K Number
K180643

Validate with FDA (Live)

Date Cleared
2018-11-23

(256 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MedCu Antibacterial Wound Dressings with Copper-Oxide are indicated for the management of wounds and to provide a physical bacterial barrier (adhesive contour type only). The dressings are applied topically and are in direct contact with the wound. The dressings are intended to be used for indications as follows:

Over-The-Counter (OTC) use indications:

Local management of superficial wounds, minor burns, abrasions and lacerations (tears).

Prescription use indications (Rx-only):

  • · Partial and full thickness wounds,
  • · Pressure ulcers (stage I-IV),
  • · Diabetic ulcers,
  • · Venous stasis ulcers,
  • · Arterial ulcers,
  • · 1st and 2nd degree burns.
  • · Surgical wounds,
  • · Vascular access or peripheral IV sites,
  • · Orthopedic external pin sites, and
  • · Acute wounds such as lacerations, abrasions, and skin tears.
Device Description

The MedCu Antibacterial Wound Dressings with Copper-Oxide are sterile, soft, single use, barrier wound dressings composed of an internal absorbent layer and one or two external nonwoven layers (only the adhesive contour model serves as physical barrier). The external layer(s) are made of Polypropylene nonwoven fabric and the internal layer is made of needle punch layer made of Polyester and Cellulose fibers. The wound dressings achieve at least 500% fluid absorbency.

Both the internal and external layers are impregnated with copper-oxide particles. The one or two external layers cover the internal layer from one or both sides, accordingly. The wound dressing size is of 10 cm x 10 cm or 5 cm x 5 cm (other variations may be included as well), and may be provided with or without an adhesive contour (adhesive border), which provides physical bacterial barrier. The wound dressings without an adhesive contour can be cut per wound shape. The Copper-Oxide ions are released within the wound dressing from the polymeric matrix of each layer in the presence of moisture. The MedCu ABWDs, reduced the tested bacteria by more than 4 log within the wound dressing based on in-vitro testing.

MedCu ABWDs absorb wound fluid, contain copper-oxide to inhibit bacterial growth within the dressing and serves as a *barrier against bacterial penetration.

*Note: Barrier against bacterial penetration is achieved only by the adhesive contour model.

The wound dressing is provided sterile and kept sterile in a sterilization pouch for its entire shelf life.

Body contact materials were evaluated for biocompatibility with accordance to FDA's Guidance document for Use of ISO 10993-1, Biological Evaluation of Medical Devices – Part 1 : Evaluation and Testing with a Risk Management Process, dated June 16, 2016 and the ISO 10993-1:2009 standard.

AI/ML Overview

The provided text describes the MedCu Antibacterial Wound Dressings with Copper-Oxide, whose 510(k) Premarket Notification (K180643) was reviewed by the FDA. The document outlines the device's indications for use, description, and performance data used to establish substantial equivalence with predicate devices.

However, the document does not contain the following information typically associated with typical diagnostic AI/ML device approval:

  • Acceptance Criteria for Performance: It describes various tests conducted but does not specify a quantitative "acceptance criterion" (e.g., sensitivity, specificity, or accuracy threshold) that the device must meet. Instead, it refers to "successfully tested," "at least 4 log reduction," and "at least 500% absorbency."
  • Study That "Proves" Device Meets Acceptance Criteria: There isn't a single, clearly defined study designed to meet a pre-set performance threshold. The performance data section lists several non-clinical tests (chemical, biological, antibacterial effectiveness, bacterial barrier, absorbency, particle analysis, leachables, MEC, packaging shelf life, sterilization, pyrogenicity) that support safety and effectiveness claims and "establish substantial equivalency."
  • Sample Size for Test Set: Not applicable as these are non-clinical, in-vitro, or bench tests, not clinical studies with a "test set" of patient data.
  • Data Provenance: Not applicable.
  • Number of Experts and Qualifications: Not applicable.
  • Adjudication Method: Not applicable.
  • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: The device is a wound dressing, not a diagnostic AI/ML device, so such a study is not relevant or included.
  • Standalone (Algorithm Only) Performance: Not applicable.
  • Type of Ground Truth: For the "antibacterial effectiveness testing," the ground truth is the measured reduction in viable bacterial titers (colony-forming units) as compared to a control or baseline. For "biocompatibility," the ground truth is established through standard toxicological and biological assays comparing the device's effects to established safety profiles.
  • Sample Size for Training Set: Not applicable as this is not an AI/ML device that requires a training set.
  • How Ground Truth for Training Set Was Established: Not applicable.

