K Number
K180545

Validate with FDA (Live)

Date Cleared
2018-05-21

(81 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VACUETTE® SAFELINK Holder with male luer lock is for use in blood collection procedures in combination with devices with a female luer port.

Device Description

The VACUETTE® SAFELINK Holder with male luer lock is a sterile, single-use, non-invasive device, which consists of a cylinder made from polypropylene plastic with a centric molded male luer lock at the top. The centric location of the male luer lock is to facilitate the connection to non-needle devices with a female luer port. A stainless steel needle with a sleeve is bonded to the inside of the holder to facilitate blood draw by puncturing the cap of an attached blood collection device. The sleeve is made from Synthetic Polyisoprene Compound.

AI/ML Overview

The provided document describes the predicate device equivalence testing for the VACUETTE® SAFELINK Holder with male luer lock (K180545). The device is a blood collection holder intended for use with female luer port devices.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

RequirementAcceptance CriteriaReported Device Performance
Liquid Leakage of rubber sleeve and ultrasonic welding area (including Kickback Testing)Equivalence to predicate device, internal specificationpass
Fitting strength between rubber sleeve and needle insertEquivalence to predicate device, internal specificationpass
Pull Out Force NeedleEquivalence to predicate device, internal specificationpass
Stress cracking and liquid leakage from fitting assembly under pressureTested in conformity to ISO 80369-7:2016pass
Air leakage into fitting assembly during aspirationTested in conformity to ISO 80369-7:2016pass
Separation force of fitting assemblyTested in conformity to ISO 80369-7:2016pass
Unscrewing torque of fitting assemblyTested in conformity to ISO 80369-7:2016pass
Resistance to overridingTested in conformity to ISO 80369-7:2016pass
Free hemoglobinBenchmark testingpass

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample sizes used for each performance test. It only states that "Performance testing was conducted to demonstrate and support our substantial equivalency claim". Given that this is a 510(k) submission for a physical medical device, the data provenance is likely from laboratory testing conducted by the manufacturer (Greiner Bio-One GmbH) or a contracted testing facility. This would typically be prospective testing conducted specifically for regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

This information is not applicable to this type of device. The "ground truth" for the performance criteria of a blood collection holder is established by engineering specifications, recognized consensus standards (like ISO 80369-7:2016), and comparison to a legally marketed predicate device. This doesn't involve human expert judgment in the way that, for example, a diagnostic imaging AI would.

4. Adjudication Method for the Test Set:

This information is not applicable. As there are no human "readers" or interpretations involved in the performance tests of this device, there is no need for an adjudication method. The results are typically quantitative measurements against defined thresholds or comparisons to the predicate.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. An MRMC study is relevant for medical devices that utilize Artificial Intelligence (AI) or assist human interpretation (e.g., radiologists reading images). The VACUETTE® SAFELINK Holder is a physical medical device and does not involve AI assistance or human interpretation in its function.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

This is not applicable. The device is not an algorithm or software. It is a mechanical device.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the performance tests of the VACUETTE® SAFELINK Holder is primarily based on:

  • Engineering Specifications: Internal design and performance requirements set by the manufacturer.
  • Recognized Consensus Standards: Specifically, ISO 80369-7:2016 for fitting assembly requirements.
  • Predicate Device Performance: Direct comparison and equivalence to the performance characteristics of the Greiner HOLDEX® Single-use Holder PP (K160532).

8. The Sample Size for the Training Set:

This is not applicable. The VACUETTE® SAFELINK Holder is a mechanical device and does not involve machine learning or AI, so there is no "training set" in this context.

9. How the Ground Truth for the Training Set was Established:

This is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 21, 2018

Greiner Bio-One NA Inc. Manfred Abel Quality System & Regulatory Affairs Manager 4238 Capital Drive Monroe, North Carolina 28110

Re: K180545

Trade/Device Name: VACUETTE® SAFELINK Holder with male luer lock Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: February 26, 2018 Received: March 1, 2018

Dear Manfred Abel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina Kiang -S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180545

Device Name

VACUETTE SAFELINK Holder with male luer lock

Indications for Use (Describe)

The VACUETTE SAFELINK Holder with male luer lock is for use in blood collection procedures in combination with devices with a female luer port.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PREMARKET NOTIFICATION 510(K) SUMMARY- K180545

1. SUBMITTERApplicant Name:Greiner Bio-One GmbH.Bad Haller Strasse 324550 Kremsmuenster,Austria
Contact person:Manfred Abel, M.S., MBAGreiner Bio-One NA Inc.4238 Capital DriveMonroe, NC 28110704 261 7823Manfred.Abel@gbo.com
Establishment registration number:8020040
Date prepared:Apr 25, 2018
2. DEVICE
Trade Name:VACUETTE® SAFELINK Holder with male luer lock
Common name:Holder
Classification:Name: Needle, hypodermic, single lumenProduct Code: FMIRegulation No: 880.5570Class: 2Review Panel: General Hospital
3. PREDICATE DEVICE
Predicate device:Greiner HOLDEX® Single-use Holder PP (K160532)
Classification:Name:Product Code:Regulation No:Class:Needle, hypodermic, single lumenFMI880.55702

