K Number
K180681

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-06-13

(90 days)

Product Code
Regulation Number
866.3390
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aptima Combo 2® Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal urogenital disease using the Panther® System as specified.

On the Panther System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens, clinician-collected gynecological specimens collected in the PreservCyt® Solution, patient-collected vaginal swab specimens, 1 and female and male urine specimens.

1 Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal and multitest swab specimen collection kits are not for home use.

Device Description

Clearance of this pre-market application will add female urine as an acceptable specimen type using the Aptima Combo 2 assay on the Panther system.

The Aptima Combo 2 Assay combines the technologies of target capture, Transcription-Mediated Amplification (TMA), and Dual Kinetic Assay (DKA). Specimens are collected and transferred into their respective specimen transport tubes. The transport solutions in these tubes release the rRNA targets and protect them from degradation during storage. When the Aptima Combo 2 Assay is performed in the laboratory, the target rRNA molecules are isolated from specimens by use of capture oligomers via target capture that utilizes magnetic microparticles. The capture oligomers contain sequences complementary to specific regions of the target molecules as well as a string of deoxyadenosine residues. A separate capture oligomer is used for each target. During the hybridization step, the sequence specific regions of the capture oligomers bind to specific regions of the target molecules. The capture oligomer:target complex is then captured out of solution by decreasing the temperature of the reaction to room temperature. This temperature reduction allows hybridization to occur between the deoxyadenosine region on the capture oligomer and the poly-deoxythymidine molecules that are covalently attached to the magnetic particles. The microparticles, including the captured target molecules bound to them, are pulled to the side of the reaction vessel using magnets and the supernatant is aspirated. The particles are washed to remove residual specimen matrix that may contain amplification reaction inhibitors. After the target capture steps are completed, the specimens are ready for amplification.

Target amplification assays are based on the ability of complementary oligonucleotide primers to specifically anneal and allow enzymatic amplification of the target nucleic acid strands. The Aptima Combo 2 Assay replicates a specific region of the 23S rRNA from CT and a specific region of the 16S rRNA from GC via DNA intermediates. A unique set of primers is used for each target molecule. Detection of the rRNA amplification product sequences (amplicon) is achieved using nucleic acid hybridization. Single-stranded chemiluminescent DNA probes, which are complementary to a region of each target amplicon, are labeled with different acridinium ester molecules. The labeled DNA probes combine with amplicon to form stable RNA:DNA hybrids. The Selection Reagent differentiates hybridized from unhybridized probe, eliminating the generation of signal from unhybridized probe. During the detection step, light emitted from the labeled RNA:DNA hybrids is measured as photon signals in a luminometer, and are reported as Relative Light Units (RLU). In DKA, differences in the kinetic profiles of the CT and GC labeled probes allow for the differentiation of signal; kinetic profiles are derived from measurements of photon output during the detection read time. The chemiluminescent detection for CT signal has very rapid kinetics and has the "flasher" kinetic type. The chemiluminescent detection reaction for GC signal is relatively slower and has the "glower" kinetic type. Assay results are determined by a cut-off based on the total RLU and the kinetic curve type.

The Aptima Combo 2 assay has been designed for and validated on the Panther system. The Panther system is an integrated hardware and software system that together with the Aptima Combo 2 assay fully automates all the steps necessary to perform the assay from sample preparation through amplification of nucleic acid, detection, data reduction and amplicon inactivation.

AI/ML Overview

Here's an analysis of the provided text regarding the Aptima Combo 2 Assay (Panther System), focusing on acceptance criteria and the supporting study:

The provided document describes a 510(k) premarket notification for the Aptima Combo 2 Assay (Panther System) to add female urine as an acceptable specimen type. The study aimed to demonstrate substantial equivalence to the predicate device, which already included other specimen types.

1. Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" in a table format with numerical targets. Instead, it presents the Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) of the device compared to a Composite Comparator Algorithm (CCA) for Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) in female urine samples. For regulatory clearance, these performance metrics are implicitly the acceptance criteria; the observed performance must be deemed sufficient for the intended use and comparable to similar marketed devices.

