K Number
K180317

Validate with FDA (Live)

Date Cleared
2018-05-04

(88 days)

Product Code
Regulation Number
888.3565
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EVOLUTION® BIOFOAM® Tibial, EVOLUTION® Revision Tibial, and EVOLUTION® Revision CCK Systems are indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

  1. Non-inflammatory degenerative joint disease: including osteoarthritis, or avascular necrosis;

  2. Inflammatory degenerative joint disease, including rheumatoid arthritis;

  3. Correction of functional deformity;

  4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

The EVOLUTION® BIOFOAM® Tibial System implants are for cementless use only.

The EVOLUTION® Revision Tibial and EVOLUTION® Revision CCK Systems are for cemented use only.

Device Description

A labeling change is taking place to include MR Conditional language and symbols to the implant components of the predicate device systems identified above. The changes only apply to the device labeling; therefore, the subject devices are identical to the predicate devices.

The non-clinical testing provided establishes the conditional safety and compatibility of the passive implants in a magnetic resonance (MR) environment. The subject devices include the following components and basic design features:

  • . Metal femoral components manufactured from cobalt chrome alloy
  • . Femoral revision accessories, including augments, manufactured from titanium alloy
  • . CCK tibial inserts manufactured from UHMWPE with locking screws and posts manufactured from titanium alloy
  • . Tibial bases manufactured from titanium alloy available in porous and non-porous versions
  • Tibial revision accessories, including keels, augments, stem extensions, and adapters, . manufactured from titanium alloy.
  • . Patellae manufactured from UHMWPE and cobalt chrome alloy
AI/ML Overview

The provided text is a 510(k) Summary for a medical device (EVOLUTION® Knee Systems – MR Labeling) and outlines nonclinical testing conducted to establish the device's conditional safety and compatibility in a magnetic resonance (MR) environment. It does not describe a study involving device performance against acceptance criteria in the usual sense of diagnostic accuracy or clinical outcomes, but rather safety and compatibility within an MR environment.

Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of clinical or diagnostic performance are not applicable to this document. I will focus on what is provided regarding the MR safety testing.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state numerical acceptance criteria in a table format for MR safety. Instead, it refers to industry standards that define safe limits. The reported "performance" is that the device met these standards and showed no safety issues.

Acceptance Criteria (Defined by Standards)Reported Device Performance
Safety and Compatibility in MR Environment (e.g., within acceptable limits for RF heating, magnetic field interactions, and image artifacts as per ASTM, NEMA standards)"The testing concluded that there are no safety issues related to magnetic field interactions under the specific conditions identified in the labeling."

2. Sample size used for the test set and the data provenance

The document states, "The tests evaluated the worst case components and constructs for RF heating, field interactions, and image artifacts." It refers to "passive implants" but does not specify the exact number of components or constructs tested.

  • Test set sample size: Not explicitly stated as a numerical count of individual devices, but "worst case components and constructs" were tested.
  • Data provenance: Nonclinical testing performed according to U.S. FDA guidance and international standards (NEMA, ASTM). The location of the testing laboratory is not specified (e.g., country of origin).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. The testing was nonclinical (MR safety testing), not involving human diagnostic interpretation or ground truth establishment by medical experts.

4. Adjudication method for the test set

This question is not applicable, as it was nonclinical testing, not involving human readers or interpretation requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study described, as this document concerns the MR safety labeling of knee implant systems, not an AI-assisted diagnostic device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This question is not applicable. The device is a knee implant system, not an algorithm.

7. The type of ground truth used

For an MR safety study, "ground truth" typically refers to the physical properties of the materials and their behavior in an MR field as measured by calibrated equipment following established standards. It's not expert consensus, pathology, or outcomes data in the clinical diagnostic sense. The ground truth is the quantitative measurements against the physical laws and established safety thresholds described by the ASTM and NEMA standards.

8. The sample size for the training set

This question is not applicable. This is not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established

This question is not applicable as there was no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" stacked on top of each other.

MicroPort Orthopedics, Inc. Grace Johnson-Bann Sr. Regulatory Affairs Specialist 5677 Airline Road Arlington, Tennessee 38002

May 4, 2018

Re: K180317

Trade/Device Name: EVOLUTION® Knee Systems – MR Labeling, EVOLUTION® BIOFOAM® Tibial System, EVOLUTION® Revision Tibial System, EVOLUTION® Revision CCK System Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, JWH Dated: February 1, 2018 Received: February 5, 2018

Dear Ms. Johnson-Bann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Katherine D. Kavlock -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K180317

Device Name

EVOLUTION® Knee Systems – MR Labeling, EVOLUTION® BIOFOAM® Tibial System, EVOLUTION® Revision Tibial System, EVOLUTION® Revision CCK System

Indications for Use (Describe)

The EVOLUTION® BIOFOAM® Tibial, EVOLUTION® Revision Tibial, and EVOLUTION® Revision CCK Systems are indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

  1. Non-inflammatory degenerative joint disease: including osteoarthritis, or avascular necrosis;

  2. Inflammatory degenerative joint disease, including rheumatoid arthritis;

  3. Correction of functional deformity;

  4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

The EVOLUTION® BIOFOAM® Tibial System implants are for cementless use only.

