K Number
K180689

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-04-12

(28 days)

Product Code
Regulation Number
876.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Padlock Clip™ is indicated for use in flexible Endoscopy and for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the wall of the gastrointestinal organs.

The Padlock Clip™ is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:

  • Endoscopic marking of lesions
  • Hemostasis for:
    • Mucosal/Submucosal defects
    • Bleeding Ulcers
    • Arteries <2mm
    • Polyps <1.5cm in diameter
    • Diverticula in the Colon
  • Closure of GI tract luminal perforations <20mm that can be treated conservatively
Device Description

The Padlock Clip™ ligation clip consists of a preloaded, radiopaque, single use, coin sized ligation clip made of superelastic shape memory alloy for tissue approximation with opening sizes of 6 to 24mm on a flexible delivery system.

The Padlock ClipTM defect closure device is mounted and secured at the distal tip on the outside surface of a flexible endoscope. The Padlock ClipTM is deployed using an independent hand control. The linking cable to the hand control is not located within the endoscopic accessory channel. The Padlock ClipTM delivery system includes a central "tissue chamber" that resides on the distal tip of the endoscope. Clinically efficacious tissue manipulation techniques may be used to pull target tissue into this "tissue chamber" to approximate a larger volume of tissue than would otherwise be approximated by deploying the clip alone. The Padlock ClipTM delivery system is compatible with flexible endoscopes with distal tip outer diameters ranging from 9.5mm to 14mm.

AI/ML Overview

Below is the requested information regarding the acceptance criteria and the study proving the device meets these criteria, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance CriteriaReported Device Performance
CytotoxicityPer ISO 10993-5Pass
SensitizationPer ISO 10993-10Pass
IrritationPer ISO 10993-10Pass
Acute Systemic ToxicityPer ISO 10993-11Pass
PyrogenicityPer ISO 10993-11Pass
Functional Testing (Deployment)DeploysPass
Functional Testing (Securing defect)Secures defectPass
Component Tensile StrengthMeets specificationPass
Component Compressive StrengthMeets specificationPass

2. Sample size used for the test set and the data provenance:

The provided text does not specify the sample sizes used for the test set or the provenance (e.g., country of origin, retrospective/prospective) of the data. The tests listed are primarily non-clinical performance and biocompatibility tests.

3. Number of experts used to establish the ground truth for the test set and their qualifications:

The provided text does not mention the use of experts to establish ground truth for a test set. The evaluation focuses on non-clinical performance and biocompatibility.

4. Adjudication method for the test set:

The provided text does not specify an adjudication method. The tests conducted are objective measurements against defined standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This document describes the modification of a medical device (Padlock Clip™) and its non-clinical performance and biocompatibility, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

A standalone performance study as described (algorithm only without human-in-the-loop) was not done. This document is for a mechanical medical device, not an algorithm.

7. The type of ground truth used:

The "ground truth" for the tests performed appears to be the established standards and specifications outlined in the ISO standards for biocompatibility tests (ISO 10993-5, 10993-10, 10993-11) and the device's internal specifications for functional and component strength tests.

8. The sample size for the training set:

A training set is relevant for machine learning algorithms. Since this submission is for a modified mechanical medical device, there is no mention of a training set or its sample size.

9. How the ground truth for the training set was established:

As there is no mention of a training set, the method for establishing its ground truth is not applicable and therefore not provided in the document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" and the words "U.S. Food & Drug Administration".

April 12, 2018

STERIS Corporation Tony Piotrkowski Senior Regulatory Affairs Manager 5976 Heisley Road Mentor. OH 44060

Re: K180689

Trade/Device Name: Padlock Clip defect closure device; Padlock Clip Pro-Select defect closure device Regulation Number: 21 CFR§ 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: II Product Code: PKL Dated: March 14, 2018 Received: March 15, 2018

Dear Tony Piotrkowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Charles Viviano -S" in a large, sans-serif font. The text is black and appears to be centered. The background is a light blue color, but it is mostly covered by the text.

