K Number
K181002

Validate with FDA (Live)

Date Cleared
2018-07-16

(91 days)

Product Code
Regulation Number
866.3215
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atellica® IM BRAHMS Procalcitonin (PCT) assay is for in vitro diagnostic use in the quantitative determination of procalcitonin in human serum and plasma (EDTA, lithium heparin, and sodium heparin) using the Atellica® IM Analyzer.

The Atellica IM BRAHMS PCT assay is intended for use, in conjunction with other laboratory findings and clinical assessments, as an aid in:

· The risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.

• Assessing the cumulative 28-day risk of all-cause mortality for patients diagnosed with severe sepsis or septic shock in the ICU or when obtained in the emergency department or other medical wards prior to ICU admission, using percent change in PCT level over time.

· Decision-making on antibiotic therapy for patients with suspected or confirmed lower respiratory tract infections (LRTI) – defined as community-acquired pneumonia (CAP), acute bronchitis, and acute exacerbation of chronic obstructive pulmonary disease (AECOPD) - in an inpatient setting or an emergency department.

  • · Decision-making on antibiotic discontinuation for patients with suspected or confirmed sepsis.
Device Description

The Atellica IM BRAHMS PCT assay is a 2-site sandwich immunoassay using direct chemiluminescent technology that uses 3 mouse monoclonal antibodies specific for PCT. The first antibody, in the Lite Reagent, is a mouse monoclonal anti-PCT antibody labeled with acridinium ester. The second and third antibodies, in the ancillary reagent, are mouse monoclonal anti-PCT antibodies labeled with fluorescein. The immunocomplex formed with PCT is captured with mouse monoclonal anti-fluorescein antibody coupled to paramagnetic particles in the Solid Phase.

A direct relationship exists between the amount of PCT present in the patient sample and the amount of relative light units (RLUs) detected by the system.

AI/ML Overview

The provided text describes the Siemens Healthcare Diagnostics Atellica IM B.R.A.H.M.S Procalcitonin (PCT) assay, an in vitro diagnostic device. The study presented focuses on the analytical and clinical performance of this assay, comparing it to a predicate device and evaluating its utility in specific clinical scenarios related to sepsis and respiratory tract infections.

Here's an analysis of the acceptance criteria and the study proving the device meets them:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are implicitly derived from the performance goals demonstrated in the analytical and clinical studies. Since this is a 510(k) submission, the primary goal is to show substantial equivalence to a legally marketed predicate device (B.R.A.H.M.S PCT sensitive KRYPTOR). The performance characteristics are presented as meeting industry standards (CLSI guidelines) and showing comparable results to the predicate.

Here's a table summarizing the analytical performance with implicit acceptance criteria (typically, these would be defined before the study in a protocol, often as a deviation tolerance from the predicate or a specific measure):

Performance CharacteristicAcceptance Criteria (Implicit/Standard Expectations)Reported Device Performance (Atellica IM BRAHMS PCT)
PrecisionLow variability (%CV) across different PCT concentrations, meeting CLSI EP05-A3 guidelines.Repeatability (Within-Run): %CVs ranging from 1.2% to 9.7% for serum samples (0.05 - 19.14 ng/mL). Within-Lab (Total Precision): %CVs ranging from 2.1% to 12.1%. Modeling analysis showed Total Error of 4%-32% at various PCT levels.
Linearity/Assay Measuring RangeDeviation from linear fit ≤ 10%; demonstrated linearity across the claimed measuring range.Linearity demonstrated in the range of 0.03 – 63.24 ng/mL. Claimed measuring range is 0.04 – 50.00 ng/mL.
Dilution RecoveryRecoveries close to 100% when samples are diluted manually and automatically.Manual dilution recoveries: 96% to 102% (mean 99%). Auto-dilution recoveries: 92% to 107%.
Hook EffectNo significant hook effect within the expected upper range of analyte concentrations.No hook effect observed up to 2000 ng/mL; samples as high as 2000 ng/mL reported > 50.00 ng/mL.
Detection Limits (LoB, LoD, LoQ)LoB < 0.03 ng/mL, LoD < 0.04 ng/mL, LoQ ≤ 0.06 ng/mL.LoB = 0.00 ng/mL. LoD = 0.03 ng/mL. LoQ = 0.04 ng/mL.
Endogenous Interference≤ 10% interference from common endogenous substances.No interference observed from various substances (Bilirubin, Cholesterol, Hemoglobin, Triglycerides, Fluorescein, Human Immunoglobulin, Total Protein) at tested concentrations. Biotin not evaluated as architecture does not use biotin:streptavidin.
Heterophile InterferenceNo significant interference from HAMA/RF.No HAMA/RF interference observed at tested concentrations.
Therapeutic Drug Interference≤ 10% interference from common therapeutic drugs.No interference observed from various therapeutic drugs at tested concentrations.
Cross ReactivityNo significant cross-reactivity with related substances.No interference observed from various related substances (e.g., Calcitonin, CGRP, therapeutic agents) at tested concentrations.
Method Comparison with Predicate Device (Correlation)High correlation and strong agreement with the predicate device (B.R.A.H.M.S PCT sensitive KRYPTOR). Explicit targets for slope, intercept, and correlation coefficient (e.g., slope near 1, intercept near 0, r > 0.95).N=522 samples (range 0.06-49.20 ng/mL): Weighted Deming Regression: Slope = 1.02 (95% CI: 0.99-1.05), Intercept = -0.02 (95% CI: -0.03 to -0.01), r = 0.98. Passing & Bablok Regression: Slope = 1.06 (95% CI: 1.04-1.09), Intercept = -0.04 (95% CI: -0.06 to -0.03), r = 0.98.
Method Comparison with Predicate Device (Concordance)High positive and negative percent agreement (PPA/NPA) at clinical cutoffs (e.g., typically >90-95%).At 0.1 ng/mL: PPA = 99.3%, NPA = 95.0%, Overall = 98.7%. At 0.25 ng/mL: PPA = 99.0%, NPA = 94.6%, Overall = 98.1%. At 0.50 ng/mL: PPA = 96.7%, NPA = 97.4%, Overall = 97.0%. At 2.0 ng/mL: PPA = 97.2%, NPA = 97.6%, Overall = 97.4%.
Matrix ComparisonNo significant matrix effect between different sample types (serum, plasma).High correlation (r=1.00) and minor differences in regression equations between Serum vs. EDTA, Li Heparin, and Na Heparin plasma samples across the range of 0.05-44.72 ng/mL.
Expected Values/Reference IntervalEstablishment of normal reference interval.99th percentile for PCT value in normal healthy subjects (N=144) was <0.05 ng/mL.
Clinical Performance (28-Day Mortality)Statistically significant association between ΔPCT and 28-day cumulative mortality; demonstration of prognostic and risk stratification utility.Binary ΔPCT (≤ 80% or > 80%) significantly associated with 28-day mortality (Fisher's Exact Test p=0.009). Cox proportional hazards regression showed HR of 1.82 (95% CI: 1.14-2.89; p=0.012) for ΔPCT positive vs. negative. Consistent utility for risk stratification based on initial PCT levels and ICU disposition.
Clinical Performance (Antibiotic Guidance)Support for use in decision-making on antibiotic therapy and discontinuation.Supported by systematic literature reviews and meta-analyses demonstrating reduction in antibiotic initiation and exposure without negative effects on mortality, complications, or length of stay.

Detailed Study Information:

  1. Sample sizes used for the test set and the data provenance:

    • Analytical Performance:
      • Precision: 5 contrived human serum samples, 80 replicates per sample.
      • Linearity: 9 samples (prepared from high and low human serum pools).
      • Dilution Recovery: 5 human serum samples.
      • Detection Limits (LoD): 360 determinations (160 blank, 200 low-level replicates).
      • Endogenous/Therapeutic Drug/Cross-Reactivity Interference: Not specified beyond "serum pools" and "serum and plasma."
      • Heterophile Interference: HAMA and RF positive patient samples + control samples.
      • Method Comparison: 623 native human serum samples with assigned values from the predicate device.
      • Matrix Comparison: 51 matched specimen sets in 4 tube types (Serum, EDTA plasma, Lithium Heparin Plasma, Sodium Heparin Plasma).
      • Expected Values/Reference Interval: 144 serum samples from normal subjects.
    • Clinical Studies:
      • 28-day Mortality: 858 adult patients (>18 years) diagnosed with severe sepsis or septic shock admitted to ICU. The analysis population comprised 598 subjects.
      • Antibiotic Therapy/Discontinuation Support: Systematic literature reviews and meta-analyses:
        • Study-level meta-analysis (Initiation/Discontinuation): 11 RCTs (4090 patients) published 2004-2016.
        • Patient-level meta-analysis (Initiation): 13 RCTs (3142 patients) published 2004-2011.
        • Patient-level meta-analysis (Discontinuation): 5 RCTs (598 patients) published 2008-2010.
    • Data Provenance:
      • Analytical Data: In-house or contracted lab studies (implied to be prospective given the detailed methodology descriptions).
      • Clinical Data: "Banked specimens that were collected from subjects... and included as part of the intention-to-diagnose population from the BRAHMS MOSES study (DEN150009)." This indicates retrospective use of previously collected clinical trial data. The original MOSES study subjects were recruited across 13 investigational sites in the US. The meta-analyses for antibiotic guidance consolidated data from previously published RCTs.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For the analytical test sets, ground truth is established by standard laboratory methods and reference materials (e.g., recombinant PCT, serum pools), not human experts.
    • For the clinical study, the ground truth for patient outcomes (e.g., severe sepsis/septic shock diagnosis, 28-day all-cause mortality, need for ICU care) was presumably established by the clinical assessments and medical records from the original BRAHMS MOSES study that provided the banked specimens. The document does not specify the number or qualifications of clinicians/experts who established these initial diagnoses and outcomes for the MOSES study, but it's implied to be standard clinical practice within a multi-site clinical trial.
    • For the meta-analyses, the "ground truth" for antibiotic management decisions and patient outcomes (mortality, LOS, etc.) derived from the various RCTs where those decisions and outcomes were prospectively recorded by the clinical teams involved in those trials.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • For the analytical studies, the "ground truth" values are quantitative measurements or characteristics of samples, typically not subject to human adjudication in the same way as, for example, image interpretation. Internal laboratory quality control and statistical methods (e.g., regression analysis, CV calculations) serve as "adjudication."
    • For the clinical study, the data were obtained from previously collected "banked specimens" and "clinical assessments" from the BRAHMS MOSES study. The document does not describe a new adjudication process for this particular 510(k) submission's analysis of this retrospective data. The rigor of the original MOSES study's data collection and endpoint ascertainment would be relevant but is not detailed here.
    • The meta-analyses relied on data from published RCTs; any adjudication would have occurred within the original trials.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This device is an in vitro diagnostic (IVD) assay that measures a biomarker (Procalcitonin) in human specimens. It is not an AI-assisted diagnostic imaging or pathology device that requires human "readers." Therefore, an MRMC study is not applicable and was not performed.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The performance characteristics (precision, linearity, detection limits, interference, method comparison) are "standalone" results of the assay itself, demonstrating its analytical performance when run on the Atellica IM Analyzer. The clinical studies demonstrate the performance of the assay's results when used in conjunction with "other laboratory findings and clinical assessments," but the assay itself generates the PCT value independently of human interpretation of that specific run.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Analytical Studies: "Ground truth" for analytical studies is typically established by meticulously prepared reference materials, spike-in methods, and comparison to a well-validated predicate device using a large sample size of native patient samples.
    • Clinical Studies:
      • For the 28-day mortality prediction, the ground truth was outcomes data (survival status at 28 days) derived from a pre-existing clinical study (BRAHMS MOSES study).
      • For antibiotic guidance, the insights were derived from meta-analyses of published Randomized Controlled Trials (RCTs). The "ground truth" here is the aggregated, prospectively collected patient outcomes (antibiotic use, duration, mortality, complications, LOS) from these trials where patients were managed either by standard care or PCT-guided protocols.
  7. The sample size for the training set:

    • This document describes a 510(k) premarket notification for an IVD assay, not an AI/ML device that typically requires a distinct "training set." The performance studies for an IVD device like this are primarily for validation and demonstrate clinical and analytical performance. There is no explicit "training set" in the context of machine learning.
    • However, if one were to consider the development of the assay's methodologies, the calibration and control materials (e.g., PCT MCM) would be used for standardization and quality control, which could be seen as analogous to internal "training" or optimization data during product development, but no specific sample size for this is detailed as "training set."
  8. How the ground truth for the training set was established:

    • As noted above, the concept of a "training set" with ground truth establishment in the ML sense is not directly applicable to this type of IVD device submission. The assay is a chemical measurement system. Its "truth" is established via analytical validation showing its accuracy, precision, and comparability to reference methods/predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 16, 2018

Siemens Healthcare Diagnostics Inc. Fatima Pacheco Regulatory Affairs Clinical Specialist 511 Benedict Ave Tarrytown, New York 10591-5097

Re: K181002

Trade/Device Name: Atellica IM BRAHMS Procalcitonin (PCT) Regulation Number: 21 CFR 866.3215 Regulation Name: Device to detect and measure non-microial analytes in human clinical specimens to aid in assessment of patients with suspected sepsis. Regulatory Class: Class II Product Code: PRI, PMT, PTF Dated: April 13, 2018 Received: April 16, 2018

Dear Fatima Pacheco:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Steven R. Gitterman -S

for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181002

Device Name

Atellica IM B.R.A.H.M.S Procalcitonin (PCT)

Indications for Use (Describe)

The Atellica® IM BRAHMS Procalcitonin (PCT) assay is for in the quantitative determination of procalcitonin in human serum and plasma (EDTA, lithium heparin) using the Atellica® IM Analyzer. The Atellica IM BRAHMS PCT assay is intended for use, in conjunction with other laboratory findings and clinical assessments, as an aid in:

· The risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.

• Assessing the cumulative 28-day risk of all-cause mortality for patients diagnosed with severe sepsis or septic shock in the ICU or when obtained in the emergency department or other medical wards prior to ICU admission, using percent change in PCT level over time.

· Decision-making on antibiotic therapy for patients with suspected or confirmed lower respiratory tract infections (LRTI) - defined as community-acquired pneumonia (CAP), acute bronchitis, and acute exacerbation of chronic obstructive

pulmonary disease (AECOPD) - in an inpatient setting or an emergency department.

· Decision-making on antibiotic discontinuation for patients with suspected or confirmed sepsis.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510K SUMMARY

Atellica® IM B.R.A.H.M.S Procalcitonin (PCT)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

1. 510(k) Number: K181002

2. Purpose for Submission:

To obtain a substantial equivalence determination for the Atellica® IM B.R.A.H.M.S Procalcitonin (PCT) assay.

3. Applicant:

Contact: Fatima Pacheco Regulatory Clinical Affairs Specialist Address: Siemens Healthcare Diagnostics Inc. 511 Benedict Ave. Tarrytown, NY 10591 Phone: (914) 374-3770

Email: fatima.pacheco@siemens-healthineers.com

Date: June 04, 2018

4. Proprietary and Established Names: Atellica® IM B.R.A.H.M.S Procalcitonin (PCT)

    1. Measurand
      Procalcitonin

6. Regulatory Information:

Regulatory Information for Atellica® IM B.R.A.H.M.S Procalcitonin (PCT)

Trade NameAtellica® IM B.R.A.H.M.S Procalcitonin (PCT)
Common NameImmunoassay, Procalcitonin
Classification NameDevice to detect and measure non-microbialanalyte(s) in human clinical specimens to aid inassessment of patients with suspected sepsis.
FDA ClassificationClass II
Review PanelMicrobiology (83)
Product CodePMT
Regulation Number21 CFR 866.3215

{4}------------------------------------------------

Regulatory Information for Atellica® IM B.R.A.H.M.S Procalcitonin Calibrator (Included in assay kit)

Trade NameAtellica® IM B.R.A.H.M.S Procalcitonin Calibrator
Common NameSingle (specified) analyte calibrators (assayed andunassayed)
Classification NameCalibrator (assayed and unassayed)
FDA ClassificationClass II (Exempt)
Review PanelClinical Chemistry (75)
Product CodeJIT
Regulation Number21 CFR 862.1150

Regulatory Information for Atellica® IM B.R.A.H.M.S Procalcitonin QC & MCM (Sold Separately)

Trade NameAtellica® IM B.R.A.H.M.S Procalcitonin QualityControl (QC)Atellica® IM B.R.A.H.M.S Procalcitonin Master CurveMaterial (QC)
Common NameSingle (specified) analyte calibrators (assayed andunassayed)
Classification NameCalibrator (assayed and unassayed)
FDA ClassificationClass II (Exempt)
Review PanelClinical Chemistry (75)
Product CodeJIT
Regulation Number21 CFR 862.1150

7. Predicate Devices:

Device Name: B.R.A.H.M.S PCT sensitive KRYPTOR 510(k) Number: DEN150009; K171338 Manufacturer: B.R.A.H.M.S GmbH (Thermo Fisher Scientific)

8. Intended Use:

Same as Indications for Use

9. Indications for Use:

The Atellica® IM BRAHMS Procalcitonin (PCT) assay is for in vitro diagnostic use in the quantitative determination of procalcitonin in human serum and plasma (EDTA, lithium heparin, and sodium heparin) using the Atellica® IM Analyzer.

The Atellica IM BRAHMS PCT assay is intended for use, in conjunction with other laboratory findings and clinical assessments, as an aid in:

· The risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.

{5}------------------------------------------------

• Assessing the cumulative 28-day risk of all-cause mortality for patients diagnosed with severe sepsis or septic shock in the ICU or when obtained in the emergency department or other medical wards prior to ICU admission, using percent change in PCT level over time.

· Decision-making on antibiotic therapy for patients with suspected or confirmed lower respiratory tract infections (LRTI) – defined as community-acquired pneumonia (CAP), acute bronchitis, and acute exacerbation of chronic obstructive pulmonary disease (AECOPD) - in an inpatient setting or an emergency department.

  • · Decision-making on antibiotic discontinuation for patients with suspected or confirmed sepsis.

