K Number
K180389

Validate with FDA (Live)

Date Cleared
2018-06-29

(136 days)

Product Code
Regulation Number
870.5800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Venera 508 Deep Vein Thrombosis (DVT) Prevention Therapy System is a portable treatment to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).

This system can be applied in either home or clinical settings to:

  • . Aid in the prevention of DVT
  • . Enhance blood circulation
  • . Diminish post-operative pain and swelling
  • . Reduce wound healing time
  • . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
  • . As a prophylaxis for DVT by persons expecting to be stationary for long periods of time
Device Description

The Venera™ 508 Deep Vein Thrombosis Prevention System is an easy to use portable pneumatic compression system that noninvasively helps prevent the onset of DVT in patients by simulating blood flow in the extremities (simulating muscle contractions). The Venera 508 system consists of a pair of pump and cuff assemblies. The device will alternatively inflate and deflate the garment (cuff) to stimulate blood flow in the extremities (muscle contraction). The device provides a 50mmHg pressure and followed by 50 seconds of deflation period once it reaches the desired pressure.

AI/ML Overview

The provided text does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the manner requested (e.g., specific metrics like sensitivity, specificity, or reader performance improvement for an AI/device). Instead, it describes a medical device (Venera 508 Deep Vein Thrombosis Prevention System) and its 510(k) submission for substantial equivalence to a predicate device.

The "Performance Tests" section lists several tests conducted, but it does not provide the acceptance criteria for these tests or the reported performance data in a quantitative way. It only states, "The conclusions drawn from the performance tests demonstrate that the device is performing as intended, and is safe and effective."

Therefore, I cannot populate the table or provide the detailed information requested regarding acceptance criteria and study particulars for a comparative effectiveness study or standalone performance. The document explicitly states: "No animal study or clinical study was conducted." This means there is no human subject study to evaluate the device's efficacy for its intended medical purpose (DVT prevention) that would typically involve a ground truth, expert readers, or statistical comparisons of performance.

Here's a breakdown of what can be extracted or inferred based on the provided text, and where gaps exist:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Not specified quantitatively in the document."The conclusions drawn from the performance tests demonstrate that the device is performing as intended, and is safe and effective." (General statement, no specific metrics provided for tests like Pressure Accuracy, Cycle Time, etc.)

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified. The document mentions "samples of the device" were used for function and mechanical testing, but the number of devices or test repetitions is not stated.
  • Data Provenance: Not specified. The tests were conducted internally by the manufacturer, Suzhou Minhua Medical Apparatus Supplies Co. Ltd. (China). The document does not indicate if external data sources or patient data were used for these performance tests. Given this is a physical medical device and not an AI/diagnostic software, the concept of "data provenance" as typically applied to image datasets is not relevant here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. The performance tests described (Pressure Accuracy, Cycle Time, etc.) relate to the device's mechanical and functional specifications, not its diagnostic or therapeutic efficacy based on patient outcomes or expert interpretation of medical data. Therefore, no "ground truth" in the clinical sense was established by experts for these engineering tests.

4. Adjudication method for the test set

  • Not applicable. Since no expert-based ground truth was established, there was no need for an adjudication method.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. The document explicitly states: "No animal study or clinical study was conducted." Therefore, no MRMC study, AI assistance evaluation, or effect size calculation was performed or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This device is not an AI algorithm. It is a physical medical device for DVT prevention. Therefore, the concept of a "standalone" algorithm performance study is not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not applicable in the clinical sense. The "ground truth" for the engineering performance tests would be the established engineering specifications and measurement standards (e.g., a calibrated pressure gauge for the "Pressure Accuracy Test").

8. The sample size for the training set

  • Not applicable. This device does not involve a "training set" as it is not an AI/machine learning system.

9. How the ground truth for the training set was established

  • Not applicable for the same reason as above.

{0}------------------------------------------------

June 29, 2018

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Suzhou Minhua Medical Apparatus Supplies Co. Ltd. Alex Wang Regulatory Affairs Specialist No.789, Wu Fang Road, Friendship Industrial Zone Song Ling Town, Wu Jiang Suzhou, 215222 China

Re: K180389

Trade/Device Name: Venera 508 Deep Vein Thrombosis (DVT) Prevention System Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: May 15, 2018 Received: June 4, 2018

Dear Alex Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Nicole M. Gillette -S 2018.06.29 08:28:57 -04'00'

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180389

Device Name

Venera 508 Deep Vein Thrombosis Prevention Therapy System

Indications for Use (Describe)

The Venera 508 Deep Vein Thrombosis (DVT) Prevention Therapy System is a portable treatment to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).

