K Number
K180229

Validate with FDA (Live)

Date Cleared
2019-01-11

(350 days)

Product Code
Regulation Number
886.1570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RESCAN 700 provides non-contact, high resolution, optical coherence tomographic (OCT) and biomicroscopic imaging of the anterior and posterior segment of the eye via an ophthalmic surgical microscope. The RESCAN 700 is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging of posterior ocular structures, including retina, macula, and optic disc, as well as imaging of anterior ocular structures, including the cornea, lens and anterior chamber angle.

RESCAN 700 uses the assistance system (CALLISTO eye) that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The assistance system allows the remote control of RESCAN 700.

Device Description

RESCAN 700 brings Spectral Domain OCT technology to the Zeiss OPMI Lumera 700 ophthalmic surgical microscope. Used in conjunction with the assistance system, CALLISTO eye, OCT images taken intra-operatively are presented on the monitor and may also be seen within the surgeon's oculars using the OPMI Lumera 700's integrated data injection system (IDIS). OCT images may be stored for subsequent retrieval using CALLISTO eye's data management system. RESCAN 700 can be controlled via the touch panel of the assistance system or via the foot control panel of an ophthalmic surgical microscope.

The proposed RESCAN 700 device and its primary functionality remain unchanged from the previously cleared device while including the following modifications:
Improvement of OCT Visualization
Increased Power of OCT Beam
Increased Scan Depth

RESCAN 700 is used with CALLISTO eye Software version 3.6.

CALLISTO eye Software version 3.6 is an assistance, information, and documentation system to support ophthalmic surgical procedures. CALLISTO eye software operates as an adjunct to the OPMI Lumera family of ophthalmic surgical microscopes to process real-time videos and OCT data (B-Scan images). CALLISTO eye must be installed on a touchscreen Panel PC; the Panel PC is offered as an accessory.

In conjunction with RESCAN 700 the following modifications were added to CALLISTO eye Software version 3.6.
Ability to record surgical videos directly to a connected USB hard drive.
The OCT function was enhanced by adding OCT XY-tracking for the Scan Location Marker. The functionality is similar to the Z-tracking function but works in the XY-plane instead of the Z-axis.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the RESCAN 700 device, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA 510(k) summary doesn't explicitly define "acceptance criteria" in a quantitative, pass/fail manner for the changes made to RESCAN 700. Instead, it describes general improvements and then presents a clinical study to confirm the acceptability of these changes. The core "acceptance criteria" for this submission would be that the modifications do not raise new issues of safety or effectiveness and that the upgraded device performs at least as well as, or better than, the predicate device, especially regarding image quality at new scan depths.

Acceptance Criteria (Implied)Reported Device Performance
OCT Visualization Improvement: Optimized noise reduction and increased OCT beam power should result in improved visual appearance and signal-to-noise ratio of live OCT images.The modifications were implemented to improve OCT visualization and remained compliant with applicable standards.
Increased Scan Depth Acceptability: New scan depths (2.9 mm, 5.8 mm) should provide clinically acceptable image quality.2.9 mm scan depth: Image quality was similar to the predicate's 2.0 mm (corneal and glaucoma subgroups), and rated slightly higher in the retinal surgery subgroup. Conclusion: Acceptable.
5.8 mm scan depth: Rated lower than 2.0 mm and 2.9 mm in corneal and glaucoma subgroups, but may provide additional spatial context at the expense of finer detail. Conclusion: Acceptable for specific use cases (additional spatial context).
Safety and Effectiveness: Modifications must not raise new issues of safety or effectiveness compared to the predicate.The modifications do not raise new issues of safety or effectiveness. The device conforms to applicable FDA recognized, international IEC, and ISO standards for performance and safety (e.g., ISO 14971, ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 62366-1, ISO 10936-2, ISO 15004-2, IEC 62471).
Software Functionality: CALLISTO eye software version 3.6 with new features should function as intended without adverse impact on RESCAN 700.Software verification and validation activities were successfully completed, proving RESCAN 700 and CALLISTO eye meet requirements and perform together as intended.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Clinical Study): 22 participants
    • 11 with corneal conditions
    • 7 with glaucoma
    • 4 with retina conditions
    • A total of 84 images were obtained.
  • Data Provenance: The document does not explicitly state the country of origin or whether the study was retrospective or prospective. Given it's a "clinical study" and "participants were enrolled," it suggests a prospective study. The location is not specified, but the manufacturer is Carl Zeiss Meditec AG, Germany, with a US contact, so it could be in either region or internationally.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Three (3) independent, masked graders.
  • Qualifications: The document does not specify the exact qualifications (e.g., "radiologist with 10 years of experience"). However, given the context of ocular surgery and OCT images, it's highly probable these were ophthalmologists or similarly qualified eye care specialists with experience in interpreting OCT images. The term "graders" typically implies expertise in evaluating clinical images.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly stated as a formal adjudication method (like 2+1, 3+1). However, the document mentions "fair to good concordance between graders," suggesting that while there were three independent graders, a formal resolution process for disagreements is not detailed. The results likely involved an analysis of the ratings from all three for each image.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

