K Number
K180866

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-08-31

(151 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CareSens S Fit Blood Glucose Monitoring System is intended for the surement of glucose in fresh capillary whole blood samples drawn from the fingertips. The CareSens S Fit Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes.

The CareSens S Blood Glucose Test Strips are with the CareSens S Fit Blood Glucose Meters to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertip.

Device Description

The CareSens S Fit BGMS is a derivation from the same design file as it predicate device, CareSens N Premier and CareSens N Premier BT Blood Glucose Monitoring System, utilizing a new test strip. This Premarket Notification (510(k) is intended to demonstrate that the candidate devices to be marketed is safe and effective as the predicate device.

The CareSens S Fit BGMS consists of a blood glucose meter, single use test strips, and control solutions with two different glucose concentrations ("Level 1" and "Level 2" ranges). The CareSens S Fit BGMS is based on an electrochemical biosensor technology (electrochemical). The System measures the glucose level in whole blood samples using a small electrical current generated in the test strips. The following items are included in the CareSens S Fit BGMS Blood Glucose Monitoring system:

  • 1 CareSens S Fit Blood Glucose Meter
  • 10 CareSens S Blood Glucose Test Strips
  • 1 Lancing device
  • 10 Lancets
  • 1 Owner's Booklet
  • 1 Quick Reference Guide
  • 1 Battery (3.0V lithium battery)

The following items are compatible with the CareSens S Fit BGMS and are available separately.

  • CareSens S Blood Glucose Test Strips
  • CareSens S Glucose Control Solution
AI/ML Overview

The provided text describes the performance testing and acceptance criteria for the CareSens S Fit Blood Glucose Monitoring System, particularly in the context of its 510(k) submission (K180866) to the FDA.

Here's an organized breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document frequently refers to "acceptance criteria" and "satisfying" or "meeting" them, particularly referencing the FDA SMBG Guidance (2016). While specific numerical acceptance criteria for each test (e.g., CV% for precision above a certain threshold) are often implied rather than explicitly stated in a single table, the key accuracy criteria from the Method Comparison/User Performance study are provided.

