K Number
K180758

Validate with FDA (Live)

Date Cleared
2018-10-24

(216 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Osteocera Resorbable Bone Substitute is intended for use as a bone void filler for voids or gaps of the skeletal system (i.e. extremities, posterolateral spine and pelvis) that are not intrinsic to the stability of the bony structure. Osteocera Resorbable Bone Substitute is indicated for use in the treatment of osseous defects created surgically or through traumatic injury. Following placement into the bony void, Osteocera Resorbable Bone Substitute resorbs and is replaced with bone during the healing process.

Device Description

Osteocera Resorbable Bone Substitute is a bioceramic medical device that consists of over 95% beta-tricalcium phosphate (ß-TCP). Osteocera is a resorbable implant intended to fill bony voids or gaps that are caused by trauma or surgery and are not intrinsic to the skeletal stability. The interconnected porous structure of Osteocera provides a matrix for bone ingrowth. Osteocera Resorbable Bone Substitute is for single use only and is supplied sterile in sticks and granules with various sizes and package volumes.

AI/ML Overview

This document describes the Osteocera Resorbable Bone Substitute, a medical device. The information provided is for regulatory clearance (510(k) submission) and focuses on demonstrating substantial equivalence to a predicate device, not on proving device performance against specific acceptance criteria in the context of a typical AI/ML medical device study.

Therefore, many of the requested elements for an AI/ML device study (like acceptance criteria for AI model performance, sample sizes for test sets, expert adjudication, MRMC studies, etc.) are not applicable or not explicitly stated in this document. The study described is an animal study to demonstrate safety and effectiveness for a bone substitute, not an AI model.

However, I can extract the relevant information from the provided text based on the nature of this device.


Device: Osteocera Resorbable Bone Substitute

Description of Use: Intended for use as a bone void filler for voids or gaps of the skeletal system (i.e. extremities, posterolateral spine and pelvis) that are not intrinsic to the stability of the bony structure. Osteocera Resorbable Bone Substitute is indicated for use in the treatment of osseous defects created surgically or through traumatic injury. Following placement into the bony void, Osteocera Resorbable Bone Substitute resorbs and is replaced with bone during the healing process.

Summary of Study Type: Animal studies were conducted to demonstrate the safety and effectiveness of the Osteocera Resorbable Bone Substitute compared to a predicate device (Synthes (USA) chronOS (K043045)). These are non-clinical studies for a material-based device, not an AI/ML algorithm.


Here's a breakdown of the requested information, indicating what is N/A (Not Applicable) or Not Explicitly Stated for this type of device and study:

1. A table of acceptance criteria and the reported device performance

For a material-based device like this, "acceptance criteria" are generally tied to demonstrating equivalence in material properties and biological performance (resorption, bone ingrowth, safety) to a legally marketed predicate device. The performance data presented focuses on demonstrating this equivalence rather than meeting pre-defined quantitative metrics in a "true positive/false positive" sense as would be found in an AI/ML context.

Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance (Summary from Animal Studies)
Biocompatibility: Meet ISO 10993 standards (e.g., non-cytotoxic, non-sensitizing, non-pyrogenic, non-hemocompatible)."Osteocera is considered biologically safe based on the device characterization and biocompatibility testing data regarding cytotoxicity, sensitization, acute systemic toxicity, subchronic toxicity, genotoxicity, pyrogenicity and hemocompatibility."
Safety & Effectiveness (Extremities): Demonstrate comparable implant resorption and new bone formation."In the distal femur implantation study, critical sized defects were created in 18 animals. Radiographic, histological, and histomorphometric analyses were performed at each time point [4, 12, and 24 weeks] to observe implant resorption and new bone formation over time... The animal studies demonstrated that Osteocera is as safe and effective as the predicate device in the indications."
Safety & Effectiveness (Posterolateral Spine): Demonstrate comparable bone graft fusion and new bone formation."In the posterolateral spine study, bone grafts were placed between the transverse processes in the paraspinal bed of 36 animals. Manual palpation, posteroanterior radiographic, histological and histomorphometric data were collected [at 4, 12, and 24 weeks]... The animal studies demonstrated that Osteocera is as safe and effective as the predicate device in the indications."

