K Number
K181087

Validate with FDA (Live)

Date Cleared
2018-12-18

(237 days)

Product Code
Regulation Number
N/A
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Apex Locator, DPEX III is a microprocessor controlled device intended for locating the apex of root canal.

Device Description

Apex Locator, DPEX III is an apex locator intended for precise localization of root canal apex. Operation of Apex Locator, DPEX III is fully automatic, no manual calibrations or adjustments are required. The measuring signal is analyzed, and automatic adjustments are made if needed. The device may operate within different conditions in the root canal. Apex Locator, DPEX III may only be used with stainless steel or nickel titanium endodontic files. The scale indication on the apex locator screen does not represent a distinct length or distance in mm or other linear units. It merely indicates the file progression towards the apex.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Apex Locator, DPEX III. It aims to demonstrate substantial equivalence to a predicate device, the RAYPEX 6.

Here's an analysis of the acceptance criteria and study information based solely on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that the acceptance criteria for performance were met by demonstrating equivalence to the predicate device through performance comparison testing.

Acceptance Criteria CategorySpecific Criteria (Implicitly based on stated equivalence)Reported Device Performance (Apex Locator, DPEX III)
Measurement PerformanceClinically acceptable measurement of apex location, equivalent to predicate device (RAYPEX 6)."The result obtained with the Apex Locator, DPEX III is clinically acceptable since they are equivalent to the results of the predicate device."
Electrical Safety (EMC)Compliance with IEC 60601-1:2005+CORR.1 (2006)+CORR.2(2007)+AM1(2012) and IEC 60601-1-2: 2014."The safety and EMC testing confirm that the subject device complies with the same FDA recognized safety and EMC standards as the primary predicate device. All of the predetermined acceptance criteria were met."
BiocompatibilityCompliance with ISO10993-5 (Cytotoxicity), ISO10993-10 (Skin Irritation), ISO10993-10 (Oral Mucosa Irritation), ISO10993-10 (Skin Sensitization)."The test result shows that the subject device complies with the same FDA recognized biocompatibility standards as predict device. All of the pre-determined acceptance criteria were met."
Sterilization ValidationCompliance with ISO 17665-1:2006, ISO 11737-2:2009, and ANSI AAMI ST79:2017 for accessories."The steam sterilization validation study supports the labeling of sterilization method for the accessories subject to be sterilized at the clinical site. All of the pre-determined acceptance criteria were met."

Note: The document does not provide quantitative specific acceptance criteria (e.g., "accuracy must be within X mm") or numerical performance data for the device. Instead, it relies on demonstrating equivalence to a legally marketed predicate device.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document mentions "using extracted tooth" for the performance comparison test, but it does not specify the sample size (i.e., number of extracted teeth) used for this test set.
  • Data Provenance: The data provenance is retrospective, as it involves testing performed by the manufacturer to demonstrate equivalence. The country of origin of the data is not explicitly stated, but the manufacturer is "Guilin Woodpecker Medical Instrument Co., Ltd." in P.R. China.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not provided in the document. The performance comparison test relies on a comparison against the predicate device, and the method for establishing "ground truth" for the apex location is not described beyond "using extracted tooth." There is no mention of human experts defining ground truth for the device's accuracy.

4. Adjudication Method for the Test Set

No adjudication method is described. Given the type of device (apex locator) and the testing described (comparison using extracted teeth), a human adjudication process by multiple experts is not indicated as part of the performance testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission." This indicates that there was no human-in-the-loop study comparing human readers with and without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The performance comparison test "using extracted tooth" functions as a standalone performance test for the device's ability to locate the apex. It evaluates the device's accuracy against the predicate without direct human interaction influencing the measurement outcome during the test itself. However, it's not an "algorithm" in the sense of a software AI model; it's a physical device measurement.

7. The Type of Ground Truth Used

The "ground truth" for the performance comparison test appears to be established by the measurement of the predicate device (RAYPEX 6). The aim was to show the new device's measurements were equivalent. For an apex locator, the true anatomical apex location in an extracted tooth would be the actual ground truth, but the document frames the comparison against the predicate as the basis for performance acceptance. It's essentially "predicate device performance" as the de facto standard for this submission.

8. The Sample Size for the Training Set

Not applicable/Not provided. This device is an electrical impedance-based apex locator, not an AI/ML device that requires a training set in the conventional sense. Its "knowledge" is embedded in its electrical design and algorithms, which are determined during engineering and design, not through a data-driven training process.

