(30 days)
SmartGrid feature is a software option that provides, upon request by user, a diagnostic radiograph image with a reduction in visible x-ray scatter similar to the effect of an anti-scatter grid.
The SmartGrid II software is designed to improve contrast and reduce the appearance of scatter in radiographic images that have been acquired without a physical grid. SmartGrid II encapsulates an algorithm for estimating and removing scatter from radiographic images.
The SmartGrid II software feature is accessible through the DirectView application software. Users will be able to select SmartGrid II processing before an image is acquired, or to change whether SmartGrid II processing is applied to a previously acquired image.
The Carestream SmartGrid II is a software option designed to reduce visible x-ray scatter in diagnostic radiograph images, similar to the effect of an anti-scatter grid. It is an improvement to the original SmartGrid software.
Here's a breakdown of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Successful installation of the SmartGrid II software component. | Installation testing was performed and verified successful installation of the SmartGrid II software. |
| Consistent production of diagnostic quality images by SmartGrid II processing. | Specific test cases were implemented and verified that SmartGrid II image processing consistently produces a diagnostic quality image. |
| Sufficient reduction of image noise and sufficient image contrast/brightness for clinical diagnosis. | Specific test cases were implemented and verified that image noise was sufficiently reduced and image contrast/brightness were sufficient for clinical diagnosis. |
| Proper performance of the SmartGrid II software feature when integrated into the Carestream application software. | Integration tests were performed and ensured proper performance of the SmartGrid II software feature when integrated into the Carestream application software. |
| No unexpected negative effects (clinical evaluation with respect to diagnostic image quality). | A reader study demonstrated that images processed with SmartGrid II were equivalent in diagnostic quality to images processed with the predicate device (original SmartGrid). Also, error handling was observed for proper function through monitoring of event logs, and Ad Hoc testing was performed to help ensure no unexpected negative effects. |
| Conformance of the device to its specifications. | Non-clinical test results demonstrated the intrinsic workflow, related performance, overall function, verification and validation of requirements for intended use, and reliability of the software. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device. |
| Substantial equivalence to the legally marketed predicate device (SmartGrid, K163157) in terms of safety and effectiveness. | Performance testing results support a substantial equivalence determination. The differences between SmartGrid II and the predicate device do not affect the intended use or alter the fundamental scientific technology. Both devices use scatter factor estimation, scatter correction, noise reduction, and image rendering. The improved algorithms in SmartGrid II for scatter estimation and correction (based on the full detector field versus collimated subregion) lead to improvements in diagnostic quality. |
2. Sample Size and Data Provenance:
- Test Set Sample Size: The document does not explicitly state the numerical sample size of images used for the clinical reader study. It mentions "existing clinical research images" were evaluated.
- Data Provenance: The document does not specify the country of origin. It indicates the study used "existing clinical research images," implying a retrospective nature where stored images were processed and then evaluated.
3. Number of Experts and Qualifications:
- Number of Experts: One board-certified radiologist was used for the clinical evaluation.
- Qualifications: Board-certified radiologist. No specific number of years of experience is provided.
4. Adjudication Method for the Test Set:
- The document describes a reader study where a single board-certified radiologist evaluated images processed with both the original SmartGrid and SmartGrid II. There is no mention of an adjudication method as only one reader was used.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, a multi-reader, multi-case comparative effectiveness study was not explicitly described. The clinical evaluation involved a single radiologist. Therefore, there is no reported effect size of human readers improving with AI vs. without AI assistance from an MRMC study in this document.
6. Standalone Performance Study (Algorithm only):
- Yes, a standalone performance evaluation of the algorithm was conducted as part of the non-clinical (bench) testing. This included verifying consistent diagnostic quality image production, sufficient noise reduction, and appropriate contrast/brightness for clinical diagnosis.
7. Type of Ground Truth Used:
- For the clinical evaluation, the ground truth was essentially based on the expert opinion/consensus (or lack thereof, since only one expert) on diagnostic image quality comparison between the two software versions (SmartGrid II vs. predicate SmartGrid). The study aimed to confirm "no unexpected negative effects with respect to diagnostic image quality" and that SmartGrid II images were "equivalent in diagnostic quality" to predicate images.
- For the non-clinical bench testing, the ground truth would be represented by the predefined specifications and requirements for image quality metrics (noise, contrast, brightness) and system functionality.
8. Sample Size for the Training Set:
- The document does not provide information regarding the sample size of the training set used for developing the SmartGrid II algorithm. It primarily focuses on the validation of the updated algorithm.
9. How the Ground Truth for the Training Set was Established:
- The document does not provide information on how the ground truth for the training set was established for the SmartGrid II algorithm. It mentions that "the SmartGrid and SmartGrid II algorithms differ slightly to further improve the scatter estimation and correction," suggesting an iterative development process that would have involved some form of ground truth or reference data, but details are not included.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and below that, the word "ADMINISTRATION" in a smaller font, also in blue.
Carestream Health, Inc. % Ms. Diane Koetter Sr. Manager, Regulatory Affairs and Quality Systems 150 Verona Street ROCHESTER NY 14608
April 13, 2018
Re: K180667
Trade/Device Name: SmartGrid Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: March 12, 2018 Received: March 14, 2018
Dear Ms. Koetter:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
(1-800-638-2041 or 301-796-7100).
