Intended Use

ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES are indicated for orthodontic movement of natural teeth.

Device Description

The ORTHODONTIC PLASTIC BRACKETS are designed to move teeth to improve their alignment. The ORTHODONTIC PLASTIC BRACKETS are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred positions can be used with bracket caps. The ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES consist of plastic orthodontic brackets, bracket caps, and buttons. The bracket caps are designed as an accessory of the orthodontic brackets for holding the archwire in the bracket's slot. Bracket caps are placed on the labial/buccal side of the brackets. The plastic buttons are designed to move teeth to improve their alignment. The plastic buttons are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred position. The button design includes a round slot for attaching wires or other appliances.

AI/ML Overview

This device, "ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES", is not an AI/ML device. Therefore, the requested information regarding acceptance criteria and study details for an AI/ML device is not applicable here.

The document describes an orthodontic product (brackets, bracket caps, and buttons) and its substantial equivalence to previously marketed predicate devices, focusing on material composition, mechanical properties, and intended use. The performance data section refers to standard material testing and conformity to ISO standards, not AI/ML algorithm performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and white and includes the words "U.S. Food & Drug Administration."

TOMY, Inc. % Ruthanne Vendy Principal Specialist Regulatory & Ouality Solutions, Inc. 2790 Mosside Blvd. #800 Monroeville, Pennsylvania 15146

October 30, 2018

Re: K180718

Trade/Device Name: Orthodontic Plastic Bracket and Accessories Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: DYW Dated: August 22, 2018 Received: August 29, 2018

Dear Ruthanne Vendy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mary S. Runner -S3

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180718

Device Name ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES

Indications for Use (Describe) ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES are indicated for orthodontic movement of natural teeth.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TOMY Inc. The logo consists of the word "TOMY" in bold, black letters, followed by the word "Inc." in a smaller font size. To the left of the word "TOMY" is a circular graphic with the word "TOMY" written inside.

510(k) Summary K180718

Date Prepared: October 25, 2018

I. SUBMITTER

TOMY, Inc. Tenko Bldg. 3-16-7 Midori-cho, Fuchu City, Tokyo 183-0006 Japan

PHONE: 81-42-363-1151 FAX #: 81-42-363-1191

Contact: Mr. Jinichi Watanabe

II. DEVICE

Primary Classification: Name of Device: ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES Common or Usual Name: Bracket, Plastic, Orthodontic Classification Name: Orthodontic plastic bracket (21 CFR 872.5470) Regulatory Class: II Product Code: DYW

III. PREDICATE AND REFERENCE DEVICES

DENTSPLY Elation MB Metal Reinforced Plastic Brackets, K092030 (Primary Predicate) CDB Corporation CDB Clip, K080906 GESTENCO INTERNATIONAL AB Orthodontic Polymer Composite Bracket, K053055 DENTSPLY NEO-CLIP Orthodontic Bracket Accessory, K042905

IV. DEVICE DESCRIPTION

The ORTHODONTIC PLASTIC BRACKETS are designed to move teeth to improve their alignment. The ORTHODONTIC PLASTIC BRACKETS are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred positions can be used with bracket caps. The ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES consist of plastic orthodontic brackets, bracket caps, and buttons. The bracket caps are designed as an accessory of the orthodontic brackets for holding the archwire in the bracket's slot. Bracket caps are placed on the labial/buccal side of the brackets. The plastic buttons are designed to move teeth to improve their alignment. The plastic buttons are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred position. The button design includes a round slot for attaching wires or other appliances.

V. INDICATIONS FOR USE: ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES are indicated for orthodontic movement of natural teeth.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for TOMY Inc. The logo consists of the TOMY logo on the left and the text "TOMY Inc." on the right. The TOMY logo is a stylized version of the company's name. The text "TOMY Inc." is in a bold, sans-serif font.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

The below tables (Tables VI-1, VI-2 and VI-3) provide a comparison of the subject devices and the predicate devices. Table VI-1 is a comparison of the brackets; Table VI-2 is a comparison of the bracket caps; and Table VI-3 is a comparison of the bracket buttons.

