K Number
K180304

Validate with FDA (Live)

Date Cleared
2018-04-20

(77 days)

Product Code
Regulation Number
884.6120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.

The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.

Device Description

The Geri Embryo Incubator is a benchtop incubator with six modular incubation chambers each with its own temperature control capability and separate gas inlet feed. Each chamber contains heating elements in its lid and base, together with an orange light source (591 nm) and camera with integrated optics that take time-lapse images of embryos and allows operators to view embryos without removing them from the incubation chamber. Inside each chamber is a filter the gas mixture entering the incubation chamber from the gas supply. The gas is supplied from standard premixed medical gas cylinders. Each chamber can also contain an optional water bottle (Geri Water Bottle) to generate humidity. The Geri Water Bottle is supplied sterile with a sterility assurance level (SAL) of 10°, and has a shelf-life of three years. The Geri Water Bottle is designed to be used with the Geri Embryo Incubator only.

The embryos are maintained in Geri Dishes that are supplied separately. The Geri Dish is an optically-clear polystyrene dish designed to be compatible with the Geri Embryo Incubator. Up to 16 embryos can be stored in one Geri Dish. Geri Dishes are supplied sterile with a sterility assurance level (SAL) of 10°, and have a shelflife of 12 months.

The Geri Embryo Incubator and Geri Dish are intended to be used together for embryo imaging purposes. The Geri Dish is not compatible with other embryo time-lapse incubators. However, other assisted reproduction culture dishes may be used in the Geri Embryo Incubator in separate non-time-lapse positions located within each chamber.

The Geri Embryo Incubator includes firmware to control the incubator settings, and software to control patient information and settings.

AI/ML Overview

This document, a 510(k) Summary for the Geri Embryo Incubator and Geri Dish, outlines the characteristics and testing for a medical device. It is not a document about an AI/ML-based medical device. Therefore, none of the specific questions regarding AI/ML device acceptance criteria, study design for AI/ML, sample sizes for AI/ML test and training sets, expert adjudication methods for AI/ML ground truth, or MRMC studies for AI assistance, can be answered from the provided text.

The document focuses on the substantial equivalence of the Geri Embryo Incubator and Geri Dish to a predicate device, with the main change being the addition of a humidity control system. The studies described are traditional non-clinical performance tests for an incubator and culture dish, such as electrical safety, EMC, cleaning validation, gas maintenance, temperature control, time-lapse imaging function, embryo development testing (mouse embryo assay), and sterilization validation.

Therefore, I cannot provide the requested information as the provided text relates to a medical device that does not appear to be an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

April 20, 2018

Genea Biomedx Pty Ltd % Roger Grav VP, Quality Assurance and Regulatory Affairs Donawa Lifescience Consulting Srl Piazza Albania, 10 00153 Rome ITALY

Re: K180304

Trade/Device Name: Geri Embryo Incubator and Geri Dish Regulation Number: 21 CFR 884.6120 Regulation Name: Assisted reproduction accessories Regulatory Class: Class II Product Code: MQG, MQK, MTX Dated: March 19, 2018 Received: March 22, 2018

Dear Roger Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Charles Viviano -S

For

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180304

Device Name

Geri Embryo Incubator and Geri Dish

Indications for Use (Describe)

The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.

The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."


Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The logo is simple and modern, with a focus on the company's name.

510(k) Summary (K180304)

I. General Information on Submitter

Submitter/Address:Genea Biomedx Pty LtdLevel 2, 321 Kent StreetSydneyNSW 2000Australia
Contact person:Ms. Kea Dent
Phone:+61 2 8484 7677
Fax:+61 2 9229 6478
Correspondent:Mr. Roger GrayVP, Quality and RegulatoryDonawa Lifescience Consulting SrlPiazza Albania 1000153 RomeItaly
Phone:+39 06 578 2665
Fax:+39 06 574 3786
Email:rgray@donawa.com

II. Date Prepared: April 19, 2018

III. General Information on Devices

Device Name:Geri Embryo Incubator and Geri Dish
Common Name:Embryo Incubator
Regulation Number:21 CFR 884.6120
Regulation Name:Assisted Reproduction Accessories
Product Code:MQG (Accessory, Assisted Reproduction), MQK (Labware, AssistedReproduction), MTX (Microscope and Microscope Accessories, Reproduction,Assisted)
Regulatory Class:II

IV. Predicate Device

Geri Embryo Incubator (K171736) manufactured by Genea Biomedx, Australia. The predicate device has not been subject to any design related recalls.