Performance MetricAcceptance Criteria (Implied/Referenced)Reported Device Performance (CareSens S Fit BGMS)
Accuracy (Method Comparison/User Performance)FDA SMBG Guidance (2016) criteria for accuracy:- For glucose concentrations < 100 mg/dL: 95% of results within ± 15 mg/dL of the reference.- For glucose concentrations ≥ 100 mg/dL: 95% of results within ± 15% of the reference.In a study with 376 subjects: - 98.1% of data were within ± 15% bias relative to the reference measurement method.- 100% of data were within ± 20% bias relative to the reference measurement method.
Accuracy at Extreme Glucose ValuesAccuracy criteria in the FDA guidance for method comparison analysis."The results satisfied the accuracy criteria in the FDA guidance for method comparison analysis." (Specifically evaluated performance for glucose > 250 mg/dL and < 80 mg/dL).
Intermediate PrecisionAcceptance criteria of intermediate precision requirements of the FDA SMBG Guidance (2016).Pooled SD of concentration interval 1 and 2 were 0.8 mg/dL and 1.5 mg/dL respectively. Pooled CV of interval 3, 4 and 5 were 2.8%, 3.2% and 4.0%. "The results have met the acceptance criteria."
Within-Run PrecisionNot explicitly stated, but "meets the acceptance criteria".Confirmed to meet the acceptance criteria.
LinearityAll three strip lots' results confirmed to be in linear relationship with the reference values.All three strip lots' results were confirmed to be in linear relationship with the reference values.
Interfering SubstancesNo significant interference at FDA recommended test concentrations."Results obtained from the primary tests were sufficient to show non-significance of the presence of these interferents at all three blood glucose concentration intervals." (52 substances tested; except gentisic acid and maltose concentrations). Hemoglobin showed no effect up to 17.7 g/dL.
Hematocrit EffectsBias within ±10 mg/dL for interval 1, and within ±8% on average (individual values within ±15%) for intervals 2-5.For interval 1: bias within ±10 mg/dL. For intervals 2-5: bias within ±8% on average, all individual values within ±15% relative to the comparator method. "Confirmed that CareSens S Fit BGMS has no interference within 20% to 60% hematocrit level."
Accelerated Stability (Open Vial)Predicted in-use stability of 5.5 months minimum.All 3 lots of test strips are predicted to have an in-use stability of 5.5 months minimum.
Test Strip Stability (Before Opening)SD and CV within acceptance criteria. Bias of meter measurement value against reference results within acceptance criteria.SD and CV values were within the acceptance criteria for all three lots at all storage conditions. Bias of meter measurement value against reference results (YSI 2300 STAT Plus) was within the acceptance criteria for all 360 data points.
Test Strip Stability (After Opening)Precision and accuracy within acceptance criteria for 1+ months of use."The strip performance with regard to precision and accuracy were within the acceptance criteria." (Tested for more than 1 month, grouped by use frequency).
Extended Open Vial StabilityNot explicitly stated as a pass/fail, but a recommendation derived from the test result.At 96 hours (4 days) at 30°C - 80% RH, measurement results started to go off acceptance criteria from day 3 (72 hours). Precision also broadened from day 3. Recommendation: "strips that have been open in the air for more than 2 days should not be used especially at high temperature and humidity conditions."
Altitude TestNo systematic response or significant effect up to 10,000 feet."Confirmed that CareSens S Fit SMBG test systems have no systematic response to altitudes and is not significantly affected by lack of atmospheric oxygen up to 10,000 feet (3,048 m) above the sea level." (Tested at sea level and simulated 5,000 feet).
Operating Condition TestSystem operates normally at claimed environmental conditions compared to YSI results."It is confirmed that CareSens S Fit SMBG test system operates normally at the claimed environmental conditions, compared to the YSI results."
Flex Study (Strip Properties)Correct display of error messages for out-of-range measurements, or no effect on results.Meters correctly displayed error messages when out of range measurements were taken, or did not affect the measurement results. (Included Samples outside Measuring Range, Short Sample Perturbation Study, and Intermittent Sampling).
Cleaning and Disinfection StudyComplete inactivation of HBV, no effect on performance or materials after 5 years of use simulation.Demonstrated complete inactivation of live virus (HBV) on meter and lancing device surfaces. Demonstrated 5 years (260 cycles) of single-patient use (pre-cleaning and disinfection) has no effect on performance or external materials.

2. Sample Size Used for the Test Set and Data Provenance

  • Method Comparison/User Performance Study:
    • Sample Size: At least 350 different subjects. The specific study conducted for CareSens S Fit had 376 subjects.
    • Data Provenance: Not explicitly stated, but the submission is from Seoul, Korea (i-SENS, Inc.), suggesting the data is likely from Korea. The study is prospective as it involves subjects participating in a user evaluation of the device.
  • Within-Run Precision Evaluation: 10 meters and 500 test strips (from 10 vials of 3 manufacturing lots) for each sample concentration.
  • Intermediate Precision Evaluation: 10 meters and 500 test strips (from 10 vials of 3 manufacturing lots) for each sample concentration.
  • Linearity Evaluation: 11 evenly spaced concentrations of blood samples.
  • Hematocrit Effect Test: Five concentration levels of blood samples at nine hematocrit levels for three lots of test strips.
  • Test Strip Stability (Before Opening): Three strip lots tested with 24 distinct samples at each testing point for 462 days. Total of 360 data points.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not mention the use of human experts to establish ground truth for the test set. Instead, it relies on a laboratory comparator method.

4. Adjudication Method for the Test Set

Not applicable, as expert adjudication was not described for establishing ground truth. The ground truth was established by a laboratory reference instrument.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study, in the sense of comparing human readers with and without AI assistance, was not explicitly done. This device is a Blood Glucose Monitoring System where the "reading" is performed by the device itself, not a human interpreter of images or complex data. The "user performance" study assesses the device's accuracy in the hands of intended users, and their ability to follow instructions, not their diagnostic performance with or without computational aid.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, numerous "standalone" tests were conducted where the device's performance was evaluated against a laboratory reference. These include:

  • Within-Run Precision
  • Intermediate Precision
  • Linearity
  • Additional Accuracy at Extreme Glucose Values
  • Interfering Substance Evaluation
  • Hematocrit Effect Test
  • Various Stability Tests (Accelerated, Before Opening, After Opening, Extended Open Vial)
  • Altitude Test
  • Operating Condition Test
  • Flex Study
  • Cleaning and Disinfection Study

The Method Comparison/User Performance study combines both standalone device accuracy (in the hands of users) with usability aspects, but the core accuracy measurement is a standalone performance metric against a reference.