Table VI-1 Comparison of DENTSPLY ELATION MB Metal Reinforced Plastic Bracket (K092030) to Subject ORTHODONTIC PLASTIC BRACKETS

ORTHODONTIC PLASTIC BRACKETS (Subject)DENTSPLY ELATION MB Metal ReinforcedPlastic Bracket (K092030)EquivalenceAnalysis
Indicationsfor UseORTHODONTIC PLASTIC BRACKETS ANDACCESSORIES are indicated for orthodonticmovement of natural teeth.Elation MB is indicated for orthodonticmovement of natural teeth.SubstantiallyEquivalent
DeviceNameDescriptionDevice: Bracket, Plastic, OrthodonticRegulation Description: Orthodontic plasticbracketDefinition: An orthodontic plastic bracket is aplastic device intended to be bonded to atooth to apply pressure to a tooth from aflexible orthodontic wire to alter its position.Device: Bracket, Plastic, OrthodonticRegulation Description: Orthodontic plasticbracketDefinition: An orthodontic plastic bracket is aplastic device intended to be bonded to a toothto apply pressure to a tooth from a flexibleorthodontic wire to alter its position.SubstantiallyEquivalent
MedicalSpecialtyProductCodeReg NumberClassRegulation Medical Specialty: DentalReview Panel: DentalProduct Code: DYWSubmission Type: 510(k)Regulation Number: 872.5470Device Class: 2Regulation Medical Specialty: DentalReview Panel: DentalProduct Code: DYWSubmission Type: 510(k)Regulation Number: 872.5470Device Class: 2
PatientContactMaterialsMade of Polycarbonate, Polyethyleneterephthalate, Polybutylene terephthalate,and Stainless Steel AISI 304Made of Polycarbonate, Polyethyleneterephthalate, Polybutylene terephthalate, andStainless Steel AISI 304SubstantiallyEquivalent
SterilityNon-SterileNon-SterileSubstantiallyEquivalent
SterilizationNot to be sterilizedNot to be sterilizedSubstantiallyEquivalent
UtilitySingle Use OnlySingle Use OnlySubstantiallyEquivalent
EnergySourceManually operated-Non-electricalManually operated—Non-electricalSubstantiallyEquivalent
VisualIdentification for toothplacementA color dot is applied to the distogingival tie-wing for visual identification.Color Codes:UR & UL CENTRALS: BLUEUR & UL LATERALS: PINKUR & UL CUSPIDS: GREENUR & UL BICUSPIDS: VIOLETLOWER ANTERIORS: ORANGELR & LL CUSPIDS: LIGHT BLUELR & LL 1st BICUSPIDS: GREYLR & LL 2nd BICUSPIDS: BLACKA color dot is applied to the distogingival tie-wing for visual identification.Color Codes:U1 BlueU2 PinkU3 GreenU4-5 VioletL1-2 OrangeL3 Light BlueL4 GreyL5 BlackSubstantiallyEquivalent
Applicationto teethBondedBondedSubstantiallyEquivalent
Manufacturing MethodInjection MoldedInjection MoldedSubstantiallyEquivalent
MechanicalLock BaseYesYesSubstantiallyEquivalent
ORTHODONTIC PLASTIC BRACKETS (Subject)DENTSPLY ELATION MB Metal ReinforcedPlastic Bracket (K092030)EquivalenceAnalysis
MetalReinforcedYesYesSubstantiallyEquivalent
Picture ofDevice(BRACKET)Image: ORTHODONTIC PLASTIC BRACKETSImage: DENTSPLY ELATION MB Metal Reinforced Plastic Bracket

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for TOMY Inc. The logo consists of the TOMY logo on the left and the text "TOMY Inc." on the right. The TOMY logo is a stylized version of the company's name. The text "TOMY Inc." is written in a bold, sans-serif font.

