Intended Use

The Beacon Advanced CO2 Advanced Energy Laser System is intended to be used in conjunction with the OmniGuide WaveGuide Fibers or the Articulated Arm to be used for the incision, ablation, vaporization and coagulation of body soft tissues including intra-oral soft tissues.

Device Description

The Beacon Advanced Energy Laser System is a carbon dioxide laser system utilizing RF excited CO2 laser tube. The device is comprised of a system console, with or without an articulated arm terminated with a handpiece or a micromanipulator, a fiber adapter, system control electronics, a touch screen providing a graphical user interface, a covered footswitch, helium gas management system (for using the OmniGuide Waveguide fibers with the system (K140378, K093451, K093251 and K070157. Located in Appendix 6), and operating software.

The laser system delivers laser energy at 10,600 nm in three different output modes: continuous, single pulse and repeat pulse and two waveforms: Continuous Wave (CW) or Super Pulse (SP). The laser energy produced by the laser tube within the laser console is delivered through an articulated arm system terminating in a handpiece or multiple-use attachments or through a fiber adaptor coupling the laser energy into a OmniGuide. The articulating arm allows the laser energy to be delivered through a focusing handpiece or a micromanipulator achieving laser beam spot sizes in the range of 0.2-0.6 mm in the operating field. The Waveguide fibers have been cleared through K140378, K093451, K093251 and K070157.

The Beacon is operated and controlled via Graphical User Interface implemented by proprietary operating software running on single board computer in the laser console.

The operator can control various aspects of the Beacon operation through the Graphical User Interface on the touch screen. The access to the Graphical User Interface and laser operation is password protected to control access. The operator can adjust the lasing modes, pulse rate and duration in case of pulsed modes, set the waveform (CW or SP) and output power. They can also select whether to use fiber or the articulating arm with the micromanipulator or handpieces or other accessories.

Material used are mainly machined or cast aluminum, stainless steel, standard optics for the transmission or the reflection of the CO2 laser wave energy. It is worth noting that all components (mirrors, lenses, fiber and articulating arm) that the COs light travels through are passive and do not alter the wavelength or any other of the fundamental properties of the CO2 Laser.

AI/ML Overview

This document (K180993) describes the OmniGuide Beacon Advanced CO2 Laser System. The acceptance criteria and the study that proves the device meets them are not presented in the format of a typical diagnostic AI/ML device study with performance metrics like sensitivity, specificity, or AUC against a ground truth.

Instead, this device is a surgical laser system, and substantial equivalence is claimed based on similar technological characteristics, intended use, and indications for use to a predicate device, as well as adherence to established electrical safety, electromagnetic compatibility, software, and risk management standards.

Therefore, the requested information elements (1-9) which are typical for diagnostic device studies (especially those involving AI/ML and human-in-the-loop performance) are not directly applicable or available in this submission.

Here's a breakdown based on the provided document:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table with quantitative acceptance criteria for diagnostic performance (e.g., sensitivity, specificity). Instead, the "performance" is demonstrated through compliance with various international standards and the assertion of substantial equivalence to a predicate device.

Acceptance Criteria (General Categories)Reported Device Performance
Risk Analysis CompliancePerformed according to IEC 14971; reviewed by an independent third party (Intertek) and found to be in compliance.
Electrical Safety & EM CompatibilityEvaluated by Intertek and found compliant with: IEC 60601-1, IEC 62366, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-6, IEC 60601-1-2 ed 4.
Software Verification & ValidationConducted per FDA Guidance and IEC 62304:2006; results found acceptable for software release.
Non-Clinical Performance (Laser Power, Beam Quality, Durability)Evaluated through testing and analysis of laser power output and beam quality. Ability to withstand variant operation, storage, and transportation tested. System testing (energy measurements, safety controls, emission indicator, switching mechanism, fiber and articulating arm, aiming beam) completed.
BiocompatibilityNot applicable (No patient contact with device components).
Substantial Equivalence to Predicate Device (K151331)Same intended use, similar indications for use, similar technological characteristics (laser type, wavelength, device design, waveforms, pulsed mode characteristics, laser beam spot sizes). Minor differences do not raise new questions of safety or efficacy.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable. The device is a surgical laser, and its performance evaluation for regulatory approval is based on engineering testing (e.g., power output stability, electrical safety) and comparison to a predicate device, not on diagnostic data interpretation by a machine learning model. There is no "test set" in the context of diagnostic data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable. There is no "ground truth" establishment by medical experts for a diagnostic test set described in this document. The "truth" for this device's performance is compliance with engineering and safety standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. There is no test set or adjudication method described for diagnostic performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This is not a diagnostic AI/ML device that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. This is a medical device (a surgical laser), not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not applicable in the context of diagnostic performance. For hardware and software, the "ground truth" is compliance with documented specifications and international standards (e.g., the laser produces the stated power output, software functions as designed).