9. How the Ground Truth for the Training Set Was Established

Not applicable/Not provided. As stated above, this is not an AI/ML device that uses a "training set" with established ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Guilin Woodpecker Medical Instrument Co., Ltd. % Charlie Mack Principal Engineer Irc 2950 E Lindrick Drive Chandler, Arizona 85249

Re: K181087

Trade/Device Name: Apex Locator, DPEX III Regulatory Class: Unclassified Product Code: LQY Dated: September 8, 2018 Received: September 19, 2018

Dear Charlie Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

December 18, 2018

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mary S. Mary S. Runner -S3 Runner -S3 Date: 2018.12.18

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181087

Device Name Apex Locator, DPEX III

Indications for Use (Describe)

Apex Locator, DPEX III is a microprocessor controlled device intended for locating the apex of root canal.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a red rectangle with rounded corners. Inside the rectangle is the logo for Woodpecker. The logo consists of a white woodpecker graphic on the left and the word "WOODPECKER" in white text on the right. There is a registered trademark symbol below the woodpecker.

Guilin Woodpecker Medical Instrument Co., Ltd.

510(k) Summary (21 CFR §807.92) 510(k) assigned number K181087

Submitter Information:

Submitter Name:Guilin Woodpecker Medical Instrument Co., Ltd.
Address:Information Industrial Park, Guilin National High-Tech Zone,
Guilin City, Guangxi Province, P.R. China 541004
Contact Person:Mr. Xunxian Wu
President
charliemack@irc-us.com
931-625-4938
Date of Preparation:December 15, 2018
Trade Name:Apex Locator, DPEX III
Common Name:Root apex locator
Regulation Number:Pre-Amendment
Regulation Name:locator, root apex
Regulatory Class:Unclassified
Product Code:LQY
Classification Panel:Dental

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for Woodpecker. The logo is white text on a red background. To the left of the text is a white outline of a woodpecker.

Primary Predicate Device

Trade Name:RAYPEX 6
510(k) Reference:K131907
Common Name:Root apex locator
Regulation Number:Pre-Amendment
Regulation Name:locator, root apex
Regulatory Class:Unclassified
Product Code:LQY
Classification Panel:Dental

Purpose of Submission

This is a new submission of an apex root locator.

Device Description

Apex Locator, DPEX III is an apex locator intended for precise localization of root canal apex.

Operation of Apex Locator, DPEX III is fully automatic, no manual calibrations or adjustments are required. The measuring signal is analyzed, and automatic adjustments are made if needed. The device may operate within different conditions in the root canal. Apex Locator, DPEX III may only be used with stainless steel or nickel titanium endodontic files. The scale indication on the apex locator screen does not represent a distinct length or distance in mm or other linear units. It merely indicates the file progression towards the apex.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for Woodpecker. The logo features a white woodpecker on the left side of the image. To the right of the woodpecker is the word "WOODPECKER" in white, bold letters. The background of the logo is red.

Indications for Use

Apex Locator, DPEX III is a microprocessor- controlled device used for locating the apex of root canal.

SUBJECTIndications for UsePRIMARYPREDICATE Device(K131907)Indications for Use
Apex Locator, DPEX III is a microprocessor-controlled device used for locating the apex ofroot canal.Rapex 6 is a microprocessor-controlleddevice used for locating the apex.

All the labeling and characteristics of the submitted Apex Locator, DPEX III is the same as the primary predicate device and most typical apex locators currently on the market. The submitted device and predicate both are used in routine procedures.

The basic design of the submitted device specifications is substantially the same as the predicate, with differences in the display, weight, and frequencies used for measurements. Testing validates that the performance of the Apex Locator, DPEX III is comparable to the predicate device.

Sterilization validation and performance testing has verified that the submitted device is safe for use.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for Woodpecker. The logo features a white woodpecker icon on the left and the word "WOODPECKER" in white on the right. The background is a red rectangle with rounded corners.

Technological Characteristics

Technological characteristics of the subject and predicate devices are equivalent. The subject device Apex Locator, DPEX III achieves its intended use based on the same technology and principles of operation as the predicate device, Rapex 6 apex locator.

CharacteristicsSubject DevicePrimary Predicate Device
Device Trade NameApex Locator, DPEX IIIApex Locator, RAYPEX 6
510K ApplicantGUILIN WOODPECKER MEDICALINSTRUMENT CO., LTDDENTSPLY International
510(K) NumberK181087K131907
Regulation NumberUnclassifiedUnclassified
Product CodeLQYLQY
Classification NameRoot Apex LocatorRoot Apex Locator
OTC or PrescriptionPrescription UsePrescription Use
Medical SpecialtyDentalDental
Indications for UseDPEX III is a microprocessor-controlleddevice used for locating the apex of rootcanal.RAYPEX 6 is a microprocessor-controlled device used for locatingthe apex.
Electrical PowerRechargeable lithium battery(3.7V)Rechargeable NiMHbatteries(2.4V)
AdapterInput: AC100-240V 50/60HzOutput: DC5V 1AInput: AC100-240V 50/60HzOutput: DC5V 1A
DisplayCustom colorized 4.5" LCD3.5" colorized TFT
Measuring VoltageNominal - doesn't exceed 200mV ACNominal - doesn't exceed 200mVAC
Frequencies usedfor measurements400Hz & 8kHz500Hz & 8kHz
Method ofcalculationThe ratio of impedance at twofrequenciesThe ratio of impedance at twofrequencies