V. Device Description

The Geri Embryo Incubator is a benchtop incubator with six modular incubation chambers each with its own temperature control capability and separate gas inlet feed. Each chamber contains heating elements in its lid and base, together with an orange light source (591 nm) and camera with integrated optics that take time-lapse images of embryos and allows operators to view embryos without removing them from the incubation chamber. Inside each chamber is a filter the gas mixture entering the incubation chamber from the gas supply. The gas is supplied from standard premixed medical gas cylinders. Each chamber can also contain an optional water bottle (Geri Water Bottle) to generate humidity. The Geri Water Bottle is supplied sterile with a sterility assurance level (SAL) of 10°, and has a shelf-life of three years. The Geri Water Bottle is designed to be used with the Geri Embryo Incubator only.

The embryos are maintained in Geri Dishes that are supplied separately. The Geri Dish is an optically-clear polystyrene dish designed to be compatible with the Geri Embryo Incubator. Up to 16 embryos can be stored in

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Genea BIOMEDX. The logo consists of a green square with rounded corners and a white curved line inside. To the right of the square, the word "Genea" is written in gray, and below it, the word "BIOMEDX" is written in green.

one Geri Dish. Geri Dishes are supplied sterile with a sterility assurance level (SAL) of 10°, and have a shelflife of 12 months.

The Geri Embryo Incubator and Geri Dish are intended to be used together for embryo imaging purposes. The Geri Dish is not compatible with other embryo time-lapse incubators. However, other assisted reproduction culture dishes may be used in the Geri Embryo Incubator in separate non-time-lapse positions located within each chamber.

The Geri Embryo Incubator includes firmware to control the incubator settings, and software to control patient information and settings.

VI. Indications for Use:

The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.

The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.

VII. Comparison of Intended Use and Technological Characteristics of the Subject Device and Predicate Device

ParameterSubject device (K180304)Predicate device (K171736)
Indications for UseSame as the predicate deviceThe Geri Embryo Incubator is intended to provide anenvironment with controlled temperature and mixed gas(CO2 and other gases) for the development of embryos.The Geri Embryo Incubator has an integrated camera andoptics for imaging and viewing embryos during incubation,for a maximum time of 120 hours.The Geri Dish is intended to be used for preparation,storage and imaging of human embryos. The Geri Dish isintended to be used only with the Geri Embryo Incubator.
Number of incubation chamberSame as the predicate deviceSix
Incubation chamber heatingSame as the predicate deviceHeated aluminium stage and lid
Incubation chamber gas supplySame as the predicate deviceFlow controlled positive pressure system with fast purgeflow rate for rapid recovery
Incubation chamber humidityDisposable water bottle togenerate humidity (~80% RH) ineach incubation chamberN/A
Gas supply typeSame as the predicate devicePremixed blend of 6% CO2, 5% O2, 89% N2 or high purity6% CO2
Gas supply pressureSame as the predicate device$150 \pm 15$ kPa ( $218 \pm 2.2$ psi)
Gas flow rate capabilitySame as the predicate device10 mL/min per chamber. Purge at 160 mL/min perchamber for 3 min
Gas flow rate accuracySame as the predicate device$\pm$ 15% of flow per chamber (normal flow) and $\pm$ 10 mL/minper chamber (purge)
HEPA FilterSame as the predicate deviceRetains 99.97% of particles >0.3 µm
Chamber temperature capabilitySame as the predicate device35.0-40.0°C in 0.1°C increments at 20-28°C. At set point of37°C, the ambient temperature range is 18-30°C
Chamber temperature accuracySame as the predicate device$\pm$ 0.2°C at calibration point
Microscope/cameraSame as the predicate deviceCMOS microscope camera (2560x1928 pixelsmonochrome )
Image resolutionSame as the predicate device2 pixels/µm
Image acquisition frequencySame as the predicate device5 min
IlluminationSame as the predicate deviceSingle orange LED (591nm)
Number of image focal planesSame as the predicate device11
Light exposureSame as the predicate device~45s/day
Alarm (for out of range)Same as the predicate device,with additional alarm for humidityChamber temperature, CO2 concentration, CO2 pressureO2 concentration, N2 pressure, Load door open >30s

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside on the left, and the text "Genea" in gray above the text "BIOMEDX" in green on the right. The logo is clean and modern, with a focus on the company name and a simple graphic element.