Based on the performance tables provided, here's a summary of the reported device performance, which likely served as the basis for acceptance:

Reported Performance of Aptima Combo 2 Assay (Panther System) for Female Urine

OrganismSymptom Statusn (valid results)Positive Percent Agreement (PPA) (95% CI)Negative Percent Agreement (NPA) (95% CI)
Chlamydia trachomatis (CT)Symptomatic137999.1% (95.0-99.8)99.8% (99.4-100)
Asymptomatic119398.5% (91.9-99.7)99.7% (99.2-99.9)
Neisseria gonorrhoeae (GC)Symptomatic138395.0% (76.4-99.1)100% (99.7-100)
Asymptomatic1196100% (70.1-100)100% (99.7-100)

Note: The confidence intervals provide the range within which the true PPA/NPA is likely to fall. For asymptomatic GC, the PPA has a wider confidence interval due to a smaller number of positive cases.

2. Sample Size and Data Provenance

  • Sample Size for Test Set:
    • Total subjects initially enrolled: 2640
    • Subjects with valid Aptima Combo 2 Assay results on Panther System: 2581
    • Evaluable subjects for performance (conclusive CCA status): 2580
    • Final Sample Size for CT performance: 2572 (1379 symptomatic, 1193 asymptomatic) after accounting for equivocal results and non-evaluable subjects.
    • Final Sample Size for GC performance: 2579 (1383 symptomatic, 1196 asymptomatic) after accounting for equivocal results and non-evaluable subjects.
  • Data Provenance: Retrospective study.
    • Specimens originated from women enrolled in a previously completed prospective study.
    • The study participants (women) were enrolled from 17 geographically and ethnically diverse US clinical sites, including family planning, academic centers, and public health clinics.

3. Number of Experts and Qualifications for Ground Truth

The document does not specify the number of experts used to establish the ground truth or their qualifications. The ground truth (Composite Comparator Algorithm, CCA) was established using multiple FDA-cleared NAATs, not directly by human experts adjudicating individual cases based on clinical information or pathology.

4. Adjudication Method for the Test Set

The adjudication method used to establish the Composite Comparator Algorithm (CCA) for the ground truth was:

  • 2 out of 3 rule: When 2 out of 3 FDA-cleared CT/GC NAATs were positive, the CCA was considered positive. When 2 out of 3 NAATs were negative, the CCA was considered negative.
  • Tie-breaker: If the results from the initial two comparator NAATs did not determine the CCA, a third FDA-cleared CT/GC NAAT was performed using remnant urine samples to determine the CCA.

This is a form of consensus-based ground truth, but using other diagnostic tests rather than direct clinical expert consensus.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. This study evaluated the standalone performance of a diagnostic assay (Aptima Combo 2 Assay) against a reference standard (CCA), not the effect of AI assistance on human readers.

6. Standalone Performance

Yes, a standalone performance study was done. The entire clinical study described evaluates the performance of the Aptima Combo 2 assay on the Panther System (the algorithm/device itself) directly against the Composite Comparator Algorithm (CCA) without human interpretation as part of the primary outcome measure. The PPA and NPA values reported are the standalone performance metrics.

7. Type of Ground Truth Used

The ground truth used was a Composite Comparator Algorithm (CCA), which was derived from the results of multiple (up to 3) FDA-cleared nucleic acid amplification tests (NAATs). While this is a common method for establishing a "gold standard" in diagnostic test evaluations, it is not direct pathology, clinical outcomes data, or expert consensus in the traditional sense of clinicians reviewing patient records or images. It establishes the "truth" based on a highly sensitive and specific panel of existing diagnostic tools.

8. Sample Size for the Training Set

The document does not provide information on the sample size for a training set. This is a diagnostic assay (a lab test), not an AI/machine learning model in the typical sense that would require a separate, explicit "training set" for model parameters. The "development" or "training" of such an assay involves reagent formulation, assay protocol optimization, and establishing cut-offs, typically done using characterized samples, but not usually reported with a distinct "training set" size in the same manner as an AI algorithm. The study described is a clinical validation or "test set" evaluation.

9. How the Ground Truth for the Training Set was Established

As mentioned above, there is no explicit "training set" described in the context of this 510(k) submission. The performance study evaluated the device against a CCA, as detailed in point 4 and 7.

