K Number
K181173

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-03-22

(324 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ELMED Laparoscopic Instruments are intended for use in a variety of laparoscopic/endoscopic, therapeutic and minimally invasive general surgical procedures for grasping, manipulating, retracting, cutting, suturing, occluding or clamping soft tissue.

The Laparoscopic Electrodes are electrosurgical devices intended to be used for a variety of general laparoscopic procedures through a port, for coagulation and transection of soft tissue by use of high frequency electric current.

Manually operated surgical instruments are available with or without monopolar high frequency connection and are designed to be used through an opening maintained by a cannula which allows inserting and removal of the instrument(s) without damaging the tissue(s).

Bipolar Forceps are designed to grasp, manipulate and coagulate selected soft tissue and are intended for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator.

Suction/ Irrigation Cannulas have an application(s) in laparoscopic procedures such as gynecologic, general, thoracic and urology and are intended to provide suction and irrigation to help flush blood and tissue debris from the operative site during laparoscopy to aid visualization.

Trocars and Trocar Sleeves are intended to create portals into an operative site for the direct visualization and/or inserting of other surgical instruments.

Needle holders are intended to grasp and manipulate needles inside the peritoneal cavity enabling a surgeon to suture internal soft tissue during a laparoscopic procedure.

Instruments should be used only by personnel completely familiar with their operation.

Device Description

These devices represent a family of Laparoscopic/Endoscopic Instruments and Accessories that consist of:

  • . Laparoscopic Monopolar electrodes
  • . Laparoscopic Bipolar forceps
  • Laparoscopic Trocars, Trocar sleeves
  • . Laparoscopic Needle Holders
  • . Laparoscopic Suction/ Irrigation Cannulas
  • . Laparoscopic Suture and Ligature Instruments
  • . Laparoscopic various manually operated surgical instruments with and without monopolar high frequency connection ( scissors, graspers/dissectors, biopsy forceps)
  • . Laparoscopic Accessories such as: Cleaning Brushes, Sealing Caps, Adapters

Various configurations, sizes and handles of the jaws (graspers, forceps, scissors, dissectors, biopsy, and punches) exist to meet the needs of surgical procedures in conjunction with a laparoscope/endoscope. The instruments have the capability for monopolar RF current, if needed. These instruments are composed of biocompatible materials (surgical grade stainless steel, PTFE, PEEK, several coatings, silicon and brass chromium plated). These materials have a long history of biocompatibility for human use at short term contact. The instruments are reusable and sold non-sterile.

AI/ML Overview

This document is an FDA 510(k) clearance letter and summary for ELMED Laparoscopic Instruments and Accessories. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of the device's diagnostic performance against specific acceptance criteria. Therefore, most of the requested information about acceptance criteria and study design for AI/diagnostic algorithms is not present in this document.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria in terms of performance metrics (like sensitivity, specificity, or accuracy) because it is a clearance for surgical instruments, not a diagnostic AI algorithm. The clearance is based on substantial equivalence for safety and effectiveness, demonstrated through non-clinical testing.

The document states:

  • "The results also demonstrate that ELMED Laparoscopic Instruments and Accessories are as safe and effective as the predicate device."
  • "ELMED Laparoscopic Instruments and Accessories are as safe and effective as the predicate device."

The types of non-clinical testing performed are listed, which act as the "acceptance criteria" for these surgical instruments:

Acceptance Criteria CategoryReported Device Performance
Mechanical TestingConforms to "Mechanical Stress Testing" requirements
Conforms to "Mechanical Push/Pull Testing" requirements
Conforms to "Drop Testing" requirements
Electrical Safety TestingConforms to "Dielectric Strength Testing"
Conforms to "Capacitive Coupling Testing"
Conforms to "Leakage Current Testing"
Electromagnetic TestingConforms to "Electromagnetic Interference/Immunity Testing"
BiocompatibilityInstruments composed of "biocompatible materials (surgical grade stainless steel, PTFE, PEEK, several coatings, silicon and brass chromium plated)" with "a long history of biocompatibility for human use at short term contact."
International StandardsConforms to IEC 60601-2-2

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. This document describes the clearance of surgical instruments, not a device relying on a test set of data for performance evaluation. The "testing" refers to physical and electrical evaluations of the instruments themselves.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. Ground truth assessment by experts in the context of data for diagnostic performance is not relevant to the clearance of these surgical instruments.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For these surgical instruments, the "ground truth" for demonstrating safety and effectiveness relies on established engineering and medical device standards (e.g., IEC 60601-2-2) and the inherent properties of the materials and design, demonstrating that the device performs its intended mechanical and electrical functions safely. There isn't a "ground truth" in the diagnostic data sense.

8. The sample size for the training set

Not applicable. There is no training set as this is not a machine learning device.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 22, 2019

ELMED Incorporated Werner Hausner President 35 N Brandon Drive Glendale Heights, Illinois 60139

Re: K181173

Trade/Device Name: ELMED Laparoscopic Instruments and Accessories Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, GEI Dated: February 20, 2019 Received: February 21, 2019

Dear Werner Hausner:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows a digital signature from Long H. Chen -S. The signature indicates that the document was digitally signed on March 22, 2019. The timestamp of the signature is 10:05:03 -04'00'. The signature is for something.

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181173

Device Name ELMED LAPAROSCOPIC INSTRUMENTS AND ACCESSORIES

Indications for Use (Describe)

The ELMED Laparoscopic Instruments are intended for use in a variety of laparoscopic/endoscopic, therapeutic and minimally invasive general surgical procedures for grasping, manipulating, retracting, cutting, suturing, occluding or clamping soft tissue.

The Laparoscopic Electrodes are electrosurgical devices intended to be used for a variety of general laparoscopic procedures through a port, for coagulation and transection of soft tissue by use of high frequency electric current.