Table VI-2 Comparison of CDB CORPORATION CDB CLIP (K080906) to Subject ORTHODONTIC PLASTIC BRACKET CAPS

ORTHODONTIC PLASTIC BRACKET CAPS(Subject)CDB CORPORATIONCDB CLIP (K080906)1EquivalenceAnalysis
Indicationsfor UseORTHODONTIC PLASTIC BRACKETS ANDACCESSORIES are indicated for orthodonticmovement of natural teeth.The CDB Clip, commonly known as anorthodontic ligating clip, is an orthodonticbracket accessory. It attaches to theorthodontic bracket to hold the archwirefirmly in the slot Together they facilitate theorthodontic movement of teeth.SubstantiallyEquivalent
DeviceNameDescriptionMedicalSpecialtyProductCodeReg NumberClassDevice: Bracket, Plastic, OrthodonticRegulation Description: Orthodontic plasticbracketDefinition: An orthodontic plastic bracket is aplastic device intended to be bonded to atooth to apply pressure to a tooth from aflexible orthodontic wire to alter its position.Regulation Medical Specialty: DentalReview Panel: DentalProduct Code: DYWSubmission Type: 510(k)Regulation Number: 872.5470Device Class: 2Device: Bracket, Ceramic, OrthodonticRegulation Description: Orthodontic plasticbracketDefinition: An orthodontic ceramic bracket isa device composed of ceramic, which isintended to be bonded to a tooth, upon whichan orthodontic wire is used to move the toothto a new postion. (the current classificationonly specifies plastic brackets. Confusionexists as to whether ceramic brackets areregulated.)Regulation Medical Specialty: DentalReview Panel: DentalProduct Code: NJMSubmission Type: 510(k)Regulation Number: 872.5470Device Class: 2ProductCodes aredifferent,however thedevices areunder thesameregulation
PatientContactMaterialsPolyoxymethylene acetyl polymerPolyoxymethylene acetyl polymerSubstantiallyEquivalent
SterilityNon-SterileNon-SterileSubstantiallyEquivalent
SterilizationNot to be sterilizedNot to be sterilizedSubstantiallyEquivalent
UtilitySingle Use OnlySingle Use OnlySubstantiallyEquivalent
EnergySourceManually operated—Non-electricalManually operated—Non-electricalSubstantiallyEquivalent
Manufacturing MethodInjection MoldedInjection MoldedSubstantiallyEquivalent
Picture ofDevice (CAP)Image: Orthodontic plastic bracket capsImage UnavailableSubstantiallyEquivalent
1 K080906 CDB Clip is identical to the Dentsply International K042905 NEO-CLIP ORTHODONTIC BRACKET ACCESSORY. There is no difference in the
ORTHODONTIC PLASTICBUTTONS (Subject)DENTSPLY ELATION MBMetal Reinforced PlasticBracket (K092030)GESTENCO INTERNATIONALAB ORTHODONTIC POLYMERCOMPOSITE BRACKET(K053055)EquivalenceAnalysis
Indicationsfor UseORTHODONTIC PLASTICBRACKETS AND ACCESSORIESare indicated for orthodonticmovement of natural teeth.Elation MB is indicated fororthodontic movement ofnatural teeth.Orthodontic PolymerComposite Bracket isintended to be bonded onteeth to aid in themovement of patient's teethduring orthodontictreatment.SubstantiallyEquivalent
DeviceNameDescriptionMedicalSpecialtyProductCodeReg NumberClassDevice: Bracket, Plastic,OrthodonticRegulation Description:Orthodontic plastic bracketDefinition: An orthodonticplastic bracket is a plasticdevice intended to be bondedto a tooth to apply pressureto a tooth from a flexibleorthodontic wire to alter itsposition.Regulation Medical Specialty:DentalReview Panel: DentalProduct Code: DYWSubmission Type: 510(k)Regulation Number: 872.5470Device Class: 2Device: Bracket, Plastic,OrthodonticRegulation Description:Orthodontic plastic bracketDefinition: An orthodonticplastic bracket is a plasticdevice intended to be bondedto a tooth to apply pressureto a tooth from a flexibleorthodontic wire to alter itsposition.Regulation Medical Specialty:DentalReview Panel: DentalProduct Code: DYWSubmission Type: 510(k)Regulation Number: 872.5470Device Class: 2Device: Bracket, Plastic,OrthodonticRegulation Description:Orthodontic plastic bracketDefinition: An orthodonticplastic bracket is a plasticdevice intended to be bondedto a tooth to apply pressureto a tooth from a flexibleorthodontic wire to alter itsposition.Regulation Medical Specialty:DentalReview Panel: DentalProduct Code: DYWSubmission Type: 510(k)Regulation Number: 872.5470Device Class: 2SubstantiallyEquivalent
PatientContactMaterialsPolycarbonate, Polyethyleneterephthalate, PolybutyleneterephthalateMade of Polycarbonate,Polyethylene terephthalate,Polybutylene terephthalate,and Stainless Steel AISI 304PolycarbonateSubstantiallyEquivalent toK092030
SterilityNon-SterileNon-SterileNon-SterileSubstantiallyEquivalent
SterilizationNot to be sterilizedNot to be sterilizedNot to be sterilizedSubstantiallyEquivalent
UtilitySingle Use OnlySingle Use OnlySingle Use OnlySubstantiallyEquivalent
EnergySourceManually operated-Non-electricalManually operated-Non-electricalManually operated-Non-electricalSubstantiallyEquivalent
Manufacturing MethodInjection MoldedInjection MoldedInjection MoldedSubstantiallyEquivalent
Picture ofDeviceImage: Orthodontic Plastic ButtonsImage: Dentsply Elation MB Metal Reinforced Plastic BracketImage: Gestenco International AB Orthodontic Polymer Composite Bracket