{0}------------------------------------------------

510(k) SUMMARY Aptima Combo 2 Assay (Panther System)

I. SUBMITTER

Hologic, Inc. 10210 Genetic Center Drive San Diego, CA 92121

Contact Person:Jeffrey Hergesheimer, MS, RACRegulatory Affairs Manager
Jeffrey.Hergesheimer@Hologic.com
Phone: (858) 410-8536
Fax: (858) 410-5557

Date Prepared: March 7, 2018

II. DEVICE

Proprietary Name of Device:Aptima Combo 2 Assay (Panther System)
Classification Name:DNA Reagents, Neisseria,
DNA Probe, Nucleic Acid Amplification, Chlamydia
Regulation Number:21 CFR 866.3390 and 866.3120
Regulatory Class:Class II
Product Code:LSL, MKZ

III. PREDICATE DEVICE

The predicate device is the Aptima Combo 2 Assay (Panther System); K132251, cleared October 17, 2013.

17, 2013.

IV. DEVICE DESCRIPTION

Clearance of this pre-market application will add female urine as an acceptable specimen type using the Aptima Combo 2 assay on the Panther system.

Principles of the Procedure

{1}------------------------------------------------

The Aptima Combo 2 Assay combines the technologies of target capture, Transcription-Mediated Amplification (TMA), and Dual Kinetic Assay (DKA). Specimens are collected and transferred into their respective specimen transport tubes. The transport solutions in these tubes release the rRNA targets and protect them from degradation during storage. When the Aptima Combo 2 Assay is performed in the laboratory, the target rRNA molecules are isolated from specimens by use of capture oligomers via target capture that utilizes magnetic microparticles. The capture oligomers contain sequences complementary to specific regions of the target molecules as well as a string of deoxyadenosine residues. A separate capture oligomer is used for each target. During the hybridization step, the sequence specific regions of the capture oligomers bind to specific regions of the target molecules. The capture oligomer:target complex is then captured out of solution by decreasing the temperature of the reaction to room temperature. This temperature reduction allows hybridization to occur between the deoxyadenosine region on the capture oligomer and the poly-deoxythymidine molecules that are covalently attached to the magnetic particles. The microparticles, including the captured target molecules bound to them, are pulled to the side of the reaction vessel using magnets and the supernatant is aspirated. The particles are washed to remove residual specimen matrix that may contain amplification reaction inhibitors. After the target capture steps are completed, the specimens are ready for amplification.

Target amplification assays are based on the ability of complementary oligonucleotide primers to specifically anneal and allow enzymatic amplification of the target nucleic acid strands. The Aptima Combo 2 Assay replicates a specific region of the 23S rRNA from CT and a specific region of the 16S rRNA from GC via DNA intermediates. A unique set of primers is used for each target molecule. Detection of the rRNA amplification product sequences (amplicon) is achieved using nucleic acid hybridization. Single-stranded chemiluminescent DNA probes, which are complementary to a region of each target amplicon, are labeled with different acridinium ester molecules. The labeled DNA probes combine with amplicon to form stable RNA:DNA hybrids. The Selection Reagent differentiates hybridized from unhybridized probe, eliminating the generation of signal from unhybridized probe. During the detection step, light emitted from the labeled RNA:DNA hybrids is measured as photon signals in a luminometer, and are reported as Relative Light Units (RLU). In DKA, differences in the kinetic profiles of the

{2}------------------------------------------------

CT and GC labeled probes allow for the differentiation of signal; kinetic profiles are derived from measurements of photon output during the detection read time. The chemiluminescent detection for CT signal has very rapid kinetics and has the "flasher" kinetic type. The chemiluminescent detection reaction for GC signal is relatively slower and has the "glower" kinetic type. Assay results are determined by a cut-off based on the total RLU and the kinetic curve type.

Assay Components

The reagents required to perform the Aptima Combo 2 assay (Panther System) are available in two kit sizes (100- and 250-tests). The kits are packaged in 3 boxes containing 11 reagents which are required for sample processing. A description of the components that are required to perform the Aptima Combo 2 assay are detailed in Table 1. In addition, there are two ancillary kits required to run the assay (Table 2), and four collection kits utilized for collection and transfer of specimens prior to testing on the Panther system (Table 3).

BoxComponents Description
1Amplification Reagent
Enzyme Reagent
Probe Reagent
Target Capture Reagent B
2Amplification Reconstitution Solution
Enzyme Reconstitution Solution
Probe Reconstitution Solution
Selection Reagent
Target Capture Reagent
3Aptima Positive Control CT/Neg Control GC
Aptima Positive Control GC/Neg Control CT

Table 1: Reagents Required to Perform the Aptima Combo 2 Assay

Table 2: Ancillary Kits Required to Perform the Aptima Combo 2 Assay

Aptima Assay Fluids Kit
Aptima Auto Detect Kit

{3}------------------------------------------------

Table 3: Specimen Collection Kits Available for Use with the Aptima Combo 2 Assay

Aptima Unisex Swab Specimen Collection Kit for Endocervical and Male Urethral Swab Specimens.

