K Number
K181150

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-05-25

(24 days)

Product Code
Regulation Number
870.4210
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ProtekDuo Mini Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.

Device Description

The ProtekDuo Mini Veno-Venous Cannula Set consists of two components: a 31 Fr. Dual Lumen Veno-Venous Cannula and a 15.5 Fr. Introducer is designed to accept a standard 0.038 inch guidewire. The ProtekDuo Mini Veno-Venous Cannula Set is intended as a single patient, single use, sterile device.

The ProtekDuo Mini Veno-Venous Cannula (Figure 5-1) consists of two distinct lumens with a wire-reinforced cannula body. The inner lumen is located entirely within the outer lumen forming two concentric channels.

The distal section (inner lumen/cannula) of the cannula body has six side holes near the distal tip opening. The proximal sections of each lumen are clear and not wire-reinforced to allow visualization of blood and to enable clamping to prevent blood flow during setup and removal of the cannula from the extracorporeal circulatory support equipment (see Figure 5-1). A non-vented barbed connector is affixed to both proximal ends (inner/distal and outer/proximal lumens) of the cannula and allow for connection of standard 3/8 inch blood circuit tubing for subsequent connection to extracorporeal circulatory support equipment.

The introducer (Figure 5-2) consists of a tube with a luer hub. The introducer fits inside the inner lumen of the cannula during insertion of the cannula/introducer assembly. The introducer is used to advance the cannula over a guidewire and facilitate cannula placement within the target vessel. The introducer has a luer hub at its proximal end to manage introducer insertion and removal from the cannula. The luer hub can also enable contrast injection, if necessary, to facilitate placement and final positioning of the cannula within the target vessel. The hemostasis cap minimizes blood loss when the cannula/introducer assembly is inserted into the target vessel. The introducer body is constructed from radiopaque material for visualization under fluoroscopy.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the ProtekDuo Mini Veno-Venous Cannula Set meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance Criterion (identical to predicate unless otherwise noted)Reported Device Performance
Comparative HemolysisNo difference in hemolysis levels between test articles and predicate controls.No difference in hemolysis levels between test articles and predicate controls.
Pressure-FlowMeasured flow rates must exceed the longer predicate at all levels of pressure difference across the cannula. Design specifications were met (specifics not detailed).Measured flow rates exceed the longer predicate at all levels of pressure difference across the cannula. Design specifications were met.
Tensile StrengthIdentical to those of the predicate.Acceptance criteria were identical to those of the predicate and were met in all tests.
Pathway IntegrityIdentical to that of the predicate.Acceptance criterion was identical to that of the predicate and was met in all tests.
Kink RadiusIdentical to that of the predicate.Acceptance criterion was identical to that of the predicate and was met in all tests.
StiffnessEstablished to ensure the cannula is stiff enough to insert while remaining flexible enough to avoid vessel injury (specific values not detailed).The cannula, the introducer, and the cannula/introducer assembly met acceptance criteria established to ensure the cannula is stiff enough to insert while remaining flexible enough to avoid vessel injury.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample sizes (number of devices, repeats per test, etc.) for the non-clinical tests conducted. The data provenance is also not explicitly stated in terms of country of origin or retrospective/prospective nature, but these are bench-top non-clinical tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

Not applicable. This is a non-clinical device performance study, not a study involving human interpretation of data where expert ground truth would be required.

4. Adjudication Method for the Test Set:

Not applicable. This is a non-clinical device performance study.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. This is a non-clinical device performance study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This is a physical medical device, not an AI algorithm. The performance described is for the device itself.

7. The Type of Ground Truth Used:

For the non-clinical tests, the "ground truth" or reference was based on established engineering principles, design specifications, and direct measurement of the device's physical properties and performance against predetermined criteria (often derived from the predicate device's performance or industry standards).

8. The Sample Size for the Training Set:

Not applicable. This is a physical medical device, not a machine learning model requiring a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. This is a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

May 25, 2018

CardiacAssist Inc Greg Johnson VP, Regulatory Affairs and Quality Assurance 240 Alpha Drive Pittsburg, Pennsylvania 15238

Re: K181150

Trade/Device Name: ProtekDuo Mini Veno-Venous Cannula Set Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, Or Tubing Regulatory Class: Class II Product Code: DWF Dated: April 30, 2018 Received: May 1, 2018

Dear Greg Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Nicole G. Ibrahim -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.

Indications for Use

510(k) Number (if known)K181150
Device NameProtekDuo Mini Veno-Venous Cannula Set
Indications for Use (Describe)The ProtekDuo Mini Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

Company of the country of the country of the country of the

{3}------------------------------------------------

TANDEM LIFE

Section 5 510(k) Summary 510(k) Special

Date: 30 April 2018

Applicant

CardiacAssist, Inc. dba TandemLife 240 Alpha Drive Pittsburgh, PA 15238 Telephone: 412-963-7770 Fax: 412-963-0800

Contact person

Greg Johnson, PhD Title: VP, Regulatory Affairs and Quality Assurance Phone: 412-963-7770 x266 e-mail: gjohnson@tandemlife.com

Device

Trade/Proprietary Name:ProtekDuo Mini Veno-Venous Cannula Set
Common Name:Veno-Venous Cannula Set
Classification Name:Cardiopulmonary bypass vascular catheter, cannula, or tubing (21 CFR 870.4210 Product Code DWF)

Predicate Device

TandemHeart 31 Fr. (31 Fr. ProtekDuo) Veno-Venous Cannula Set (K160257) (marketed under Trade Name: 31 Fr. ProtekDuo)

Device Description

The ProtekDuo Mini Veno-Venous Cannula Set consists of two components: a 31 Fr. Dual Lumen Veno-Venous Cannula and a 15.5 Fr. Introducer is designed to accept a standard 0.038 inch guidewire. The ProtekDuo Mini Veno-Venous Cannula Set is intended as a single patient, single use, sterile device.

