K Number
K180481

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-06-29

(126 days)

Product Code
Regulation Number
876.5880
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Servator B SALF Solution is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.

Device Description

Servator B SALF Solution is a clear to light yellow, single use only, sterile (by steam sterilization) and non-pyrogenic for hypothermic flushing and preservation of organs.

The formulation includes soluble colloids, buffers, sodium and potassium salts, redox stabilizers, and phosphoric compounds to aid tissue viability by enabling regeneration of adenosine triphosphate (ATP).

The primary containers used for the device are PVC free bags 1000ml and 2000ml; therefore, they are free of phthalates. The bags must be chilled to between 2° and 6° C (36° to 43° F) prior to use. The solution may be used without any point of use filtration.

The Servator B SALF Solution is intended for perfusion and flushing donor kidneys, liver and pancreas immediately before or immediately after removal from a dead donor or immediately after removal from a living donor. The solution remains in the organ vessels during hypothermic storage and transportation. Servator B SALF Solution is meant for the cold storage of the organ and is not indicated for hypothermic preservation with continuous mechanical perfusion.

AI/ML Overview

The provided text describes the "Servator B SALF Solution," an organ preservation solution, and its substantial equivalence to a predicate device, CoStorSol® (Belzer UW). The document is a 510(k) summary for FDA clearance.

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of acceptance criteria and the reported device performance

The document does not present a formal table of "acceptance criteria" in the typical sense of numerical thresholds for clinical performance metrics (e.g., sensitivity, specificity for a diagnostic device). Instead, the acceptance criteria are implicitly the demonstration of substantial equivalence to the predicate device, CoStorSol®, across various technological characteristics and non-clinical performance aspects. The "reported device performance" is the result of tests confirming this equivalence.

Here's a table based on the provided "Table of Comparison" and "Non-Clinical Performance Data" section, framed as a comparison to achieve substantial equivalence:

Characteristic/TestAcceptance Criterion (Implicit: Substantial Equivalence to Predicate)Reported Servator B SALF Solution Performance
Indications for UseSame as predicateSame as predicate (flushing and cold storage of kidney, liver and pancreas)
Intended UseSame as predicateSame as predicate (Organ storage and preservation for transplantation)
Composition (Key Chemicals)Identical to predicate formulationSame chemical composition as predicate
Material Container/BagPVC Free / BagsPVC Free / Bags
Protecting Overwrap BagYesYes
Single Use OnlyYesYes
Bag Connections1 flip off, 1 needle point1 flip off, 1 needle point
Sodium Concentration29 mEq/L29 mEq/L
Potassium Concentration125 mEq/L125 mEq/L
pH7.4 at 20°C7.4 at 20°C
Osmolality320 mOsm/Kg320 mOsm/Kg
Sterilization MethodValidated according to ISO standardsSteam (ISO 17665-1 compliant)
Shelf Life24 Months24 Months
BiocompatibilityMeet ISO 10993 series standardsPassed Cytotoxicity, Irritation, System Toxicity, and Haemocompatibility (ISO 10993 series)
Sterility ValidationMeet ISO 17665-1 and ASTM F1980-16Passed according to ISO 17655-1 and ASTM F1980-16
Chemical ComparisonsDemonstrate equivalence to predicateDemonstrated substantial equivalence
Leachables PerformanceDemonstrate equivalence to predicateDemonstrated substantial equivalence

2. Sample size used for the test set and the data provenance

The document explicitly states: "Performance Testing...was completed as a direct comparison between the subject and predicate device. The chemical comparisons and leachables performance testing demonstrated the substantial equivalence of this device to the predicate."

However, it does not specify the sample size for these tests (e.g., how many batches were tested for chemical composition, how many samples for leachables) nor the data provenance (e.g., country of origin, retrospective or prospective). Given the nature of the device (a solution), the "test set" would refer to specific batches or formulations tested in a laboratory setting, not patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. For a 510(k) submission concerning a medical solution, ground truth is typically established through analytical chemical testing and standardized biological assays, not human expert consensus on interpretations of complex data like medical images.

4. Adjudication method for the test set

This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers interpreting data, especially in diagnostic imaging. For a chemical solution, the assessment relies on objective measurements and established scientific protocols.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was done or is relevant for this device. This is a medical solution for organ preservation, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a solution, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for demonstrating substantial equivalence for this device relies on:

  • Analytical Chemistry and Physicochemical Properties: Direct measurement of concentrations (Na+, K+), pH, osmolality, and confirmation of chemical composition.
  • Sterility Testing: Conformance to ISO and ASTM standards.
  • Biocompatibility Testing: Conformance to ISO 10993 series standards.
  • Leachables Testing: To ensure no harmful substances migrate from the packaging into the solution.
  • Functional Equivalence: Implicitly, the solution's performance in preserving organs is assumed to be equivalent to the predicate due to identical composition and physical properties.