Aptima Vaginal Swab Specimen Collection Kit

Aptima Multitest Swab Specimen Collection Kit For Aptima Combo 2 Assay. Aptima Multitest Swab Specimen Collection Kit has been validated for the collection of vaginal swab specimen. Aptima Urine Specimen Collection Kit for Male and Female Urine

Specimens

Aptima Specimen Transfer Kit

Instrumentation

The Aptima Combo 2 assay has been designed for and validated on the Panther system. The Panther system is an integrated hardware and software system that together with the Aptima Combo 2 assay fully automates all the steps necessary to perform the assay from sample preparation through amplification of nucleic acid, detection, data reduction and amplicon inactivation.

V. INDICATIONS FOR USE

Intended Use

The Aptima Combo 2® Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal urogenital disease using the Panther® System as specified.

On the Panther System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens, clinician-collected gynecological specimens collected in the PreservCyt® Solution, patient-collected vaginal swab specimens, 1 and female and male urine specimens.

1 Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal and multitest swab specimen collection kits are not for home use.

{4}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

A comparison of the Aptima Combo 2 assay (Panther System) with the addition of the female urine specimen type to the predicate device, Aptima Combo 2 assay (Panther System) (K132251), is summarized in Table 4 (similarities) and Table 5 (differences).

Table 4: Similarities Between Predicate Device and Subject Device
ItemAptima Combo 2 Assay(Panther System)(Predicate Device, K132251)Aptima Combo 2 Assay(Panther System)(Subject Device, K180681)
Technology Principleof OperationTarget Capture (TC), Transcription-Mediated Amplification (TMA),Hybridization Protection Assay (HPA)Same
PlatformAutomated Panther SystemSame
FunctionDetection and differentiation of rRNAfrom Chlamydia trachomatis andNeisseria gonorrhoeaeSame
Organisms DetectedChlamydia trachomatis (CT) and/orNeisseria gonorrhoeae (GC)Same
Patient PopulationSymptomatic and asymptomaticindividualsSame

Table 4: Similarities Between Predicate Device and Subject Device

Table 5: Differences Between Predicate Device and Subject Device

ItemAptima Combo 2 Assay(Panther System)(Predicate Device, K132251)Aptima Combo 2 Assay(Panther System)(Subject Device, K180681)
Intended UseThe Aptima Combo 2 Assay is a targetamplification nucleic acid probe testthat utilizes target capture for the invitro qualitative detection anddifferentiation of ribosomal RNA(rRNA) from Chlamydia trachomatis(CT) and/or Neisseria gonorrhoeae(GC) to aid in the diagnosis ofchlamydial and/or gonococcalurogenital disease using the PantherSystem as specified. On the PantherSystem, the assay may be used to testthe following specimens fromsymptomatic and asymptomaticindividuals: clinician-collectedendocervical, vaginal and maleThe Aptima Combo 2 Assay is a targetamplification nucleic acid probe testthat utilizes target capture for the invitro qualitative detection anddifferentiation of ribosomal RNA(rRNA) from Chlamydia trachomatis(CT) and/or Neisseria gonorrhoeae(GC) to aid in the diagnosis ofchlamydial and/or gonococcalurogenital disease using the PantherSystem as specified. On the PantherSystem, the assay may be used to testthe following specimens fromsymptomatic and asymptomaticindividuals: clinician-collectedendocervical, vaginal and male

{5}------------------------------------------------

ItemAptima Combo 2 Assay(Panther System)(Predicate Device, K132251)Aptima Combo 2 Assay(Panther System)(Subject Device, K180681)
urethral swab specimens, clinician-collected gynecological specimenscollected in the PreservCyt Solution,patient-collected vaginal swabspecimens1 and male urine specimens.1Patient-collected vaginal swab specimens arean option for screening women when a pelvicexam is not otherwise indicated. The vaginalswab specimen collection kit is not for homeuse.urethral swab specimens, clinician-collected gynecological specimenscollected in the PreservCyt Solution,patient-collected vaginal swabspecimens,1 and female and male urinespecimens.1Patient-collected vaginal swab specimens arean option for screening women when a pelvicexam is not otherwise indicated. The vaginaland multitest swab specimen collection kit isnot for home use.
Specimen TypesFemale specimens:Vaginal swab Endocervical swab ThinPrep in PreservCyt solutionMale Specimens:Urethral Swab UrineFemale specimens:Vaginal swab Endocervical swab ThinPrep in PreservCyt solution UrineMale Specimens:Urethral Swab Urine