7. The Type of Ground Truth Used

The primary ground truth used for performance evaluation, particularly for accuracy, was:

  • Laboratory Reference Instrument: The YSI 2300 STAT Plus glucose analyzer was consistently used as the reference equipment (comparator method) for measuring glucose levels in blood samples.

8. The Sample Size for the Training Set

The document does not specify a separate training set for an algorithm in the way that would typically be described for AI/machine learning models. The CareSens S Fit BGMS is an electrochemical biosensor system, implying its performance is largely governed by its hardware and electrochemical design, rather than a machine learning algorithm trained on a large dataset. The development and calibration of such a system would involve extensive testing and refinement, but this is not typically referred to as a "training set" in the context of this type of medical device's regulatory submission.

9. How the Ground Truth for the Training Set was Established

As no specific "training set" in the AI/ML sense is mentioned, this question is not directly applicable. The device's calibration and design would be based on established glucose measurement principles and validated against reference methods during its development phase.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is the FDA acronym in blue, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.

August 31, 2018

i-SENS, Inc. Joon Jung RA Team Manager 43, Banpo-daero 28-gil, Seocho-gu Seoul. 06646 Kr

Re: K180866

Trade/Device Name: CareSens S Fit Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: March 26, 2018 Received: April 2, 2018

Dear Joon Jung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180866

Device Name

CareSens S Fit Blood Glucose Monitoring System

Indications for Use (Describe)

The CareSens S Fit Blood Glucose Monitoring System is intended for the surement of glucose in fresh capillary whole blood samples drawn from the fingertips. The CareSens S Fit Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes.

The CareSens S Blood Glucose Test Strips are with the CareSens S Fit Blood Glucose Meters to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertip.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for i-sens. The logo is in blue, with a green dot above the "i". The text is sans-serif and bolded.

i-SENS, Inc. 43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea

510(k) Summary

(As required by 21 CFR 807.92)

Introduction:This summary of 510(k) information is being submitted in accordancewith requirements of 21 CFR Part 807.92.
Type of 510(k):Traditional 510(k)Assigned 510(k) Number: K180866
Submitter Information:i-SENS, Inc.43, Banpo-daero 28-gil, Seocho-gu, Seoul, KoreaTel.) +82-2-910-0453Fax) +82-2-942-2514e-mail: jhjung@i-sens.comContact Person: Joon Ho Jung
Prepared Date:July 26, 2018
Device NameCareSens S Fit Blood Glucose Monitoring System
Classification NameBlood Glucose Test System
Product CodeClassificationRegulation SectionPanel
NBWClass II21 CFR 862.1345Clinical Chemistry 75
Blood Glucose Test System
Predicate DeviceCareSens N Premier and CareSens N Premier BT Blood Glucose MonitoringSystem (K170614).
Device DescriptionThe CareSens S Fit BGMS is a derivation from the same design file as itpredicate device, CareSens N Premier and CareSens N Premier BT BloodGlucose Monitoring System, utilizing a new test strip. This PremarketNotification (510(k) is intended to demonstrate that the candidate devices tobe marketed is safe and effective as the predicate device.
The CareSens S Fit BGMS consists of a blood glucose meter, single use teststrips, and control solutions with two different glucose concentrations("Level 1" and "Level 2" ranges). The CareSens S Fit BGMS is based on anelectrochemical biosensor technology (electrochemical). The Systemmeasures the glucose level in whole blood samples using a small electricalcurrent generated in the test strips. The following items are included in theCareSens S Fit BGMS Blood Glucose Monitoring system:
- 1 CareSens S Fit Blood Glucose Meter- 10 CareSens S Blood Glucose Test Strips- 1 Lancing device- 10 Lancets

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for i-sens. The "i" is blue with a green dot above it. The word "sens" is also in blue. The logo is simple and modern.

i-SENS, Inc. 43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea Tel: +82-2-916-6191 Fax: +82-2-942-2514

  • 1 Owner's Booklet -
  • -1 Quick Reference Guide
  • 1 Battery (3.0V lithium battery) -

The following items are compatible with the CareSens S Fit BGMS and are available separately.

