Intended Use

The LYFT device is intended to be used:

LED functional mode

To emit energy in the red and blue region of the spectrum, specifically to treat mild to moderate acne on the face. To emit energy in the red region of the spectrum for the treatment of periorbital wrinkles.

Micro Vibration functional mode

As an electrically powered device intended for medical purposes to relieve minor aches and pains.

Device Description

The LYFT is a hand-held, rechargeable Li-Polymer battery powered device used for the treatment of mild to moderate acne, preorbital wrinkles, and the relief of minor aches and pains. The LYFT devices combines two (2) functional modes of operation, LED and micro vibration. The LED functional mode provides narrow bandwidth spectral output in the blue (430 ± 5 nm) and red (625 ± 5 nm) ranges. The micro vibration functional mode operates at a fixed frequency of 3 MHz ± 5%. The LYFT device uses a supplied power adaptor and is rechargeable.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the device LYFT and assesses its substantial equivalence to a predicate device, Ultra Renew Plus (K132833). However, the document does not contain specific acceptance criteria, reported device performance data, results from a clinical study for performance, nor details about sample sizes, ground truth establishment, or expert qualifications that would be typically found in a clinical study report proving a device meets acceptance criteria.

The document primarily focuses on non-clinical testing for safety and technological equivalence.

Here's an analysis based on the absence of the requested information in the provided text:


1. A table of acceptance criteria and the reported device performance

No explicit acceptance criteria or reported device performance for clinical effectiveness are provided in the document. The tables present technological characteristics for comparison with the predicate device, not performance against clinical endpoints.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. The document does not describe a clinical performance study with a test set. The tests mentioned are non-clinical (e.g., in vitro, electromagnetic compatibility).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The document does not describe a clinical performance study that would require expert-established ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical performance study is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The LYFT device is a light therapy and micro-vibration device, not an AI or imaging diagnostic tool that would typically involve an MRMC study with human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The LYFT device is a physical therapeutic device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. As no clinical performance study for effectiveness is detailed, there's no mention of ground truth for such a study. The "ground truth" for the non-clinical tests would be the established ISO or IEC standards.

8. The sample size for the training set

Not applicable. The document does not describe a machine learning algorithm or a clinical study that would have a training set.

9. How the ground truth for the training set was established

Not applicable. As there's no training set mentioned, this information is not provided.


Summary of Device and Evidence Provided in the Document:

The provided document is a 510(k) summary for the LYFT device. It focuses on demonstrating substantial equivalence to a predicate device (Ultra Renew Plus K132833) rather than proving the device meets clinical performance acceptance criteria through a standalone study.

The device has two functional modes:

  • LED functional mode:
    • To emit energy in the red and blue region of the spectrum to treat mild to moderate acne on the face.
    • To emit energy in the red region of the spectrum for the treatment of periorbital wrinkles.
  • Micro Vibration functional mode:
    • As an electrically powered device intended for medical purposes to relieve minor aches and pains.

The "study that proves the device meets the acceptance criteria" in this context refers to the non-clinical testing performed to establish safety and technological equivalence, as listed in the "Summary of Non-Clinical Test Reports."

Non-Clinical Tests Performed (acting as evidence for substantial equivalence rather than clinical performance):

  • In Vitro Cytotoxicity: Performed according to ISO 10993-5 standards.
  • Skin Sensitization Test and Skin Irritation Tests: Performed according to ISO 10993-10 standards.
  • Electromagnetic Compatibility Test: Performed according to IEC 60601-1-2 standards.
  • Electrical Safety and Essential Performance Test: Performed according to IEC 60601-1; IEC 60601-2-57; IEC 60601-1-11 standards.
  • Usability: Performed according to IEC 62366-1.
  • Software Verification and Validation Test: Performed according to FDA “Guidance for Pre Market Submissions and for Software Contained in Medical Device.”

These tests demonstrate the device's safety and adherence to recognized standards, which are part of the criteria for substantial equivalence to a legally marketed predicate device. The document explicitly states: "the test results show that the subject device is substantially equivalent to the predicate devices in the market."

The comparison table (Table 1 and the subsequent table on page 5) highlights the technological similarities between LYFT and the predicate Ultra Renew Plus to support the claim of substantial equivalence.

Technological Characteristics Comparison (instead of performance criteria):

Technological CharacteristicsLYFTUltra Renew Plus (Predicate K132833)
Indication for UseFull face wrinkles and moderate to mild acne. Relieve minor aches and pains.Full face wrinkles and moderate to mild acne. Relieve minor aches and pains.
Power SourceAC to DC and Rechargeable batteryAC to DC
Wavelengths430 and 625nm450 and 650nm
Effective power for blue and red LED light0.9W and 0.8W(Not specified for predicate in this table)
Micro Vibration Function Mode3 MHz ± 5%3 MHz ± 5%
LightsourceLED/infrared bulbsLED/infrared bulbs
HandheldYESYES

