K Number
K180287

Validate with FDA (Live)

Device Name
GSI Corti
Date Cleared
2018-05-11

(99 days)

Product Code
Regulation Number
874.1050
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GSI Corti series is indicated for testing of cochlear function in infants, children and adults by measuring otoacoustic emissions (OAEs). The OAEs are generated by a series of clicks that are directed into the ear canal. Otoacoustic emissions are low level audio-frequency sounds that are produced by the cochlea as part of the normal-hearing process. Available evidence suggests that otoacoustic emissions are generated by the cochlea's outer hair cells and that the presence of OAEs. is an indication that the outer hair cells are vidence indicates that these emissions normally occur with normal hearing, or at most, mild hearing loss (usually 30-40 dB HL). The majority of hearing-impaired individuals will be identified by a simple OAE test.

Device Description

Corti test system provides rapid measurement and documentation of Distortion Product Otoacoustic Emissions (DPOAEs) or Transient Evoked Otoacoustic Emissions (TEOAEs) at several frequencies.

The system consists of the instrument, probe, printer, single-use eartips replaceable probe tubes and other accessories. The Corti instrument contains the hardware and software for generating the test stimuli, measuring and displaying the OAEs, and storing the results until they are printed. The plastic housing contains circuit boards that provide the signal processing and display the test results. The instrument also contains a rechargeable lithium-ion battery to power the device. The instrument uses a liquid-crystal display (LCD) and three light-emitting diodes (LEDs) to provide a visual display of test status to the operator. Four push buttons located on the keypad of the device allow the user to control testing and printing, and to reset test protocols.

The Probe houses the speaker and microphone which produce test stimuli and measure the sound pressure level (SPL) present in the sealed ear canal. Interface of the instrument to the ear canal is accomplished through disposable eartips, which fit onto the probe tube. The disposable eartips are color coded to facilitate easy selection by size.

Distortion Product Otoacoustic Emissions (DPOAEs) are acoustic signals that can be detected in the ear canal of a person with normal outer hair cell function, subsequent to stimulation of the auditory system with a pair of pure tones at frequencies f1 and f2. The resulting emission of interest is the distortion product tone at the frequency 2f1-f2.

The Corti instrument generates a series of test tones, directs them into the ear canal, and then measures the level of the DPOAE tone generated by the cochlea. By using different test frequencies, the Corti device provides an estimate of outer hair cell function over a wide range of frequencies.

Transient Evoked Otoacoustic Emissions (TEOAEs) are acoustic signals that can be detected in the ear canal of a person with normal outer hair cell function, subsequent to stimulation of the auditory system with a series of wideband clicks.

The Corti instrument generates a series of clicks, directs them into the ear canal, and then analyzes the spectrum of the returning signal, separating the noise and emission. By using band pass filters, the Corti device provides an estimate of outer hair cell function over a wide range of frequencies

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the GSI Corti device, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria in a table format, nor does it provide specific numerical performance values for the GSI Corti that are directly compared against such criteria. Instead, it states that the device was validated to ensure it provides equivalent diagnostic results to an "equivalent device" (the ERO-SCAN predicate device).

The "criteria" are implied by adherence to international standards and the demonstration of "substantial equivalence."

Acceptance Criteria (Implied)Reported Device Performance
Diagnostic Equivalence to Predicate Device (ERO-SCAN)Device provides equivalent diagnostic results to the ERO-SCAN.
Conformity with Essential Principles for hearing diagnosisDemonstrated conformity.
Performance and Safety Claims met and documentedMet and documented.
Risk Management (identification, addressing, evaluation)All risks identified, addressed, and evaluated; acceptable and weighted.
Adherence to OAE International Standards (IEC 60645 series)Meets requirements from IEC 60645 series.
Adherence to EMC and Safety Standards (IEC 60601-1 series)Complies with IEC 60601-1 series.
Software Validation (per FDA Guidance, "moderate" level of concern)Software verification testing conducted and documentation provided.
Auditory Impedance Testing Characteristics and Safety Systems comparable to predicateCompared and found comparable.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: "A selection of test subjects" with a targeted population of approximately 80% normal hearing and 20% having a range of impairment (from complete impairment to mild hearing impairment). The exact number of subjects is not specified but is referred to as "the subject ears."
  • Data Provenance: Not explicitly stated, but it's a clinical validation study, implying prospective data collection in a controlled environment. The country of origin for the data is not mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The study compares the GSI Corti against a predicate device (ERO-SCAN), suggesting that the "ground truth" might be established by the results of the predicate device or by standard audiological assessments. However, the details of how the truth was established (e.g., specific expert audiologists or independent clinical diagnoses) are missing.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This is not an MRMC study or an AI-assisted interpretation device. The GSI Corti is an audiological device that measures otoacoustic emissions. The study compares the performance of the new device to a predicate device, not the performance of human readers with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in a way. The "performance" being evaluated is of the device itself (GSI Corti) in measuring OAEs and providing diagnostic results, in comparison to another device (ERO-SCAN). There isn't a human interpreting the device's output and that output then being compared to a human baseline. The device's output is directly compared to the predicate device's output.

