K Number
K180835

Validate with FDA (Live)

Date Cleared
2019-02-08

(315 days)

Product Code
Regulation Number
862.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intended for the in vitro quantitative measurement of acetaminophen in serum, lithium heparin plasma. Measurement of acetaminophen is used in the diagnosis and treatment of acetaminophen overdose toxicity.

Device Description

The SEKURE Acetaminophen L3K assay is an enzymatic, spectrophotometric assay for the measurement of acetaminophen concentration in serum, lithium heparin plasma and sodium heparin plasma. The assay consists two working reagents, an enzyme reagent and a color reagent. The enzyme reagent contains acyl amidohydrolase, which cleaves the amide bond of the acetaminophen, forming p-aminophenol which then reacts with the 2,5- dimethylphenol (contained the color reagent) in the presence of manganese. The product of that reaction causes increased absorbance at 605 nm which is directly proportional to the acetaminophen concentration in the sample. Testing is performed on open system clinical chemistry analyzers, such as the Hitachi 717 (K872494) in conjunction with a calibrator (510(k) exempt) which is included and controls which are provided separately.

AI/ML Overview

The provided document is a 510(k) Pre-market Notification for a medical device called the "SEKURE Acetaminophen L3K Assay." This document describes the analytical studies conducted to demonstrate the device's performance and substantial equivalence to a predicate device, rather than a study involving human readers or AI assistance in image interpretation. Therefore, many of the requested points related to AI, human reader performance, expert consensus, and MRMC studies are not applicable to this type of device submission.

Here's an analysis based on the available information:

Device Type: In vitro diagnostic device (IVD) for quantitative measurement of acetaminophen in biological samples.

Focus of the Document: Analytical performance studies (precision, analytical sensitivity, linearity, interference, method comparison, matrix comparison) to demonstrate the assay's accuracy and reliability.

Information NOT Applicable to this Document Type:

  • Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective) - Data provenance is typically not detailed for IVD analytical studies in this manner; samples are clinical specimens or prepared materials.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience) - Ground truth for IVDs is established by reference methods or gravimetric preparation, not expert review of images.
  • Adjudication method (e.g. 2+1, 3+1, none) for the test set - Not applicable for IVD analytical studies.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance - Not applicable; this is not an AI-assisted diagnostic imaging device.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done - Not applicable; this is not an AI algorithm.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc) - Ground truth is established by reference methods, gravimetric preparation, or theoretical values.
  • The sample size for the training set - Not applicable; this device does not use an AI training set.
  • How the ground truth for the training set was established - Not applicable.

Acceptance Criteria and Reported Device Performance

The device is an in vitro diagnostic assay, and its performance is evaluated through various analytical studies. The acceptance criteria are therefore analytical performance specifications, not clinical outcomes directly.

Here's a table summarizing the acceptance criteria and a selection of reported performance data from the document:

Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance CriteriaReported Device Performance (Example Data)Pass/Fail
Precision (Within Laboratory %CV)≤ 4% CV (for Control 1, Lot 1 example)Lot 1, Control 1: 3.8% (Acetaminophen 68 µmol/L)Pass
Lot 1, Unaltered P1: 1.4% (Acetaminophen 170.3 µmol/L)Pass
Limit of Blank (LoB)Maximal value across three lots. Specific criteria not explicitly stated but implied by calculation.1.0 µmol/L (maximal value across 3 lots)Pass
Limit of Detection (LoD)Maximal value across three lots. Specific criteria not explicitly stated but implied by calculation.2.4 µmol/L (maximal value across 3 lots)Pass
Limit of Quantitation (LoQ)Lowest acetaminophen concentration at which %TE was ≤25%8 µmol/L (based on lowest concentration where Total Error (TE) for each lot was ≤25%, e.g., Lot 1 @ 8umol/L was 20.41% TE)Pass
Linearity/Assay Reportable RangeDeviation ±10% from theoretical valuesFor 100 µmol/L (theoretical), Lot 1: -1.2% deviation; For 2500 µmol/L, Lot 1: 0.6% deviation. All reported deviations within ±10% across lots and concentrations.Pass
Analytical Specificity (Interference) - EndogenousSignificant interference defined as percent difference > ±10% or 8 µmol/L from controlHemoglobin: No significant interference up to 1000 mg/dL (155 µmol/L) at various acetaminophen levels. Conjugated Bilirubin: No significant interference up to 40 mg/dL.Pass
Analytical Specificity (Interference) - Exogenous DrugsSignificant interference defined as percent difference > ±10% or 8 µmol/L from controlTheophylline: No significant interference at 222 µmol/L. Salicylate: No significant interference at 4.34 mmol/L.Pass
Method Comparison (Slope)1.0 ± 0.1 (comparing to predicate device)Lot A: 0.974 (Deming), 0.975 (Passing-Bablok); Lot B: 0.984 (Deming); Lot C: 1.009 (Deming). All within the target range.Pass
Method Comparison (% Bias)± 5.0% (comparing to predicate device)Lot A: -2.27%; Lot B: -1.38%; Lot C: 1.55%. All within the target range.Pass
Method Comparison (Correlation Coefficient)≥ 0.975 (comparing to predicate device)Lot A: 0.99999; Lot B: 0.99992; Lot C: 0.9998. All met the criterion.Pass
Matrix Comparison (Slope vs. Serum)1.0 ± 0.1SST: 1.012; Lithium Heparin: 1.015; PST: 1.009; Sodium Heparin: 1.012; Barricor: 1.004. All met the criterion.Pass
Matrix Comparison (% Bias vs. Serum)± 5.0%SST: 1.1%; Lithium Heparin: 1.2%; PST: 0.8%; Sodium Heparin: 1.1%; Barricor: 0.2%. All met the criterion.Pass
Matrix Comparison (Correlation Coefficient vs. Serum)≥ 0.975All tested tubes (SST, Lithium Heparin, PST, Sodium Heparin, Barricor) showed a correlation coefficient of 1.000.Pass

Study that Proves the Device Meets the Acceptance Criteria

The study that proves the device meets the acceptance criteria is a series of Non-Clinical Performance Data studies, as detailed in the "510(k) Summary" document. These are analytical studies, typically following CLSI (Clinical and Laboratory Standards Institute) guidelines, to characterize the performance of the in vitro diagnostic assay.

1. Sample Sized Used for the Test Set and Data Provenance:

  • Precision: 80 measurements per sample/control per lot (assayed in duplicate twice a day for 20 days). This included two unaltered patient serum samples, two spiked patient serum samples, and three levels of controls. Data is likely from laboratory samples, not specified by country.
  • Analytical Sensitivity (LoB/LoD): 60 measurements per lot (five blank samples and five low concentration samples in quadruplicate over three operating days).
  • Analytical Sensitivity (LoQ): 40 replicates per low concentration sample per lot (five low concentration samples tested in 40 replicates over five runs across three operating days).
  • Linearity/Assay Reportable Range: 4 replicates per sample per lot (nine internally prepared samples across the measuring range).
  • Analytical Specificity (Interference): 5 replicates per interferent concentration per acetaminophen concentration (tested at 3-4 acetaminophen concentrations).
  • Method Comparison: 105 patient specimens, tested in duplicate, over seven operating days. Samples were distributed evenly throughout the assay range.
  • Matrix Comparison: 40 matched sets of patient specimens.

Data Provenance: The document does not explicitly state the country of origin for patient samples or whether they were retrospective or prospective. However, for analytical performance, samples are typically collected from a pool of patient samples or are internally prepared clinical matrix materials.

2. Number of Experts and Qualifications:

  • Not applicable. This device is an analytical chemistry assay, not one that relies on human experts for interpretation or ground truth establishment. The performance is assessed against quantitative analytical standards and a predicate device.

