K Number
K180768

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-12-14

(266 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

Shoulder: Rotator Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 SwiveLock only).

Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.

Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.

Hip: Capsular repair, acetabular labral repair.

Device Description

The Arthrex SwiveLock Anchor is a two-component, knotless suture anchor comprised of an eyelet and a hollow anchor body. The SwiveLock Anchor is premounted on a driver with the anchor body and eyelet physically separated on the driver shaft. Arthrex 510(k) cleared suture may also be provided with the device.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device: Arthrex SwiveLock Anchors. It is primarily focused on demonstrating substantial equivalence to predicate devices for the purpose of a regulatory clearance for an expanded indication (quadriceps tendon repair).

Crucially, this document does not contain information about an AI/ML-driven device or study that would involve acceptance criteria for algorithms, human-in-the-loop performance, or the other specific details requested in your prompt (e.g., sample size for test/training sets, expert consensus, MRMC studies, ground truth establishment methods for AI).

The "Performance Data" section of this 510(k) summary refers to:

  • Static pull-out testing of the Arthrex SwiveLock Anchor and a predicate device.
  • Cyclic pull-out testing of the subject device after degradation, compared to a non-resorbable predicate.
  • Bacterial endotoxin testing (pyrogen limit specifications).

These are biomechanical and biocompatibility tests for a physical implantable device, not performance metrics for an AI algorithm.

Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves an AI device meets acceptance criteria based on this document. The document describes the regulatory submission for a physical medical device (suture anchors) and its associated mechanical and biological performance, not an AI product.

{0}------------------------------------------------

December 14, 2018

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

Arthrex Inc. Jessica L. Singelais Regulatory Affairs Specialist 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K180768

Trade/Device Name: Arthrex SwiveLock Anchors Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: November 7, 2018 Received: November 8, 2018

Dear Ms. Singelais:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Laurence D. Coyne -S

  • For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180768

Device Name Arthrex SwiveLock Anchors

Indications for Use (Describe)

The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

Shoulder: Rotator Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 SwiveLock only).

Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.

Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.

Hip: Capsular repair, acetabular labral repair.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary or 510(k) Statement

Date PreparedDecember 14, 2018
SubmitterArthrex Inc.1370 Creekside BoulevardNaples, FL 34108-1945
Contact PersonJessica L. SingelaisRegulatory Affairs Specialist1-239-598-4302, ext. 73091Jessica.singelais@arthrex.com
Name of DeviceArthrex SwiveLock Anchor
Common NameSuture Anchor
Product CodeMAI, MBI
Classification Name21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener21 CFR 888.3030: Single/multiple component metallic bone fixation appliancesand accessories
Regulatory ClassII
Predicate DeviceK151342: Arthrex SwiveLock AnchorsK171141: Arthrex SwiveLock AnchorsK150768: Zimmer Biomet JuggerKnot Soft Anchors
Purpose of SubmissionThis traditional 510(k) premarket notification is submitted to obtain quadricepstendon repair indications for the Arthrex SwiveLock Anchors originally clearedunder K151342 and K171141.
Device DescriptionThe Arthrex SwiveLock Anchor is a two-component, knotless suture anchorcomprised of an eyelet and a hollow anchor body. The SwiveLock Anchor ispremounted on a driver with the anchor body and eyelet physically separated onthe driver shaft. Arthrex 510(k) cleared suture may also be provided with thedevice.
Indications for UseThe Arthrex SwiveLock Suture Anchors are intended for fixation of suture (softtissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip inthe following procedures:Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair,Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair,Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles TendonRepair, Hallux Valgus Reconstruction, Mid-foot Reconstruction,Metatarsal Ligament Repair/Tendon Repair and Bunionectomy. Knee: Medial Collateral Ligament Repair, Lateral Collateral LigamentRepair, Patellar Tendon Repair, Posterior Oblique Ligament Repair,Illiotibial Band Tenodesis, and Quadriceps Tendon Repair. Secondary oradjunct fixation for ACL/PCL reconstruction or repair (4.75-5.5 SwiveLockonly). Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or RadialCollateral Ligament Reconstruction. Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar orRadial Collateral Ligament Reconstruction, Lateral Epicondylitis repair. Hip: Capsular repair, acetabular labral repair.
Performance DataStatic pull-out testing was conducted on the Arthrex SwiveLock Suture Anchor
and the Zimmer Biomet JuggerKnot Soft Anchor. In addition, cyclic pull-outtesting was performed on the subject device after 16 weeks of degradation andcompared to the non-resorbable predicate device. Results demonstrate that theArthrex SwiveLock Anchor performs statistically equivalent to the predicatedevice (K150768 – Zimmer Biomet JuggerKnot Soft Anchor).Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate thatthe device meets pyrogen limit specifications.
ConclusionThe Arthrex SwiveLock Anchor is substantially equivalent to the predicate devicein which the basic design features and intended uses are the same. Anydifferences between the proposed device and the predicate device areconsidered minor and do not raise questions concerning safety or effectiveness.
Based on the indications for use, technological characteristics, and the summaryof data submitted, Arthrex Inc. has determined that the proposed device issubstantially equivalent to the currently marketed predicate devices.