Given these limitations from the provided text, I can only populate the table with the performance data and claims mentioned, and clarify that other requested information is not present or relevant for this type of device according to the document.


1. Table of Acceptance Criteria and Reported Device Performance

As specific acceptance criteria (e.g., "must achieve X% sensitivity") are not explicitly stated in the document for the overall device performance, the table reflects the performance claims or results reported from the non-clinical tests.

Acceptance Criterion (Implicit/Claimed)Reported Device Performance (from K180643)
BiocompatibilitySuccessfully evaluated for biocompatibility with accordance to FDA's Guidance document for Use of ISO 10993-1, Biological Evaluation of Medical Devices – Part 1 and ISO 10993-1:2009 standard, including: Exhaustive extractions, FTIR, GC-MS, ICP-OES, UPLC-MS, Cytotoxicity, Skin irritation and Intracutaneous reactivity, Acute systemic toxicity, Sensitization (Maximization), Wound healing study on large animal (porcine).
Antibacterial EffectivenessAchieves more than 4 log reduction of viable titers of 6 bacteria (E. coli, E. faecalis, E. aerogenes, K. pneumoniae, S. epidermis, and MRSA) within the wound dressing, based on in-vitro testing per AATCC 100, following real-time aging.
Bacterial Barrier EffectivenessDemonstrated for the device intended use duration (adhesive contour model only).
Absorbance CapacityAchieves at least 500% absorbency of the wound dressing weight. (Reported 2 layers: ~79.4 oz/yd²; 3 layers: ~75.6 oz/yd²). Note: The 500% absorbency is a general claim, while the oz/yd² are specific measurements provided in the comparative table.
Copper-Oxide Particle IntegrityAnalyzed for size, shape, and distribution using ICP, SEM, EDS, and DLS. Confirmed copper oxide particles were not released (leachables) using DLS method. Minimal Effective Concentration (MEC) study evaluated for 4 log reduction.
Shelf LifePackaging shelf life validated using accelerated aging method according to ISO 11607-1.
SterilizationValidated using EtO half cycle method according to ISO 11135:2014, including EtO/ECH residuals according to ISO 10993-7:2008.
PyrogenicityPyrogenicity test conducted using LAL method and Material-Mediated Pyrogen study.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not applicable in the context of typical AI/ML diagnostic device test sets. The document describes non-clinical laboratory tests (e.g., in-vitro antibacterial tests, chemical analyses, biological evaluations on material, pig wound healing study). Specific sample sizes for each of these non-clinical tests (e.g., number of bacterial replicates, number of animals in the wound healing study) are not detailed in the provided summary.
  • Data Provenance: The tests are described as in-vitro and laboratory-based studies performed by the manufacturer (MedCu Technologies, Ltd.). The wound healing study was conducted on a "large animal (porcine)." No human clinical data or geographical origin of patient data is mentioned in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Not applicable. This device is a wound dressing, and the performance evaluation relies on laboratory test standards and biological assays, not interpretation by human experts to establish a "ground truth" for a diagnostic task.

4. Adjudication Method for the Test Set

  • Not applicable. (No expert adjudication involved for this type of non-clinical testing).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Not applicable. This is a medical device (wound dressing), not a diagnostic AI/ML system that would typically involve human readers interpreting medical images or data.

6. Standalone (Algorithm Only Without Human-in-the Loop Performance) Study

  • Not applicable. This is a medical device (wound dressing), not an AI/ML algorithm.

7. The Type of Ground Truth Used

  • Biocompatibility: Ground truth is established by standardized biological and chemical assays that measure cellular response, toxicity, irritation, sensitization, and systemic effects according to ISO and FDA guidelines. For the wound healing study, the ground truth would be the observation of wound healing processes in the animal model.
  • Antibacterial Effectiveness: The ground truth is the direct measurement of viable bacterial titers (colony-forming units, CFU) after exposure to the device, compared to control groups. A reduction of "more than 4 log" signifies a 99.99% reduction in bacteria.
  • Bacterial Barrier: Ground truth is the measured ability of the dressing to prevent bacterial passage.
  • Absorbency: Ground truth is the direct measurement of fluid absorption by weight.
  • Material Characterization (Copper Oxide particles, leachables): Ground truth is established by analytical chemistry techniques (ICP, SEM, EDS, DLS) and direct measurement of concentrations or particle properties.

