K Number
K180952

Validate with FDA (Live)

Device Name
Ceramic Bracket
Date Cleared
2019-06-07

(422 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ceramic Bracket is indicated for orthodontic movement of natural teeth.

Device Description

Ceramic bracket are designed to move teeth to improve their alignment. Ceramic bracket are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred position. Ceramic bracket are intended for use in affixed to a tooth so that pressure can be exerted on the teeth. Use bonding supplies to bond the bracket on to a tooth. Use bonding supplies to bond the bracket on to a tooth. Ceramic bracket are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred position.

AI/ML Overview

This is a 510(k) premarket notification for a medical device called "Ceramic Bracket." The document describes the device, its intended use, and argues for its substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a numerical or pass/fail threshold manner, as might be seen for diagnostic device performance. Instead, it relies on demonstrating compliance with relevant ISO standards and comparability to the predicate device. The device is a physical orthodontic bracket, not a diagnostic algorithm.

Therefore, I will present the performance tests conducted and the general conclusion for device acceptance.

Acceptance Criteria (Implied by Standards)Reported Device Performance
Compliance with ISO 27020 (General Orthodontic Brackets)Device meets performance specifications of ISO 27020:2010
BiocompatibilityEvaluated and tested according to ISO 10993 standards: - ISO 10993-3 (Genotoxicity, Carcinogenicity, Reproductive toxicity) - ISO 10993-5 (In vitro cytotoxicity) - ISO 10993-10 (Irritation and skin sensitization) - ISO 10993-11 (Systemic toxicity) - ISO 10993-12 (Sample preparation and reference materials)
Mechanical Function (Shear Bond Strength)Tested, results compared to ISO 6872:2015, ISO 29022:2013, ISO 27020:2010, ISO 6474-2:2012, and clinical literature data.
Mechanical Function (Torque Strength)Tested, results compared to ISO 6872:2015, ISO 29022:2013, ISO 27020:2010, ISO 6474-2:2012, and clinical literature data.
Structural IntegrityTested, results compared to ISO 6872:2015, ISO 29022:2013, ISO 27020:2010, ISO 6474-2:2012, and clinical literature data.

2. Sample size used for the test set and the data provenance

  • Sample Size: The document does not specify the exact sample sizes used for each mechanical or biocompatibility test. It indicates that "Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use" and lists the types of tests.
  • Data Provenance: Not specified in terms of country of origin. The testing was conducted to comply with international ISO standards, suggesting tests were performed in accordance with these globally recognized methodologies. The testing appears to be based on physical device samples rather than clinical data sets. It is a retrospective evaluation against established standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to this type of device and study. The "ground truth" for a physical device like an orthodontic bracket is its physical and biological performance as measured against engineering and biocompatibility standards, not an expert-annotated dataset. The standards themselves are developed by expert consensus in the relevant fields (dentistry, materials science, biocompatibility).

4. Adjudication method for the test set

Not applicable. The "test set" here refers to physical samples of the device being evaluated through standardized laboratory tests, not a dataset requiring human adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an orthodontic bracket, not an AI-powered diagnostic tool. Therefore, no MRMC study or AI assistance evaluation was performed or required.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is an orthodontic bracket, not an algorithm.

7. The type of ground truth used

The "ground truth" for this device's performance evaluation is established by:

  • International Standards: Specific ISO standards (e.g., ISO 27020, ISO 6872, ISO 29022, ISO 6474-2 for mechanical properties; ISO 10993 series for biocompatibility).
  • Clinical Literature Data: Used as a reference for comparison in mechanical testing.

Essentially, the device's characteristics are measured and compared against predefined thresholds and benchmarks set by these standards and literature, rather than against a human expert's "ground truth" for a case.

8. The sample size for the training set

Not applicable. Since this is a physical medical device and not an AI/algorithm, there is no "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 7, 2019

MEM Dental Technology Co., Ltd Anita Chen Official Correspondent 2F., No.22, Ln.31, Sec.1, Huandong Rd., Xinshi Dist. Tainan City, 74146 Tw

Re: K180952

Trade/Device Name: Ceramic Bracket Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: April 20, 2019 Received: May 6, 2019

Dear Anita Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Srinivas Nandkumar, Ph.D. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180952

Device Name Ceramic Bracket

Indications for Use (Describe)

Ceramic Bracket is indicated for orthodontic movement of natural teeth.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) summary information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92

The assigned 510(k) Number:K180952
1.Submitter
Mailing AddressMEM Dental Technology Co., Ltd
2F., No.22, Ln.31, Sec.1, Huandong Rd.,
Xinshi Dist., Tainan City 74146, Taiwan
Phone: +886 6 5053939
Establishment Registration No.: 3009011037
Contact PersonAnita Chen
Phone:+886(0) 939-855-759
Fax:+886 6 5051212
E-mail:m9104303@gmail.com
Date Prepared2019.06.06
  • Device Name 2
Proprietary Name:Ceramic Bracket
Common or Usual NameBracket, Ceramic, Orthodontic
Classification NameBracket Orthodontic Plastic Bracket (21 CFR872.5470)
Product CodeNJM
DeviceCeramic Bracket
CFR ClassificationCFR Part 872.5470
Device ClassII
Classification PanelDental
Primary Predicate
510(k) numberK160615