2. Sample sizes used for the test set and the data provenance

  • Test Set Sample Size:
    • Distal Femur Study (Extremities): 18 New Zealand white rabbits
    • Posterolateral Spine Study: 36 New Zealand white rabbits
  • Data Provenance:
    • Country of Origin: Not specified, but likely where the animal studies were conducted.
    • Retrospective or Prospective: Prospective (controlled animal studies).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This is an animal study evaluating direct biological response to a material implant, not an AI model requiring human expert ground truth for interpretation of images or other data. "Ground truth" would be established via direct measurements (histology, histomorphometry, radiography).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. As above, no human reader adjudication is involved in establishing the "truth" for direct biological and anatomical assessments in an animal study of an implant.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the animal studies, the "ground truth" was established through:
    • Radiographic analysis: Imaging assessments.
    • Histological analysis: Microscopic examination of tissue samples.
    • Histomorphometric analysis: Quantitative measurement of tissue structures from histological slides.
    • Manual palpation (for the posterolateral spine study, to assess fusion).
    • These methods directly assess the biological outcome (bone formation, resorption, fusion) after implant placement.

8. The sample size for the training set

  • Not Applicable. This is not an AI/ML device, so there is no "training set."

9. How the ground truth for the training set was established

  • Not Applicable. This is not an AI/ML device, so there is no "training set" or corresponding ground truth establishment process in that context.

{0}------------------------------------------------

October 24, 2018

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Wiltrom Corporation Limited Yi-Chun Su Director 1F, No. 26, Section 2, Shengyi Road Zhubei City, Hsinchu County 30261 Taiwan (Republic of China)

Re: K180758

Trade/Device Name: Osteocera Resorbable Bone Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: September 14, 2018 Received: September 17, 2018

Dear Yi-Chun Su:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K180758

Device Name

Osteocera Resorbable Bone Substitute

Indications for Use (Describe)

Osteocera Resorbable Bone Substitute is intended for use as a bone void filler for voids or gaps of the skeletal system (i.e. extremities, posterolateral spine and pelvis) that are not intrinsic to the stability of the bony structure. Osteocera Resorbable Bone Substitute is indicated for use in the treatment of osseous defects created surgically or through traumatic injury. Following placement into the bony void, Osteocera Resorbable Bone Substitute resorbs and is replaced with bone during the healing process.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

PSC Publishing Services (301) 443-6740 EF

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Wittrom, a health partner. The word "wittrom" is written in a bold, sans-serif font, with the "i" dotted with a circle. Below the word "wittrom" is the tagline "Your Health Partner" in a smaller, sans-serif font. The logo is black and white.

510(k) Summary

I. SubmitterWiltrom Corporation LimitedNo. 221, Sec. 1, Chung Hsing Rd.Chutung, Hsinchu County 31053Taiwan
Contact InformationYi-Chun Su, Director1F, No. 26, Sec. 2, Shengyi Rd.Zhubei City, Hsinchu County 30261TaiwanTEL: +886-3-6107168FAX: +886-3-6580006
Date PreparedMarch 12, 2018

III. Predicate Device Synthes (USA) chronOS (K043045)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Wittrom. The logo is black and consists of the word "wittrom" in a bold, sans-serif font. Above the "i" is a black dot. Below the word "wittrom" is the phrase "Your Health Partner" in a smaller, sans-serif font.

IV. Device Description

Osteocera Resorbable Bone Substitute is a bioceramic medical device that consists of over 95% beta-tricalcium phosphate (ß-TCP). Osteocera is a resorbable implant intended to fill bony voids or gaps that are caused by trauma or surgery and are not intrinsic to the skeletal stability. The interconnected porous structure of Osteocera provides a matrix for bone ingrowth. Osteocera Resorbable Bone Substitute is for single use only and is supplied sterile in sticks and granules with various sizes and package volumes.

V. Indication for Use

Osteocera Resorbable Bone Substitute is intended for use as a bone void filler for voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis) that are not intrinsic to the stability of the bony structure. Osteocera Resorbable Bone Substitute is indicated for use in the treatment of osseous defects created surgically or through traumatic injury. Following placement into the bony void, Osteocera Resorbable Bone Substitute resorbs and is replaced with bone during the healing process.