3. Adjudication Method:

  • None. Not applicable for analytical performance studies of an IVD assay.

4. MRMC Comparative Effectiveness Study:

  • No. Not applicable as this is not an imaging device or an AI-assisted diagnostic device.

5. Standalone Performance:

  • Yes (in the context of an IVD). The performance data presented (e.g., precision, analytical sensitivity, linearity, interference) represent the standalone performance of the SEKURE Acetaminophen L3K Assay itself, without a "human-in-the-loop" in the artificial intelligence sense. The assay quantitatively measures acetaminophen concentration.

6. Type of Ground Truth Used:

  • Reference Methods / Gravimetric Preparation / Theoretical Values / Predicate Device Comparison.
    • Precision: Relative to the mean of repeat measurements.
    • Analytical Sensitivity: Derived statistically from blank and low-level samples.
    • Linearity: Compared to theoretical concentrations of gravimetrically prepared standards.
    • Interference: Compared to control samples without interferent.
    • Method Comparison: Compared to measurements obtained using a legally marketed predicate device (SEKURE Acetaminophen L3K Assay, K081938). This is a common form of "ground truth" for demonstrating substantial equivalence for IVDs.
    • Matrix Comparison: Compared to serum samples (considered the reference matrix).

7. Sample Size for the Training Set:

  • Not applicable. This device is an in vitro diagnostic assay, not an AI algorithm that requires a training set.

8. How the Ground Truth for the Training Set Was Established:

  • Not applicable. See above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

February 8, 2019

Sekisui Diagnostics PEI Inc. Pennv White Regulatory Affairs Manager 70 Watts Avenue Charlottetown, PE Canada, C1E 2B9

Re: K180835

Trade/Device Name: SEKURE Acetaminophen L3K Assay Regulation Number: 21 CFR 862.3030 Regulation Name: Acetaminophen test system Regulatory Class: Class II Product Code: LDP Dated: December 21, 2018 Received: December 26, 2018

Dear Penny White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180835

Device Name SEKURE Acetaminophen L3K Assay

Indications for Use (Describe)

Intended for the in vitro quantitative measurement of acetaminophen in serum, lithium heparin plasma. Measurement of acetaminophen is used in the diagnosis and treatment of acetaminophen overdose toxicity.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR §807.92. This is a Traditional 510(k).

The assigned 510(k) number:K180835
Applicant Information and Date [807.92(a)(1)]
Applicant Name and Address:SEKISUI DIAGNOSTICS P.E.I. INC.
70 Watts Avenue, Charlottetown, PE
Canada, C1E 2B9
Establishment Registration Number: 8020316
Application correspondent:Penny White
Regulatory Affairs Manager
902-628-0934
Penny.white@sekisuidiagnostics.com

Date Summary prepared:

February 8, 2019

Device Name and Classification [807.92(a)(2)]

Trade NameCommon NameClassification NameClassificationProductCode
SEKUREAcetaminophenL3K AssayAcetaminophenAcetaminophen TestSystemClass II21 CFR 862.3030LDP

Identification of Legally Marketed Predicate Devices [807.92(a)(3)]

Predicate DevicePredicate 510(k) Number
SEKURE Acetaminophen L3K AssayK081938

{4}------------------------------------------------

Device Description [807.92(a)(4)]

Trade NameDevice Description
SEKUREAcetaminophenL3K AssayThe SEKURE Acetaminophen L3K assay is an enzymatic, spectrophotometric assay for the measurement of acetaminophen concentration in serum, lithium heparin plasma and sodium heparin plasma. The assay consists two working reagents, an enzyme reagent and a color reagent. The enzyme reagent contains acyl amidohydrolase, which cleaves the amide bond of the acetaminophen, forming p-aminophenol which then reacts with the 2,5- dimethylphenol (contained the color reagent) in the presence of manganese. The product of that reaction causes increased absorbance at 605 nm which is directly proportional to the acetaminophen concentration in the sample. Testing is performed on open system clinical chemistry analyzers, such as the Hitachi 717 (K872494) in conjunction with a calibrator (510(k) exempt) which is included and controls which are provided separately.