  • MRMC Study: Yes, a type of MRMC study was conducted. The study compared the visualization quality of OCT images from two RESCAN 700 systems (predicate vs. modified with new software/capabilities) using three independent graders across multiple cases (22 participants, 84 images).
  • Effect Size of Human Readers' Improvement with AI vs. without AI Assistance: This specific information is not provided. The study aimed to compare image quality between different device versions and scan depths, not explicitly to measure the improvement of human readers with AI assistance versus without. The CALLISTO eye system provides "assistance" (non-diagnostic video documentation, image capture, remote control, and new "Cockpits" for information visualization), but the study's focus was on the inherent image quality differences produced by the hardware/software modifications, not on a human-AI interactive improvement metric.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: No, a standalone algorithm performance study was not explicitly described for diagnostic purposes. The CALLISTO eye is an "assistance system" that provides "non-diagnostic video documentation and image capture." The "improvement of OCT Visualization" involved optimized standard image processing methods for noise reduction, which could be considered an algorithmic enhancement, but its performance was evaluated via the clinical study where human graders assessed the resulting images. There is no mention of the device making automated diagnoses or measurements that were evaluated in a standalone manner against a ground truth.

7. The Type of Ground Truth Used

  • Ground Truth: The clinical utility and quality of the images were assessed by expert consensus/opinion (three independent, masked graders). There's no mention of pathology, long-term outcomes data, or other objective measures being used for ground truth in this specific image quality study.

8. The Sample Size for the Training Set

  • Training Set Sample Size: This information is not provided in the document. The general description focuses on the verification and validation of the modified device, not on the development or training of specific machine learning models within the device that would require a distinct training set. The "optimized image processing methods for noise reduction" likely involved internal development and testing, but details on a formal training set for this are not disclosed.

9. How the Ground Truth for the Training Set Was Established

  • Training Set Ground Truth: As the training set size is not provided, how its ground truth was established is also not described. If "optimized image processing methods" involved machine learning or AI, the ground truth for such internal training would have typically involved expert-labeled images or simulations. However, this level of detail is absent from the 510(k) summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

January 11, 2019

Carl Zeiss Meditec AG % Mandy Ambrecht Staff Regulatory Affairs Specialist Carl Zeiss Meditec, Inc. 5160 Hacienda Drive Dublin, CA 94568

Re: K180229

Trade/Device Name: RESCAN 700, CALLISTO eye Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO Dated: December 7, 2018 Received: December 10, 2018

Dear Mandy Ambrecht:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours.

Bradley S. Cunningham -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180229

Device Name RESCAN 700

Indications for Use (Describe)

RESCAN 700 provides non-contact, high resolution, optical coherence tomographic (OCT) and biomicroscopic imaging of the anterior and posterior segment of the eye via an ophthalmic surgical microscope. The RESCAN 700 is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging of posterior ocular structures, including retina, macula, and optic disc, as well as imaging of anterior ocular structures, including the cornea, lens and anterior chamber angle.