Manually operated surgical instruments are available with or without monopolar high frequency connection and are designed to be used through an opening maintained by a cannula which allows inserting and removal of the instrument(s) without damaging the tissue(s).

Bipolar Forceps are designed to grasp, manipulate and coagulate selected soft tissue and are intended for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator.

Suction/ Irrigation Cannulas have an application(s) in laparoscopic procedures such as gynecologic, general, thoracic and urology and are intended to provide suction and irrigation to help flush blood and tissue debris from the operative site during laparoscopy to aid visualization.

Trocars and Trocar Sleeves are intended to create portals into an operative site for the direct visualization and/or inserting of other surgical instruments.

Needle holders are intended to grasp and manipulate needles inside the peritoneal cavity enabling a surgeon to suture internal soft tissue during a laparoscopic procedure.

Instruments should be used only by personnel completely familiar with their operation.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(k) Summary

These 510(k) summary is being submitted in accordance with 21 CFR §807.92.

Type of submission: Traditional

The assigned 510 (k) number is: K181173

A. Company Contact

ELMEDINCORPORATED 35 N Brandon Drive Glendale Heights, IL 60139

  • B. Submitter Information: Werner Hausner
    President ELMEDINCORPORATED werner@elmed.com

C. Identification of theDevice:

Proprietary Name:ELMED Laparoscopic Instruments and Accessories
Common Name:Laparoscopic Instrument and Accessory
Classification Name:Laparoscope, General and Plastic surgery
Product Code:GCJ
Secondary Product Code:GEI
Regulatory Class:Class II 21 CFR 876.1500
Panel:General and Plastic Surgery

D. Identification of the Predicate Device

Legally Marketed Equivalent Devices

ManufacturerBrand Name510(k)Number
JAKOUBEKMEDIZINTECHNIK GMBHTROCAR & TROCAR SLEEVES, LAPAROSCOPES,LAPAROSCOPIC FORCEPS, & LAPAROSCOPICINSTRUMENTSK990785
HANS HERMANN GMBHHANS HERMANN LAPAROSCOPES AND ACCESSORIESK051610

Table 1: Predicative Device Information

{4}------------------------------------------------

E. Device Description:

These devices represent a family of Laparoscopic/Endoscopic Instruments and Accessories that consist of:

  • . Laparoscopic Monopolar electrodes
  • . Laparoscopic Bipolar forceps
  • Laparoscopic Trocars, Trocar sleeves
  • . Laparoscopic Needle Holders
  • . Laparoscopic Suction/ Irrigation Cannulas
  • . Laparoscopic Suture and Ligature Instruments
  • . Laparoscopic various manually operated surgical instruments with and without monopolar high frequency connection ( scissors, graspers/dissectors, biopsy forceps)
  • . Laparoscopic Accessories such as: Cleaning Brushes, Sealing Caps, Adapters

Various configurations, sizes and handles of the jaws (graspers, forceps, scissors, dissectors, biopsy, and punches) exist to meet the needs of surgical procedures in conjunction with a laparoscope/endoscope. The instruments have the capability for monopolar RF current, if needed. These instruments are composed of biocompatible materials (surgical grade stainless steel, PTFE, PEEK, several coatings, silicon and brass chromium plated). These materials have a long history of biocompatibility for human use at short term contact. The instruments are reusable and sold non-sterile.

F. Intended Use and Indication for Use of the subject device(s):

The ELMED Laparoscopic Instruments are intended for use in a variety of laparoscopic/endoscopic, therapeutic and minimally invasive general surgical procedures for grasping, dissecting, manipulating, retracting, suturing, occluding or clamping soft tissue.

The Laparoscopic Electrodes are electrosurgical devices intended to be used for a variety of general laparoscopic procedures through a port, for coagulation and transection of soft tissue by use of high frequency electric current. Manually operated surgical instruments are available with or without monopolar

high frequency connection and are designed to be used through an opening maintained by a cannula which allows inserting and removal of the instrument(s) without damaging the tissue(s).

{5}------------------------------------------------

Bipolar Forceps are designed to grasp, manipulate and coagulate selected soft tissue and are intended for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator.

Suction/ Irrigation Cannulas have an application(s) in laparoscopic procedures such as gynecologic, general, thoracic and urology and are intended to provide suction and irrigation functions to help flush blood and tissue debris from the operative site during laparoscopy to aid visualization.

Trocars and Trocar Sleeves are intended to create portals into an operative site for the direct visualization and/or inserting of other surgical instruments. Needle holders are intended to grasp and manipulate needles inside the peritoneal cavity enabling a surgeon to suture internal soft tissue during a laparoscopic procedure.

Instruments should be used only by personnel completely familiar with their operations.

G. Non-clinical Testing

The devices conform to IEC 60601-2-2 . Testing conducted on the instruments also included but was not limited to:

    1. Mechanical Stress Testing
    1. Mechanical Push/Pull Testing
    1. Drop Testing
    1. Dielectric Strenght Testing
  • Capacitive Coupling Testing 5.
    1. Leakage Current Testing
    1. Electromagnetic Interference/Immunity Testing

The results also demonstrate that ELMED Laparoscopic Instruments and Accessories are as safe and effective as the predicate device.

H. Clinical Testing

There were no prospective clinical studies to support the medical device as the indications for use, technology: materials of construction are virtually identical to the predicate device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

I. Substantial Equivalence/ Conclusion

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device.

The differences between the predicates and the modified design do not raise any new risks of safety or efficacy. The changes are minor, so the subject devices are as safe and effective as the predicate devices, perform similarly to other legally marketed predicate devices for laparoscopic/endoscopic and minimally invasive procedure. ELMED Laparoscopic Instruments and Accessories are as safe and effective as the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.