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for Woodpecker. The logo features a stylized woodpecker bird in white, with the word "WOODPECKER" in white text to the right of the bird. The logo is set against a red background. There is a registered trademark symbol below the bird.

CharacteristicsSubject DevicePrimary Predicate Device
AccessoryMeasuring wire, File clip, Lip hook, Touch probe, Adapter, TesterMeasuring wire, File clip, Lip hook, Touch probe, Adapter, Tester
Weight385g350g
Operating EnvironmentHospital environment, clinics or dental offices by qualified dental personnel.Hospital environment, clinics or dental offices by qualified dental personnel
ButtonThree push buttons:1. ON/OFF2. Sound adjustment3. Demonstration functionOne push button:ON/OFF-Touchscreen
Sound IndicationPiezo transducer with sound level control (high, medium, low, mute)A speaker which enables:-the sound level adjustment from mute to the high sound level-tone selection
Automatic Turn off functionThe device turns off automatically after 5 minutes of the idle state.The device turns off automatically after 5 minutes of an idle state.
CalibrationNot requiredNot required
BiocompatibilityComplies with ISO10993-1Complies with ISO10993-1

The submitted Apex Locator DPEX III uses the same technology to determine the apex of the root, using two frequencies that are alternated, with the ratios of the impedance measurement to determine the apex of the root canal. Both the subject device and the predicate Rapex 6 use a sound alert when the file is near the apex. The Apex Locator DPEX III uses a different high frequency, but testing reveals that there is no performance difference between the predicate and submitted device.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for Woodpecker. The logo features a white woodpecker on the left side of the image. To the right of the woodpecker is the word "WOODPECKER" in white, all on a red background.

Summary of Performance Tests

The following performance testing was performed to demonstrate substantial equivalence of the subject device Apex Locator DPEX III and applicable standards:

Non-clinical Test

Performance Comparison Test:

The comparison testing on Apex Locator. DPEX III and the predicate device (RAYPEX 6 (K131907) was performed to compare the measurement performance of locating apex root using extracted tooth. The result obtained with the Apex Locator, DPEX III is clinically acceptable since they are equivalent to the results of the predicate device.

Safety and EMC Test:

    1. IEC 60601-1: 2005+CORR.1 (2006)+CORR.2(2007)+AM1(2012) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
    1. IEC 60601-1-2: 2014 Medical devices part 1-2: General requirements for basic safety and essential performance – Collateral standards: electromagnetic compatibility – Test and requirements

The safety and EMC testing confirm that the subject device complies with the same FDA recognized safety and EMC standards as the primary predicate device. All of the predetermined acceptance criteria were met.

Biocompatibility Test:

    1. Cytotoxicity Tests (ISO10993-5)
    1. Skin Irritation Test (ISO10993-10)
    1. Oral Mucosa Irritation Test (ISO10993-10)
    1. Skin Sensitization Test (ISO10993-10)

The test result shows that the subject device complies with the same FDA recognized biocompatibility standards as predict device. All of the pre-determined acceptance criteria were met.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a logo for Woodpecker. The logo is red with white text and a white woodpecker. The woodpecker is on the left side of the logo, and the text "WOODPECKER" is on the right side of the logo.

Sterilization Validation:

    1. ISO 17665-1:2006, Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices (Sterility)
    1. ISO 11737-2:2009, Sterilization of medical devices -- Microbiological methods --Part 2: Tests of sterility performed in the validation of a sterilization process. (Sterility)
    1. ANSI AAMI ST79:2017, Comprehensive guide to steam sterilization and sterility assurance in health care facilities

The steam sterilization validation study supports the labeling of sterilization method for the accessories subject to be sterilized at the clinical site.

All of the pre-determined acceptance criteria were met.

Clinical Test

No clinical study is included in this submission.

Summary of Substantial Equivalence

Based on the indications for use, technological characteristics, and performance testing, the subject Apex Locator, DPEX III has demonstrated to be substantially equivalent to the primary predicate Rapex 6 Apex Locator.

N/A