10. Special Conditions for use statement(s):

For Prescription Use Only

11. Warnings and Precautions for Test Interpretation:

The Atellica® IM BRAHMS Procalcitonin (PCT) assay is not indicated to be used as a stand-alone diagnostic assay and should be used in conjunction with clinical signs and symptoms of infection and other diagnostic evidence.

Decisions regarding antibiotic therapy should NOT be based solely on PCT concentrations.

PCT results should always be interpreted in the context of the clinical status of the patient and other laboratory results. Changes in PCT levels for the prediction of mortality, and overall mortality, are strongly dependent on many factors, including pre-existing patient risk factors and clinical course.

The need to continue ICU care at Day 4 and other covariates, such as age and Sequential Organ Failure Assessment (SOFA) score, are also significant predictors of 28-day cumulative mortality risk.

PCT levels may not be elevated in patients infected by certain atypical pathogens, such as Chlamydophila pneumoniae and Mycoplasma pneumoniae.

Certain patient characteristics, such as severity of renal failure or insufficiency, may influence PCT values and should be considered as potentially confounding clinical factors when interpreting PCT values.

The safety and performance of PCT-quided therapy for individuals younger than 18 years of age, pregnant women, immunocompromised individuals, or those on immunomodulatory agents, was not formally analyzed in the supportive clinical trials.

Increased PCT levels may not always be related to systemic infection. These conditions include, but are not limited to:

  • · Patients experiencing major trauma and/or recent surgical procedure, including extracorporeal circulation or burns:
  • · Patients under treatment with OKT3 antibodies, OK-432, interleukins, TNF-alpha, and other drugs stimulating the release of pro-inflammatory cytokines or resulting in anaphylaxis:
  • · Patients diagnosed with active medullary C-cell carcinoma, small cell lunq carcinoma, or bronchial carcinoid:

{6}------------------------------------------------

  • · Patients with acute or chronic viral hepatitis and/or decompensated severe liver cirrhosis (Child-Pugh Class C);
  • Patients with prolonged or severe cardiogenic shock, prolonged severe organ perfusion anomalies, or after resuscitation from cardiac arrest;
  • · Patients receiving peritoneal dialysis or hemodialysis treatment;
  • · Patients with biliary pancreatitis, chemical pneumonitis, or heat stroke;
  • · Patients with invasive fungal infections (such as candidiasis and aspergillosis) or acute attacks of plasmodium falciparum malaria; and
  • Neonates during the first 2 days of life.

12. Special Instrument Requirement:

For use on the Atellica® IM Analyzer

13. Device Description:

The Atellica IM BRAHMS PCT assay is comprised of the following reagents:

ComponentVolumeIngredients
Atellica IM BRAHMS PCT Primary Reagent ReadyPack (assay kit)
PCT Lite Reagent5.0 mL/packMouse monoclonal anti-PCT antibody (~0.5 μg/mL) labeled with acridinium ester in protein buffer; bovine serum albumin; surfactant; preservatives
PCT Solid Phase Reagent10.0 mL/packMouse monoclonal anti-fluorescein antibody coated paramagnetic particles (~0.15 mg/mL) in buffer; surfactant; preservatives
PCT Ancillary Reagent4.5 mL/packMouse monoclonal anti-PCT antibody (~13.3 μg/mL) labeled with fluorescein in protein buffer; bovine serum albumin; surfactant; preservatives
PCT Calibrator (Kitted-included in assay kit)
PCT Low and High Calibrators2.0 mL/vialLyophilized; after reconstitution, recombinant PCT; equine serum; preservatives
Atellica IM BRAHMS PCT Quality Control (sold separately)
PCT Low and High Quality Controls2.0 mL/vialLyophilized; after reconstitution recombinant PCT; equine serum; preservatives
Atellica IM BRAHMS PCT Master Curve Material (sold separately)
PCT MCM1-51.0 mL/vialLyophilized; after reconstitution, various levels of recombinant PCT; equine serum; preservatives

{7}------------------------------------------------

14. Substantial Equivalence Information

The following table describes the similarities and differences between the Atellica IM BRAHMS PCT (Candidate Device) and the B.R.A.H.M.S PCT sensitive KRYPTOR® (Predicate Device).

Trade NameCandidate DevicePredicate Device
Atellica® IM B.R.A.H.M.SProcalcitonin (PCT)B.R.A.H.M.S PCT sensitiveKRYPTOR®
Intended UseThe Atellica® IM BRAHMSProcalcitonin (PCT) assay is forin vitro diagnostic use in thequantitative determination ofprocalcitonin in human serumand plasma (EDTA, lithiumheparin, and sodium heparin)using the Atellica® IM Analyzer.The Atellica IM BRAHMSPCT is intended for use, inconjunction with otherlaboratory findings and clinicalassessments as an aid in:The risk assessment ofcritically ill patients on theirfirst day of intensive care unit(ICU) admission forprogression to severe sepsisand septic shock.Assessing the cumulative 28-day risk of all-cause mortalityfor patients diagnosed withsevere sepsis or septic shockin the ICU or when obtainedin the emergency departmentor other medical wards priorto ICU admission, usingpercent change in PCT levelover time.Decision making on antibiotictherapy for patients withsuspected or confirmed lowerrespiratory tract infections(LRTI) - defined ascommunity-acquiredpneumonia (CAP), acutebronchitis, and acuteexacerbation of chronicobstructive pulmonarydisease (AECOPD) – in aninpatient setting or anThe B·R·A·H·M·S PCTsensitive KRYPTOR® is animmunofluorescent assay usingTime-Resolved AmplifiedCryptate Emission (TRACE®)technology to determine theconcentration of PCT(procalcitonin) in human serumand EDTA or heparin plasma.The B·R·A·H·M·S PCTsensitive KRYPTOR® isintended to be performed onthe B·R·A·H·M·S KRYPTOR®analyzer family.Used in conjunction with otherlaboratory findings and clinicalassessments, B·R·A·H·M·SPCT sensitive KRYPTOR® isintended for use as follows:to aid in the risk assessment ofcritically ill patients on their firstday of ICU admission forprogression to severe sepsisand septic shock,to determine the change in PCTlevel over time as an aid inassessing the cumulative 28-day risk of all-cause mortalityfor patients diagnosed withsevere sepsis or septic shock inthe ICU or when obtained in theemergency department or othermedical wards prior to ICUadmission,to aid in decision making onantibiotic therapy, for inpatientsor patients in the emergencydepartment with suspected orconfirmed lower respiratorytract infections (LRTI) – definedas community-acquiredpneumonia (CAP), acutebronchitis and acute
emergency department.Decision making on antibioticdiscontinuation for patients withsuspected or confirmed sepsis.exacerbation of chronicobstructive pulmonary disease(AECOPD),to aid in decision making onantibiotic discontinuation forpatients with suspected orconfirmed sepsis.
AnalyteProcalcitonin (PCT)Same
AutomatedAutomated assaySame
MeasurementQuantitativeSame
Sample TypeHuman Serum, Plasma (EDTA,Lithium, Heparin, SodiumHeparin)Human Serum, Plasma ( EDTAand Heparin)
Assay MeasuringInterval0.04 - 50.00 ng/mL0.02 - 50 µg/L
Operating PrincipleSandwichSandwich
TechnologyDirect ChemiluminescenttechnologyImmunofluorescence TRACE®technology
InstrumentAtellica IM AnalyzerKRYPTOR® Test System
Sample Volume100 µL50 µL
CalibratorsAtellica IM B.R.A.H.M.S PCTCalibrators (lyophilized)2 levels (Low and High); Afterreconstitution, recombinant PCT;equine serum; preservativesB.R.A.H.M.S PCT sensitiveKRYPTOR® Calibrator:1 vial of lyophilizedrecombinant PCT indefibrinated human plasma(range 22.50-27.50µg/L)
ControlsAtellica IM B.R.A.H.M.S PCTQuality Control (lyophilized)2 levels (Low and High); Afterreconstitution, recombinant PCT;equine serum; preservativesB.R.A.H.M.S PCT sensitiveKRYPTOR® QC Kit:6 vials (3 of each control):Control 1 (0.20-0.40µg/L)Control 2 (8.00-12.00µg/L)

{8}------------------------------------------------

{9}------------------------------------------------

Standard/Guidance Document Referenced 15.

The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission:

  • 트 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline -Third Edition (CLSI EP05-A3, 2014; Recognition Number 7-251)
  • Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical . Approach; Approved Guideline (CLSI EP06-A, 2003; Recognition Number 7-193)
  • . Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition (CLSI EP07-A2, 2005; Recognition Number 7-127)
  • . Measurement Procedure Comparison And Bias Estimation Using Patient Samples --Third Edition (CLSI EP09-A3, 2013; Recognition Number 7-245)
  • . Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline -- Second Edition (CLSI EP17-A2, 2012; Recognition Number 7-233)
  • Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition (CLSI EP28-A3c - formerly C28-A3c, 2010; Recognition Number 7-224)
  • . Medical devices – Application of risk manaqement to medical devices (ANSI/AAMI/ISO 14971:2007/(R)2010; Recognition Number 5-70)

16. Test Principle

The Atellica IM BRAHMS PCT assay is a 2-site sandwich immunoassay using direct chemiluminescent technology that uses 3 mouse monoclonal antibodies specific for PCT. The first antibody, in the Lite Reagent, is a mouse monoclonal anti-PCT antibody labeled with acridinium ester. The second and third antibodies, in the ancillary reagent, are mouse monoclonal anti-PCT antibodies labeled with fluorescein. The immunocomplex formed with PCT is captured with mouse monoclonal anti-fluorescein antibody coupled to paramagnetic particles in the Solid Phase.