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Clinical Performance

Clinical Study 3. Female urine Specimen Clinical Study

A retrospective study was conducted that used results and remnant urine samples from women enrolled for a previously completed prospective study of the Aptima Combo 2 assay on the Panther system. Specimens originated from symptomatic and asymptomatic women (n=2640) enrolled from 17 geographically and ethnically diverse US clinical sites, including family planning, academic center, and public health clinics. In the previous study, all female urine samples were tested with the Aptima Combo 2 assay on the Panther system, and with two FDAcleared comparator nucleic acid amplification tests (NAATs). For the current study, previous study testing results were re-analyzed to assess the performance of the Aptima Combo 2 assay on the Panther system in urine samples relative to an FDA-agreed upon composite comparator

{6}------------------------------------------------

algorithm (CCA). The CCA was determined for Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) separately using urine test results from up to 3 FDA-cleared CT/GC NAATs. When 2 out of 3 of the NAATs were positive, the CCA was considered positive. When 2 out of 3 of the NAATs were negative, the CCA was considered negative. If the two comparator NAAT results for urine samples did not determine the CCA, additional (tie-breaker) testing was performed with a third FDA-cleared CT/GC NAAT using remnant urine samples from the previous study to determine the CCA. Results from testing female urine samples with the Aptima Combo 2 assay on the Panther system were compared to the CCA results to establish the positive percent agreement (PPA) and negative percent agreement (NPA) estimates, with corresponding 2-sided 95% Score confidence intervals (CI). Analyses were performed separately for each CT and GC. Results in urine samples by Aptima Combo 2 Assay (Panther System) were also compared with the vaginal and endocervical swab specimens using the patient infected status (PIS) algorithm. In Clinical Study 3, female urine detected 8.3% fewer CT infections than vaginal and endocervical swab specimens and 12.9% fewer GC infections than vaginal swab specimens and 15.2% fewer GC infections than endocervical swab specimens when compared using the PIS algorithm.

Of the 2640 subjects enrolled, 42 subjects were withdrawn. Of the 2598 non-withdrawn subjects 2581 had urine samples tested with the Aptima Combo 2 Assay on the Panther System in accordance with the Aptima Combo 2 Assay package insert instructions. Seventeen subjects had urine samples that were withdrawn or not collected (missing both CT and GC Aptima Combo 2 Assay (Panther System) results). Samples with initial invalid, equivocal, or error results were retested. All 2581 samples had final valid results after required retesting. One sample had a repeat CT equivocal result and one sample had a repeat GC equivocal result.

Of the 2581 subjects that had valid Aptima Combo 2 Assay (Panther System) results, 2580 subjects had a conclusive CT and/or GC composite comparator status and were evaluable for performance; one subject had unknown composite comparator status for both CT and GC and was not evaluable. One evaluable subject had a final equivocal CT result (negative GC result), and one evaluable subject had a final equivocal GC result (negative CT result). Demographics of the 2580 evaluable subjects is presented in Table 6.

{7}------------------------------------------------

N (%)
Total2580 (100)
Symptom Status
Asymptomatic1196 (46.4)
Symptomatic1384 (53.6)
Age Group
16 to 17 years47 (1.8)
18 to 20 years346 (13.4)
21 to 30 years1350 (52.3)
31 to 40 years550 (21.3)
>40 years287 (11.1)

Table 6: Summary of Subject Symptom Status and Age for Samples in the Aptima Combo 2 Assay Evaluation

Of the 2580 evaluable samples tested with the Aptima Combo 2 assay on the Panther system,

0.4% (10/2580) were positive for both CT and GC, 6.7% (172/2580) were positive for CT only, and 0.7% (18/2580) were positive for GC only.

Performance characteristics for detection of CT and GC in female urine samples were calculated separately (see Table 7 and Table 8).