Review Panel:

4. DEVICE DESCRIPTION

The VACUETTE® SAFELINK Holder with male luer lock is a sterile, single-use, non-invasive device, which consists of a cylinder made from polypropylene plastic with a centric molded male luer lock at the top. The centric location of the male luer lock is to facilitate the connection to non-needle devices with a female luer port. A stainless steel needle with a sleeve is bonded to the inside of the holder to facilitate

General Hospital

{4}------------------------------------------------

blood draw by puncturing the cap of an attached blood collection device. The sleeve is made from Synthetic Polyisoprene Compound.

5. INDICATION FOR USE

The VACUETTE® SAFELINK Holder with male luer lock is for use in blood collection procedures in combination with devices with a female luer port.

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device and the predicate device have the same fundamental technological characteristics, and intended use.

VACUETTE® SAFELINK HolderGreiner HOLDEX Single-use
with male luer lockHolder PP (predicate device)
FDA statusUnder reviewK160532
ClassificationClass IIClass II
Regulation880.5570880.5570
Product CodeFMIFMI
Intended UseThe VACUETTE® SAFELINK HolderThe HOLDEX® Single-Use Holder
with male luer lock is for use in bloodPP is for use in venous blood
collection procedures in combinationcollection procedures in
with devices with a female luer port.combination with blood collection
devices with a female luer port.
ManufacturerGreiner Bio-One (GBO)Greiner Bio-One (GBO)
SterilizerMediscan GmbHMediscan GmbH
Needle InsertPolypropylenePolypropylene
SleeveSynthetic Polyisoprene Compound.Synthetic Polyisoprene
Does not contain dry natural rubberCompound. Does not contain dry
natural rubber
NeedleStainless SteelStainless Steel
LuerCentric male luer lockEccentric male slip luer
Individual packagingOPP Flat film (Width 120 mm,OPP Flat film (Width 120 mm,
(blister)Thickness 25 µm)Thickness 25 µm)
Inner Carton (containingCarton (181x181x171 mm), Sort:Carton (181x181x171 mm), Sort:
100 Blisters)T20E with E-FluteT20E with E-Flute
Outer Carton (containingCarton (380x380x360 mm), Sort:Carton (380x380x360 mm), Sort:
8 inner cartons)SO53/2 with C-FluteSO53/2 with C-Flute
SterilizationE-BeamE-Beam
Sterility Assurance Level10-610-6
Storage4 - 25 C°4 - 25 C°
Shelf life36 months36 months
UseSingle use onlySingle use only

The indications of use of the VACUETTE® SAFELINK Holder with male luer lock under review are substantially equivalent to the predicate device. The used materials for all components, all packaging levels, and the production and sterilization processes of the device under review are the same as for the predicate device. The only difference is that the design under review has a centric male luer lock, which as benchmark testing indicates, is equivalent to the predicate device and does not raise new question for safety and effectiveness.

Based on the device features, principles of operation and technological characteristics, the VACUETTE® SAFELINK with male luer lock is substantially equivalent to the predicate device.

{5}------------------------------------------------

7. PERFORMANCE DATA

The VACUETTE® SAFELINK with male luer lock is manufactured and tested in conformity with internal specifications and recognized consensus standards. Performance testing was conducted to demonstrate and support our substantial equivalency claim that the device performs as intended and in an equivalent manner as the predicate device.

RequirementAcceptanceResultspass/fail
Liquid Leakage of rubber sleeveand ultrasonic welding area(seaming seal area) includingKickback TestingEquivalence to predicate device, internalspecificationpass
Fitting strength between rubbersleeve and needle insertEquivalence to predicate device, internalspecificationpass
Pull Out Force NeedleEquivalence to predicate device, internalspecificationpass
Stress cracking and liquidleakage from fitting assemblyunder pressureTested in conformity to ISO 80369-7:2016pass
Air leakage into fitting assemblyduring aspirationTested in conformity to ISO 80369-7:2016pass
Separation force of fittingassemblyTested in conformity to ISO 80369-7:2016pass
Unscrewing torque of fittingassemblyTested in conformity to ISO 80369-7:2016pass
Resistance to overridingTested in conformity to ISO 80369-7:2016pass
Free hemoglobinBenchmark testingpass

8. CONCLUSION

Results from the design verification show that the VACUETTE® SAFELINK holder with male luer lock meets established performance criteria and performance is as intended and substantially equivalent to the predicate device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).