Intended Use [807.92(a)(5)]

For the in vitro quantitative measurement of acetaminophen in serum, lithium heparin plasma and sodium heparin plasma. Measurement of acetaminophen is used in the diagnosis and treatment of acetaminophen overdose toxicity.

{5}------------------------------------------------

CharacteristicPredicate DeviceModified Device
SEKURE Acetaminophen L3K AssaySEKURE Acetaminophen L3K Assay
Intended UseFor the in vitro quantitative measurementof acetaminophen in serum and plasma.Measurement of acetaminophen is usedin the diagnosis and treatment ofacetaminophen overdose toxicity.SimilarFor the in vitro quantitative measurementof acetaminophen in serum, lithiumheparin plasma and sodium heparinplasma. Measurement of acetaminophenis used in the diagnosis and treatment ofacetaminophen overdose toxicity.
PlatformHitachi 717 (K872494)SAME
MethodologyEnzymatic (Acyl amidohydrolase) andSpectrophotometric (2,5-dimethylphenolchromophore)SAME
Specimen typeSerum and lithium heparin plasmaSimilarSerum, lithium heparin plasmaandsodium heparin plasma
Measuringinterval0.6 to 377.5 µg/mL (4 to 2500 µmol/L)Similar17.4 to 377.5 µg/mL (115 to 2500µmol/L)

Technological Similarities and Differences to the Predicate [807.92(a)(6)]

{6}------------------------------------------------

Summary of Non-Clinical Performance Data [807.92 (b)(1), 807.92 (b)(3)|

Precision

Testing was conducted according to CLSI EP05-A3. Two unaltered patient serum samples, two spiked patient serum samples and three levels of controls were assayed in duplicate twice a day for 20 days on the Hitachi 717 with daily calibration. Each sample was evaluated using three lots of Acetaminophen L3K Assay. The mean, standard deviation and coefficient of variation (%CV) were calculated using EP Evaluator. The acetaminophen data in umol/L are summarized in the following table.

LotSampleNMean Acetaminophen (µmol/L)RepeatabilityWithin LaboratoryAcceptance CriteriaPass/Fail
1Control 180681.82.72.53.8≤ 4% CVPass
1Control 2802384.41.96.1Pass
1Control 38076413.41.814.31.9Pass
4Unaltered P180170.32.11.22.41.4Pass
4Unaltered P280195.52.41.22.61.4Pass
4Spiked 1801295.911.50.919.71.5Pass
4Spiked 2802278.617.30.851.02.2Pass
2Control 180681.11.72.43.5Pass
2Control 2802383.91.75.82.4Pass
2Control 38076414.21.914.21.9Pass
5Unaltered P180169.51.50.92.11.2Pass
5Unaltered P280195.42.61.32.61.3Pass
5Spiked 1801294.312.81.021.41.7Pass
5Spiked 2802281.515.90.752.32.3Pass
3Control 180681.42.02.43.5Pass
3Control 2802384.01.65.82.4Pass
3Control 38076413.61.714.21.8Pass
6Unaltered P180175.01.60.92.01.2Pass
6Unaltered P280202.02.71.33.01.5Pass
6Spiked 1801339.313.81.022.51.7Pass
6Spiked 2802357.917.50.750.02.1Pass

{7}------------------------------------------------

Analytical Sensitivity

Limit of Blank (LoB) and Limit of Detection (LoD)

Limit of Blank and Limit of Detection testing was based on CLSI EP17-A2. Testing was conducted using three lots of SEKURE Acetaminophen L3K reagent and two lots of calibrator on the Hitachi 717 analyzer. Testing involved analyzing five blank samples and five low concentration samples in quadruplicate over three operating days, producing a total of 60 measurements on each lot of reagent. The LoB for each lot was determined using a non-parametric approach to determine the rank position value at the 95th percentile. The LoD for each lot was determined using a parametric approach using the pooled SDs across low level samples.