fundamental technology. The CDB Clip is made of the same material and has the same overall intended use.

{6}------------------------------------------------

Table VI-3 Comparison of GESTENCO INTERNATIONAL AB ORTHODONTIC POLYMER COMPOSITE BRACKET (K053055) to Subject ORTHODONTIC PLASTIC BRACKET BUTTONS

VII. PERFORMANCE DATA

ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES conform to the following Recognized Consensus

ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES TOMY, Inc.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for TOMY Inc. The logo consists of the word "TOMY" in a bold, sans-serif font, followed by "Inc." in a smaller font. To the left of the word "TOMY" is a stylized graphic, possibly representing the company's initials.

Standards:

  • . ISO 27020 First edition 2010-12-15 "Dentistry - Brackets and Tubes for use in Orthodontics"
  • ISO 7405: 2008 "Dentistry Evaluation of biocompatibility of medical devices used in dentistry" O
  • ISO 10993-1:2009 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"

Additionally, the ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES were successfully tested for Tensile Strength, Elongation, and Shear Bond Strength. The tensile and elongation tests were performed to determine no significant deterioration with aging. New and aged brackets exceeded the criteria of ≥50 MPa for tensile strength and ≥ 100% elongation. The average shear bond strengths were greater than 100 N for wet and dry conditions, which exceeded the internal test criteria for shear bond strength.

All the components and manufacturing processes found in the ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES are identical to the materials used in legally marketed devices (DENTSPLY ELATION MB Metal Reinforced Plastic Bracket [K092030] and DENTSPLY NEO-CLIP ORTHODONTIC BRACKET ACCESSORY [K042905]) and/or were found acceptable for dental use. ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES are the same and/or of the equivalent composition as the predicate devices. Therefore, further biocompatibility testing is not necessary.

No animal studies or clinical testing were required for these devices.

VIII. CONCLUSIONS

There are no substantial differences in terms of composition and mechanical properties between the ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES and the predicate devices. The indications for use are comparable and any differences in technological characteristics do not raise any new questions of substantial equivalence. Therefore, we believe that the information provided herein demonstrates that the ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES are substantially equivalent to the predicate and reference devices in design, principle of performance, and intended use.

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.