8. The sample size for the training set

This information is not applicable. This device does not have a "training set" in the context of machine learning.

9. How the ground truth for the training set was established

This information is not applicable. This device does not have a "training set" in the context of machine learning.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 28, 2018

OmniGuide, Inc Sam Wade Vice President, Regulatory Affairs and Quality Assurance 4 Maguire Road Lexington, Massachusetts 02421

Re: K180993

Trade/Device Name: Beacon Advanced CO2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 13, 2018 Received: April 16, 2018

Dear Sam Wade:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Jennifer R. Stevenson -53 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180993

Device Name Beacon CO2 Advanced Energy Laser System

Indications for Use (Describe)

The Beacon Advanced Energy CO2 Laser System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues in the following specialties: Dermatology, General Surgery, Gynecology, Head & Neck Surgery, Neurosurgery, Orthopedic Surgery, in soft tissue, Otolaryngology, Plastic & Reconstructive Surgery, Podiatry, Urology.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 7: 510(k) Summary

General Provisions

510(k) Owner's Name:OmniGuide, Inc.
Registration Number:3005350457
Address:4 Maguire RoadLexington, MA 02421
Contact Person:Samuel D. WadeVice President, Quality Assurance and Regulatory Affairs
Phone Number:(617) 551 4810, Cell: (978) 844-7515
Fax Number:(617) 551-8445
Classification Name:Powered Laser Surgical Instrument
Regulation:21 CFR §878.4810
Regulatory Class:II
Product Code:GEX
Proprietary Name:Beacon Advanced Energy Laser System
Common Name:Powered Laser Surgical Instrument

Name of Predicate Device(s)

  • K151331 The UltraPulse system (UltraPulse and UltraPulse Duo models, members of the ● modified Lumenis Family of UltraPulse SurgiTouch CO2 Surgical Lasers)

Name of Reference Device(s)

  • K093251 OmniGuide BeamPath®FELS 25A, CO2 Laser System ●
  • K100384 Lumenis FiberLase CO2 Laser WaveGuide ●
  • K945648 TTI Medical AccuBeam COs Laser Handpiece ●
  • K864378 TTI CO2 Laser Micromanipulator ●

Indications for Use

The Beacon Advanced Energy CO2 Laser System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues in the following specialties: Dermatology, General Surgery, Gynecology, Head & Neck Surgery, Neurosurgery, Oral Surgery, Orthopedic Surgery in soft tissue, Otolaryngology (ENT), Plastic & Reconstructive Surgery, Podiatry, Urology

Intended Use

The Beacon Advanced CO2 Advanced Energy Laser System is intended to be used in conjunction with the OmniGuide WaveGuide Fibers or the Articulated Arm to be used for the incision, ablation, vaporization and coagulation of body soft tissues including intra-oral soft tissues.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for OmniGuide. The logo consists of two concentric blue circles on the left side of the word "OmniGuide". The word "OmniGuide" is written in a sans-serif font, with the "O" in "OmniGuide" being partially enclosed by the two concentric circles. There is a trademark symbol to the right of the word "OmniGuide".

Device Description

The Beacon Advanced Energy Laser System is a carbon dioxide laser system utilizing RF excited CO2 laser tube. The device is comprised of a system console, with or without an articulated arm terminated with a handpiece or a micromanipulator, a fiber adapter, system control electronics, a touch screen providing a graphical user interface, a covered footswitch, helium gas management system (for using the OmniGuide Waveguide fibers with the system (K140378, K093451, K093251 and K070157. Located in Appendix 6), and operating software.

The laser system delivers laser energy at 10,600 nm in three different output modes: continuous, single pulse and repeat pulse and two waveforms: Continuous Wave (CW) or Super Pulse (SP). The laser energy produced by the laser tube within the laser console is delivered through an articulated arm system terminating in a handpiece or multiple-use attachments or through a fiber adaptor coupling the laser energy into a OmniGuide. The articulating arm allows the laser energy to be delivered through a focusing handpiece or a micromanipulator achieving laser beam spot sizes in the range of 0.2-0.6 mm in the operating field. The Waveguide fibers have been cleared through K140378, K093451, K093251 and K070157.

The Beacon is operated and controlled via Graphical User Interface implemented by proprietary operating software running on single board computer in the laser console.

The operator can control various aspects of the Beacon operation through the Graphical User Interface on the touch screen. The access to the Graphical User Interface and laser operation is password protected to control access. The operator can adjust the lasing modes, pulse rate and duration in case of pulsed modes, set the waveform (CW or SP) and output power. They can also select whether to use fiber or the articulating arm with the micromanipulator or handpieces or other accessories.

Material used are mainly machined or cast aluminum, stainless steel, standard optics for the transmission or the reflection of the CO2 laser wave energy. It is worth noting that all components (mirrors, lenses, fiber and articulating arm) that the COs light travels through are passive and do not alter the wavelength or any other of the fundamental properties of the CO2 Laser.