LotLimit of Blank(µmol/L)Limit of Detection(µmol/L)
11.02.38
21.01.95
31.02.23

The stated LoB and LoD are the maximal values across the three lots. The LoB is 1.0 µmol/L and the LoD is 2.4 umol/L.

Limit of Quantitation (LoQ)

Limit of Quantitation testing was based on CLSI EP17-A2. Testing was conducted using three lots of SEKURE Acetaminophen L3K reagent and two lots of calibrator on the Hitachi 717 analyzer. Five low concentration samples were tested in forty replicates over five runs across three operating days on each reagent lot. The stated LoQ is the highest value across the three lots. The mean, standard deviation, bias and %TE (calculated using the Westgard model) were determined for each low level sample.

TheoreticalAcetaminophen(µmol/L)NLot 1Lot 2Lot 3Total ErrorLimit
Mean(µmol/L)TotalError (%)Mean(µmol/L)TotalError (%)Mean(µmol/L)TotalError (%)
124012.4815.8911.9316.3612.4515.04
104010.5318.839.9019.0010.4318.49
8408.2820.417.7023.527.9022.28≤25%
6406.0522.125.5334.055.9021.35
4404.0025.323.3554.733.8035.38

The LoQ for each reagent lot was determined as the lowest acetaminophen concentration at which the %TE was ≤25%. The claimed LoQ for the assay is the maximal value obtained across the three lots. The LoQ claimed is 8 umol/L.

{8}------------------------------------------------

Linearity/Assay Reportable Range

A linearity study was performed based on guidance from CLSI EP06-A. Linearity testing was conducted using three lots of SEKURE Acetaminophen L3K reagent and two lots of calibrator. Internally prepared linearity assessment material comprised nine samples from 100 to 2700 umol/L acetaminophen. Each sample was assayed in quadruplicate on each reagent lot. The mean and deviation of the measured mean from theoretical values were calculated using EP Evaluator.

TheoreticalAcetaminophen(µmol/L)NLot 1Lot 2Lot 3AcceptanceCriteriaPass/Fail
Observed(µmol/L)Deviation(%)Observed(µmol/L)Deviation(%)Observed(µmol/L)Deviation(%)
100498.8-1.2%99.3-0.7%101.31.3%±10%Pass
1154116.31.1%115.50.4%118.53.0%Pass
1254125.30.2%124.5-0.4%128.83.0%Pass
2504251.00.4%251.50.6%260.34.1%Pass
5004489.5-2.1%493.0-1.4%509.82.0%Pass
100041015.81.6%1022.82.3%1048.84.9%Pass
2000420180.9%2020.01.0%2079.54.0%Pass
2500425150.6%2501.30.0%2587.33.5%Pass
270042690.3-0.4%2686.5-0.5%2784.53.1%Pass

The linearity data met acceptance criteria and support an assay measuring range up to 2500 umol/L.

Traceability, Stability, Expected Values (controls calibrators or methods)

Value Assignment

The SEKURE Acetaminophen Calibrator is prepared gravimetrically from USP grade reference acetaminophen material to a concentration of 151 ug/mL (1000 umo/L). The calibrator preparation is verified against the master calibrator (an internal reference standard). Two lots of Acetaminophen Calibrator were verified and met acceptance criteria.

Stability

Shelf-life claims based on testing performed according to CLSI EP25-A. Stability testing supports a shelf life of 12 months.

{9}------------------------------------------------

Analytical Specificity (Endogenous and Exogenous Interference)

Endogenous Interference

An interference study was performed based on CLSI EP07-A2 to assess common or known substances that could interfere with the Acetaminophen L3K Assay. Interference testing was conducted using three lots of SEKURE Acetaminophen L3K reagent. All potential interferents were tested at acetaminophen concentrations of 33, 100, 199 and 900 umol/L in replicates of five. Significant interference was identified as a percent difference greater than ±10% or 8 µmol/L from control.