This system can be applied in either home or clinical settings to:

  • . Aid in the prevention of DVT
  • . Enhance blood circulation
  • . Diminish post-operative pain and swelling
  • . Reduce wound healing time
  • . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
  • . As a prophylaxis for DVT by persons expecting to be stationary for long periods of time

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510K Summary

Submitter:

Suzhou Minhua Medical Apparatus Supplies Co. Ltd.

No.789, Wu Fang Road, Friendship Industrial Zone, Song Ling Town,

Wu Jiang District, Su Zhou City, Jiangsu Province, China

Postal Code: 215222

Tel: +86-0512-63091066

Contact Person: Alex Wang

Date Prepared: Jan 15, 2018

Device:

Common Names:Intermittent Pneumatic Compression Device
Proprietary Name:Venera™ 508 Deep Vein Thrombosis Prevention System
Regulatory Class:II
Product Code:JOW

Predicate Devices:

The Venera 508 Deep Vein Thrombosis Prevention System is equivalent to the following:

Predicate DeviceManufacturer510(k)#
Cirona 6300 Deep VeinThrombosis(DVT) Prevention SystemDevon Medical ProductsK151189

Device Description

The Venera™ 508 Deep Vein Thrombosis Prevention System is an easy to use portable pneumatic compression system that noninvasively helps prevent the onset of DVT in patients by

{4}------------------------------------------------

simulating blood flow in the extremities (simulating muscle contractions). The Venera 508 system consists of a pair of pump and cuff assemblies. The device will alternatively inflate and deflate the garment (cuff) to stimulate blood flow in the extremities (muscle contraction). The device provides a 50mmHg pressure and followed by 50 seconds of deflation period once it reaches the desired pressure.

Indications for Use:

The Venera™ 508 Deep Vein Thrombosis Prevention System is a portable treatment system intended to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).

This system can be applied in either home or clinical settings to:

  • Aid in the prevention of DVT .
  • . Enhance blood circulation
  • . Diminish post-operative pain and swelling
  • . Reduce wound healing time
  • . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
  • As a prophylaxis for DVT by persons expecting to be stationary for long periods of time

Technological Characteristics:

Below is a table of comparison for the technological characteristics against the predicate device:

PredicateCirona 6300 DVT System
Indication for UseS
ComponentsS
MaterialSE
BiocompatibilitySE
Principle of OperationSE
Treatment parametersS
User InterfaceS

*SE – Substantial Equivalent *S - Same

{5}------------------------------------------------

The manufacturer believes that the technological characteristics of the Venera™ 508 Deep Vein Thrombosis Prevention System are substantially similar to those of the predicate device. Venera 508 DVT Prevention System has similar components to its predicate device and identical principles of operation.

Performance Tests

To verify that the device design met its function and performance requirements, samples of the device underwent function and mechanical testing.

The following tests were conducted:

Function Performance Tests
Pressure Accuracy Test Report
Pressure Switch Test Report
Venera 508 Cycle Time Test Report
Alarm Function Test Report
Battery Life Test Report

The conclusions drawn from the performance tests demonstrate that the device is performing as intended, and is safe and effective.

Biocompatibility

Biocompatibility test evaluation for Venera 508 DVT Prevention System is done in accordance to the FDA Good Laboratory Practice quidelines. The following tests were done:

  • Cytotoxicity い
  • Sensitization ﻬﺎ ﻓﻬﻮ ﺍﻟﻤﺪﻳﻨﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ
  • Irritation い

Sterilization and Shelf Life

Sterilization and shelf life are not applicable to Venera 508 DVT Prevention System.

Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety and EMC tests were conducted according to the following standards:

  • IEC 60601-1: 2005 + A1:2012 Medical Electrical Equipment Part 1: General requirements , for basic safety and essential performance

{6}------------------------------------------------

  • IEC 60601-1-2: 2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
  • IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment Part 1-6: General requirements ﻬﺎ ﻓﻬﻮ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ for safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices
  • IEC 60601-1-11: 2015 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

Software Verification and Validation

Software verification and validation were conducted and documentation is provided. The software was considered as a Moderate level of concern, since a failure or latent flaw in the software could directly or indirectly result in minor injury to the patient or operator.

Animal Study and Clinical Study

No animal study or clinical study was conducted.

Statement of Substantial Equivalence

The Venera 508 Deep Vein Thrombosis Prevention System is substantially equivalent in its intended use, technology, function, operating parameters to predicate devices that are currently commercially available and in distribution, and does not raise any new concerns regarding safety or effectiveness.

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).