8. The Sample Size for the Training Set

  • Not applicable. This is not an AI/ML device.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below.

November 23, 2018

MedCu Technologies, Ltd. % Tali Hazan RA Consultant Talmed Ltd. MP Upper Galilee Ramot Naftali, 1383000 IL

Re: K180643

Trade/Device Name: MedCu Antibacterial Wound Dressings with Copper-Oxide (MedCu ABWDs) Regulatory Class: Unclassified Product Code: FRO Dated: October 20, 2018 Received: October 23, 2018

Dear Tali Hazan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kimberly Ferlin -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180643

Device Name

MedCu Antibacterial Wound Dressings with Copper-Oxide (also identified as MedCu ABWDs)

Indications for Use (Describe)

The MedCu Antibacterial Wound Dressings with Copper-Oxide are indicated for the management of wounds and to provide a physical bacterial barrier (adhesive contour type only). The dressings are applied topically and are in direct contact with the wound. The dressings are intended to be used for indications as follows:

Over-The-Counter (OTC) use indications:

Local management of superficial wounds, minor burns, abrasions and lacerations (tears).

Prescription use indications (Rx-only):

  • · Partial and full thickness wounds,
  • · Pressure ulcers (stage I-IV),
  • · Diabetic ulcers,
  • · Venous stasis ulcers,
  • · Arterial ulcers,
  • · 1st and 2nd degree burns.
  • · Surgical wounds,
  • · Vascular access or peripheral IV sites,
  • · Orthopedic external pin sites, and
  • · Acute wounds such as lacerations, abrasions, and skin tears.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "MEDCu" in a stylized font. The letters "MED" are in blue, while the letters "Cu" are in brown and slightly overlap the "D". The font is bold and sans-serif, giving the word a modern and clean look. The overall design is simple yet eye-catching, with the contrasting colors adding visual interest.

510(K) SUMMARY FOR MEDCU ANTIBACTERIAL WOUND DRESSINGS WITH COPPER-OXIDE (ABWDS)

DATE PREPARED: November 11, 2018

  • 510(K) OWNER NAME 1.
Company Name:MedCu Technologies Ltd.
Address:23 David Elazar St.Herzliya 4627703Israel
Phone:+972-546-611287
E-mail:gadib@medcu.com (Dr. Gadi Borkow)
Contact PersonName:Tali Hazan, RA Consultant
Address:Ramot-Naftali, 1383000Israel
Phone:+972-505-292304
Fax:+972-72-2448981
E-mail:tali.hazan@talmed.co.il

2. DEVICE NAME

Common/Usual Name:Wound Dressings
Proprietary/Trade name:MedCu Antibacterial Wound Dressings with Copper-Oxide (also identified as MedCu ABWDs)

Classification: The MedCu Antibacterial Wound Dressings with Copper-Oxide are unclassified medical devices under the following classification names:

Classification NameProduct Code21 CFR Ref.Panel
Dressing, Wound, DrugFROPre-Amendment,UnclassifiedGeneral andPlastic Surgery

{4}------------------------------------------------

MEDCU

3. PREDICATE DEVICES

MedCu Antibacterial Wound Dressings with Copper-Oxide are substantially equivalent to the following Predicate Devices:

  • 3.1 Argentum Medical's Silverlon® Wound Pad Dressing & Silverlon® Island Wound Dressing, cleared under 510(k) number K143001 at November 7, 2014;
  • 3.2 ProCyte Corporation's Lamin® Wet Dressing (Copper-Saline), cleared under 510(k) number K964468 at February 3, 1997.

DEVICE DESCRIPTION 4.

The MedCu Antibacterial Wound Dressings with Copper-Oxide are sterile, soft, single use, barrier wound dressings composed of an internal absorbent layer and one or two external nonwoven layers (only the adhesive contour model serves as physical barrier). The external layer(s) are made of Polypropylene nonwoven fabric and the internal layer is made of needle punch layer made of Polyester and Cellulose fibers. The wound dressings achieve at least 500% fluid absorbency.