Table 7: Performance of Aptima Combo 2 Assay on the Panther System Compared to Composite Comparator Algorithm Results for Detection of CT in Female Urine Samples

SymptomStatusnCCA+AC2+CCA-AC2+CCA-AC2-CCA+AC2-PPA(95% CI)2NPA(95% CI)2
Symptomatic13791092312674199.1 (95.0-99.8)99.8 (99.4-100)
Asymptomatic11936535112461198.5 (91.9-99.7)99.7 (99.2-99.9)
  • = positive, - = negative, AC2 = Aptima Combo 2 assay on the Panther system, CCA = composite comparator algorithm, NPA= negative percent agreement, PPA= positive percent agreement

4 Includes equivocal results from Panther AC2 testing. Equivocal results from AC2 testing are considered indeterminate; a new specimen should be collected.

2 Score Confidence Interval

3 Of the 2 symptomatic subjects with CCA-negative results, both had positive CT vaginal swab sample results in both reference NAATs.

{8}------------------------------------------------

SymptomStatusnCCA+AC2+CCA-AC2+CCA-AC2-CCA+AC2-PPA(95% CI)2NPA(95% CI)2
------------------------------------------------------------------------------------------------------------------

4 Of the 1267 symptomatic subjects with CCA-negative results, 38 had at least one positive CT vaginal swab sample result by a reference NAAT; one or more vaginal swab sample reference results were not available for an additional 11 symptomatic subjects. The remaining 1218 symptomatic subjects had negative vaginal swab sample reference results.

5 Of the 3 asymptomatic subjects with CCA-negative results, 1 had positive CT vaginal swab sample results in both reference NAATs; the remaining 2 asymptomatic subjects had negative vaginal swab sample reference results.

6 Of the 1124 asymptomatic subjects with CCA-negative results, 20 had at least one positive CT vaginal swab sample result by a reference NAATs; one or more vaginal swab sample reference results were not available for an additional 11 asymptomatic subjects. The remaining 1093 asymptomatic subjects had negative vaginal swab sample reference results.

Table 8: Performance of Aptima Combo 2 Assay on the Panther System Compared to Composite Comparator Algorithm Results for Detection of GC in Female Urine Samples

SymptomStatusnCCA+AC2+CCA-AC2+CCA-AC2-CCA+AC2-1PPA(95% CI)2NPA(95% CI)2
Symptomatic138319013633195.0 (76.4-99.1)100 (99.7-100)
Asymptomatic119690118740100 (70.1-100)100 (99.7-100)
  • = positive. - = negative. AC2 = Aptima Combo 2 assay on the Panther system. CCA = composite comparator algorithm, NPA= negative percent agreement, PPA= positive percent agreement

1 Includes equivocal results from Panther AC2 testing. Equivocal results from AC2 testing are considered indeterminate; a new specimen should be collected.

2 Score Confidence Interval

3 Of the 1363 symptomatic subjects with TN (CCA-negative) results, 5 had at least one positive GC vaginal swab sample result by a reference NAATs; one or more reference results were not available for an additional 11 symptomatic subjects. The remaining 1347 symptomatic subjects had negative vaginal swab sample reference results.

4 Of the 1187 asymptomatic subjects with TN (CCA-negative) results, 6 had at least one positive GC vaginal swab sample result by a reference NAAT; one or more vaginal swab sample reference results were not available for an additional 11 asymptomatic subjects. The remaining 1170 asymptomatic subjects had negative vaginal swab sample reference results.

VIII. CONCLUSIONS

The clinical study results demonstrate that the Aptima Combo 2 assay on the Panther system is

substantially equivalent to the predicate device.

{9}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

Indications for Use (Describe)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue. The logos are simple and professional, and they are easily recognizable.

June 13, 2018

Hologic, Inc. Jeffrey Hergesheimer, MS, RAC Regulatory Affairs Manager 10210 Genetic Center Drive San Diego, California 92121

Re: K180681

Trade/Device Name: Aptima Combo 2 Assay (Panther System) Regulation Number: 21 CFR 866.3390 Regulation Name: Neisseria spp. direct serological test reagents Regulatory Class: Class II Product Code: LSL, MKZ Dated: March 14, 2018 Received: March 15, 2018

Dear Jeffrey Hergesheimer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{11}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

§ 866.3390

Neisseria spp. direct serological test reagents.(a)
Identification. Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identifyNeisseria spp. from cultured isolates. Additionally, some of these reagents consist ofNeisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence ofNeisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusNeisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.(b)
Classification. Class II (performance standards).