  • CareSens S Blood Glucose Test Strips -
  • -CareSens S Glucose Control Solution

Intended Use CareSens S Fit BGMS:

The CareSens S Fit Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. The CareSens S Fit Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes.

The CareSens S Blood Glucose Test Strips are for use with the CareSens S Fit Blood Glucose Meters to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertip.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for i-sens. The "i" is blue with a green dot above it. The word "sens" is in blue.

i-SENS, Inc. 43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea

Comparison to the Predicate Device

The similarities and differences between the predicate device (K160742) and the candidate device are highlighted in Tables 1 and 2 below.

Table 1: Similarities between the Candidate and Predicate Device

CharacteristicPredicate DeviceCareSens N Premier Blood GlucoseMonitoring System, K170614Candidate DeviceCareSens S Fit Blood Glucose MonitoringSystem
Intended UseThe CareSens N Premier Blood GlucoseMonitoring System is intended for thequantitative measurement of glucose in freshcapillary whole blood samples drawn from thefingertips. The CareSens N Premier BloodGlucose Monitoring System is intended forself-testing outside the body (for in vitrodiagnostic use) by people with diabetes at homeas an aid to monitor the effectiveness ofdiabetes control. The system is intended to beused by a single person and should not beshared. It is not intended for use on neonatesand is not for the diagnosis or screening ofdiabetes.The CareSens N Blood Glucose Test Strips arefor use with the CareSens N Premier BloodGlucose Meter to quantitatively measureglucose in fresh capillary whole blood samplesdrawn from the fingertip.Same
Common nameSystem, test, blood glucose, over the counterSame
Test PrincipleElectro-chemical reaction. The glucose metermeasures electrical current generatedbyenzyme using the glucose as substrate insample.Same
EnzymeGlucose Oxidase- Glucose Oxidase (Aspergillus sp.): 2.7 units- Hexaamineruthenium(III) chloride: 45.7 µg- Other ingredients: 1.6 µgSame
MeasurementPrincipleAmperometric methodSame
Sample typeFresh capillary whole bloodSame
CalibrationPlasma-equivalentSame
Coding systemNo coding required (Automatic codeidentification)Same
Sample volume(μl)0.5Same
Measurement unitmg/dLSame
OperatingHumidity10~90%Same
Control LevelsTwo Levels (A and B)Same
Memory capacityUp to 1,000 test resultsSame
Test result averagerange1, 7,14, 30 and 90 days(Pre-meal, Post-meal, Fasting and Total)Same
HypoglycemiaindicatorYesSame
Expiration dateindicatorYesSame
DatacommunicationUSB cableSame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for i-sens. The logo is blue with a green dot above the i. The text is sans-serif and the dot is a perfect circle.

i-SENS , Inc. i-SENS , Inc.

43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea

Table 2: Differences between the Candidate and Predicate Device
---------------------------------------------------------------------
CharacteristicPredicate DeviceCareSens N Premier Blood GlucoseMonitoring System, K170614Candidate DeviceCareSens S Fit Blood Glucose MonitoringSystem
Test time (sec.)56
Test range (mg/dL)20-60040-600
OperatingTemperature42.8-111.2°F50-104 °F
Power SourceTwo 3.0V lithium batteries (CR2032)One 3.0V lithium battery (CR2032)
Hematocrit range(%)15-6520-60
HyperglycemiaindicatorNoYes

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for i-sens. The "i" has a green dot above it, and the rest of the word is in blue. The font is sans-serif and appears to be bolded.

i-SENS. Inc. 43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea Tel: +82-2-916-6191 Fax: +82-2-942-2514

Performance Test Summary

Within-Run Precision Evaluation

Within-run precision is determined using venous whole blood samples altered by spiking, and by allowing glycolysis. For each sample concentration, 10 meters and 500 test strips from 10 vials of 3 manufacturing lots are used. Test strips are taken from the same vial for each glucose concentration range, the mean value, standard deviation (with 95% confidence intervals) and percent CV are calculated. It was confirmed that the CareSens S Fit SMBG test system meets the acceptance criteria.