Benjamin R. Fisher, Ph.D. For Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K180689

Device Name

Padlock Clip defect closure device Padlock Clip Pro-Select defect closure device

Indications for Use (Describe)

The Padlock Clip™ is indicated for use in flexible Endoscopy and for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the gastrointestinal organs.

The Padlock Clip™ is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:

  • · Endoscopic marking of lesions
  • · Hemostasis for:
    • o Mucosal/Submucosal defects
    • o Bleeding Ulcers
    • o Arteries <2mm
    • o Polyps <1.5cm in diameter
    • o Diverticula in the Colon

· Closure of GI tract luminal perforations <20mm that can be treated conservatively

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TERIS, with the registered trademark symbol next to it. Below the logo is a graphic of several blue horizontal lines. In the upper right corner of the image, the text "K180689 Page 1 of 5" is visible.

510(k) Summary For Padlock Clip Pro-Select™ defect closure device and Padlock Clip™ defect closure device

STERIS Corporation 5960 Heisley Road Mentor, OH 44060

Anthony Piotrkowski Contact: Senior Regulatory Affairs Manager Tel: 440-392-7458 Fax: 440-357-9198 Email: tony_piotrkowski@steris.com

Summary Date: April 11, 2018

Premarket Notification Number: K180689

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

K180689 STERIS Special 510(k) PREMARKET NOTIFICATION Modification to Padlock Clip and Padlock Clip Pro-Select defect closure devices

1. Device Name

Trade Name:Padlock Clip defect closure devicePadlock Clip Pro-Select defect closure device
Device Classification:Class II
Common/usual Name:Ligator clip
Classification Name:Hemorrhoidal ligator
Classification Number:21 CFR 876.4400
Product Code:PKL

2. Predicate Device

K120814 Aponos Medical, Padlock Clip

3. Description of Device

The Padlock Clip™ ligation clip consists of a preloaded, radiopaque, single use, coin sized ligation clip made of superelastic shape memory alloy for tissue approximation with opening sizes of 6 to 24mm on a flexible delivery system.

4. Intended Use

The Padlock Clip™ is indicated for use in flexible Endoscopy and for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the wall of the gastrointestinal organs.

The Padlock Clip™ is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:

  • · Endoscopic marking of lesions
  • · Hemostasis for:
    • o Mucosal/Submucosal defects
    • o Bleeding Ulcers
    • o Arteries <2mm
    • o Polyps <1.5cm in diameter
    • o Diverticula in the Colon
  • · Closure of GI tract luminal perforations <20mm that can be treated conservatively

{5}------------------------------------------------

5. Description of Technological Similarities and Differences

The Padlock Clip™ is identical to the predicate device except for the specific modification described in this submission. The differences between the proposed and predicate devices are limited to the addition of an alternate supplier for the stainless steel, PTFE-coated stainless steel and black fluorocarbon components of the device and these proposed changes raise no new concerns of safety and effectiveness when compared to the predicate device.