A direct relationship exists between the amount of PCT present in the patient sample and the amount of relative light units (RLUs) detected by the system.

{10}------------------------------------------------

17. Performance Characteristics

1. Analytical Performance

A. Precision

The precision of the Atellica IM B.R.A.H.M.S PCT assay precision study was performed in accordance with CLSI EP05-A3 - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition. Wayne, PA: Clinical and Laboratory Standards Institute; 2014.

A single site study was conducted with a panel of five (5) contrived human serum samples, calibrators, and controls were tested. The panel of human samples was prepared by spiking normal serum samples with recombinant PCT spiking material. Samples were frozen in aliquots prior to the study. Each sample was assayed in two (2) replicates per run, with two (2) runs per day separated by at least 2 hours, for 20 days yielding a total of 40 runs, and 80 replicates per sample. For each analyte concentration level, the mean value with variance components (standard deviation and %CV) was determined. The following results were obtained:

SampleNo. ofreplicatesMean(ng/mL)Repeatability(Within-Run)Within-Lab(Total Precision)
SD%CVSD%CV
Serum 1800.050.009.70.0112.1
Serum 2800.270.001.80.012.8
Serum 3800.750.011.20.022.1
Serum 4801.520.021.40.042.6
Serum 58019.140.291.50.432.2
Control 1800.230.013.40.015.6
Control 2807.850.111.40.607.7

A modeling analysis was conducted to estimate the %Bias, %CV, and % Total Error at each medical decision point. %Bias was calculated relative to reference values derived using the slope and the intercept derived experimentally from the method comparison study between Atellica IM BRAHMS PCT and BRAHMS sensitive KRYPTOR. The %CV measurement was derived using the precision profile: the %Total Error was calculated as 1.65*(%CV) + (%Bias). The results from each of three regressions (%Bias, %CV, and %TE) are summarized in the table below.

PCT Level(ng/mL)Bias(%)CV(%)Total Error(%)
0.1022%6.132%
0.256%2.611%
0.52%1.44%
2.0-1%0.41%

{11}------------------------------------------------

B. Linearity/Assay Measuring Range

Linearity of the Atellica IM B.R.A.H.M.S PCT assay was performed according to CLSI EP06-A - Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach: Approved Guideline. The study was performed using 9 samples spanning the assay range, prepared using high and low human serum pools. Testing was performed on one (1) Atellica IM Analyzer with 2 PCT reagent lots, 3 replicates per level. The mean dose value of each sample was used in data analysis. The results were analyzed by % Deviation from Linear Fit.

Regression analysis using the first, second, and third order models were conducted to assess the linearity of the assay. Regression statistics (i.e., deviation from linearity for non-linear pools) at all levels tested demonstrated a ≤10% deviation from linear fit.

Linearity was confirmed in the range of 0.03 – 63.24 ng/mL. The measuring range claim for the Atellica IM BRAHMS PCT assay is 0.04 – 50.00 ng/mL. For PCT concentrations greater than 50.00 ng/mL, the measurement range can be extended up to 1000 ng/mL by 1:20 dilution of the sample.

C. Dilution Recovery

To evaluate the degree of bias introduced when the instrument auto-dilution is used on samples within the assay measuring range or above.

Five (5) human serum samples spiked with recombinant PCT stock. The samples were diluted to 1:20 with Atellica IM Multi-Diluent 1 both manually and automatically onboard the system. Recoveries ranged from 96% to 102% with a mean of 99%.

The samples were diluted (1:20) with Atellica IM Multi-Diluent 1 using the auto-dilution protocol. Recoveries ranged from 92% to 107%.

D. Hook Effect

A study was performed to evaluate the hook effect. In this assay, patient samples with procalcitonin concentrations as high as 2000 ng/mL will report > 50.00 ng/mL.

E. Detection Limits

Detection capability was determined in accordance with CLSI Document EP17-A2. The assay is designed to have a limit of blank (LoB) < 0.03 ng/mL, a limit of detection (LoD) < 0.04 ng/mL, and a limit of quantitation (LoQ) ≤ 0.06 ng/mL. Representative detection capability data are presented below.

The LoB corresponds to the highest measurement result that is likely to be observed for a blank sample. The LoB of the Atellica IM BRAHMS PCT assay is 0.00 ng/mL.

The LoD corresponds to the lowest concentration of procalcitonin that can be detected with a probability of 95%. The LoD for the Atellica IM BRAHMS PCT assay is 0.03 ng/mL and was determined using 360 determinations, with 160 blank and 200 low-level replicates, and LoB of 0.00 ng/mL.

The LoQ corresponds to the lowest amount of procalcitonin in a sample at which the within-lab CV is ≤ 20%. The LoQ of the Atellica IM BRAHMS PCT assay is 0.04 ng/mL, and was determined using multiple patient samples in the interval 0.01-0.05 ng/mL. All samples were assayed in duplicate in each of 2 runs per day using 2 reagent lots, over a period of 20 days.

{12}------------------------------------------------

F. Endogenous Interference

Endogenous interference studies were performed according to CLSI EP07-A2. The assay was designed to have ≤ 10% interference from the following substances. Interference (bias) is the difference in the results between the control sample (does not contain the interferent) and the test sample (contains the interferent) expressed in percent. Potential interference from the following substances was tested by adding these substances to serum pools containing PCT ranging from 0.32-2.59 ng/mL.

The interference with biotin was not evaluated since Atellica IM BRAHMS PCT reagents architecture does not use biotin:streptavidin. Therefore interference with biotin is not possible.

Endogenous SubstanceConcentration Tested(mg/dL)Results
Bilirubin (Conjugated)40 mg/dLNo interference observed
Bilirubin (Unconjugated)40 mg/dLNo interference observed
Cholesterol400 mg/dLNo interference observed
Hemoglobin500 mg/dLNo interference observed
Triglycerides1000 mg/dLNo interference observed
Fluorescein0.1 µg/mLNo interference observed
Human Immunoglobulin3.6 g/dLNo interference observed
Total Protein (as low as )3.5 g/dLNo interference observed
Total Protein (as high as)up to 12.0 g/dLNo interference observed

G. Heterophile Interference

Testing was performed to determine whether the presence of HAMA (Human anti-mouse antibody) may interfere with Atellica IM BRAHMS PCT assay results.

HAMA and RF positive patient samples were divided in half and spiked with PCT to a "Level 1" (0.4-0.7 ng/mL) and a "Level 2" (1.5-2.0 ng/mL) concentrations. A control sample negative for HAMA and RF ("serum control" and "plasma control") was spiked with equal amounts of PCT. Acceptance criteria were met for the HAMA spiking study. The presence of HAMA in the samples did not influence the performance of the Atellica IM BRAHMS PCT assay. No HAMA/RF interference was observed for PCT analyte with the presence of HAMA at the concentrations tested.

H. Therapeutic Drug Interference

Therapeutic drug interference studies were performed according to CLSI EP07-A2. The assay was designed to have ≤ 10% interference from the following substances. Interference (bias) is the difference in the results between the control sample (does not

{13}------------------------------------------------

contain the interferent) and the test sample (contains the interferent) expressed in percent. Potential interference from the following substances was tested by adding these substances to serum pools containing PCT ranging from 0.42-1.99 ng/mL. All the therapeutic drugs evaluated did not affect the test performance at concentrations evaluated.

DrugConcentration Tested(mg/dL)Results
Acetaminophen20No interference observed
Acetylsalicylic Acid100No interference observed
Alcohol405No interference observed
Azithromycin1.17No interference observed
Caffeine6No interference observed
Celecoxib24No interference observed
Cetirizine HCl0.36No interference observed
Dextramethorphan0.14No interference observed
Doxycycline5No interference observed
Epinephrine1.79No interference observed
Fentanyl1No interference observed
lbuprofen50No interference observed
Levofloxacin1.75No interference observed
Loratadine0.03No interference observed
Nicotine0.1No interference observed
Oxymetazoline HCl0.01No interference observed
Phenylephrine0.02No interference observed
Prednisolone0.3No interference observed
Salmeterol0.006No interference observed
Tiotropium0.0022No interference observed

l. Cross Reactivity

Potential cross-reactivity of drugs and metabolites were evaluated in accordance with CLSI document EP07-A2. The following substances were evaluated for cross-reactivity and do not interfere with the Atellica IM BRAHMS PCT assay when present in serum and plasma at the concentrations indicated. Testing was performed with serum, EDTA plasma, sodium heparin, and lithium heparin plasma matrices. The results are presented in the table below:

SubstanceConcentration Tested(mg/dL)Results
Calcitonin-human8No interference observed
Calcitonin salmon30No interference observed
Calcitonin eel30No interference observed
α-CGRP30No interference observed
β-CGRP30No interference observed
Katacalcin30No interference observed
Cefotaxim900,000No interference observed
Dobutamine11,200No interference observed

{14}------------------------------------------------

SubstanceConcentration Tested(mg/dL)Results
Dopamine130,000No interference observed
Furosemide20,000No interference observed
Heparin40,000No interference observed
Imipenem1,180,000No interference observed
Noradrenaline2000No interference observed
Vancomycin3,500,000No interference observed

J. Method Comparison with Predicate Device

The method comparison study was performed according to the quidance of CLSI Guideline EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples.