Intermediate Precision Evaluation

Intermediate precision is determined using control solutions. For each sample concentration, 10 meters and 500 test strips from 10 vials of 3 manufacturing lots are used. Test strips are taken from the same vial for each meter. For each glucose concentration range, the mean value, standard deviation (with 95% confidence intervals) and percent CV are calculated. The pooled SD of concentration interval 1 and 2 were 0.8 mg/dL and 1.5 mg/dL respectively, and pooled CV of interval 3, 4 and 5 were 2.8%, 3.2% and 4.0%. The results have met the acceptance criteria of intermediate precision requirements of the FDA SMBG Guidance (2016).

Linearity Evaluation

The linearity study includes an evaluation of 11 evenly spaced concentrations of blood samples, and the results are analyzed according to "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach", CLSI document EP6-A. Linearity test was conducted using meters with altered firmware that shows numerical response at concentrations that would normally display "Low" and "High" error messages at the two extreme ends of concentrations. The correlation coefficient, v-intercept, slope of the regression line were obtained, and by fitting a regression line to the data, the best polynomial fit of the data were determined. All of the three strip lots' results were confirmed to be in linear relationship with the reference values.

Method Comparison/User Performance

Following the FDA SMBG Guidance, at least 350 different subjects participated in the user evaluation of the Self-Monitoring Blood Glucose Test Systems in comparison to YSI 2300 STAT Plus glucose analyzer as a reference equipment. The method comparison/user evaluation study is designed for the assessment of both system accuracy in the hands of the intended users as well as other aspects to support lay-use, such as a labeling assessment and usability. A Method Comparison/User Evaluation study was conducted for the CareSens S Fit SMBG test system, in which 376 subjects participated. The measurement results demonstrated that 98.1% of data were within ± 15% and 100% of data were within ± 20% bias relative to the reference measurement method. For the questionnaire, more than 95% of subjects have given scores of 3 and above to each of the questions. It is concluded that most of the intended users will be able to understand the user manual and will be able to perform self blood glucose testing with no difficulty.

Additional Accuracy at Extreme Glucose Values Report

Additional accuracy study is conducted to provide a robust evaluation of SMBG performance in the extreme upper (greater than 250 mg/dL) and lower ends (less than 80 mg/dL) of the claimed measuring range.

For the evaluation of the CareSens S Fit SMBG Test System's performance in the extreme upper and lower

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the i-sens logo. The "i" is blue with a green dot above it. The word "sens" is in blue.

i-SENS. Inc. 43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea Tel: +82-2-916-6191 Fax: +82-2-942-2514

ends, the measurement results of modified capillary blood samples were analyzed. The results satisfied the accuracy criteria in the FDA guidance for method comparison analysis.

Interfering Substance Evaluation

The effect of potentially interfering endogenous and exogenous substances and conditions such as icterus, lipemia and the effect of common medications are tested to demonstrate that these substances do not interfere with the CareSens S Fit SMBG test system. Interference testing is performed using samples containing various glucose concentrations to evaluate clinically relevant decision points

52 potentially interfering substances were studied for their interfering effects on the blood glucose measurements of CareSens S Fit SMBG. All of the substances were tested at FDA recommended test concentrations except for gentisic acid and maltose. The results obtained from the primary tests were sufficient to show non-significance of the presence of these interferents at all three blood glucose concentration intervals. However, dose response test was performed for Hemoglobin and the maximum concentration at which these substances have no effect on the CareSens S Fit SMBG test system was 17.7 g/dL.

Hematocrit Effect Test

The hematocrit interference is tested at five glucose intervals. Specific percentages of hematocrit is achieved by manipulating the plasma to packed cell ratio following centrifugation. Hematocrit levels tested span the claimed range in 5% intervals.