7. The Type of Ground Truth Used

The "ground truth" for the clinical validation was effectively the diagnostic results obtained from the predicate device (ERO-SCAN) and the known hearing status of the subjects (approximately 80% normal hearing, 20% impaired). The study aimed to determine if the GSI Corti provided "equivalent diagnostic results" to the ERO-SCAN, suggesting the ERO-SCAN's output served as the reference for equivalence.

8. The Sample Size for the Training Set

This information is not applicable or not provided. This is a device validation study, not a machine learning model training study. The device's algorithms are built into the hardware/software and are not "trained" on a dataset in the typical sense of AI/ML.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable or not provided as it's not a machine learning model training study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 11, 2018

Grason-Stadler, Inc. Amy Yanta Director of Regulatory Affairs 10395 West 70th Street Eden Prairie, MN 55344

Re: K180287

Trade/Device Name: GSI Corti Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO Dated: February 1, 2018 Received: February 1, 2018

Dear Amy Yanta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours,

Srinivas Nandkumar -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180287

Device Name GSI Corti

Indications for Use (Describe)

The GSI Corti series is indicated for testing of cochlear function in infants, children and adults by measuring otoacoustic emissions (OAEs). The OAEs are generated by a series of clicks that are directed into the ear canal. Otoacoustic emissions are low level audio-frequency sounds that are produced by the cochlea as part of the normal-hearing process. Available evidence suggests that otoacoustic emissions are generated by the cochlea's outer hair cells and that the presence of OAEs. is an indication that the outer hair cells are vidence indicates that these emissions normally occur with normal hearing, or at most, mild hearing loss (usually 30-40 dB HL). The majority of hearing-impaired individuals will be identified by a simple OAE test.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Grason-Stadler. The logo features a stylized sound wave graphic in light blue on the left. To the right of the sound wave is the text "gsi" in bold, black font. Below "gsi" is the text "Grason-Stadler" in a smaller, thinner, black font.

510(k) Summary as required by 21 CFR 807.92.

Administrative Information

  • Submitter: Grason-Stadler Inc. 10395 West 70th Street Eden Prairie, MN 55344 Tel: 952-947-6097 Fax: 952-278-4401
    Contact Person: Amy Yanta Director of Regulatory Affairs 10393 West 70th Street Eden Prairie, MN 55344 952-947-6097 amyy@diagnostic-group.us

Date Summary Prepared: January 29, 2018

Device Identification

Trade Name: GSI Corti Common Name: audiometer, otoacoustic emission device Device Classification Name: Audiometer Device classification: Class II Panel: Ear Nose & Throat Classification Regulation: 874.1050 Product Code: EWO Predicate Device 1: ER36 Series OAE Test System, cleared on 03/26/2015 via K150491

Device Description

Corti test system provides rapid measurement and documentation of Distortion Product Otoacoustic Emissions (DPOAEs) or Transient Evoked Otoacoustic Emissions (TEOAEs) at several frequencies.

The system consists of the instrument, probe, printer, single-use eartips replaceable probe tubes and other accessories. The Corti instrument contains the hardware and software for generating the test stimuli, measuring and displaying the OAEs, and storing the results until they are printed. The plastic housing contains circuit boards that provide the signal processing and display the test results. The instrument also contains a rechargeable lithium-ion battery to power the device. The instrument uses a liquid-crystal display (LCD) and three light-emitting diodes (LEDs) to provide a visual display of test status to the operator. Four push buttons located on the keypad of the device allow the user to control testing and printing, and to reset test protocols.