A total of 623 native human serum samples with assigned BRAHMS Kryptor does values (0-1000 ng/mL) were tested with the Atellica IM BRAHMS PCT assay. Samples > 50.00 ng/mL were auto-diluted (1:20) by the Atellica IM Analyzer.

Two separate data analysis for this was performed, one with analysis of all samples with BRAHMS PCT sensitive KRYPTOR assay with sample range 0.06-49.20 ng/mL (N=522) and one with an extended analysis including all samples tested (N=623). Weighted Deming and Passing & Bablok regression analyses including slope and intercept with 95% Cl were calculated for N=522 samples. Regression analysis comparing the Atellica IM BRAHMS PCT assay results to the predicate assay results are summarized in table below and illustrated in Figures below.

ParameterWeighted DemingRegressionPassing & BablokRegression
N522522
Slope1.021.06
95% CI two sided0.99 to 1.051.04 to 1.09
Intercept-0.02-0.04
95% Cl two sided-0.03 to -0.01-0.06 to -0.03
Correlation coefficient ( r )0.980.98
Sample Range (ng/mL)0.06-49.200.06-49.20

{15}------------------------------------------------

Image /page/15/Figure/0 description: The image is a scatter plot comparing two sets of data, labeled as 'Rep 1 of atellica im' on the y-axis and 'Rep 1 of brahms kryptor' on the x-axis, both ranging from 0 to 60. The plot includes individual data points represented by triangles, along with two trend lines: an 'Identity Line' shown as a dashed line and a 'Weighted Deming Fit' shown as a solid line. The data points generally cluster around the trend lines, indicating a positive correlation between the two sets of data.

Figure 1: Weighted Deming Regression Analysis Atellica IM BRAHMS PCT vs. Predicate

Figure 2: Passing & Bablock Regression Analysis Atellica IM BRAHMS PCT vs. Predicate

Image /page/15/Figure/3 description: The image is a scatter plot comparing "Rep 1 of atellica im" on the y-axis and "Rep 1 of brahms kryptor" on the x-axis. The plot includes data points represented by triangles, along with two lines: an "Identity Line" and a "Passing & Bablok Fit" line. The data points are clustered along the lines, indicating a positive correlation between the two variables. The axes range from 0 to 60.

{16}------------------------------------------------

Data was further analyzed for concordance at the 0.1, 0.25, 0.5, and 2.0 ng/mL clinical cutoffs. From the total N=623 samples tested, the percent agreement between the Atellica IM BRAHMS PCT and the predicate BRAHMS sensitive KRYPTOR PCT at the cutoffs obtained are presented in Tables below.

Atellica IM BRAHMS PCTBRAHMS sensitive KRYPTORTotal
> 0.1 ng/mL≤ 0.1 ng/mL
> 0.1 ng/mL5394543
≤ 0.1 ng/mL47680
Total54380623

Atellica IM BRAHMS PCT vs Predicate at 0.1 ng/mL cutoff

Positive percent agreement (PPA) = 539/543 (99.3%); 95% two sided Cl 98.1%-99.7% Negative percent agreement (NPA) = 76/80 (95.0%); 95% two sided Cl 87.8%-98.0% Overall percent agreement = 98.7%; 95% Confidence Interval: 97.5%-99.4%

Atellica IM BRAHMS PCT vs Predicate at 0.25 ng/mL cutoff

BRAHMS sensitive KRYPTOR
Atellica IM BRAHMS PCT> 0.25 ng/mL≤ 0.25 ng/mLTotal
> 0.25 ng/mL4887495
≤ 0.25 ng/mL5123128
Total493130623

Positive percent agreement (PPA) = 488/493 (99.0%); 95% two sided Cl 97.7%-99.6% Negative percent agreement (NPA) = 123/130 (94.6%); 95% two sided Cl 89.3%-97.4% Overall percent agreement = 98.1%; 95% Confidence Interval: 96.7%-98.9%

Atellica IM BRAHMS PCT vs Predicate at 0.50 ng/mL cutoff

Atellica IM BRAHMS PCTBRAHMS sensitive KRYPTORTotal
> 0.50 ng/mL≤ 0.50 ng/mL
> 0.50 ng/mL4145419
≤ 0.50 ng/mL14190204
Total428195623

Positive percent agreement (PPA) = 414/428 (96.7%); 95% two sided Cl% 94.6-98.0% Negative percent agreement (NPA) =190/195 (97.4%); 95% two sided Cl 94.1%-98.8% Overall percent agreement = 97.0%; 95% Confidence Interval: 95.3%-98.0%

Atellica IM BRAHMS PCT vs Predicate at 2.0 ng/mL cutoff

Atellica IM BRAHMS PCTBRAHMS sensitive KRYPTORTotal
> 2.0 ng/mL≤ 2.0 ng/mL
> 2.0 ng/mL2798287
≤ 2.0 ng/mL8328336
Total287336623

Positive percent agreement (PPA) =279/287 (97.2%); 95% two sided Cl 94.6%-98.6% Negative percent agreement (NPA) = 328/336 (97.6%); 95% two sided Cl 95.4%-98.8% Overall percent agreement = 97.4%: 95% Confidence Interval: 95.9%-98.4%

{17}------------------------------------------------

Atellica IMBRAHMS PCT(ng/mL)≤ 0.10> 0.10 - ≤ 0.25> 0.25 - ≤ 0.50> 0.50 - ≤ 2.0> 2.0Total
≤ 0.1076400080
> 0.10 - ≤ 0.2543970050
> 0.25 - ≤ 0.5005555065
> 0.50 - ≤ 2.000141198141
> 2.00008279287
Total804876132287623

Cross Tabulation of all Samples (n=623) Atellica IM BRAHMS PCT vs Predicate

K. Matrix Comparison

Specimen equivalency was determined using the Passing-Bablok regression model in accordance with CLSI Document EP09-A3. Fifty-one (51) matched specimen sets in 4 tube types (Serum, EDTA plasma, Lithium Heparin Plasma, Sodium Heparin Plasma) were obtained from a vendor and spiked with equal amounts of recombinant PCT stock. Each sample as tested in singlicate per sample matrix and compared to the serum tube. Regression analysis demonstrated that no significant matrix effect between Serum, EDTA plasma, Lithium Heparin Plasma, Sodium Heparin Plasma tubes. The following results were obtained.

SpecimenType(x)ComparisonSample Type(y)NSample Range(ng/mL)Regression EquationCorrelationCoefficient
SerumEDTA510.05—44.72y = 1.04x + 0.03 ng/mL1.00
SerumLi Heparin510.05—44.72y = 1.05x + 0.02 ng/mL1.00
SerumNa Heparin510.05—44.72y = 1.03x + 0.02 ng/mL1.00

L. Expected Values/ Reference Interval

To establish the reference interval of a normal population, one-hundred and forty-four (144) serum samples from normal subjects were tested using the Atellica IM BRAHMS PCT assay on the Atellica IM Analyzer according to the CLSI Guideline EP28-A3c. In a population of N=144 self-reported healthy subjects, the PCT value corresponding to the 99t0 percentile was <0.05 ng/mL.

Age RangeNEthnicity
African AmericanAsianCaucasianHispanicOther
<60 years13259144253
≥60 years1221810

{18}------------------------------------------------

2. Clinical Studies

The Atellica IM BRAHMS PCT clinical performance study was conducted with banked specimens that were collected from subjects(>18 years of age) diagnosed with severe sepsis or septic shock, admitted to the ICU and included as part of the intention-todiagnose population from the BRAHMS MOSES study (DEN150009).

Cumulative 28-day mortality in patients with Severe Sepsis and Septic Shock a. · APCT ≤ 80%

A decrease in the PCT levels below or equal to 80% defines a positive ΔPCT test result representing a higher risk for 28-day all-cause mortality of patients diagnosed with severe sepsis or septic shock.

ΔΡCT > 80% .

A decrease in the PCT levels of more than 80% defines a negative ΔPCT test result representing a lower risk for 28-day all-cause mortality of patients diagnosed with severe sepsis or septic shock.

The combination of the first PCT level (≤ 2.0 ng/mL or > 2.0 ng/mL) at initial diagnosis of severe sepsis or septic shock with the patient's clinical course and the change in PCT level over time until Day 4 provides important additional information about the mortality risk.

b. Risk Assessment for progression to Severe Sepsis and Septic Shock

  • PCT > 2.0 ng/mL: A PCT level above 2.0 ng/mL on the first day of ICU . admission is associated with a high risk for progression to severe sepsis and/or septic shock.
  • . PCT < 0.5 ng/mL: A PCT level below 0.5 ng/mL on the first day of ICU admission is associated with a low risk for progression to severe sepsis and/or septic shock.

PCT levels < 0.5 ng/mL do not exclude an infection, because localized infections (without systemic signs) may also be associated with such low levels.