Both the internal and external layers are impregnated with copper-oxide particles. The one or two external layers cover the internal layer from one or both sides, accordingly. The wound dressing size is of 10 cm x 10 cm or 5 cm x 5 cm (other variations may be included as well), and may be provided with or without an adhesive contour (adhesive border), which provides physical bacterial barrier. The wound dressings without an adhesive contour can be cut per wound shape. The Copper-Oxide ions are released within the wound dressing from the polymeric matrix of each layer in the presence of moisture. The MedCu ABWDs, reduced the tested bacteria by more than 4 log within the wound dressing based on in-vitro testing.

MedCu ABWDs absorb wound fluid, contain copper-oxide to inhibit bacterial growth within the dressing and serves as a *barrier against bacterial penetration.

*Note: Barrier against bacterial penetration is achieved only by the adhesive contour model.

The wound dressing is provided sterile and kept sterile in a sterilization pouch for its entire shelf life.

Body contact materials were evaluated for biocompatibility with accordance to FDA's Guidance document for Use of ISO 10993-1, Biological Evaluation of Medical Devices – Part 1 : Evaluation and Testing with a Risk Management Process, dated June 16, 2016 and the ISO 10993-1:2009 standard.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "MEDCU" in a stylized font. The letters "MED" are in a bright blue color, while the letters "CU" are in a copper color. The "C" and "U" overlap the "D" in the word.

5. INTENDED USE

The MedCu Antibacterial Wound Dressings with Copper-Oxide are indicated for the management of wounds and to provide a physical bacterial barrier (adhesive contour type only). The dressings are applied topically and are in direct contact with the wound. The dressings are intended to be used for indications as follows:

Over-The-Counter (OTC) use indications:

Local management of superficial wounds, minor burns, abrasions and lacerations (tears).

Prescription use indications (Rx-only):

  • Partial and full thickness wounds, ●
  • Pressure ulcers (stage I-IV),
  • Diabetic ulcers, ●
  • Venous stasis ulcers, ●
  • Arterial ulcers, ●
  • 1st and 2nd degree burns,
  • Surgical wounds,
  • Vascular access or peripheral IV sites. ●
  • Orthopedic external pin sites, and ●
  • Acute wounds such as lacerations, abrasions, and skin tears. ●

TECHNOLOGICAL CHARACTERISTICS 6.

The MedCu non-adherent Antibacterial Wound Dressings with Copper-Oxide are impregnated with copper-oxide particles, which serve as an antibacterial agent to achieve log reduction of at least 4 log of wide range of bacteria within the wound dressing.

The device containing adhesive contour also serves as a bacterial barrier to confer additional protection to the wound.

The wound dressings' non-adherent external layer reduces pain during dressing removal.

7. PERFORMANCE DATA

MedCu Antibacterial Wound Dressings with Copper-Oxide have been successfully tested through laboratory non-clinical tests to support the safety and effectiveness of the device and the determination of substantial equivalence with predicate devices. MedCu conducted the following tests:

..............................................................................................................................................................................

{6}------------------------------------------------

EDCu

  • 7.1 Chemical and biological tests according to ISO 10993-1, ISO 10993-12, ISO 10993-18 and FDA Guidance document for Biological Evaluation June 2016: Exhaustive extractions, FTIR, GC-MS, ICP-OES, UPLC-MS, Cytotoxicity, Skin irritation and Intracutaneous reactivity, Acute systemic toxicity, Sensitization (Maximization), Wound healing study on large animal (porcine).
  • 7.2 Antibacterial effectiveness testing per AATCC 100 following real-time aging products showing more than 4 log reduction of viable titers of 6 bacteria for the device intended use duration. The bacteria tested are: E. coli, E. faecalis, E. aerogenes, K. pneumoniae, S. epidermis and MRSA.
  • 7.3 Bacterial barrier effectiveness for the device intended use duration (adhesive contour model only).
  • 7.4 Absorbance capacity of at least 500% of the wound dressing weight.
  • 7.5 Copper Oxide particles size and shape analyses using ICP (Inductively Coupled Plasma), SEM (Scanning Electron Microscopy) and EDS (Energy Dispersive X-ray Spectrometry) analyses.
  • 7.6 Copper Oxide particle size distribution using DLS (Dynamic Light Scattering) method.
  • Copper Oxide leachables in order to determine if copper oxide particles were 7.7 released from the antibacterial wound dressing, using DLS method.
  • 7.8 MEC (Minimal Effective Concentration) study evaluating the minimal effective concentration required to achieve at least 4 log reduction of the tested microorganisms.
  • 7.9 Packaging shelf life using accelerated aging method according to ISO 11607-1.
  • 7.10 Sterilization validation using EtO half cycle method according to ISO 11135:2014, including EtO/ECH residuals according to ISO 10993-7:2008.
  • 7.11 Pyrogenicity test using LAL method and Material-Mediated Pyrogen study.