3

{4}------------------------------------------------

Device nameOrthodontic Ceramic Brackets 1.1
ManufacturerTOMY, INC6 tenko blg. 3-16-7midoricho, fuchu city, JP 183-0006
4 Device DescriptionCeramic bracket are designed to move teeth toimprove their alignment. Ceramic bracket are bondedto natural teeth by dental professionals to connectwith orthodontic wires to cause tooth movement to amore preferred position.Ceramic bracket are intended for use in affixed to atooth so that pressure can be exerted on the teeth. Usebonding supplies to bond the bracket on to a tooth.Use bonding supplies to bond the bracket on to atooth. Ceramic bracket are bonded to natural teethby dental professionals to connect with orthodonticwires to cause tooth movement to a more preferredposition.
5. Indication for Use:Ceramic bracket is indicated for orthodonticmovement of natural teeth.
6. TechnologicalA comparison of the device features, intended use,
Characteristics andand other information demonstrates that the
Substantial EquivalenceCeramic bracket is substantially equivalent to the
Comparison with Predicate:predicate device as summarized in Table 1.

Table 1

{5}------------------------------------------------

ManufacturerMEM Dental Technology Co., Ltd.Tomy, Inc.
Subject DevicePrimary Predicate Device
Device nameCeramic bracketOrthodontic Ceramic Brackets 1.1
510(k) NumberK180952K160615
Indication for UseCeramic bracket is indicated fororthodontic movement ofnatural teeth.Orthodontic Ceramic Brackets1.1 are indicated for orthodonticmovement of natural teeth.
Classification nameOrthodontic Plastic BracketOrthodontic Plastic Bracket
Rag Number872.5470872.5470
Product CodeNJMNJM
ClassIIII
Medical SpecialtyDentalDental
LigationNon-Self-LigatingNon-Self-Ligating
Self-LigatingSelf-Ligating
BaseMechanical Lock BaseMechanical Lock Base
SterilityNon-SterileNon-Sterile
MaterialsPolycrystalline (translucent) aluminaPolycrystalline (translucent) alumina
UtilitySingle-use onlySingle-use only
DesignArchwire slot, tiewings for ligature andidentification marks for placementHooks for ligation, for additional toothmovementMolded ceramic body with roundedcorners and e Slot to hold orthodonticwires edgesArchwire slot, tiewings for ligature andidentification marks for placementHooks for ligation, for additional toothmovementMolded ceramic body with roundedcorners and e Slot to hold orthodonticwires edges

{6}------------------------------------------------

ManufacturerMEM Dental Technology Co., Ltd.Tomy, Inc.
Subject DevicePrimary Predicate Device
Device nameCeramic bracketOrthodontic Ceramic Brackets 1.1
510(k) NumberK180952K160615
Picture of DeviceImage: Non-Self-LigatingImage: Self-LigatingImage: Non-Self-Ligating Mandibular BicuspidImage: Non-Self-Ligating for all other TeethImage: Self-Ligating Mandibular BicuspidImage: Self-Ligating for all other Teeth

7. Performance Testing

Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the device:

· ISO 27020 First edition 2010-12-15 Dentistry - Brackets and Tubes for use in Orthodontics

Ceramic Bracket.

All of the components found in the Ceramic Bracket have been used in legally marketed devices, such as the primary predicate Orthodontic Ceramic Brackets 1.1.

Biocompatibility Testing

{7}------------------------------------------------

The biocompatibility evaluation and testing of the Product name was conducted in accordance with the following standards and guidance, as recognized by the FDA:

  • . FDA Draft Guidance - Use of International Standard ISO- 10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing", dated 04-23-2013
  • . ISO 10993-3,Biological evaluation of medical device-Part 3: Test for genotoxicity, carcinogenicity and reproductive toxicity.
  • . ISO 10993-5, Biological evaluation of medical device-Part 5: Tests for in vitro cytotoxicity.
  • . ISO 10993-10. Biological evaluation of medical device-Part 10: Tests for irritation and skin sensitization.
  • ISO 10993-11, Biological evaluation of medical device-Part 11: Tests for systemic toxicity.
  • . ISO 10993-12, Biological evaluation of medical device-Part 12: Sample preparation and reference materials.

Sterilization Status

The Ceramic Bracket are provided non-sterile and are not processed prior to use.

Mechanical testing

Ceramic bracket's mechanical function including Shear Bond Strength, Torque Strength test and structure integrity were tested. The results were compared to recognized standards (ISO 6872:2015, ISO 29022:2013, ISO 27020:2010, ISO 6474-2:2012) and clinical literature data to demonstrate substantial equivalence.

No animal studies or clinical testing have been required for these devices.

8. Conclusion

Based on the intended use and/or indications for use, technological characteristics, performance testing and comparison to the predicate device, the Ceramic bracket is substantially equivalent.

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.