Technological Characteristics

The Beacon has the same intended use and similar indications for use, technological characteristics and operating principles as K151331 The UltraPulse system (UltraPulse and UltraPulse Duo models, members of the modified Lumenis Family of UltraPulse SurgiTouch CO2 Surgical Lasers). The design and components are very similar to the predicate devices have the Graphical User Interface on a touch screen in the front of the device has an articulating arm and an adapter port for the attachment of fibers. Both devices have the laser energy generated within the laser console and delivered to either the articulated arm or the fiber adapter. Each device has an aiming beam to assist in the proper delivery of the treatment beams but aiming beam does not deliver any therapeutic energy. Both devices have similar key design aspects including spot sizes, laser wavelength, ranges of laser operating parameters such as pulse widths and output power. The differences are minor and do not raise any new issues of safety or efficacy of the Beacon device as shown in the performance testing results.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for OmniGuide. The logo consists of the word "OmniGuide" in a simple, sans-serif font. To the left of the word is a graphic of two concentric circles, with the outer circle being slightly thicker than the inner circle. The color of the logo is a light blue.

Performance Data

Risk Analysis

Risk analysis was performed according to IEC 14971 and the FDA Guidance Document: Application of risk management to medical devices and reviewed by an independent third party (Intertek) and found to be in compliance.

Electrical Safety and Electromagnetic Compatibility and other standard testing

Evaluation of the device was conducted by an independent testing laboratory (Intertek) and was found in compliance:

  • IEC 60601-1, Medical electrical equipment, Part 1 General requirements for basic safety and essential performance.
  • IEC 62366 Medical devices- Application of usability engineering to medical devices
  • IEC 60601-2-22 Third Edition 2007-05, Medical electrical equipment-Part2: Particular ● requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
  • IEC 60825-1 Ed. 2.0 (2007), Safety of Laser Products- Part 1: Equipment classification and requirements
  • IEC 62304:2006 Medical Device software- Software life cycle processes
  • IEC 14971:2007/(R) 2010 Medical Devices- Application of Risk Management to Medical Devices
  • IEC 60601-1 Medical Electrical Equipment, Part 1-6 General requirement for Safety: Collateral ● Standard Usability
  • IEC 60601-1-2 ed 4. EMC

Non-Clinical Performance Data

The Beacon Advanced Energy CO2 Laser System performance characteristics have been evaluated through testing and analysis of laser power output and beam quality. The ability of new device to withstand variant operation, storage and transportation was tested. The system testing (energy measurements, safety controls, emission indicator, switching mechanism, fiber and articulating arm and aiming beam was also completed. This type of testing complies with the respective section of the FDA Guidance on the Content and Organization of a Premarket Notification for a Medical Laser (1995) and is similar to the predicate device tests. The performance of the Beacon Advanced Energy CO2 Laser System and related parameters of the predicate device are substantially equivalent.

Software verification and validation testing was conducted, and documentation provided as recommended by FDA's "Guidance for the Content of Premarket Submissions Contained in Medical Devices" and IEC 62304:2006 Medical Device Software. Device software verification and validation results were found acceptable for software release

Clinical Performance Data

Formal clinical trials were not deemed necessary as the device is using the same technology and intended use as the predicate device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for OmniGuide. The logo consists of the word "OmniGuide" in a simple, sans-serif font. To the left of the word is a circular graphic made up of three concentric circles. The logo is clean and professional in appearance.

Biocompatibility

No part of the laser contacts the patient therefor no biocompatibility testing was necessary

Summary of Substantial Equivalence

The Beacon Advanced Energy CO2 Laser System and the predicate device have the same intended use as surgical lasers with similar indications for use. The Beacon Advanced Energy CO2 Laser System presents similar technological characteristics as its predicate device, including the laser type, wavelength, device design, waveforms and pulsed mode characteristics, laser beam spot sizes. Although there might be minor differences in the details of the device design, they are not sufficient to raise new questions of safety or efficacy.

Conclusion

The Beacon Advanced Energy CO2 Laser System is as safe and effective as the Lumenis UltraPulse and UltraPulse Duo Laser Family (K151331). The new device has similar intended use and indications for use, technological characteristics and the same principle of operation as the predicate device. The minor differences in the indications for use do not alter the intended surgical use of the device and do not affect the safety or efficacy when used as labeled. The minor design differences raise no new issues of safety or efficacy. Thus, the Beacon Advanced Energy CO2 Laser and the Lumenis UltraPulse and UltraPulse Duo are substantially equivalent.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for OmniGuide. The logo consists of the word "OmniGuide" in a simple, sans-serif font. The word is accompanied by a circular design to the left, which is made up of concentric rings in a light blue color. There is a trademark symbol to the right of the word "OmniGuide".

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.