Westgard<br>%CV Run to<br>Run<br>Requirement | Run to<br>Run<br>%CV | Run to<br>Run<br>SD | %CV<br>Pass<br>(Y/N) |
|--------------------------------------------------|-------------------------------|----------------------------------------------|----------------------|---------------------|----------------------|----------------------------------------------|----------------------|---------------------|----------------------|
| 81 | 81.1 | | 0.05 | 0.04 | Y | | 0.04 | 0.03 | Y |
| 115 | 114.87 | | 0.06 | 0.07 | Y | | 0.05 | 0.06 | Y |
| 135 | 134.97 | 0.6 | 0.05 | 0.06 | Y | 0.7 | 0.04 | 0.06 | Y |
| 150 | 150.02 | | 0.06 | 0.09 | Y | | 0.05 | 0.07 | Y |
| 180 | 181.05 | | 0.07 | 0.13 | Y | | 0.07 | 0.12 | Y |
Figure 1.2 - Mean, Standard Deviation & %CV for all 80 Results for Potassium
| Potassium Precision Data Summary Table | | | | | | | | | |
|-----------------------------------------------------|-------------------------------|----------------------------------------------|----------------------|---------------------|----------------------|----------------------------------------------|----------------------|---------------------|----------------------|
| Target<br>Concentration<br>of Potassium<br>in mEq/L | Actual<br>Mean<br>in<br>mEq/L | Westgard<br>%CV Within<br>Run<br>Requirement | Within<br>Run<br>%CV | Within<br>Run<br>SD | %CV<br>Pass<br>(Y/N) | Westgard<br>%CV Run to<br>Run<br>Requirement | Run to<br>Run<br>%CV | Run to<br>Run<br>SD | %CV<br>Pass<br>(Y/N) |
| 1.5 | 1.49 | | 0.52 | 0.01 | Y | | 0.37 | 0.01 | Y |
| 3.0 | 2.98 | | 0.22 | 0.01 | Y | | 0.19 | 0.01 | Y |
| 5.8 | 5.79 | 4.6 | 0.12 | 0.01 | Y | 5.6 | 0.1 | 0.01 | Y |
| 7.5 | 7.49 | | 0.11 | 0.01 | Y | | 0.09 | 0.01 | Y |
| 10.0 | 10.04 | | 0.58 | 0.06 | Y | | 0.46 | 0.05 | Y |
Figure 1.3 - Mean, Standard Deviation & %CV for all 80 Results for Chloride
| Chloride Precision Data Summary Table | | | | | | | | | |
|----------------------------------------------------|-------------------------------|----------------------------------------------|----------------------|---------------------|----------------------|----------------------------------------------|----------------------|---------------------|----------------------|
| Target<br>Concentration<br>of Chloride in<br>mEq/L | Actual<br>Mean<br>in<br>mEq/L | Westgard<br>%CV Within<br>Run<br>Requirement | Within<br>Run<br>%CV | Within<br>Run<br>SD | %CV<br>Pass<br>(Y/N) | Westgard<br>%CV Run to<br>Run<br>Requirement | Run to<br>Run<br>%CV | Run to<br>Run<br>SD | %CV<br>Pass<br>(Y/N) |
| 60 | 60.22 | | 0.52 | 0.31 | Y | | 0.48 | 0.29 | Y |
| 90 | 90.08 | | 0.1 | 0.09 | Y | | 0.07 | 0.07 | Y |
| 101 | 100.77 | 1.2 | 0.48 | 0.49 | Y | 1.5 | 0.43 | 0.43 | Y |
| 112 | 112 | | 0.08 | 0.09 | Y | | 0.06 | 0.06 | Y |
| 130 | 129.98 | | 0.06 | 0.08 | Y | | 0.05 | 0.48 | Y |
# Acceptance Criteria
- The Within-Run Precision and Within-Laboratory Precision results should be less than ● the noted requirements from Westgard for percent imprecision.
{6}------------------------------------------------
# b. Linearity/Reportable Range
### Linearity
Testing for Linearity and Random Error was accomplished by following the procedures and protocols outlined in the NCCLS document EP6-A, Vol. 23, No. 16; Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach: Approved Guideline
QC samples will be performed before daily testing AND with each run. QC results must be within the established range before testing can begin.
| Measurand (in<br>mmol/L) | Slope | Intercept | R2 | Sample Range<br>Tested (in<br>mmol/L) | Claimed Measuring<br>Range (in mmol/L) |
|--------------------------|--------|-----------|--------|---------------------------------------|----------------------------------------|
| Sodium | 0.9987 | 1.1021 | 0.9995 | 80-180 | 80-180 |
| Potassium | 1.0048 | 0.0237 | 0.9999 | 1.5-10 | 1.5-10 |
| Chloride | 0.9814 | 1.3911 | 0.9998 | 60-140 | 60-140 |
### Linearity Results:
#### c. Analytical Specificity
### Interfering Substances
Testing for Interference was accomplished by following the procedures and protocols outlined in the CLSI document EP07-A2; Interference Testing in Clinical Chemistry; Approved Guideline; Second Edition.
The following concentrations of endogenous substances were shown not to interfere with Sodium and Chloride. Hemoglobin- 500mg/dL Bilirubin (conjugated and unconjugated)- 342umol/L Triglycerides- 37mmol/L
The following concentrations of endogenous substances were shown not to interfere with Potassium. Bilirubin (conjugated and unconjugated)- 342umol/L Triglycerides- 37mmol/L
### Acceptance Criteria - Analyte Target Concentration ≤ ± 10% bias.
No bias was seen to be greater than 10%, thus all potential endogenous substance interference is considered to be within acceptable range. Hemolysis is a known contributing factor to abnormally high potassium levels, therefore hemolyzed samples should not be used. Other exogenous substances studies will be referenced to studies done by Young, et al.
{7}------------------------------------------------
#### d. Comparison Studies
Testing for Comparisons and Bias was accomplished by following the procedures and protocols outlined in the CLSI document EP09-A3; Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Third Edition.
#### Data Examination
- a. Visual examination of the data is performed to look for outliers.
- b. A correlation coefficient was calculated for each analyte.
- c. Slope and Intercept were also calculated for each analyte.
# Table 2.1 - IR-1200 and Hitachi 717 Comparison Study Data Summary
| Sodium | |
|-------------------------|----------|
| Correlation Coefficient | 0.992246 |
| Slope | 0.990656 |
| Intercept | 0.943789 |
| Potassium | |
| Correlation Coefficient | 0.993021 |
| Slope | 0.985861 |
| Intercept | -0.10291 |
| Chloride | |
| Correlation Coefficient | 0.993346 |
| Slope | 0.998305 |
| Intercept | 0.122558 |
### 11. Conclusions
The new device is substantially equivalent to the predicate because it has the same intended use and has the same or similar technological characteristics that do not raise new types of questions of safety and effectiveness.