Substance TestedHighest Tested Concentration with NoSignificant InterferenceAcetaminophen Concentration
Hemoglobin200 mg/dL (31 µmol/L)4.5 µg/mL (30 µmol/L)
Hemoglobin200 mg/dL (31 µmol/L)14.0 µg/mL (93 µmol/L)
Hemoglobin400 mg/dL (62 µmol/L)28.4 µg/mL (188 µmol/L)
Hemoglobin1000 mg/dL (155 µmol/L)130 µg/mL (861 µmol/L)
Conjugated Bilirubin2 mg/dL (23.7 µmol/L)4.7 µg/mL (31 µmol/L)
Conjugated Bilirubin2 mg/dL (23.7 µmol/L)14.8 µg/mL (98 µmol/L)
Conjugated Bilirubin32 mg/dL (364 µmol/L)17.4 µg/mL (115 µmol/L)
Conjugated Bilirubin40 mg/dL (475 µmol/L)30.2 µg/mL (200 µmol/L)
Conjugated Bilirubin40 mg/dL (475 µmol/L)144 µg/mL (957 µmol/L)
UnconjugatedBilirubin2 mg/dL (34.2 µmol/L)5.0 µg/mL (33 µmol/L)
UnconjugatedBilirubin4 mg/dL (68 µmol/L)14.8 µg/mL (98 µmol/L)
UnconjugatedBilirubin32 mg/dL (547 µmol/L)17.4 µg/mL (115 µmol/L)
UnconjugatedBilirubin24 mg/dL (410 µmol/L)30.2 µg/mL (200 µmol/L)
UnconjugatedBilirubin40 mg/dL (475 µmol/L)144 µg/mL (958 µmol/L)
Ascorbic Acid3000 µg/dL (170 µmol/L)4.5 µg/mL (30 µmol/L)
Ascorbic Acid3000 µg/dL (170 µmol/L)16.2 µg/mL (107 µmol/L)
Ascorbic Acid3000 µg/dL (170 µmol/L)32.5 µg/mL (215 µmol/L)
Ascorbic Acid3000 µg/dL (170 µmol/L)142.8 µg/mL (946 µmol/L)
N-Acetylcysteine1500 mg/L4.5 µg/mL (30 µmol/L)
N-Acetylcysteine1500 mg/L16.6 µg/mL (110 µmol/L)
N-Acetylcysteine1500 mg/L32.5 µg/mL (215 µmol/L)
N-Acetylcysteine1500 mg/L139.2 µg/mL (922 µmol/L)
Intralipid600 mg/dL [1800 mg/dL (20 mmol/L)Simulated Triglycerides]4.8 µg/mL (32 µmol/L)
Intralipid1000 mg/dL [3000 mg/dL (34 mmol/L)Simulated Triglycerides]15.1 µg/mL (100 µmol/L)
Intralipid1000 mg/dL [3000 mg/dL (34 mmol/L)Simulated Triglycerides]30.7 µg/mL (203 µmol/L)
Intralipid1000 mg/dL [3000 mg/dL (34 mmol/L)Simulated Triglycerides]135.6 µg/mL (898 µmol/L)

{10}------------------------------------------------

Exogenous Drug Interference

Potential drug interference testing was performed based on CLSI EP07-A2 to assess common or known drugs that could interfere with the Acetaminophen L3K Assay. All potential drug interferents were tested at acetaminophen concentrations of approximately 33, 100, 199 and 900 umol/L in replicates of five. Significant interference was identified as a percent difference greater than ±10% or 8 umol/L from control.