The Probe houses the speaker and microphone which produce test stimuli and measure the sound pressure level (SPL) present in the sealed ear canal. Interface of the instrument to the ear canal is

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a logo for Grason-Stadler. The logo features the letters "gsi" in a bold, sans-serif font, with the "g" slightly offset to the left and a blue sound wave graphic next to it. Below the letters "gsi" is the name "Grason-Stadler" in a smaller, sans-serif font.

accomplished through disposable eartips, which fit onto the probe tube. The disposable eartips are color coded to facilitate easy selection by size.

Distortion Product Otoacoustic Emissions (DPOAEs) are acoustic signals that can be detected in the ear canal of a person with normal outer hair cell function, subsequent to stimulation of the auditory system with a pair of pure tones at frequencies f1 and f2. The resulting emission of interest is the distortion product tone at the frequency 2f1-f2.

The Corti instrument generates a series of test tones, directs them into the ear canal, and then measures the level of the DPOAE tone generated by the cochlea. By using different test frequencies, the Corti device provides an estimate of outer hair cell function over a wide range of frequencies.

Transient Evoked Otoacoustic Emissions (TEOAEs) are acoustic signals that can be detected in the ear canal of a person with normal outer hair cell function, subsequent to stimulation of the auditory system with a series of wideband clicks.

The Corti instrument generates a series of clicks, directs them into the ear canal, and then analyzes the spectrum of the returning signal, separating the noise and emission. By using band pass filters, the Corti device provides an estimate of outer hair cell function over a wide range of frequencies

Device Intended Use

The GSI Corti is a test instrument that measures otoacoustic emissions in infants, children, and adults.

The GSI Corti series is indicated for testing of cochlear function in infants, children and adults by measuring otoacoustic emissions (OAEs). The OAEs are generated by a series of clicks that are directed into the ear canal.

Otoacoustic emissions are low level audio-frequency sounds that are produced by the cochlea as part of the normal-hearing process. Available evidence suggests that otoacoustic emissions are generated by the cochlea's outer hair cells and that the presence of OAEs is an indication that the outer hair cells are viable. Clinical evidence indicates that these emissions normally occur with normal hearing, or at most, mild hearing loss (usually 30-40 dB HL). The majority of hearing-impaired individuals will be identified by a simple OAE test.

Technological Characteristics

A comparison between the subject and predicate device shows that the technological characteristics and indications for use are equivalent. The device employs similar technology to accomplish the same tasks as the predicate. A detailed table is provided below.