8. The sample size for the training set

Not applicable. This device is a medical solution, not based on an AI algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 21, 2019

S.A.L.F. S.p.A. % Joyce St. Germain Regulatory Dept. Manager The 510k Consulting, LLC 1449 Springleaf Dr. Ormond Beach, FL 32174

Re: K180481

Trade/Device Name: Servator B SALF Solution Regulation Number: 21 CFR& 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: KDN Dated: May 25, 2018 Received: May 31, 2018

Dear Joyce St. Germain:

This letter corrects our substantially equivalent letter of June 29, 2018.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gema Gonzalez -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180481

Device Name Servator B SALF Solution

Indications for Use (Describe)

Servator B SALF Solution is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a stylized key design. The key is composed of two intertwined lines, one in gold and the other in black. The top of the key forms a diamond shape, while the bottom forms the key's stem and teeth.

The 510k Consulting, LLC

1449 Springleaf Dr., Ormond Beach, FL 32174

Key to success in obtaining your medical device clearance

510(k) Summary

Submitter/Applicant

S.A.L.F. S.p.A Via Marconi, 2, Cenate Sotto, BG, Italy Phone: +39-035-940097 Contact: Dr. Carmelo Gagliano (Quality Manager) carmelo.gagliano@salfspa.it

Date Prepared: June 28, 2018

Preparer/Consultant

The 510k Consulting, LLC 1449 Springleaf Dr. Ormond Beach, FL 32174 904-477-3203 Phone: Contact: Joyce St. Germain, Regulatory Consultant (joyce510kfda(@gmail.com)

Device Classification

Trade/Model Names:Servator B SALF Solution
Common Name:Organ perfusion and preservation solution
Regulation Name:Isolated kidney perfusion and transport system andaccessories
Regulation Number:21 CFR 876.5880
Product Code:KDN
Regulatory Class:II
Medical Specialty:Gastroenterology/Urology Panel

Predicate Device

The subject device claims equivalence to the following legally marketed predicate:

510(k) Number:K091245
Date ClearedJuly 17, 2009
Trade Name:CoStorSol / Brand Name: Belzer UW
Common Name:Organ perfusion and preservation solution

{4}------------------------------------------------

Classification Name:System, Perfusion, Kidney
Regulation Name:Isolated kidney perfusion and transport system and
accessories
Regulation Number:21 CFR 876.5880
Product Code:KDN
Regulatory Class:II
Medical Specialty:Gastroenterology/Urology Panel

Indications for Use

Servator B SALF Solution is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.

Intended Use

Organ storage and preservation for transplantation.

Device Description

Servator B SALF Solution is a clear to light yellow, single use only, sterile (by steam sterilization) and non-pyrogenic for hypothermic flushing and preservation of organs.

The formulation includes soluble colloids, buffers, sodium and potassium salts, redox stabilizers, and phosphoric compounds to aid tissue viability by enabling regeneration of adenosine triphosphate (ATP).

The primary containers used for the device are PVC free bags 1000ml and 2000ml; therefore, they are free of phthalates. The bags must be chilled to between 2° and 6° C (36° to 43° F) prior to use. The solution may be used without any point of use filtration.

The Servator B SALF Solution is intended for perfusion and flushing donor kidneys, liver and pancreas immediately before or immediately after removal from a dead donor or immediately after removal from a living donor. The solution remains in the organ vessels during hypothermic storage and transportation. Servator B SALF Solution is meant for the cold storage of the organ and is not indicated for hypothermic preservation with continuous mechanical perfusion.

Comparison of Technological Characteristics with Predicate

The indications for use and intended use of the subject and predicate devices are identical.

The technologies are substantially equivalent as the composition of both solutions are identical.

{5}------------------------------------------------

The subject and predicate devices are both supplied in bags with overbags for single use.

The subject and predicate devices are both supplied sterile.

Tests were performed in order to confirm the equivalence between the subject and predicate devices. The following table compares technological and other characteristics of the subject and predicate device.

