K Number
K180668

Validate with FDA (Live)

Date Cleared
2018-11-15

(246 days)

Product Code
Regulation Number
876.1075
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This instrument has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration(FNA) of submucosal and extramural lesions of the gastrointestinal tract.

Device Description

The Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N (aka ViziShot 2) is used in conjunction with Olympus ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA). The subject devices consist of a handle section and insertion. The handle section is connected to the endoscope's instrument channel port via the single use adapter biopsy valve (MAJ-1414). The insertion section is composed of the sheath, needle and stylet. The needle tube is stored in the sheath and extended from the sheath to puncture the target tissue to collect specimen by moving the needle slider on the handle. A syringe is attached to the aspiration port on the handle section to aspirate the specimen. The needle is dimpled for echo enhancement. The subject devices will be sold with or without medallion syringe manufactured by Merit Medical, which consists of VACLOK Syringe and Stopcock.

AI/ML Overview

This is a submission for a medical device called the "Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N". This document is a 510(k) summary, which means it's a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device.

Here's an analysis of the provided information concerning acceptance criteria and the study that proves the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly provide a table of acceptance criteria with corresponding performance results for each criterion in a summarized format. Instead, it lists the types of performance data collected and states that these tests were conducted "to ensure that the subject devices perform as intended and meet design specifications." The overall conclusion is that the device "raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, efficacy and performance."

Based on the "Performance testing - Bench" section, we can infer some of the areas where acceptance criteria would have been applied:

Acceptance Criterion (Inferred from Test Conducted)Reported Device Performance (Implied from Summary)
Inserting into endoscopePerformed as intended and met design specifications.
Flexibility of the insertion portionPerformed as intended and met design specifications.
Piercing capabilityPerformed as intended and met design specifications.
Ultrasound visibilityPerformed as intended and met design specifications.
Needle extraction and retractionPerformed as intended and met design specifications.
Aspiration capabilityPerformed as intended and met design specifications.
Withdrawal from endoscopePerformed as intended and met design specifications.
Locking force of handle portionPerformed as intended and met design specifications.
Limitation of needle depthPerformed as intended and met design specifications.
Sterility (Sterilization/Shelf life testing)Met requirements in accordance with FDA Guidance and ASTM F1980-16. Real-time aging is ongoing.
BiocompatibilityMet requirements in accordance with FDA Guidance and ISO 10993-1.
Risk ManagementRisk analysis conducted, and design verification tests identified and performed according to ISO 14971:2007.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes for each of the performance tests (e.g., how many needles were tested for piercing or ultrasound visibility). It only states that tests were "conducted."

The provenance of the data is in-house testing conducted by Olympus Medical Systems Corp. or its contractors. The studies appear to be bench tests and lab-based assessments (e.g., sterilization, biocompatibility, mechanical performance). There is no mention of human clinical data or patient data, so the concepts of "country of origin of the data," "retrospective or prospective," and "test set" in the context of clinical data do not apply here.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Given that the performance data is entirely based on bench testing (mechanical, biocompatibility, sterilization), the concept of "experts establishing ground truth for a test set" in a clinical diagnostic sense (e.g., for image interpretation) is not applicable. The "ground truth" for these engineering tests would be established by validated test methods, reference standards, and established physical/chemical properties, rather than expert human interpretation.

4. Adjudication Method for the Test Set

As the performance data is derived from bench testing, an adjudication method like 2+1 or 3+1 (typically used for resolving discrepancies in expert interpretations of clinical data) is not relevant. The results of bench tests would be determined by objective measurements against predefined specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No MRMC comparative effectiveness study was done. This device is a physical medical instrument (an aspiration needle) and does not involve AI or human interpretation in a diagnostic workflow where such a study would be applicable.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This device is not an algorithm or AI-based system, so a standalone performance study in that context is not applicable.

7. The Type of Ground Truth Used

The ground truth for the performance testing is based on:

  • Design Specifications: The device was tested to perform "as intended and meet design specifications."
  • International Standards: Compliance with standards like ISO 10993-1, ISO 14971, ASTM F1980-16, and various ISO standards for sterilization and packaging.
  • Validated Test Methods: The bench tests themselves would have employed validated methods to objectively measure parameters like flexibility, piercing force, aspiration volume, etc.

8. The Sample Size for the Training Set

This device does not utilize a "training set" in the context of machine learning or AI. The design and manufacturing processes are refined through engineering development and quality control, not through iterative training on a dataset.

9. How the Ground Truth for the Training Set Was Established

Since there is no training set in the AI sense, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 15, 2018

Olympus Medical Systems Corp. % Sheri L. Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034-0610

Re: K180668

Trade/Device Name: Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: II Product Code: FCG Dated: October 11, 2018 Received: October 15, 2018

Dear Sheri L. Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Glenn B. Bell -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K180668

Device Name

Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N

Indications for Use (Describe)

This instrument has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration(FNA) of submucosal and extramural lesions of the gastrointestinal tract.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are thick and sans-serif, giving the word a strong and modern appearance. A thin yellow line is visible beneath the word, adding a subtle accent to the logo.