The EVOLUTION® Revision Tibial and EVOLUTION® Revision CCK Systems are for cemented use only.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

|× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle containing a red diamond shape. To the right of the square is the word "MicroPort" in blue, with the word "Orthopedics" underneath in red. Below that is the phrase "Full Function. Faster." in blue.

510(K) SUMMARY

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the EVOLUTION® Knee Systems MR Labeling.

Submitted by:MicroPort Orthopedics Inc.5677 Airline Rd, Arlington TN, 38002Phone: 866-872-0211Fax: 855-446-2247
Date:May 2, 2018
Contact Person:Grace Johnson-BannSr. Regulatory Affairs Specialist
Proprietary Name of Modified Device:EVOLUTION® Knee Systems – MR LabelingEVOLUTION® BIOFOAM® Tibial SystemEVOLUTION® Revision Tibial SystemEVOLUTION® Revision CCK System
Common Name:Cementless Knee Systems, Cemented Revision Knee Systems, MR Labeling
Classification Name and Reference:21 CFR 888.3565 - Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesisClass II21 CFR 888.3560 – Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesisClass II
Subject Product Code and Panel Code:Orthopedics/87/MBH, JWH
Predicate Devices:EVOLUTION® BIOFOAM® Tibial System (K152298)EVOLUTION® Revision Tibial System (K162026)EVOLUTION® Revision CCK System (K171389)
Reference Device:MPO Total Knee Systems MR Labeling (K152631)

MicroPort Orthopedics BV Hoogoorddreef 5, 1101 BA Amsterdam, The Netherlands +31 20 545 01 00

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a logo for MicroPort Orthopedics. The logo features a blue square with rounded corners on the left side, inside of which is a white circle with a red diamond shape in the center. To the right of the square is the word "MicroPort" in blue, with the word "Orthopedics" in red underneath. Below the words is the phrase "Full Function, Faster" in blue.

DEVICE INFORMATION

A. Intended Use

The EVOLUTION® BIOFOAM® Tibial System, EVOLUTION® Revision Tibial System, and EVOLUTION® Revision CCK System are indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

    1. Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis;
    1. Inflammatory degenerative joint disease, including rheumatoid arthritis;
    1. Correction of functional deformity;
    1. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

The EVOLUTION® BIOFOAM® Tibial System is for use without bone cement.

The EVOLUTION® Revision Tibial and CCK Systems are for cemented use only.

B. Device Description

A labeling change is taking place to include MR Conditional language and symbols to the implant components of the predicate device systems identified above. The changes only apply to the device labeling; therefore, the subject devices are identical to the predicate devices.

The non-clinical testing provided establishes the conditional safety and compatibility of the passive implants in a magnetic resonance (MR) environment. The subject devices include the following components and basic design features:

  • . Metal femoral components manufactured from cobalt chrome alloy
  • . Femoral revision accessories, including augments, manufactured from titanium alloy
  • . CCK tibial inserts manufactured from UHMWPE with locking screws and posts manufactured from titanium alloy
  • . Tibial bases manufactured from titanium alloy available in porous and non-porous versions
  • Tibial revision accessories, including keels, augments, stem extensions, and adapters, . manufactured from titanium alloy.
  • . Patellae manufactured from UHMWPE and cobalt chrome alloy

C. Substantial Equivalence Information

The scope of this Premarket Notification only includes labeling changes for the subject devices. The design features and materials of the subject devices are identical to those of the predicate devices. The indications for use of the subject devices are also identical those of the predicate devices. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this

Premarket Notification.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners, inside of which is a white circle containing a red star-like shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Underneath "Orthopedics" is the tagline "Full Function. Faster."

D. Nonclinical Testing

Nonclincal testing was conducted to establish the conditional safety and compatibility of the passive implants in a magnetic resonance (MR) environment according to the recommendations provided in the Guidance for Industry and Food and Drug Administration Staff entitled "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment", issued by the US FDA on December 11, 2014. Testing was also conducted according to the following standards:

  • . NEMA MS 8-2008 Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems
  • . ASTM F2052-15 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
  • . ASTM F2119-07(2013) Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants
  • . ASTM F2182-11a Standard Test Method For Measurement of Radio Frequency Induced Heating On Or Near Passive Implants During Magnetic Resonance Imaging
  • . ASTM F2213-06(2011) Standard Test Method For Measurement Of Magnetically Induced Torque On Medical Devices In The Magnetic Resonance Environment
  • ASTM F2503-13 Standard Practice For Marking Medical Devices And Other Items For Safety In The Magnetic Resonance Environment

The test determined the effects of MRI on the implants and the effects of the implants on the MR image quality. The tests evaluated the worst case components and constructs for RF heating, field interactions, and image artifacts. The testing concluded that there are no safety issues related to magnetic field interactions under the specific conditions identified in the labeling.

E. Clinical Testing

Clinical data was not provided for the subject devices.

F. Conclusion

The information provided in this Premarket Notification adequately supports the labeling change for the subject devices and the substantial equivalence of the subject devices to the predicate devices.

N/A