The Atellica IM BRAHMS PCT assay was evaluated for the prediction of cumulative 28-day all-cause mortality in a study of 858 adult patients diagnosed with severe sepsis or septic shock recruited across 13 investigational sites in the US. The analysis population (598 subjects) comprised 44% female and 56% male patients with a mean age of 64 years. About half of the patients had severe sepsis (51%) vs. septic shock (49%). Infections were mainly community-acquired (91%).

The binary test result (ΔΡCT > 80% or ≤ 80%) was significantly associated with 28-day cumulative mortality (vital status on day 28) (Two-sided Fisher's Exact Test p-value = 0.009). Adjusted for ICU vs. non-ICU patient subgroups (based on hospital location at Day 4 after initial diagnosis), the association remained significant (Cochran-Mantel-Haenszel Test p-value = 0.044). In each binary ΔPCT subgroup, the 28-day cumulative mortality rate was stratified by need to continue ICU care on Day 4 and/or the selection of Day 0 vs. Day 1 as the baseline measurement day for the ΔΡCT calculation as follows:

{19}------------------------------------------------

Prediction performance of binary ΔPCT stratified by ICU care on Day 4.

28-Day Mortality Risk Stratified by Patient Location on Day 4 ΔΡCT > 80% = Test Negative

ΔPCT ≤ 80% = Test Positive
28-Day Mortality RiskPrognostic Accuracya
Day 4 PatientLocationPCTIntervalΔPCT > 80%(95% CI)bΔPCT ≤ 80%(95% CI)Sensitivity (%)(95% CI)Specificity (%)(95% CI)
ICUΔPCT4.021.129.975.833.8
(12.2-30.0)(23.2-36.6)(65.7-85.9)(26.9-40.7)
ΔPCT4.119.031.177.336.1
(10.4-27.6)(24.2-37.9)(67.1-87.5)(29.1-43.1)
Non-ICUΔPCT4.05.710.773.741.3
(1.7-9.8)(6.4-15.0)(57.1-90.3)(35.5-47.2)
ΔPCT4.17.49.567.938.5
(2.7-12.0)(5.5-13.6)(50.6-85.2)(32.7-44.3)

a Prognostic accuracy refers to how accurate the ΔΡCT (≤ 80% vs. > 80%) can predict mortality risk.

b Confidence interval.

Additional stratification of patients based on absolute initial PCT levels (> or ≤ 2.0 ng/mL) at Day 0 (ΔΡCT4.0) or Day 1 (ΔΡCT4.1) revealed subgroups with particularly reduced or elevated mortality risk considering their hospital disposition on Day 4. Mortality risk and prognostic performance are given for the following subgroups in the tables below:

  • Patients with PCT > 2.0 ng/mL at Day 0 (or Day 1) receiving ICU care on Day 4. 1.
  • Patients with PCT ≤ 2.0 ng/mL at Day 0 (or Day 1) receiving ICU care on Day 4. 2.
  • Patients with PCT > 2.0 ng/mL at Day 0 (or Day 1) without ICU care on Day 4. 3.
  • Patients with PCT ≤ 2.0 ng/mL at Day 0 (or Day 1) without ICU care on Day 4. ব

28-Day Mortality Risk Stratified by Patient Location on Day 4

Absolute PCT Value on Day 0

ΔΡCT > 80% = Test Negative

ΔΡCT ≤80% = Test Positive

Day 4 PatientLocationPCT at Day0ΔPCT > 80%(95% CI)bΔPCT ≤ 80%(95% CI)Sensitivity (%)(95% CI)Specificity (%)(95% CI)
ICU≤ 2.0 ng/mL12.8(0.0-36.7)23.2(13.9-32.5)94.8(84.9-100.0)10.3(2.4-18.2)
> 2.0 ng/mL22.0(12.5-31.6)35.2(25.8-44.6)68.6(55.7-81.5)46.9(37.8-55.9)
Non-ICU≤ 2.0 ng/mL0.0(0.0-21.8)8.1(3.5-12.7)100.0(71.5-100.0)10.5(4.9-16.2)
> 2.0 ng/mL6.4(1.9-11.0)16.1(6.8-25.3)57.6(33.7-81.4)67.2(59.5-74.9)

ര Prognostic accuracy refers to how accurate the ΔΡCT (≤ 80% vs. > 80%) can predict mortality risk.

b Confidence interval.

{20}------------------------------------------------

28-Day Mortality Risk Stratified by Patient Location on Day 4

Absolute PCT Value on Day 1 ΔΡCT > 80% = Test Negative

ΔPCT ≤ 80% = Test Positive
Day 4 PatientLocationPCT at Day 128-Day Mortality RiskPrognostic Accuracya
ΔPCT > 80%(95% CI)bΔPCT ≤ 80%(95% CI)Sensitivity (%)(95% CI)Specificity (%)(95% CI)
ICU≤ 2.0 ng/mL20.2(0.0-57.2)26(16.2-35.8)95.2(86.0-100.0)6.9(0.0-14.3)
> 2.0 ng/mL19.0(10.1-27.9)35.0(25.5-44.5)69.8(56.4-83.1)49.9(41.1-58.6)
Non-ICU≤ 2.0 ng/mL0.1(0.0-26.5)6.4(2.3-10.4)99.9(66.4-100.0)8.5(3.2-13.7)
> 2.0 ng/mL8.2(3.1-13.3)16.4(7.2-25.5)53.4(31.1-75.8)65.7(57.7-73.7)

a Prognostic accuracy refers to how accurate the ΔΡCT (≤ 80% vs. > 80%) can predict mortality risk.

b Confidence interval.

Image /page/20/Figure/5 description: The image shows the text "Survival until Day 28, PCT >2.0 ng/mL at Day 0, ICU Day 4 (n=176)". The text describes survival until day 28 with a procalcitonin (PCT) level greater than 2.0 ng/mL. The data was collected at day 0, and the intensive care unit (ICU) stay was day 4. The sample size is 176.

Image /page/20/Figure/6 description: This image is a survival plot showing the proportion of patients surviving over a period of 28 days. The x-axis represents time in days, ranging from 0 to 28. The y-axis represents the proportion of patients surviving, ranging from 0% to 100%. Two survival curves are plotted: one for patients who tested negative (ΔPCT > 80%) and another for patients who tested positive (ΔPCT ≤ 80%). The survival rate is higher for patients who tested negative.

{21}------------------------------------------------

Image /page/21/Figure/0 description: This image is a survival plot showing the proportion of patients surviving over time. The x-axis represents time in days, ranging from 0 to 28. There are two survival curves, one for patients who tested negative (ΔPCT > 80%) and another for patients who tested positive (ΔPCT ≤ 80%). The survival rate is higher for patients who tested negative compared to those who tested positive.

Survival until Day 28, PCT >2.0 ng/mL at Day 0, non-ICU Day 4 (n=183)

Time (days)

Survival until Day 28, PCT ≤ 2.0 ng/mL at Day 0, ICU Day 4 (n=89)

Image /page/21/Figure/4 description: This image is a survival plot showing the proportion of patients surviving over time in days. The x-axis represents time in days, ranging from 0 to 28, and the y-axis represents the proportion of patients surviving, ranging from 0% to 100%. There are two lines on the plot, one representing patients who tested negative (ΔPCT > 80%) and the other representing patients who tested positive (ΔPCT ≤ 80%). The survival rate is higher for patients who tested negative.

{22}------------------------------------------------

Image /page/22/Figure/0 description: This image is a survival plot showing the proportion of patients surviving over a period of 28 days. The x-axis represents the number of days, ranging from 0 to 28. The y-axis represents the proportion of patients surviving, ranging from 0% to 100%. There are two lines on the plot, one green and one red, representing two different test groups: test negative (ΔPCT > 80%) and test positive (ΔPCT ≤ 80%).

Survival until Day 28, PCT ≤ 2.0 ng/mL at Day 0, non-ICU Day 4 (n=150)

Time (days)

Time-to-event analysis illustrated by the Kaplan-Meier curves above shows that patients in the ICU, or with an initial PCT value > 2.0 ng/mL, had a lower survival probability (higher cumulative mortality risk) from study Day 4 until the end of follow-up time (28 days) when the APCT test result was positive compared to when the APCT result was negative (patient subgroups according to hospital location on Day 4 and initial PCT level). A generally lower mortality rate was observed in the non-ICU subgroup.

The mortality ratios for ΔΡCT positive vs. ΔΡCT negative patient subgroups were:

  • 1.92 for patients with PCT > 2.0 ng/mL at Day 0 receiving ICU care on Day 4 .
  • . 2.11 for patients with PCT ≤ 2.0 ng/mL at Day 0 receiving ICU care on Day 4
  • . 2.78 for patients with PCT > 2.0 ng/mL at Day 0 without ICU care on Day 4
  • . ---* for patients with PCT ≤ 2.0 ng/mL at Day 0 without ICU care on Day 4

*Mortality ratio cannot be calculated due to 0 fatalities in the APCT negative group. Mathematically, it tends towards infinite

Based on relative mortality ratios, a decrease by more than 80% from Day 0 (or Day 1) to Day 4 constitutes a lower risk for mortality within 28 days compared to smaller declines in each subgroup. For the prediction of absolute mortality risks, ICU disposition at Day 4 and initial PCT concentrations should be considered:

. An initial PCT level ≤ 2.0 ng/mL on Day 0 followed by a PCT decline of more than 80% until Day 4 indicates an almost 2-fold lower cumulative 28-day mortality risk (12.8%) for patients with severe sepsis or septic shock who are still in the ICU by Day 4 compared to those patients with an initial PCT level > 2.0 ng/mL (22.0%). Regardless of the initial PCT level, patients in the ICU on Day 4 that do not decline by more than 80% in PCT plasma concentration from Day 0 to Day 4 have an even higher mortality risk of 23.2-35.2%.