ParameterSubject device (K180304)Predicate device (K171736)
Culture dish (Geri Dish)Same as the predicate device16-well culture dish (40-80 $\mu$ l fill volume/well) made ofpolystyrene. Each incubation chamber can hold one dish

The subject and predicate devices have the same Intended Use for preparing and storing human embryos undergoing time-lapse imaging procedures. They also have the same technological characteristics, except that the subject device has an additional humidity control system, including a water container, humidity sensor, and software component to manage this new feature. The addition of humidification system to the device does not raise different questions of safety and effectiveness as compared to the predicate device. In addition, similar humidification systems are present on other cleared devices of this type.

VIII. Summary of Non-Clinical Performance Testing

  1. The following studies that were provided in the predicate submission (K171736) were leveraged in the current submission as the additional humidity control system did not impact the ability of these studies to support substantial equivalence:

Testing for Geri Embryo Incubator

  • Electrical safety testing per IEC 61010-1:2010
  • Electromagnetic Compatibility testing per IEC 61326-1:2012 ●
  • . Cleaning and disinfection validation testing per FDA Guidance Document, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling: Guidance for Industry and Food and Drug Administration" issued on March 17, 2015
  • Bench performance tests that met design specifications described in Section VII, as follows: ●
    • o Gas maintenance testing to ensure that gas flow rates, gas recovery rates and CO2 levels in chambers meet design specifications
    • o Temperature control testing to ensure that culture chamber temperature conditions were within defined ranges and were stable over time
    • Time-lapse testing to demonstrate that all design specifications of time-lapse function are met O (dish presence detection, dish position accuracy, image quality and resolution, imaging of embryos in all wells on Geri Dishes within time and focal plane settings)
  • Embryo development test demonstrating that ≥70% mouse 1-cell embryos developed to blastocyst on ● day 5 following exposure to the maximum time lapse exposure conditions
  • . Embryo light energy exposure testing comparing the potential for light toxicity associated with timelapse imaging under worst-case exposure conditions (maximum light exposure over the full duration of use [120 hours]) to that of total light exposure during in assisted reproduction procedures (IVF and ICSI) using typical laboratory microscopes. Results showed total light energy exposure following timelapse imaging should not exceed levels associated with typical assisted reproduction imaging procedures.

Testing for Geri Dish

  • . Radiation sterilization validation testing per ISO 11137-2:2009
  • Package integrity testing:
    • Dye penetration testing per ASTM F1929-12 O
    • Peel strength testing ASTM F88/F88M-09 o
  • Endotoxin testing per USP <85> and ANSI/AAMI ST72: 2011. The testing demonstrated that the ● device met the specification of <20 EU/device.
  • Mouse embryo assay (MEA):

One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Genea BIOMEDX. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The green square is on the left side of the logo, and the text is on the right side.

hours were assessed in comparison with the control group. The testing demonstrated that the device met acceptance criterion of "1-cell MEA ≥80% embryos developed to blastocyst in 96 hours."

  • Shelf-life testing (accelerated) demonstrating that visual requirements, media fill volume, time-lapse . alignment and MEA met design specifications before and after aging
  1. The following studies on the subject device have been conducted to support the addition of the humidity control system to the Geri Embryo Incubator:

Testing for Geri Embryo Incubator:

  • Software verification and validation testing that met the requirements of the FDA Guidance Document, ● "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005
  • Humidity control testing to ensure that the humidity levels in chambers meet design specifications

Testing for Geri Water Bottle:

  • . Radiation sterilization validation testing per ISO 11137-2:2013
  • . Package integrity testing:
    • Dye penetration testing per ASTM F1929-15 O
    • Seal strength testing per ASTM F88/F88M-15 o

IX. Conclusion

The subject and predicate devices have the same intended use and fundamental technological characteristics. The differences in technological characteristics between subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.

§ 884.6120 Assisted reproduction accessories.

(a)
Identification. Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova).
(2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures.
(3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation.
(4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature.
(5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
(b)
Classification. Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.