Tests results support all MedCu's antibacterial wound dressing labeling claims in order to establish substantial equivalency.

8. SUBSTANTIAL EQUIVALENCE

MedCu Antibacterial Wound Dressings with Copper-Oxide is substantial equivalent to the predicate devices selected in terms of indication for use, technology, performances, patient population and nature of body contact.

………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………

4

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the word "MEDCu" in a stylized font. The letters "MED" are in a bright blue color, while the letters "Cu" are in a brown color. The "Cu" overlaps with the "D" in "MED", creating a visual effect.

The substantial equivalent decision was received based on the following comparison with the predicate devices:

FeatureSilverlon® Wound PadDressing & Silverlon®Island Wound DressingPrimary PredicateDevice Cleared underK143001ProCyte Corp. LaminⓇWet Dressing(Copper Saline)Reference PredicateDevice Cleared underK964468MedCu AntibacterialDressing withCopper-Oxide- Subject Device -
Indication for useSilverlon® Island WoundDressing and Silverlon®Wound Pad Dressing aremulti-layer,sterile, non-adherent,antimicrobial barrierwound dressingsfor The Over-The-Counter Indications:Local management ofsuperficial woundsminor burns abrasionsand lacerations.and for PrescriptionIndications:Under the supervision ofa healthcare professionalSilverlon® Island WoundDressingsand Silverlon® WoundPad Dressings areintended for up to 7 dayuse for wounds suchas vascular access orperipheral IV sites,orthopedic external pinsites, wound drain sites,surgical wounds (donorand graft sites, incisions),partial to full thicknessdermalulcers (stage I-IVpressure sores, venousstasis ulcers, arterialThe device is a hydrogelfor the dressing andmanagement of bothgraft recipient and donorsites, postoperativeincisions, pressureulcers, diabetic ulcers,stasis ulcers, 1st and 2nddegree burns, arterialulcers, pressure sores,and skin conditionsassociated withperistomal care. Thedressing is intended tocover a wound or burnon a patient's skin,provide a moist woundenvironment, absorbwound exudate, andprotect against abrasion,friction, desiccation, andcontamination.The MedCuAntibacterial WoundDressings with Copper-Oxide are indicated forthe management ofwounds and to provide aphysical bacterialbarrier. The dressingsare applied topically andare in direct contact withthe wound. Thedressings are intended tobe used for indicationsas follows:Over-The-Counter(OTC) use indications:Local management ofsuperficial wounds,minor burns, abrasionsand lacerations (tears).Prescription useindications (Rx-only):Partial and full thicknesswounds, pressure ulcers(stage I-IV), diabeticulcers,venous stasis ulcers,arterial ulcers, 1st and 2nddegree burns, surgicalwounds, vascular accessor peripheral IV sites,orthopedic external pinsites, and acute woundssuch as lacerations,
FeatureSilverlon® Wound Pad Dressing & Silverlon® Island Wound DressingPrimary Predicate Device Cleared under K143001ProCyte Corp. Lamin® Wet Dressing(Copper Saline) Reference Predicate Device Cleared under K964468MedCu Antibacterial Dressing with Copper-Oxide - Subject Device -
Prescription / OTCBoth prescription and OTCUnknownBoth prescription and OTC
Design4 (Pad) or 5 (Island) Layers:Layer 1 - no-adherent knitted nylon fiber coated with metallic silver Layer 2 - polyethylene film used to bond the nylon substrate to the Layer 3; Layer 3 - absorptive laminate pad Layer 4 - polyethylene film that bonds the pad to Layer 5, Layer 5 – adhesive polyester fabric1- Layer of cotton-blend non-woven gauze saturated with copper saline2- or 3-layer: Internal layer made of absorbent cellulose and polyester layer containing copper-oxide one or two external non-adherent nonwoven polypropylene layers containing copper-oxide
DimensionsMultiple sizes including 10 cm x 10 cm and 5 cm x 5 cmUnknown10 cm x 10 cm, or 5 cm x 5 cm
MaterialNylon, polyethylene, polyesterCellulosePolypropylene, Polyester, Cellulose
Antimicrobial agentSilver / Silver-oxideCopper-salineCopper-oxide
BiocompatibilityBiocompatibleBiocompatibleBiocompatible
SterilizationSterile (EtO)SterileSterile (EtO)
Principle of operationAntimicrobial effect is achieved via the released silver ions within the dressing in moist environmentCopper saline dressing provides a moist wound environment and protects the woundAntibacterial effect is achieved via the released copper ions within the dressing in moist environment
TechnologyOnly the layer in contact with the skin is made of nylon fiber covered (coated) with silverCopper saline wet dressingAll layers of the wound dressing are impregnated with copper-oxide
FeatureSilverlon® Wound PadDressing & Silverlon® Island Wound DressingPrimary PredicateDevice Cleared underK143001ProCyte Corp. Lamin®Wet Dressing(Copper Saline)Reference PredicateDevice Cleared underK964468MedCu AntibacterialDressing withCopper-Oxide- Subject Device -
Adhesive / non-adhesive contourBothNoBoth
Recommendeddressing changeUp to 7 daysUnknownUp to 7 days
Absorbency65 oz/yd²Absorbs wound exudate2 layers: ~79.4 oz/yd²3 layers: ~75.6 oz/yd²
Single UseYesYesYes