{23}------------------------------------------------

. An initial PCT level > 2.0 ng/mL that does not decline by more than 80% until Day 4 signals that such patients remain at a high mortality risk (16.1%) even when they are no longer receiving ICU care on Day 4. Mortality was otherwise observed between 0.0-8.1% for patients discharged from the ICU by Day 4.

Performance of ΔPCT from Day 0 to Day 4 (≤ 80% vs. > 80%) as a prognostic for 28-day cumulative risk of mortality was quantified by Cox proportional hazards regression analysis with a hazard ratio of 1.82 (95% Cl: 1.14-2.89; p-value = 0.012). That is, the relative risk of cumulative 28-day mortality is about 1.8-fold higher if an individual tests positive for ΔΡCT (≤ 80%) than if an individual tests neqative (> 80%). As a comparison, the table below lists the univariate hazard ratios for other clinical factors evaluated as separate predictors of mortality in the study population.

Hazardp-Value
FactorComparisonRatio Per-ProtocolPer-Protocol
ΔPCT4.0≤ 80% vs. > 80%1.82 (1.14-2.89)0.012
ΔPCT4.1≤ 80% vs. > 80%1.72 (1.08-2.75)0.023
APACHEDifference of 5 Units1.36 (1.22-1.53)< 0.001
Max SOFADifference of 3 Units1.73 (1.50-2.00)< 0.001
Antibiotic AdequacyNo vs. Yes1.59 (1.00-2.53)0.051
Sepsis SeveritySeptic Shock vs. SevereSensis1.19 (0.80-1.76)0.386
Biol. Infection TypeGram Pos vs. Gram Neg0.83 (0.48-1.45)0.522
Biol. Infection TypeOther vs. Gram Neg0.99 (0.63-1.54)0.960
Biol. Infection TypeFungal vs. Gram Neg2.44 (0.87-6.84)0.090
Clinical Infection TypeNosocomial vs. Community0.76 (0.35-1.64)0.481
Positive Blood CultureYes vs. No1.05 (0.69-1.58)0.834
Baseline PCT> 2.0 ng/mL vs. ≤ 2.0 ng/mL1.62 (1.04-2.50)0.031
AgeDifference of 5 Years1.16 (1.08-1.24)< 0.001
GenderMale vs. Female0.95 (0.64-1.40)0.782
ICU Care on Day 4Yes vs. No3.45 (2.24-5.31)< 0.001

Univariate Hazard Ratios for 28-Dav All-Cause Mortality of ΔΡCT and Clinical Covariates

ΔPCT from Day 0 (or Day 1) to Day 4 remains a prognostic parameter for the risk of cumulative 28-day mortality in patients diagnosed with severe sepsis or septic shock even when the hazard ratio is adjusted for other mortality predictors in Cox multiple rearession models. The relative mortality risk estimates for ΔΡCT and selected predictors are given below with 95% confidence intervals. For continuous predictors, the hazard ratio (HR) was calculated for 1 standard deviation (SD) change in the predictor. For binary predictors, the risk estimate compares the hazards for the 2 binary results.

{24}------------------------------------------------

Hazard Ratio (95% Confidence Interval)
ModelBinary PredictorsContinuous Predictors (HR per 1 SD)
ΔPCTIntervalScore +CovariatesaΔPCT (≤ 80%vs. > 80%)Day 4 PatientLocation(ICU vs. Non-ICU)APACHE(1SD =8.13)maxSOFA(1SD =3.98)Age(1SD = 16.18)
Day 0untilDay 4APACHE1.91(1.09–3.35)2.65(1.65–4.24)1.23(0.98–1.55)1.59(1.27–1.98)
max SOFA1.75(1.01–3.02)1.69(1.02–2.78)1.96(1.53–2.52)1.69(1.35–2.10)
Day 1untilDay 4APACHE1.91(1.12–3.28)2.64(1.65–4.22)1.27(1.02–1.59)1.58(1.26–1.97)
max SOFA1.73(1.02–2.95)1.74(1.06–2.84)1.95(1.52–2.50)1.67(1.34–2.09)

Hazard Ratios for APCT and Selected Predictors from Multivariate Cox Regression Models

The models also included the following predictors (HR results not shown): Antibiotic Adequacy, Sepsis Severity,
Biological Infection Type, Clinical Infection Type, Positive B analysis, missing values for predictors were multiple imputed assuming they were Missing at Random (MAR) with the multiple imputations combined according to Rubin's rules.

The change of PCT over time can also be described by the ratio of PCT values from Day 4 and Day 0 (or Day 1):

PCT ratio = PCT Day 4/PCT Day 0 (or Day 1)

A decline of ΔΡCT = 80% translates into a PCT ratio of 0.2. The PCT ratio has values larger than 0.2 when the ΔΡCT decline is < 80%, which is associated with a higher risk for cumulative 28-day all-cause mortality in patients diagnosed with severe sepsis or septic shock. Likewise, a PCT ratio < 0.2 indicates a lower risk for mortality within 28 days. On a continuous scale, the relative mortality risk for such patients is higher the PCT ratio. The following table lists the hazard ratios for an increase by the factor 2 in PCT ratio (that is, the relative increase in mortality risk for a patient with any given PCT ratio compared to a patient with a 2-fold lower PCT ratio).

{25}------------------------------------------------

Hazard Ratios for APCT and Selected Predictors from Multivariate Cox Regression Models

ModelHazard Ratio (95% Confidence Interval)BinaryPredictor
Continuous Predictors(HR per 2-fold increase in PCT ratio or per equivalent in SD)
ΔPCTIntervalCovariatesaScore +PCT ratio(2-foldincrease)(SD equivalent)bAPACHE(SD equivalent)bmax SOFA(SD equivalent)bAge(SDequivalent)bDay 4PatientLocation(ICU vs.Non-ICU)
Day 0untilDay 4APACHE1.29(1.14–1.45)1.07 (0.95–1.21)1.31(1.16–1.47)2.56(1.59–4.11)
Day 0untilDay 4max SOFA1.22(1.08–1.38)1.38 (1.20–1.58)1.34(1.19–1.51)1.68(1.02–2.77)
Day 1untilDay 4APACHE1.35(1.17–1.57)1.19 (1.02–1.40)1.40(1.19–1.64)2.56(1.60–4.10)
Day 1untilDay 4max SOFA1.29(1.11–1.51)1.58 (1.32–1.89)1.46(1.25–1.71)1.74(1.06–2.86)

a The models also included the following predictors (HR results not shown): Antibiotic Adequacy, Sepsis Severity, Biological Infection Type, Positive Blood Culture, PCT on Day 0, and Gender.
In the analysis, missing values for predictors were multiple imputed assuming the Rando (MAR), with the multiple imputations combined according to Rubin's rules.

b A unit change of ΔΡCT on log-2-scale corresponded to 0.54 SD of ΔPCT from Day 0 until Day 4 (0.71 SD for ΔPCT from Day 1 until Dav 4). Accordingly, the reported ΔPCT hazard ratios refer to an increase of ΔPCT by a factor of 2. For comparability, hazard ratios of the other continuous predictors were estimated for the same fractional SDs, that is, 0.54 or 0.71, respectively

Cumulative 28-day all-cause mortality did not differ significantly for male vs. female patients. Demographics with outcome information are shown below:

Comparison of Clinical Covariates by 28-Day Outcome

AllDeadAlive% Dead
VariableClassNaNN%
GenderFemale2644621817.4
Male3345527916.5
Age, years (categorized)≤ 30391382.6
> 30, ≤ 45454418.9
> 45, ≤ 557486610.8
> 55, ≤ 651492612317.4
> 65, ≤ 751252110416.8

{26}------------------------------------------------

VariableClassAllNaDeadNAliveN% Dead%
Ethnicity> 751664112524.7
EthnicityAfrican-American2023217015.8
EthnicityAsian7070.0
EthnicityCaucasian3626429817.7
EthnicityHispanic2351821.7
EthnicityOther4040.0
Sepsis Classification on AdmissionSeptic Shock2915323818.2
Sepsis Classification on AdmissionSevere Sepsis3074825915.6
Baseline PCT, ng/mL< 0.5114169814.0
Baseline PCT, ng/mL≥ 0.5, ≤ 2.01141210210.5
Baseline PCT, ng/mL> 2.03386627219.5
Baseline PCT, ng/mLMissing3272521.9

a Number of patients.

  • c. Decision-making on antibiotic therapy for patients with suspected or confirmed LRTI
    • PCT < 0.10 nq/mL .

Antibiotic therapy strongly discouraged.

  • PCT 0.10-0.25 ng/mL ●
    Antibiotic therapy discouraged.

  • PCT 0.26-0.50 ng/mL .
    Antibiotic therapy encouraged.