Substance TestedConcentration with No Significant Interference
Theophylline222 µmol/L
Phenylbutazone2.89 mmol/L
Ibuprofen2425 µmol/L
Imipramine2.5 µmol/L
Acetylsalicylic Acid6.51 mmol/L
Levodopa25.3 µmol/L
Ampicillin152 µmol/L
Doxycycline67.5 µmol/L
Amitriptyline3.61 µmol/L
Metronidazole701 µmol/L
Cefoxitin1546 µmol/L
Cyclosporin10.0 µmol/L
Methyldopa71 µmol/L
Rifampicin78.1 µmol/L
Salicylate4.34 mmol/L

Summary of Method Comparison

Method Comparison with Predicate Device

Method comparison testing was conducted based on CLSI EP09-A3 on the Roche Hitachi 717 analyzer using the modified SEKURE Acetaminophen L3K reagent a as compared to the predicate Acetaminophen L3K reagent. Testing was completed on 105 patient specimens in duplicate over seven operating days. The samples were distributed evenly throughout the assay range (115 umol/L - 2500 umol/L, 17.4 ug/mL -377.5 ug/mL). Each sample was tested using three lots of the modified reagent and one lot of the predicate reagent.

The SEKURE Acetaminophen L3K assay is designed to have a slope of 1.0 ± 0.1 and a correlation coefficient (t) of ≥ 0.975 when compared to the predicate method with Deming regression.

{11}------------------------------------------------

Method Comparison Summary

LotSampleRangeSlope% BiasCorrelation Coefficient
TargetDemingPassingBablokPass/FailTargetResultPass/FailTargetResultPass/Fail
ಗಿ115.5 -2446.5umol/L1.0±0.10.9740.975Pass± 5.0%-2.27%Pass> 0.9750.99999Pass
B0.9840.988Pass-1.38%Pass0.99992Pass
C1.0091.009Pass1.55%Pass0.9998Pass

Matrix Comparison

A matrix comparison study was performed to qualify plasma as a specimen for analysis with the SEKURE Acetaminophen L3K Assay. The study was performed using one lot of reagent and one lot of calibrator on one Hitachi 717 instrument. The acetaminophen concentrations spanned the assay range. A set of forty matched sets were collected and tested.

The following tube types were compared to serum tubes (without gel in plastic) and recovered within the stated acceptance criteria and are therefore suitable specimens for analysis:

  • Serum with gel in plastic (SST)
  • Lithium heparin with gel in plastic (PST)
  • Lithium heparin without gel in plastic ●
  • Lithium heparin Barricor tube ●
  • Sodium heparin without gel in plastic

Matrix Comparison Summary

TubeTypeSampleRangeSlope% BiasCorrelation Coefficient
TargetRegularDemingPassing-BablokPass/FailTargetResultPass/FailTargetResultPass/Fail
SST115 – 2322 µmol/L1.0 ± 0.11.0121.0121.007Pass± 5.0%1.1%Pass≥ 0.9751.000Pass
LithiumHeparin1.0151.0151.012Pass1.2%Pass1.000Pass
PST1.0091.0091.000Pass0.8%Pass1.000Pass
SodiumHeparin1.0121.0121.006Pass1.1%Pass1.000Pass
Barricor1.0041.0051.000Pass0.2%Pass1.000Pass

Clinical studies

(clinical sensitivity, clinical specificity, other clinical supportive data)

Not Applicable

{12}------------------------------------------------

Expected values/Reference Range

The package insert contains the following reference ranges cited from literature. (Tietz Textbook of Clinical Chemistry, Second Edition, pp 1168, 2212, W.B. Saunders Company, Philadelphia (1994))

Therapeutic concentration: 10-30 µg/mL (66-199 µmol/L) Toxic concentration: > 200 µg/mL (1324 µmol/L) These values are suggested guidelines. It is recommended that each laboratory establish its own expected range.

Proposed Labeling

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10

Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 862.3030 Acetaminophen test system.

(a)
Identification. An acetaminophen test system is a device intended to measure acetaminophen, an analgestic and fever reducing drug, in serum. Measurements obtained by this device are used in the diagnosis and treatment of acetaminophen overdose.(b)
Classification. Class II.