{8}------------------------------------------------

MEDCu

б

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the word "MEDCu" in a stylized font. The letters "MED" are in a bright blue color, while the letters "Cu" are in a reddish-brown color. The "D" and "C" overlap, creating a visual effect where the colors blend together. The overall design is simple and modern.

Substantial Equivalence Discussion

As can be seen in the above comparison table, same claims are covered under the indications for use statement of MedCu proposed device as compared to both predicate devices.

Certain differences exist between the subject and predicate devices in terms of number of the dressing layers and antibacterial agent (silver used for Silverlon, while Lamin and MedCu are copper based).

Unlike Silverlon in MedCu ABWD the antibacterial agent exists in all layers (except for the adhesive contour layer, when such layer is included). This difference does not affect the intended purpose of the device or its safe use. The successful biocompatibility tests confirmed that the existence of the antibacterial agent throughout the entire dressing does not affect the safe use of the device. The ABWDs have an antibacterial effect within the dressing. This claim, which narrows the cleared Silverlon antimicrobial claim. does not affect the safe use of the device.

The predicate dressings' materials consist of biocompatible polymeric materials with and without cellulose. MedCu ABWD includes biocompatible polymeric materials and cellulose in all subject models. This difference does not alter the intended purpose of the device or its safe use since the intended use of both devices remains the same while MedCu chose to provide ABWDs only as absorbent wound dressings. The device was evaluated under full absorbency capacity simulating worst case clinical use and performed as intended.

All three dressings (two predicates and the subject device) absorb wound exudates, are provided sterile, biocompatible and intended for single use only.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the word "MEDCu" in a stylized font. The letters "MED" are in a bright blue color, while the letters "Cu" are in a reddish-brown color. The "Cu" letters are slightly overlapping the "D" letter, creating a visual effect.

The similarities and differences were thoroughly discussed within the 510(k) submission and it was concluded that the differences do not critically affect the intended clinical use and do not raise new questions of safety and effectiveness with regards to the proposed device.

In light of the above, we have concluded that substantial equivalence of the MedCu ABWD with the predicate devices was established.

9. CONCLUSIONS

After considering all substantial equivalency features of the predicate devices against the subject device, and after discussing similarities and differences of these features, we have concluded that MedCu ABWD (subject device) meets all proposed labeling claims and specification requirements. Certain differences between the predicate devices and the subject device were discussed and addressed within the 510(k) submission and found to be not significant in terms of safety and effectiveness. The evaluation of our device performances demonstrates that it is as safe and as effective as the predicate devices, and therefore the requirements of substantial equivalency are met.

……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………

N/A