  • PCT >0.50 ng/mL
    Antibiotic therapy strongly encouraged.

d. Sepsis Antibiotic Discontinuation:

  • . ДРСТ > 80% Antibiotic therapy may be discontinued
  • . PCT ≤ 0.50 ng/mL Antibiotic therapy may be discontinued

Two systematic literature reviews were performed to produce both study- and patientlevel meta-analyses, which are studies that combine and contrast data from multiple sources to identify patterns among study results (FDA public docket FDA-2016-N-2880).

The study-level meta-analysis used aggregate descriptive information extracted from publications, and the patient-level meta-analysis used aggregate patient-level data from

{27}------------------------------------------------

the raw dataset of each study. Each meta-analysis used random-effects models and calculated point estimates, differences, odds ratios (OR), interquartile ranges (IQRs), and 95% confidence intervals as appropriate. The endpoints evaluated were: proportion of subjects initiating antibiotics, duration of antibiotic therapy, exposure to antibiotics, length of hospital stay, mortality, and complications (patient-level only).

The study-level meta-analysis encompassed 11 randomized control trials (RCTs), which were published between 2004–2016, and included 4090 patients.

The patient-level meta-analysis encompassed 13 RCTs, which were published between 2004-2011 and included 3142 patients as listed below.

PublicationNaPCT Device
Bouadma, 2010630B·R·A·H·M·S PCT sensitive KRYPTOR®
Briel, 2008300B·R·A·H·M·S PCT sensitive KRYPTOR
Burkhardt, 2010550B·R·A·H·M·S PCT sensitive KRYPTOR
Christ-Crain, 2004243B·R·A·H·M·S PCT sensitive KRYPTOR
Christ-Crain, 2006302B·R·A·H·M·S PCT sensitive KRYPTOR
Hochreiter, 2009110B·R·A·H·M·S PCT LIA®
Kristoffersen, 2009223B·R·A·H·M·S PCT sensitive KRYPTOR
Long, 2011172B·R·A·H·M·S PCT sensitive KRYPTOR
Long, 2009127B·R·A·H·M·S PCT LIA
Nobre, 200879B·R·A·H·M·S PCT sensitive KRYPTOR
Schroeder, 200927B·R·A·H·M·S PCT LIA
Schuetz, 20091381B·R·A·H·M·S PCT sensitive KRYPTOR
Stolz, 2007226B·R·A·H·M·S PCT sensitive KRYPTOR

a Number of patients.

These meta-analyses concluded that PCT-guided antibiotic therapy resulted in:

  • 19.2% reduction in relative antibiotic initiation for all patients. ●
  • 38% reduction in overall antibiotic exposure (that is, total days of antibiotic therapy) for inpatients.
  • . 51% reduction in overall antibiotic exposure (that is, total days of antibiotic therapy) for patients who presented to the emergency department and other associated clinics, but were not admitted.
  • . 2.9-day reduction in antibiotic duration [1.25-day reduction in study-level].
  • 3.6-day reduction in total antibiotic exposure [2.79-day reduction in study-level].
  • . No negative effects in regards to mortality, complications, or length of stay.

{28}------------------------------------------------

ParameterStandard Care TherapyPCT-Guided Therapy
NaN (%) or Days, median (IQR)NN (%) or Days, median (IQR)
Initiation of antibiotics16061420 (88.4%)15361096 (71.4%)
Duration of antibiotics142010 (7, 12)10967 (4, 10)
Total exposure of antibiotics16069 (8, 12)15365 (0, 8)
30-day mortality1606119 (7.4%)1536103 (6.7%)
Complications1606339 (21.1%)1536276 (18.0%)
Hospital length of stay15836 (0, 13)15087 (0, 12)

Overview of the Patient-Level Meta-Analysis

a Number of patients included.

Decision-Making on Antibiotic Discontinuation for Suspected or Confirmed Septic Patients

Two systematic literature reviews were performed along with study- and patient-level meta-analyses, which are studies that combine and contrast data from multiple sources to identify patterns among study results. The study-level meta-analysis used aggregate descriptive information extracted from publications, and the patient-level meta-analysis used aggregate patient-level data from the raw dataset of each study. Each metaanalysis used random-effects models and calculated point estimates, differences, odds ratios (OR), interquartile ranges (IQRs), and 95% confidence intervals as appropriate (see tables below). The endpoints evaluated were: duration of antibiotic therapy (study level only), exposure to antibiotics (patient-level only), length of ICU stay, length of hospital stay (patient-level only), and mortality.

The study-level meta-analysis encompassed 10 RCTs, which were published between 2008-2016 and included 3489 patients. (See FDA public docket FDA-2016-N-2880).

PublicationNa, bPCT Device
Bouadma, 2010630B·R·A·H·M·S PCT sensitive KRYPTOR
Hochreiter, 2009110B·R·A·H·M·S PCT LIA
Nobre, 200879B·R·A·H·M·S PCT sensitive KRYPTOR
Schroeder, 200927B·R·A·H·M·S PCT LIA
Stolz, 2007101B·R·A·H·M·S PCT sensitive KRYPTOR

The above patient-level meta-analysis encompassed 5 RCTs, which were published between 2008-2010 and included 598 patients as listed below.

a Number of patients.

b Patients that did not classify as sepsis were removed prior to analysis (185 for PCT group and 164 for control group), leaving 598 patients.

{29}------------------------------------------------

Using this subset of meta-analyses, it was concluded that PCT-guided antibiotic therapy resulted in:

  • 1.5-day reduction in antibiotic duration. ●
  • . 3.2-day reduction in total antibiotic exposure.
  • 23% reduction in overall antibiotic exposure (that is, total days of antibiotic therapy) for patients who presented to the emergency department and other associated clinics, but were not admitted.
  • . No negative effects in regards to mortality, hospital length of stay, or ICU length of stay.
ParameterNaStandard Care TherapyN (%) or Days, median (IQR)PCT-Guided TherapyN (%) or Days, median (IQR)
Total exposure of antibiotics31112 (8, 18)287 8 (5, 15)
30-day mortality31174 (23.8%)287 57 (19.9%)
Hospital length of stay28823 (13, 38)259 21 (11, 37)
ICU length of stay31112 (6, 22)287 12 (6, 23)

Overview of the Patient-Level Meta-Analysis

a Number of patients included.

Recommendations for Laboratory Reports for Initiation and Discontinuation:

Procalcitonin Calculator is available at www.BRAHMS-PCT-The Change in Calculator.com.

It is suggested to report the absolute PCT values (single or serial). For serial PCT values, the report should also indicate if the ΔΡCT (%) was ≤ 80% or > 80%. The laboratory report should include a reference or a link to the Atellica IM BRAHMS PCT package insert for a guided interpretation of the test results.

3. Traceability and Value Assignment

The Atellica IM BRAHMS PCT assay is standardized using a reference preparation antigen as master standard. Assay standardization is verified through regression analysis of a PCT-positive patient population. Assigned values for calibrators and controls are traceable to this standardization.

4. Stability

Reagent Stability - Shelf-life

To establish the shelf-life claims for the Atellica IM BRAHMS PCT reagents and Calibrator (kit), the readents and calibrators stored at 2-8°C and were tested with various samples at multiple time points. Sets of calibrators were also stored at or below -40°C and were tested at the same time points.

Testing demonstrated that the Atellica IM BRAHMS PCT reagents are stable up to 12 months when properly stored at 2-8°C. Unopened Atellica IM BRAHMS PCT reagents and calibrators are stable at 2-8°C until expiration date printed on the product packaging.

Reagent Stability - On-Board

{30}------------------------------------------------

The onboard stability of the Atellica IM BRAHMS PCT reagents is 60 days onboard, pack calibration interval of 35 days, and lot calibration interval of 82 days.

Calibrator Stability- Shelf-life

Testing demonstrated that the Atellica IM BRAHMS PCT reagents are stable up to 12 months when properly stored at 2-8°C. Lyophilized calibrators are stable until the expiration date on the product when stored at 2-8°C.

Calibrator Stability- Reconstituted In use (2-8°C/OBS)

Reconstituted calibrators are stable for 24 hours at 2-8°C or for 8 hours at room temperature (package insert claims).

Specimen Stability

  • a. Room temperature
    Studies demonstrated that serum samples specimens are stable for 8 hours at room temperature.

  • b. Refrigerated (2-8°C)
    Studies demonstrated that serum samples are stable for 24 hours at 2-8°C.

  • c. Freeze/Thaw cycles (≤ -20°C)
    The recommended specimen handling are provided below.

  • Do not use samples that have been stored at room temperature for longer than 8 hours.

  • Tightly cap and refrigerate specimens at 2-8°C if the assay is not completed within 8 hours.

  • Freeze samples at ≤ -20°C if the sample is not assayed within 48 hours.

  • · Freeze samples up to 5 times, and mix thoroughly after thawing.

Additional sample stability studies were performed on the MOSES clinical samples as described in DEN150009.

18. Proposed Labeling

The labeling is sufficient and it meets the requirements of 21 CFR Parts 801 and 809, as applicable and the special controls for this device under 21 CFR 866.3215.

19. Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence determination.

§ 866.3215 Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.

(a)
Identification. A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
e.g., age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
e.g., collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.