VI. Comparison of Technological Characteristics with the Predicate Device

The subject device has the same intended use and technological principles as the predicate device. The devices are both ß-TCP bone substitutes with porous structure. The interconnected pores provide a scaffold for bone ingrowth, and the biodegradable ß-TCP allows replacement of the bone substitute by natural bone. The subject device has higher porosity and similar pore size distribution to that of the predicate device. The devices are to be used in non-load bearing indications and internal fixation techniques may often be recommended to stabilize the operating area. Both devices are biocompatible and are sterilized by irradiation.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Wittrom, a health partner. The word "wittrom" is written in a bold, sans-serif font, with a small circle above the "i". Below the word "wittrom" is the tagline "Your Health Partner" in a smaller, sans-serif font. The logo is simple and modern, and the tagline conveys the company's mission.

Subject DevicePredicate Device
Trade NameOsteocera Resorbable Bone SubstituteSynthes (USA) chronOS (K043045)
Intended UseTo be used as bone void filler and to provide a matrix for bone ingrowthTo be used as bone void filler and to provide a matrix for bone ingrowth
Anatomical SiteBony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis)Bony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis)
Indications for Use• Bony voids or gaps that are not intrinsic to the stability of the bony structure• Bony defects created surgically or through traumatic injury• Bony voids or gaps that are not intrinsic to the stability of the bony structure• Bony defects created surgically or through traumatic injury
Compositionβ-TCP conforming to ASTM F1088β-TCP conforming to ASTM F1088
Design• Biodegradable ceramic• Interconnected porous structure• Biodegradable ceramic• Interconnected porous structure
Performance• Possessing osteoconductivity to promote bone formation• Resorbing and being replaced with bone during the healing process• Possessing osteoconductivity to promote bone formation• Resorbing and being replaced with bone during the healing process
Porosity75-85%60-70%
Pore Size300-600 μm100-500 μm
Physical FormStick, granulesBlocks, wedges, granules and cylinders
DimensionStick• 5 x 5 x 10 mm• 5 x 5 x 20 mmGranules• 0.25-0.5 mm• 0.5-1.0 mm• 1.0-2.0 mm• 2.0-3.0 mmBlock• 5 x 5 x 10 mm• 12.5 x 12.5 x 10 mm• 20 x 20 x 10 mmGranules• 0.5-0.7 mm• 0.7-1.4 mm• 1.4-2.8 mm• 2.8-5.6 mm
BiocompatibilityEstablishedEstablished
SterilitySterile (SAL:10⁻⁶)Sterile (SAL:10⁻⁶)
SterilizationGamma irradiationGamma irradiation

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Wittrom. The logo is black and consists of the word "wittrom" in a bold, sans-serif font. Above the "i" is a black dot. Below the word "wittrom" is the phrase "Your Health Partner" in a smaller, sans-serif font.

VII. Performance Data

The following performance data were provided in support of the substantial equivalence.

Biocompatibility testing

ISO 10993 standards and the FDA guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process' " were applied to evaluate the biocompatibility of Osteocera Resorbable Bone Substitute. Osteocera is considered biologically safe based on the device characterization and biocompatibility testing data regarding cytotoxicity, sensitization, acute systemic toxicity, subchronic toxicity, genotoxicity, pyrogenicity and hemocompatibility.

Animal testing

The performance of Osteocera in the extremities and posterolateral spine indications was studied in the distal femur and the posterolateral spine of New Zealand white rabbit, respectively. Skeletally matured animals were used in these studies. Evaluation was carried out at three time points: 4, 12 and 24 weeks post-implantation. In the distal femur implantation study, critical sized defects were created in 18 animals. Radiographic, histological, and histomorphometric analyses were performed at each time point to observe implant resorption and new bone formation over time. In the posterolateral spine study, bone grafts were placed between the transverse processes in the paraspinal bed of 36 animals. Manual palpation, posteroanterior radiographic, histological and histomorphometric data were collected. The animal studies demonstrated that Osteocera is as safe and effective as the predicate device in the indications.

VIII. Conclusion

Osteocera Resorbable Bone Substitute has the same intended use and indications for use as the predicate device, Synthes (USA) chronOS (K043045). Based on the material, technological characteristics and non-clinical data, Osteocera Resorbable Bone Substitute is substantially equivalent to the predicate device.

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.