Five concentration levels of blood samples at nine hematocrit levels were tested for three lots of test strips. For all of the nine samples in interval 1, the bias of measurement was within ±10 mg/dL. For all of the 36 samples in intervals 2 ~ 5, the bias of measurements was within ±8% on average, all individual values were within ±15% relative to the comparator method. Hence, it has been confirmed that CareSens S Fit BGMS has no interference within 20% to 60% hematocrit level.

Accelerated Stability Test

The purpose of the accelerated stability test using open vial test strips is to estimate the lifetime of test strips after opening the vials. The strip vials are stored at several high temperatures, and the vials are opened at every testing point. The Arrhenius equation is used to calculate for the stability test protocol was designed from documentation CLSI EP25-A. 'Evaluation of Stability of In Vitro Diagnostic Reagents: Approved Guideline. All 3 lots of test strips are predicted to have in-use stability of 5.5 months minimum. The real-time stability test for open vial strip lots is ongoing and the shelf life study will be conducted for more than 25 months.

Test Strip Stability Test (Before Opening)

The purpose of shelf-life test before opening vial (shelf life stability) is to study the stability of test strip under recommended storage conditions before the strip bottles have been opened. Precision and accuracy evaluation is conducted once in every 1~3 month's interval using for 25 months. At every testing point, strip vials are removed from the designated storage conditions. Accuracy evaluation is performed and analyzed by computing measurement bias from the reference values (YSI 2300 STAT Plus glucose analyzer) and precision study is analyzed by computing the SD and CV. Precision evaluation by control solution is conducted over 5 days.

The test strip stability study was conducted for 462 days (before opening vial) using three strip lots at five storage

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for i-sens. The logo is blue, except for two green dots that are part of the "i". The text is sans-serif and the logo is simple and modern.

i-SENS, Inc. 43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea Tel: +82-2-916-6191 Fax: +82-2-942-2514

conditions. Precision study was assessed by evaluating the SD and CV using blood samples measurement values as well as control solutions measurement values. The SD and CV values were within the acceptance criteria for all the three lots at all storage conditions. Accuracy study was also conducted. At each testing point, 24 distinct samples were measured by CareSens S Fit SMBG test system and for all of the 360 data points, the bias of meter measurement value against the reference results (YSI 2300 STAT Plus) was within the acceptance criteria.

Test Strip Stability Test (After Opening)

The purpose of shelf-life test after opening vial (in-use stability) is to study the stability of test strip under recommended storage conditions after the strip bottles have been opened. At every testing point, strip vials are removed from the designated storage conditions. Accuracy evaluation is performed and analyzed by computing measurement bias from the reference values (YSI 2300 STAT Plus glucose analyzer) and precision study is analyzed by computing the SD and CV. Precision evaluation by control solution is conducted over 5 days.

Stability of open vial test strip have been conducted by grouping the strip use frequency into the four categories. The tests have been conducted for more than 1 month and, the strip performance with regard to precision and accuracy were within the acceptance criteria. The test for open vial strip lots is ongoing

Extended Open Vial Stability Evaluation

Strip performance assessment mimics the use of test strips from vials that have been left completely open when stored at different combinations of temperature and humidity spanning the recommended storage of extended open vial strip stability test is to determine the stability duration of test strips with the cap vial left off for an extended period of time. At every time points, precision evaluation is performed using control solutions and whole blood samples. Accuracy is also evaluated using spiked or glycolyzed whole blood samples to span the claimed measuring range, and the SMBG results be compared to values obtained with the comparator method.

Extended open vial stability test was conducted for 96 hours (4 days) at 5 combined temperature and humidity conditions. For the accuracy as the exposure time increased, the measurement results tended to result in higher measurement results. The measurement results started to go off the acceptance criteria starting from day 3 (72 hours) at 30℃ - 80% RH conditions. For the precision assessment as well, the closeness of data points started to broaden from day 3. Therefore, it is recommended that the strips that have been open in the air for more than 2 days should not be used especially at high temperature and humidity conditions.