DescriptionER36 Series OAE Test System(K150491)GSI CortiEquivalency
TypeAudiometerAudiometersame
RegulationNumber21 CFR 874.1050(otoacoustic emission device)21 CFR 874.1050(otoacoustic emission device)same
ClassificationProduct CodeEWOEWOGrason-Stadler
Regulatory ClassClass IIClass IIsame
Indications for UseThe ER36 is a test instrumentthat measures otoacousticemissions in infants, children,and adults.The GSI Corti is a test instrumentthat measures otoacousticemissions in infants, children,and adults.same
The ER36 series is indicated fortesting of cochlear function ininfants, children and adults bymeasuring otoacoustic emissions(OAEs). The OAEs are generatedby a series of clicks that aredirected into the ear canal.The GSI Corti series is indicatedfor testing of cochlear functionin infants, children and adults bymeasuring otoacousticemissions (OAEs). The OAEs aregenerated by a series of clicksthat are directed into the earcanal.same
Otoacoustic emissions are lowlevel audio frequency soundsthat are produced by the cochleaas part of the normal-hearingprocess. Available evidencesuggests that otoacousticemissions are generated by thecochlea's outer hair cells and thatthe presence of OAEs is anindication that the outer haircells are viable. Clinical evidenceindicates that these emissionsnormally occur with normalhearing, or at most, mild hearingloss (usually 30-40 dB HL). Themajority of hearing-impairedindividuals will be identified by asimple OAE test.Otoacoustic emissions are lowlevel audio frequency soundsthat are produced by thecochlea as part of the normal-hearing process. Availableevidence suggests thatotoacoustic emissions aregenerated by the cochlea's outerhair cells and that the presenceof OAEs is an indication that theouter hair cells are viable.Clinical evidence indicates thatthese emissions normally occurwith normal hearing, or at most,mild hearing loss (usually 30-40dB HL). The majority of hearing-impaired individuals will beidentified by a simple OAE test.same
TargetPopulationInfants, children, and adultsInfants, children, and adultssame
OAE Specifications
OAEMeasurementTypeDistortion-Product OAETransient-Evoked OAEDistortion-Product OAETransient-Evoked OAEsame
DPOAEFrequencyRange1.0 kHz to 12kHz1.0 kHz to 12kHz
DPOAE StimulusLevel40 to 65 dB SPL40 to 65 dB SPLsame
DPOAEAutomaticPass/Fail TestYesYessame
DPOAECalibrationIn-Ear self-calibrationIn-Ear self-calibrationsame
DPOAE f1/f2Frequency Ratio1.221.22same
DPOAE AnalysisBands4 to 6 bands available4 to 6 bands availablesame
TEOAEFrequencyRange500 Hz to 4000 Hz500 Hz to 4000 Hzsame
TEOAE StimulusLevelAdjustable, 70 to 85 dB SPLAdjustable, 70 to 85 dB SPLsame
TEOAEAutomaticPass/Fail TestYesYessame
TEOAECalibrationIn-Ear self-calibrationIn-Ear self-calibrationsame
TEOAE AnalysisBands66same
Instrument Characteristics and Specifications
Instrument TypeHandheldHandheldsame
Hand heldHousing Shape2.5in x 6in7 in. x 2.75 in x 1.25 inLarger devicefor betterhandle by user
HandheldHousingMaterialPC-ABSPC+ABS CX7240Same (materialname morespecific)
Housing of EarProbeMachined AluminumMachined Aluminumsame
User Controls4 button directional keyboard4 button directional keyboardsame
4 ButtonNavigationSwitchSurface mount tactile switchesSurface mount tactile switchessame
Cable-ProbeConnectorProprietary durable polarizedconnectorProprietary durable polarizedconnectorsame
Test RecordsStored inInternalMemoryUp to 250 testsUp to 250 testssame
Memory TypeFlash EEPROMFlash EEPROMsame
Power SupplyRechargeable Lithium-IonBattery. Not accessible by user.Rechargeable Lithium-IonBattery. Not accessible by user.same
Maximum90 dB SPL90 dB SPLsame
Test ProbeTypesCabled probeCabled probesame
User IndicatorsMonochrome or Color Character-based LCD display; 4 LEDindicatorsOLED DisplayProvides asharper imagefor user
Connection toComputerUSB interfaceUSB A to micro-B interfaceSame (Morespecificdescription ofUSB)
PrinterBluetooth wireless connection toAC or battery-powered thermalprinterBluetooth wireless connectionto AC or battery-poweredthermal printersame
TestPerformanceAAMIANSIS3.6:2010Specifications for AudiometersIEC 60645-1:2012:Electroacoustics - AudiometricEquipment - Part 1: Pure-ToneAudiometersIEC 60645-3:2007:Electroacoustics - AudiometricEquipment - Part 3: Test Signalsof Short Duration, Second EditionIEC 60645-6:2009:Electroacoustics - AudiometricEquipment - Part 6: Instrumentsfor the Measurement ofOtoacoustic Emissions, FirstEdition 2009-04AAMIANSIS3.6:2010IEC 60645-1:2012IEC 60645-3:2007IEC 60645-6:2009same
Operational andSafetyAAMIANSI 60601-1:2005(R):2012: Medical ElectricalEquipment, Part 1:GeneralRequirements for Safety, ThirdEditionAAM IANSI IEC62366:2007:(R)2013:MedicalDevices - Application of UsabilityEngineering to Medical Devices,Second Edition 2007-03-01ISO 14971:2007: Medical Devices,Application of Risk Managementto Medical Devices, SecondEdition 2007-03-01IEC 60601-1:2005+A1:2012(E)ANSI/AAMI 60601-1:2005/(R)2012AAMI/ANSI/IEC 62366:2015ISO 14971:2007UL Registered, File E486032Same

Equivalence Predicate Chart 1:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Grason-Stadler. The logo features a blue sound wave graphic on the left, followed by the letters "gsi" in bold, black font. Below "gsi" is the text "Grason-Stadler" in a smaller font size.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Grason-Stadler (GSI). The logo features a stylized sound wave graphic in light blue, followed by the letters "gsi" in bold, black font. Below "gsi" is the text "Grason-Stadler" in a smaller, black font.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Grason-Stadler. The logo features a stylized sound wave graphic in light blue on the left. To the right of the sound wave is the text "gsi" in bold, black font, with "Grason-Stadler" in a smaller font size underneath.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Grason-Stadler, a company that specializes in audiology equipment. The logo features the letters "gsi" in a bold, sans-serif font, with the "g" stylized to include sound waves emanating from it. Below the letters is the name "Grason-Stadler" in a smaller font. The logo is simple and modern, and the use of sound waves in the "g" suggests the company's focus on audiology.

