K Number
K180740

Validate with FDA (Live)

Device Name
VitriGuard
Manufacturer
Date Cleared
2018-05-04

(43 days)

Product Code
Regulation Number
884.6160
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

Device Description

VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. It is single-use, disposable, and provided sterile. The device consists of a two-piece polystyrene assembly that includes a hexagonalshaped stick and cap. As part of the vitrification procedure, the embryos to be stored are loaded on the tip, and capped prior to plunging the device in liquid nitrogen and for subsequent storage. The tip has a trough area for loading, maintaining, and securing the embryos. The stick and cap include a taper design that creates a seal when assembled. Markings at the end of the stick and the tip of the device provide visual aid for proper device orientation.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance CriteriaReported Device Performance
Cooling rateNot explicitly stated-2,271°C/min
Warming rateNot explicitly stated36,377°C/min
SterilizationSAL 10⁻⁶Radiation, SAL 10⁻⁶
Endotoxin≤2.0 EU/device≤2.0 EU/device
Mouse Embryo Assay (MEA)≥80% blastocyst at 96hMet the predetermined acceptance criteria (≥80% blastocyst at 96h)

Note on Cooling/Warming rates: While listed as "Specifications," the document does not explicitly state them as "acceptance criteria" where a threshold needs to be met for the study. They are presented as inherent characteristics of the device. However, for the purpose of demonstrating device performance, they are included here. The MEA is the only one explicitly stated as meeting "predetermined acceptance criteria."

2. Sample size used for the test set and the data provenance

  • Sample size: Not explicitly stated. The text mentions "VitriGuard devices" and "One-cell mouse embryos" without specifying a number per test.
  • Data provenance: Not explicitly stated. However, given it's a pre-market submission, the study was likely conducted specifically for this submission and would be considered prospective for the device's regulatory approval. The study uses mouse embryos, which are a biological model, not human data from a specific country.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for the Mouse Embryo Assay (MEA) is established based on the biological outcome of embryo development (percent blastocyst formation), not through expert consensus or interpretation of images/data. Therefore, no "experts" were required to establish ground truth in this context.

4. Adjudication method for the test set

Not applicable. As noted above, the ground truth for the MEA is a direct biological outcome, not a subjective assessment requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is not an AI-assisted diagnostic tool; it is a physical cryopreservation storage device. Therefore, an MRMC study is not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is not an algorithm or AI system.

7. The type of ground truth used

The ground truth for the key performance test (Mouse Embryo Assay) was biological outcome data, specifically the successful development of mouse embryos to the blastocyst stage.

8. The sample size for the training set

Not applicable. This device is a physical product, not a machine learning model. There is no "training set" in this context.

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth established for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

May 4, 2018

ORIGIO a/s % Roaida Johnson Director, RA New Product Development CooperSurgical, Inc. 95 Corporate Drive Trumbull, CT 06611

Re: K180740

Trade/Device Name: VitriGuard™ Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MOK Dated: April 24, 2018 Received: April 25, 2018

Dear Roaida Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the overall impression is clean and professional.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

510(k) Number (if known)K180740
Device NameVitriGuard™
Indications for Use (Describe)VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRASstaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)Page 1 of 1
-----------------------------------

PSC Publishing Services (301) 443-6740

000009 EF

»

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for "origio", which is written in a stylized purple font. To the right of the logo, it says "a CooperSurgical company" in a smaller, also purple, font. The logo is simple and modern.

510(k) Summary K180740

Submitter Information

Company Name:ORIGIO a/s
Company Address:Knardrupvej 22760 MåløvDenmark
Telephone: +45 46 79 02 00
Fax: +45 46 79 03 02

Submission Correspondent Information

Company Name:CooperSurgical Inc.
Company Address:95 Corporate DriveTrumbull, CT 06611
Telephone:203-601-5200 Ext. 3325
Fax:203-601-9870

Contact Person: Roaida Johnson

Date Prepared: May 3, 2018

Device Information

Trade Name:VitriGuard™
Common Name:Cryopreservation Storage Device
Regulation Number:21 CFR 884.6160
Regulation Name:Assisted Reproduction Labware
Product Code:MQK (Labware, Assisted Reproduction)
Regulatory Class:II

Predicate Device Information

ORIGIO a/s VitriGuard (K162833).

The predicate device has not been subject to a design-related recall.

Device Description

VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. It is single-use, disposable, and provided sterile. The device consists of a two-piece polystyrene assembly that includes a hexagonalshaped stick and cap. As part of the vitrification procedure, the embryos to be stored are loaded on the tip, and capped prior to plunging the device in liquid nitrogen and for subsequent storage. The tip has a trough

{4}------------------------------------------------

area for loading, maintaining, and securing the embryos. The stick and cap include a taper design that creates a seal when assembled. Markings at the end of the stick and the tip of the device provide visual aid for proper device orientation.

Product specifications are listed in Table 1 below:

Table 1: Product Specifications

ParameterSpecification
Cooling rate-2,271°C/min
Warming rate36,377°C/min
SterilizationRadiation, SAL 10-6
Endotoxin≤2.0 EU/device
Mouse Embryo Assay (MEA)≥80% blastocyst at 96h

Indications for Use

VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

Substantial Equivalence Discussion

Table 2 provides a comparison of the subject and predicate device.

PropertySubject VitriGuardPredicate VitriGuard
510(k) NumberK180740K162833
Indications for UseVitriGuard is a cryopreservationstorage device that is intended for usein vitrification procedures to containand maintain human 4-8 cell andblastocyst stage embryosSame
FundamentalScientific TechnologyThe device is composed of a two-pieceassembly that includes a stick and cap.As part of the vitrification procedure,the embryos to be stored are loaded onthe tip of the stick and capped forsubsequent storage.Same
MaterialPolystyrene (clear, blue, green, yellow,orange, pink, lime green, and purple)Polystyrene (clear, blue, green,and yellow)
Black marker bandsBlack marker bands
Sterilization MethodRadiationSame
Number of UsesSingle Use, DisposableSame
PackagingBox of 20 (individually pouched) orbox of 50 (5 per pouch)Box of 20 (individuallypouched)
Shelf LifeOne YearThree Years

Table 2: Comparison of the Subject VitriGuard to the Predicate VitriGuard

The subject and predicate devices have intended use and are technologically comparable. Differences between the predicate and subject device includes the subject device having additional color

{5}------------------------------------------------

options of the stick component, a package configuration including five VitriGuard devices in a pouch, and a reduced shelf-life as compared to the predicate device. These differences do not raise different questions of safety and effectiveness as compared to the predicate device.

Non-Clinical Performance Testing

As part of demonstrating substantial equivalence to the predicate, the following tests were performed:

  • . MEA was performed on VitriGuard devices that were unaged and after one year of accelerated aging per ASTM F1980-07(2011). One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed in comparison with the control group. The subject device met the predetermined acceptance criteria (see Table 1).
  • Risk assessment to apply sterilization validation and bioburden testing on the predicate device to the subject device
  • Shipping and Distribution Testing per ISTA 3A 2008
  • . Package Integrity Testing Following Accelerated Aging
    • ASTM F1980-07(2011) o
    • O ASTM F1929-15
    • ASTM F88/F88M-15 O
    • o ASTM F1886/F1866M-09(2013)

Conclusion

The results of the performance testing provided demonstrated that the subject devices are as safe and effective as the predicate devices and supports substantial equivalence.

§ 884.6160 Assisted reproduction labware.

(a)
Identification. Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.(b)
Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.