October 10, 2018

5. 510(k) Summary

5.1 GENERAL INFORMATION

  • 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
  • Sheri L. Musgnung ■ Contact Person: Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung(@olympus.com

5.2 DEVICE IDENTIFICATION

■ Device NameSingle Use Aspiration Needle
■ Model NameNA-U401SX-4025, NA-U401SX-4025N
■ Common NameAspiration Needle
■ Regulation Number21 CFR 876.1075
■ Regulation NameGastroenterology-urology biopsy instrument
■ Regulatory ClassII
■ Product CodeFCG
■ Classification PanelGastroenterology/Urology

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The word is positioned above a thick, horizontal, yellow line. The font is sans-serif and the letters are closely spaced, giving the word a strong, unified appearance. The overall design is simple and clean, with a focus on the brand name.

K180668

5.3 PREDICATE DEVICE

1) Predicate device

Table 12-1 Primary predicate device for NA-U401SX-4025/NA-U401SX-4025N

Device name510(k) Submitter510(k) No.
NA-U401SX-4022(Single Use Aspiration NeedleNA-U401SX)OLYMPUS MEDICALSYSTEMS CORP.K160098

2) Reference device

Table 12-2 Reference device for NA-U401SX-4025/NA-U401SX-4025N

Device name510(k) Submitter510(k) No.
ECHO-HD-25-EBUS-O(Echotip Ultra Endobronchial HighDefinition Ultrasound Biopsy Needle)Cook Ireland Ltd.K160229

5.4 DEVICE DESCRIPTION

1) General Description of the subject device

The Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N (aka ViziShot 2) is used in conjunction with Olympus ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA).

The subject devices consist of a handle section and insertion. The handle section is connected to the endoscope's instrument channel port via the single use adapter biopsy valve (MAJ-1414). The insertion section is composed of the sheath, needle and stylet. The needle tube is stored in the sheath and extended from the sheath to puncture the target tissue to collect specimen by moving the needle slider on the handle. A syringe is attached to the aspiration port on the handle section to aspirate the specimen. The needle is dimpled for echo enhancement.

The subject devices will be sold with or without medallion syringe manufactured by Merit Medical, which consists of VACLOK Syringe and Stopcock.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. There is a yellow line underneath the word. The registered trademark symbol is located to the right of the letter "S".

2) Principle of Operation

Ultrasound emitted by ultrasound transducer of ultrasound endoscope diffusely reflects at dimples on the echo-enhanced region of needle tube. The ultrasound transducer detects the part of ultrasound which diffusely reflects and the echo-enhanced region is indicated on ultrasound image.

The needle tube and stylet are extracted from the sheath by moving the needle slider to distal side and is drawn into the sheath by moving the needle slider in the counter direction.

The subject devices are used with specified ultrasound endoscopes for aspirating tissue or cells by piercing the target area in ultrasound image and attaching the medallion syringe to the aspiration port on the handle section.

5.5 INDICATIONS FOR USE

This instrument has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the gastrointestinal tract.

5.6 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE

The Single Use Aspiration Needles, NA-U401SX-4025/NA-U401SX-4025N have the same technological characteristics and design as the predicate device except for the following new features:

  • Needle width
  • Needle shape
  • Needle tip shape
  • Sheath composition
  • New colored sheath (NA-U401SX-4025 only)

All other technological characteristics of both the subject and predicate devices are identical.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The letters are closely spaced and fill most of the upper portion of the image. Below the word "OLYMPUS" is a horizontal bar in a gradient of yellow and gold. The bar is positioned directly beneath the text, providing a visual base for the logo.

K180668

5.7 PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

1) Sterilization/Shelf life testing

Sterilization/shelf life life testing for Single Use Aspiration Needle. NA-U401SX-4025/NA-U401SX-4025N was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".

Accelerated aging test for the NA-U401SX-4025/NA-U401SX-4025N was conducted in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. The real-time aging test for three-years will be performed to demonstrate longer stability and support the results of the accelerated aging test.

2) Biocompatibility testing

Biocompatibility testing for the the Single Use Aspiration Needle, NA-U401SX-4025/NA-U401SX-4025N were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1:2009, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process''.

3) Performance testing - Bench

Bench testing for the Single Use Aspiration Needle,

NA-U401SX-4025/NA-U401SX-4025N as listed below was conducted to ensure that the subject devices perform as intended and meet design specifications.

  • Inserting into endoscope
  • Flexibility of the insertion portion
  • Piercing
  • Ultrasound visibility
  • Needle extraction and retraction
  • Aspiration
  • Withdrawal from endoscope
  • Locking force of handle portion
  • Limitation of needle depth

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are a sans-serif font and are evenly spaced. There is a yellow line underneath the word. The registered trademark symbol is in the upper right corner of the image.

4) Risk analysis

Risk analysis for the Single Use Aspiration Needle, NA-U401SX-4025/NA-U401SX-4025N was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analvsis assessment.

The following standards have been applied to the Single Use Aspiration Needle, NA-U401SX-4025/NA-U401SX-4025N.

  • ISO 10993-1: 2009 - AAMI/ANSI/ISO 10993-5: 2009 - ISO 10993-10: 2010 - ISO 10993-11: 2006 - ISO 11135: 2014 - ISO 10993-7: 2008+Cor1 - ISO 11607-1: 2006+A1 - ISO 11607-2: 2006+A1 - ASTM F1980-16 - ISO 14971: 2007

5.8 CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate devices, the NA-U401SX-4025/ NA-U401SX-4025N raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, efficacy and performance.

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.