Altitude Test

High altitude often involves extremes of temperature and humidity and can result in changes to hematocrit and blood pressure. The altitude effects study compares results from whole blood samples with the SMBG results at different high-altitude conditions relative to the comparator method. The studies also include a pressure change. Altitude pressure changes are accomplished by simulating increasing altitudes and atmospheric conditions in a pressurized chamber.

The altitude effects test was conducted at sea level (0 feet), at simulated altitudes of 5,000 feet. It was confirmed that CareSens S Fit SMBG test systems have no systematic response to altitudes and is not significantly affected by lack of atmospheric oxygen up to 10,000 feet (3,048 m) above the sea level.

Operating Condition Test

Measurements are made on whole blood samples under various operating temperature and humidity conditions. Tested temperature and humidity ranges not only cover the operating ranges specified in the device labeling, but

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for i-sens. The logo is written in a sans-serif font, with the "i" in green and the rest of the word in blue. There is a green dot above the "i" and another green dot to the left of the "s".

i-SENS, Inc. 43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea Tel: +82-2-916-6191 Fax: +82-2-942-2514

also the ranges outside of the claimed operating ranges. Meter measurement results are compared to values obtained with the reference method (YSL). Operating conducted using meters with altered firmware that shows numerical response at temperature conditions that would normally display "Er3" error messages outside of operating conditions. It is confirmed that CareSens S Fit SMBG test system operates normally at the claimed environmental conditions, compared to the YSI results.

Altitude Effects Evaluation

High altitude often involves extremes of temperature and humidity and can result in changes to hematocrit and blood pressure. The altitude effects study compares results from whole blood samples with the SMBG results at different high-altitude conditions relative to the comparator method. The studies also include a pressure change. Altitude pressure changes are accomplished by simulating increasing altitudes and atmospheric conditions in a pressurized chamber. It was confirmed that CareSens S Fit SMBG test systems have no systematic response to altitudes and is not significantly affected by lack of atmospheric oxygen up to 10,000 feet (3,048 m) above the sea level.

Flex Study (Strip Properties)

Flex studies are designed to test some of the factors that may contribute to erroneous results when used in home use settings rather than in professional healthcare settings. The SMBG is also demonstrated through operating conditions test, altitude effects test, and stability tests. Other potential sources of errors are tested in this set of tests including Samples outside the Measuring Range. Short Sample Perturbation Study, and Intermittent Sampling. In all of the tests, the meters correctly displayed Error messages when out of range measurements were taken, or did not affect the measurement results.

Cleaning and Disinfection Study

The device is intended for single patient home use. Disinfection studies were performed on the surface areas of meter and lancing device by an outside commercial testing service to evaluate effectiveness of disinfectant, CLOROX GERMICIDAL Wipes (EPA Reg. No: 67619-12), in preventing the spread of blood-borne pathogens, using hepatitis B virus (HBV). The results demonstrated complete inactivation of live virus inoculated on the materials of the meter and lancing device. We have also demonstrated that 5 years (260 each of pre-cleaning and disinfection cycles for meter and lancing device) of single patient device use has no effect on the performance and the external materials of the meter and lancing device.

Data demonstrating substantial equivalence

The candidate device was tested in accordance with FDA Guidance for Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use. Analytical performance testing included method comparison, repeatability, and intermediate precision testing, etc. A user performance evaluation assessed accuracy of results and usability of the device in the hands of intended users. The candidate device performed similarly to both the predicate device as well as to a laboratory comparator method, the Yellow Springs Instrument (YSI). All testing demonstrated safety and effectiveness of the candidate device and substantial equivalence to the predicate device. Therefore, there are no substantive differences between the products defined in this 510(k) submission and the predicate device.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for i-sens. The "i" is blue with a green dot above it. The rest of the word, "-sens", is also in blue. The logo is simple and modern.

i-SENS, Inc. 43, Banpo-daero 28-gil, Seocho-gu 06646 Seoul, Korea Tel: +82-2-916-6191 Fax: +82-2-942-2514

Based on the submitted information in this premarket notification, the candidate Conclusion devices are substantially equivalent to the predicate device. Further, the candidate devices have met the performance, safety, and effectiveness of the device for its intended use.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.