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
Sincerely,
Michael D'Hara For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K180667
Device Name SmartGrid
Indications for Use (Describe)
"SmartGrid feature is a software option that provides, upon request by user, a diagnostic radiograph image with a reduction in visible x-ray scatter similar to the effect of an anti-scatter grid."
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Carestream
"510(k) Summary"
| 510(k) Owner Name: | Carestream Health, Inc. |
|---|---|
| 510(k) Owner Address: | 150 Verona Street |
| Rochester, New York 14608 | |
| 510(k) Owner Phone: | 585 627-6505 |
| 510(k) Owner Fax: | 585 627-8802 |
| Contact Person & Info: | Diane Koetter |
| Sr. Manager, Regulatory Affairs and Quality Systems | |
| diane.koetter@carestream.com | |
| 585-627-6505 | |
| Date Summary Prepared: | April 11, 2018 |
| Device Trade Name: | SmartGrid |
| Device Common Name: | Flat Panel Digital Detector System |
| Classification Name: | Stationary x-ray system |
| Device Class: | II |
| Device Code: | MQB |
| Regulation Number: | 21 CFR 892.1680 |
| Predicate Device: | SmartGrid |
| Manufactured by: Carestream Health, Inc. | |
| 510(k) No.: K163157 (March 21, 2017) |
Please Note: The marketing name for the modified SmartGrid device is not expected to change. The modified device is referred to as "SmartGrid II" within this submission for clarity.
Device Description:
The SmartGrid II software is designed to improve contrast and reduce the appearance of scatter in radiographic images that have been acquired without a physical grid. SmartGrid II encapsulates an algorithm for estimating and removing scatter from radiographic images.
The SmartGrid II software feature is accessible through the DirectView application software. Users will be able to select SmartGrid II processing before an image is acquired, or to change whether SmartGrid II processing is applied to a previously acquired image.
{4}------------------------------------------------
Indications for Use / Intended Use:
The Indications for Use for the device, as described in its labeling, are:
"SmartGrid feature is a software option that provides, upon request, a diagnostic radiograph image with a reduction in visible x-ray scatter similar to the effect of an anti-scatter grid."
The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above.
The Indications for Use for the subject device is the same as for the predicate device and the intended use remains unchanged. Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination.
Comparison of Technological Characteristics:
Based upon information provided within this submission, we believe that the SmartGrid II software is substantially equivalent to the legally marketed SmartGrid (predicate device).
SmartGrid II has the same technological characteristics as SmartGrid (predicate device). Both are optional software components that allow for reduction in visible x-ray scatter in radiographic images.
Both devices use:
- Scatter Factor Estimation ●
- Scatter Correction
- Noise Reduction and Image Rendering .
The predicate device differs from the investigational device in the methods used for Scatter Estimation and Scatter Correction. For instance, the original SmartGrid feature estimated and corrected scatter based on the exposed area within the collimation field. SmartGrid II estimates and corrects scatter based on the full detector field rather than the collimated subregion. In addition, the SmartGrid and SmartGrid II algorithms differ slightly to further improve the scatter estimation and correction.
Discussion of Testing:
The performance characteristics and operation / usability of the SmartGrid II software were evaluated in non-clinical (bench) testing. These studies demonstrated the intended workflow, related performance, overall function, verification and validation of requirements for intended use, and reliability of the software. Non-clinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.
{5}------------------------------------------------
Specifically, the following "bench" testing was performed as described below:
- Installation testing was performed to verify successful installation of the SmartGrid ● II software component onto the representative system.
- Specific test cases were implemented to verify that SmartGrid II image processing . consistently produces a diagnostic quality image
- . Specific test cases were implemented to verify that image noise was sufficiently reduced and image contrast/brightness were sufficient for clinical diagnosis
- . Regression testing was performed using manual and automated test case execution
- Integration tests were performed to ensure proper performance of the SmartGrid II ● software feature when integrated into the Carestream application software
- Error handling was observed for proper function through monitoring of event logs
- Ad Hoc testing was performed as considered applicable to help ensure no . unexpected negative effects
In addition to bench testing, existing clinical research images processed with both the original SmartGrid and SmartGrid II software were evaluated by a board certified radiologist. This clinical evaluation was intended to confirm no unexpected negative effects with respect to diagnostic image quality associated with the use of the improved algorithm. The reader study demonstrated that images processed with SmartGrid II were equivalent in diagnostic quality to images processed with the predicate device, SmartGrid.
Summary:
The differences between SmartGrid II and the predicate device do not affect the intended use of the device or alter the fundamental scientific technology of the device. Performance testing results support a substantial equivalence determination of SmartGrid II to the predicate device.
FDA Reference Materials:
The following FDA guidance documents were consulted in the development of the SmartGrid II software feature:
- . General Principles of Software Validation – Final Guidance for Industry and FDA Staff; issued 01/11/2002
- Off-The-Shelf Software Use in Medical Devices – Guidance for Industry and FDA Reviewers, and Compliance; issued 09/09/1999
- Content of Premarket Submissions for Management of Cybersecurity in Medical ● Devices - Guidance for Industry and Food and Drug Administration Staff; issued 10/02/2014
- . Deciding When to submit a 510(k) for a Software Change to an Existing Device -Guidance for Industry and Food and Drug Administration Staff; issued 10/25/2017
- Guidance for the Content of Premarket Submissions for Software Contained in . Medical Devices: issued 05/11/2005
N/A