Table of Comparison

Subject DevicePredicate DeviceComparison
DeviceServator B SALFCoStorSol®NA
510k NumberK180481K091245NA
ManufacturerS.A.L.F. S.p.A.(Italy)Preservation Solutions, Inc.,WI, USANA
Classification & Product Code876.5880; KDN876.5880; KDNSame
Device Classification NameIsolated kidney perfusion and transport system and accessoriesIsolated kidney perfusion and transport system and accessoriesSame
Device DescriptionSystem, Perfusion, KidneySystem, Perfusion, KidneySame
Common NameOrgan perfusion and preservation solutionOrgan perfusion and preservation solutionSame
Indications for UseServator B SALF Solution is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.CoStorSol® is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.Same

Table B.1 -- Technological Comparison

{6}------------------------------------------------

Intended for UseOrgan storage andpreservation fortransplantationOrgan storage andpreservation fortransplantationSame
Meets UNOS PolicyYESYESSame
MaterialContainer/BagPVC Free / BagsPVC Free / BagsSame
ProtectingOverwrap BagYESYESSame
Single Use OnlyYESYESSame
Bag Connections1 flip off, 1 needlepoint1 flip off, 1 needlepointSame
Used for OrganTransplants ofUsed for kidneys,liver and pancreasUsed for kidneys,liver and pancreasNA
Product StateLiquid - SolutionLiquid - SolutionSame
Model NumbersBag1000ml /SERVB10DMABag 2000ml /SERVB20DMBag 1000ml /Model # unknownDifferent ModelNumbers - due todifferentmanufacturers
ConfigurationsBox containing 10bags of 1000mlPVC free.Box containing 5bags of 2000mlPVC free.Bags 1000 ml,and carton of 10Same,Subject devicehas additionalpackagingavailable
SterilizationMethodSteam -Sterilizationprocesses validatedSolution isasepticallyprocessed and thetubing used for theDifferent;however bothpassed accordingto ISO standards
according to ISO17665-1aseptic filling issteam sterilized.The bags aregamma sterilized.Sterilizationprocesses werevalidated accordingto ISO 17665-1 orUSP Section<1211>, asappropriate.
SodiumConcentration29 mEq/L29 mEq/LEquivalent
PotassiumConcentration125 mEq/L125 mEq/LEquivalent
pH7.4 at 20°C7.4 at 20°CEquivalent
Osmolality320 mOsm/Kg320 mOsm/KgEquivalent
ManufactureStandards ofConformityISO 9001:2008ISO 13485:2003GMP CertificationUnknownSubject meetscurrent standards
Shelf Life24 Months24 MonthsSame
Device Standards ofConformitySubject devicepassed accordingto ISO standardsUnknownSubject devicepassed accordingto ISO standards
ISO StandardsISO 10993-110993-210993-410993-510993-1010993-1110993-1210993-18ISO 11607-111607-2ISO 11737-1ISO 14001ISO 14971ISO15189ISO 10993 seriesISO 17665USP <788>USP <1211>Subject andPredicate passedstandardized tests

{7}------------------------------------------------

{8}------------------------------------------------

ISO15223-1
ISO 17025
ISO 17665-1
ASTM F1980-16
USP <71>
USP <85>
USP <39>
USP <1225>

The above comparison shows the subject and predicate devices are substantially equivalent in technology characteristics. The differences are highlighted and those differences do not make the subject device any less safe and effective as the predicate device. The differences are due to different manufacturers with different model numbers and the subject device has additional configuration of packaging that the predicate doesn't have.

Non-Clinical Performance Data

The following performance data is provided in support of the substantial equivalence determination. All tests performed are included in this submission.

Biocompatibility... is required for this device. The Cytotoxicity, Irritation, System Toxicity and Haemocompatibility tests were all performed of the ISO 10993 series and all are listed above in the Table of Comparison. The subject device passed all biocompatibility test standards.

Sterilization and Shelf Life . . . is required for the subject device. The Validation of Sterility was performed and the results passed according to ISO 17655-1 and ASTM F1980-16. Shelf life for the subject and predicate device is the same at 24 months.

Electrical Safety and EMC . . . testing was not applicable for this device.

Software ... was not applicable for this device.

Performance Testing...was completed as a direct comparison between the subject and predicate device. The chemical comparisons and leachables performance testing demonstrated the substantial equivalence of this device to the predicate.

Conclusion

The subject and predicate devices have the same indications for use and the same intended use. Both devices are substantially equivalent in design, materials, packaging and other technological characteristics and performance (since they have, in fact, the same chemical composition). The Servator B SALF Solution does not raise any questions regarding safety and effectiveness and is equivalent to the predicate device. The non-clinical data supports and demonstrates the safety of the device.

{9}------------------------------------------------

The conclusion is that Servator B SALF Solution warrants a finding of substantial equivalence to the legally marketed CoStorSol® by Preservations Solutions, Inc., and therefore, should have clearance for premarket activities in the United States.

§ 876.5880 Isolated kidney perfusion and transport system and accessories.

(a)
Identification. An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.(b)
Classification. Class II (performance standards).