UL Registered, File E359876

Summary of Non-Clinical Testing

Design verification and validation were performed according to current standards for OAE to assure the device meets its performance specifications. EMC and Safety was performed in compliance with recognized standards IEC 60601-1 series, Medical Electrical Equipment – General requirements for basic safety and essential performance. The product meets the requirements from the international standard for OAE measurements IEC 60645 series. Software verification testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in medical Devices." The software for this device was considered as a "moderate" level of concern since a failure in the device may result in unacceptable loud stimuli which might affect the hearing of the patient. Additionally, a malfunction in the device may delay the proper diagnostic of the hearing problem in patients. The detailed information about the validation and verification of PASS/REFER for the OAE is provided in the GSI Corti Manual, e.g., PASS/REFER Criteria, Sensitivity and Specificity etc. Based on the fulfillment of the international standards for OAE, we believe the device is safe and effective. The auditory impedance testing characteristics and safety systems were compared and found to be comparable.

Summary of Clinical Testing

The clinical validation was performed to determine substantial equivalence in the performance of the brand name GSI Corti to the ERO-SCAN. These instruments were used to perform Distortion-Product Otoacoustic Emissions (DPOAE) and Transient-Evoked Otoacoustic Emissions (TEOAE). The individual test results were evaluated for overall test result and signal-to-noise ratio at each frequency. The entire data set was evaluated for emission level, signal-to-noise ratio, as well as the noise floor. The diagnostic performance of the device was evaluated on a selection of test subjects. A population of approximately 80% normal hearing was targeted for the study population, with the remaining approximately 20% having a range of impairment from complete impairment (cavity) to mild hearing impairment. For each subject, the following tests were performed. These tests were performed in an alternating sequence, as arbitrarily determined by the clinical investigator. Each set of tests were performed in their entirety at each of the scheduled test times, on each of the two days.

DPOAE: 4 Frequency (2.0 kHz to 5.0 kHz(default DP screening protocol)) – 65/55 dB SPL Primaries – 4 Second Averaging Time

DPOAE: 6 Frequency (1.5 kHz to 6 kHz(default DP diagnostic protocol)) – 65/55 dB SPL Primaries – 4 Second Averaging Time

TEOAE: 6 Frequency (700 Hz to 4 kHz (default diagnostic TE protocol)) – 83 dB SPL Click Stimulus – 64 Second Maximum Time

TEOAE: 6 Frequency (1.5 kHz to 4 kHz (default TE screening protocol)) – 83 dB SPL Click Stimulus – 64 Second Maximum Time

The results were analyzed to determine if the device provided equivalent diagnostic results on the subject ears as another equivalent device. The results of the validation are considered evidence, when

{9}------------------------------------------------

510(k) Number: K180287 GSI Corti

Image /page/9/Picture/1 description: The image shows the logo for Grason-Stadler, also known as GSI. The logo features a stylized blue sound wave graphic on the left, followed by the letters "gsi" in bold, black font. Below "gsi" is the text "Grason-Stadler" in a smaller font size.

combined with the verification testing, of the ability of the device to meet the requirements associated with the indications for use.

Based on the data and rationales in the clinical validation report it is concluded that for the indication for diagnose and evaluation of hearing;

  • The devices demonstrates conformity with the essential principles
  • Performance and safety claims have been met and documented
  • All risks regarding indications for use and clinical data have been identified, addressed and evaluated
  • . Risks associated with the use of the devices are acceptable and weighted

Conclusion

We have compared the intended use and performance characteristics with the predicate device. The Corti was tested according to current standards and the differences found between the devices were related to minor differences of functionality, not in relation to safety and efficiency. The differences found between the devices were related to minor differences of functionality such as the Corti has a different splash screen on the device and user has option to recharge the batteries utilizing the device cradle in addition to using the charging cable. The Corti conforms to the current standards. After analyzing bench testing, safety, EMC, and software validation (with risk analysis) testing we conclude that the Corti is found to be substantially equivalent to the predicate device in technological characteristics and indications for use.

§ 874.1050 Audiometer.

(a)
Identification. An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.(b)
Classification. Class II. Except for the otoacoustic emission device, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, if it is in compliance with American National Standard Institute S3.6-1996, “Specification for Audiometers,” and subject to the limitations in § 874.9.