Image /page/3/Figure/13 description: The image shows a technical drawing of a medical device. The device has barbed connectors on one end and insertion depth markers labeled 36, 34, and 32. There are also tantalum markers and a wire-reinforced area labeled with numbers 18, 16, and 14.

Figure 5-1. ProtekDuo Mini Veno-Venous Cannula

{4}------------------------------------------------

TANDEM LIFE

Section 5 510(k) Summary 510(k) Special

The ProtekDuo Mini Veno-Venous Cannula (Figure 5-1) consists of two distinct lumens with a wire-reinforced cannula body. The inner lumen is located entirely within the outer lumen forming two concentric channels.

The distal section (inner lumen/cannula) of the cannula body has six side holes near the distal tip opening. The proximal sections of each lumen are clear and not wire-reinforced to allow visualization of blood and to enable clamping to prevent blood flow during setup and removal of the cannula from the extracorporeal circulatory support equipment (see Figure 5-1). A non-vented barbed connector is affixed to both proximal ends (inner/distal and outer/proximal lumens) of the cannula and allow for connection of standard 3/8 inch blood circuit tubing for subsequent connection to extracorporeal circulatory support equipment.

The introducer (Figure 5-2) consists of a tube with a luer hub. The introducer fits inside the inner lumen of the cannula during insertion of the cannula/introducer assembly. The introducer is used to advance the cannula over a guidewire and facilitate cannula placement within the target vessel. The introducer has a luer hub at its proximal end to manage introducer insertion and removal from the cannula. The luer hub can also enable contrast injection, if necessary, to facilitate placement and final positioning of the cannula within the target vessel. The hemostasis cap minimizes blood loss when the cannula/introducer assembly is inserted into the target vessel. The introducer body is constructed from radiopaque material for visualization under fluoroscopy.

Image /page/4/Figure/5 description: The image shows a medical device, specifically an introducer needle, with labels pointing to its components. The "Introducer Hub" is labeled at the left end of the needle, which is the part that connects to other medical equipment. The "Hemostasis Cap" is labeled near the hub, indicating a component that helps control bleeding during insertion.

Figure 5-2: 15.5 Fr. Introducer

Indications for Use

The ProtekDuo Mini Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.

Comparison of Technological Characteristics

ProtekDuo Mini Veno-Venous Cannula Set is an extension to the ProtekDuo Veno-Venous Cannula Family. It is identical to the TandemHeart 31 Fr. (31 Fr. ProtekDuo) Veno-Venous Cannula Set, except the distal lumen is approximately 14.2 cm shorter in length. The introducer is also approximately 14.2 cm shorter. It is designed for the same

{5}------------------------------------------------

TANDEM LIFE

Section 5 510(k) Summary 510(k) Special

intended use as the TandemHeart 31 Fr. (31 Fr. ProtekDuo) Veno-Venous Cannula Set. All materials and methods of manufacture are identical.

Summary of Non-clinical Testing

Testing of the ProtekDuo Mini Veno-Venous Cannula Set included comparative hemolysis, pressure-flow testing, tensile strength, pathway integrity, kink radius, and stiffness.

TestMethodConclusion
ComparativeHemolysisSide-by-side comparison withpredicate of bench tophemolysis levels over 6 hours.No difference in hemolysis levelsbetween test articles and predicatecontrols.
Pressure-FlowMeasure pressure losses acrosscannula at different flow rates.Measured flow rates exceed thelonger predicate at all levels ofpressure difference across thecannula. Design specificationswere met.
Tensile StrengthPull testing of both cannula andintroducerAcceptance criteria were identicalto those of the predicate and weremet in all tests.
Pathway IntegrityPressure testingAcceptance criterion was identicalto that of the predicate and was metin all tests.
Kink RadiusFlow rate reduction caused byspecified minimum bend radiuswas measuredAcceptance criterion was identicalto that of the predicate and was metin all tests.
StiffnessDeflection testing to measureforce required to bend cannulaa specified distanceThe cannula, the introducer, and thecannula/introducer assembly metacceptance criteria established toensure the cannula is stiff enough toinsert while remaining flexibleenough to avoid vessel injury.

Conclusion

The ProtekDuo Mini Veno-Venous Cannula Set is made from the same materials, using the same manufacturing processes and sterilization techniques as the predicate device. Testing demonstrated that hemolysis, tensile strength, pathway integrity, and kink radius are identical to the predicate. Stiffness and pressure-flow properties differed as expected for the shorter length cannula and met established acceptance criteria. The ProtekDuo Mini Veno-Venous Cannula Set is determined to be substantially equivalent to the predicate TandemHeart 31 Fr. (31 Fr. ProtekDuo) Veno-Venous Cannula Set.

{End of Section}

§ 870.4210 Cardiopulmonary bypass vascular catheter, cannula, or tubing.

(a)
Identification. A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.(b)
Classification. Class II (performance standards).