Device Comparison Table

FeatureProposedPadlock ClipK120814Aponos Padlock ClipComparison
Indicationsfor UseThe Padlock Clip™ isindicated for use in flexibleEndoscopy and for thecompression of tissue in thegastrointestinal tract, forhemostasis or for treatinglesions of the wall of thegastrointestinal organs.The Padlock Clip™ isindicated for clip placementwithin the gastrointestinal(GI) tract for the purpose of:Endoscopic marking oflesions Hemostasis for: Mucosal/Submucosaldefects Bleeding Ulcers Arteries <2mm Polyps <1.5cm indiameter Diverticula in the Colon Closure of GI tractluminal perforations<20mm that can betreated conservativelyThe Aponos MedicalPadlock Clip™ is indicatedfor use in flexible Endoscopyand for the compression oftissue in the gastrointestinaltract, for hemostasis or fortreating lesions of the wall ofthe gastrointestinal organs.The Padlock Clip™ isindicated for clip placementwithin the gastrointestinal(GI) tract for the purpose of:Endoscopic marking oflesions Hemostasis for: Mucosal/Submucosaldefects Bleeding Ulcers Arteries <2mm Polyps <1.5cm indiameter Diverticula in the Colon Closure of GI tractluminal perforations<20mm that can betreated conservativelyIdenticalRemovedAponosMedicalfrom theIndicationssince thedevice waspurchasedby STERISCorporation
FeatureProposedPadlock ClipK120814Aponos Padlock ClipComparison
OperatingPrinciples/TechnologyThe Padlock ClipTM defectclosure device is mounted andsecured at the distal tip on theoutside surface of a flexibleendoscope. The Padlock ClipTMis deployed using anindependent hand control. Thelinking cable to the hand controlis not located within theendoscopic accessory channel.The Padlock ClipTM deliverysystem includes a central "tissuechamber" that resides on thedistal tip of the endoscope.Clinically efficacious tissuemanipulation techniques may beused to pull target tissue into this"tissue chamber" to approximatea larger volume of tissue thanwould otherwise beapproximated by deploying theclip alone. The Padlock ClipTMdelivery system is compatiblewith flexible endoscopes withdistal tip outer diameters rangingfrom 9.5mm to 14mm.The Padlock ClipTM defectclosure device is mounted andsecured at the distal tip on theoutside surface of a flexibleendoscope. The Padlock ClipTMis deployed using anindependent hand control. Thelinking cable to the hand controlis not located within theendoscopic accessory channel.The Padlock ClipTM deliverysystem includes a central "tissuechamber" that resides on thedistal tip of the endoscope.Clinically efficacious tissuemanipulation techniques may beused to pull target tissue into this"tissue chamber" to approximatea larger volume of tissue thanwould otherwise beapproximated by deploying theclip alone. The Padlock ClipTMdelivery system is compatiblewith flexible endoscopes withdistal tip outer diameters rangingfrom 9.5mm to 14mm.Identical
ClipRadiopaque, single use, coinsized clip made ofsuperelastic shape memoryalloy (Nitinol®)Radiopaque, single use, coinsized clip made ofsuperelastic shape memoryalloy (Nitinol®)Identical
HousingPolypropylenePolypropyleneIdentical
StainlessSteelType 304Type 302Type 303Type 18-2Type 304Type 302Type 303Type 18-2Same gradeof steeldifferentsupplier
SecuringstrapBlack fluorocarbonBlack fluorocarbonSamematerialdifferentsupplier
UsageSingle UseSingle UseIdentical
SterilityEO SterilizationEO SterilizationIdentical
Shelf Life2 years2 yearsIdentical

Table 5-1. Processor Device Comparison Table

{6}------------------------------------------------

K180689 STERIS Special 510(k) PREMARKET NOTIFICATION Modification to Padlock Clip and Padlock Clip Pro-Select defect closure devices

Table 5-2 summarizes the verification activity that was performed with its respective acceptance criteria to ensure that this modification does not affect the safety or effectiveness of the Padlock Clip defect closure device.

{7}------------------------------------------------

K180689 STERIS Special 510(k) PREMARKET NOTIFICATION Modification to Padlock Clip and Padlock Clip Pro-Select defect closure devices

TestAcceptance CriteriaResult
CytotoxicityPer ISO 10993-5Pass
SensitizationPer ISO 10993-10Pass
IrritationPer ISO 10993-10Pass
Acute Systemic ToxicityPer ISO 10993-11Pass
PyrogenicityPer ISO 10993-11Pass
Functional TestingDeploysPass
Secures defectPass
Component Tensile StrengthMeets specificationPass
Component Compressive StrengthMeets specificationPass
Table 5-2. Summary of verification activities.
------------------------------------------------------

6. Conclusion

Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as the legally marketed predicate device (K120814), Class II (21 CFR 876.4400), product code PKL.

§ 876.4400 Hemorrhoidal ligator.

(a)
Identification. A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.(b)
Classification. Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.