The conclusion states that based on this analysis, the LYFT "raises no different questions of safety or effectiveness and has been found to be substantially equivalent to the predicate devices."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 30, 2018

Uplevity, Inc. % Marc Sanchez Regulatory Counsel Contract In-House Counsel and Consultants, LLC (d/b/a FDA Atty) 1717 Pennsylvania Ave., Ste. 1025 Washington, District of Columbia 27517

Re: K180888

Trade/Device Name: LYFT Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: August 1, 2018 Received: August 1, 2018

Dear Marc Sanchez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/ofpmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K180888

Device Name LYFT

Indications for Use (Describe) The LYFT device is intended to be used:

LED functional mode

To emit energy in the red and blue region of the spectrum, specifically to treat mild to moderate acne on the face. To emit energy in the red region of the spectrum for the treatment of periorbital wrinkles.

Micro Vibration functional mode

As an electrically powered device intended for medical purposes to relieve minor aches and pains.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

UPLEVITY, Inc.

5. 510(K) Summary

The following information is provided as required by 21 CFR 807.92 for the LYFT 510(k) premarket notification.

Applicant:UPLEVITY, Inc.40 Marshfield RdNiantic, CT 06357Establishment Registration: 3013515062
Manufacturer:Shenzhen Perfect Idea Technology Co., Ltd.4f, Building 1, Fuhua Industrial Park, Tangwei, Fuyong, Bao'anShenzhen Guangdong, 518103 CHINAEstablishment Registration: Awaiting Assignment of Registration Number
Contact:Marc C. Sanchez, Esq.Contract In-House Counsel and Consultants, LLC53516 Bickett Chapel Hill NC 27517Ph: 202.765.4491E-mail: msanchez@fdaatty.com
Date of Submission:August 29, 2018
Proprietary Name:LYFT
Common Name:Over-The-Counter Powered Light Based Laser For Acne
Regulation Number:21 CFR 878.4810
Regulatory Class:Class II
Product Code:OLP

Predicate Device(s): Ultra Renew Plus (K132833)

Device Description:

The LYFT is a hand-held, rechargeable Li-Polymer battery powered device used for the treatment of mild to moderate acne, preorbital wrinkles, and the relief of minor aches and pains. The LYFT devices combines two (2) functional modes of operation, LED and micro vibration. The LED functional mode provides narrow bandwidth spectral output in the blue (430 ± 5 nm) and red (625 ± 5 nm) ranges. The micro vibration functional mode operates at a fixed frequency of 3 MHz ± 5%. The LYFT device uses a supplied power adaptor and is rechargeable.

{4}------------------------------------------------

UPLEVITY, Inc.

Intended Use: The LYFT device is intended to be used:

  • LED functional mode
    • O To emit energy in the red and blue region of the spectrum, specifically to treat mild to moderate acne on the face.
    • O To emit energy in the red region of the spectrum for the treatment of periorbital wrinkles.
  • Micro Vibration functional mode
    • O As an electrically powered device intended for medical purposes to relieve minor aches and pains.

Summary of Non-Clinical Test Reports

The following tests were performed on the LYFT device and the test results show that the subject device is substantially equivalent to the predicate devices in the market.

In Vitro CytotoxicityISO 10993-5 for In Vitro Cytotoxicity standards
Skin Sensitization Test and SkinIrritation TestsISO 10993-10 Tests for Irritation and Skin sensitization
Electromagnetic Compatibility TestIEC 60601-1-2 standards
Electrical Safety and EssentialPerformance TestIEC 60601-1; IEC 60601-2-57; IEC 60601-1-11 standards
UsabilityIEC 62366-1
Software Verification and ValidationTestFDA “Guidance for Pre Market Submissions and for SoftwareContained in Medical Device”

Summary of Technological Similarities/Differences

The LYFT emits visible and/or infrared energy and micro vibration functional mode operates at a fixed frequency the same the Ultra Renew Plus. Both devices use similar outputs and have similar technical characteristics. The intended use is the same on both devices. There may be minor differences in labeling but this does not raise any new questions of safety or effectiveness.

Technological Characteristics
Device Name/ModelLYFTUltra Renew Plus
510(k) Numbert/b/dK132833

Table 1

{5}------------------------------------------------

UPLEVITY, Inc.

Traditional 510k Submission LYFT

Indication for UseFull face wrinklesand moderate to mildacne. Relieve minoraches and pains.Full face wrinklesand moderate tomild acne. Relieveminor aches andpains.
Power SourceAC to DC andRechargeable batteryAC to DC
Wavelengths430 and 625nm450 and 650nm
Effective power for blueand red LEd light0.9W and 0.8W
Micro Vibration FunctionMode3 MHz ± 5%3 MHz ± 5%
LightsourceLED/infrared bulbsLED/infrared bulbs
HandheldYESYES

Conclusion

Therefore, taking into consideration Table 1 for Substantial Equivalence, an analysis of safety, indications, intended uses, performance, and technological properties, the LYFT raises no different questions of safety or effectiveness and has been found to be substantially equivalent to the predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.