RESCAN 700 uses the assistance system (CALLISTO eye) that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The assistance system allows the remote control of RESCAN 700.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY (per 21 CFR §807.92)

RESCAN 700

GENERAL INFORMATION

Manufacturer:Carl Zeiss Meditec AGGoeschwitzer Strasse 51-5207745 Jena, Germany+49 7364 206985+49 7364 954432
Contact Person:Mandy AmbrechtSenior Staff Regulatory Affairs SpecialistCarl Zeiss Meditec, Inc.5160 Hacienda DriveDublin, California 94568(925) 557-4561 (phone)(925) 557-4259 (fax)
Date Summary Prepared:January 3, 2019
Classification nameOphthalmoscope
Classification886.1570
ClassII
Product CodeOBO
Trade / Proprietary NameRESCAN 700
Common NameTomography, Optical coherence
Primary Predicate DeviceCompany:Device:Carl Zeiss Meditec AGRESCAN 700, K141844
Secondary Predicate DeviceCompany:Device:Carl Zeiss Meditec AGCIRRUS photo, K112184

{4}------------------------------------------------

INDICATIONS FOR USE (21 CFR §807.92(a)(5))

RESCAN 700

RESCAN 700 provides non-contact, high resolution, optical coherence tomographic (OCT) and biomicroscopic imaging of the anterior and posterior segment of the eye via an ophthalmic surgical microscope. The RESCAN 700 is indicated for in vivo viewing, axial cross sectional. and three-dimensional imaging of posterior ocular structures, including retina, macula, and optic disc, as well as imaging of anterior ocular structures, including the cornea, lens and anterior chamber angle.

RESCAN 700 uses the assistance system (CALLISTO eye) that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The assistance system allows the remote control of RESCAN 700.

DEVICE DESCRIPTION (21 CFR §807.92(a)(4))

RESCAN 700

RESCAN 700 brings Spectral Domain OCT technology to the Zeiss OPMI Lumera 700 ophthalmic surgical microscope. Used in conjunction with the assistance system, CALLISTO eye, OCT images taken intra-operatively are presented on the monitor and may also be seen within the surgeon's oculars using the OPMI Lumera 700's integrated data injection system (IDIS). OCT images may be stored for subsequent retrieval using CALLISTO eye's data management system. RESCAN 700 can be controlled via the touch panel of the assistance system or via the foot control panel of an ophthalmic surgical microscope.

Product Update: RESCAN 700

The proposed RESCAN 700 device and its primary functionality remain unchanged from the previously cleared device while including the following modifications:

Improvement of OCT Visualization ●

To improve the visual appearance of the live OCT images, the standard image processing methods for noise reduction were optimized.

. Increased Power of OCT Beam

The power of the OCT Beam was increased from 490 µW to 620 µW to improve the signal-to-noise ratio.

Increased Scan Depth ●

The scan depth was increased to provide images at a scan depth of 2.9 mm or 5.8 mm.

{5}------------------------------------------------

RESCAN 700 is used with CALLISTO eye Software version 3.6.

CALLISTO eye Software version 3.6 is an assistance, information, and documentation system to support ophthalmic surgical procedures. CALLISTO eye software operates as an adjunct to the OPMI Lumera family of ophthalmic surgical microscopes to process real-time videos and OCT data (B-Scan images). CALLISTO eye must be installed on a touchscreen Panel PC; the Panel PC is offered as an accessory.

In conjunction with RESCAN 700 the following modifications were added to CALLISTO eye Software version 3.6.

  • Ability to record surgical videos directly to a connected USB hard drive. ●
  • The OCT function was enhanced by adding OCT XY-tracking for the Scan Location Marker. The functionality is similar to the Z-tracking function but works in the XY-plane instead of the Z-axis.

Table 5-1 presents the functions offered by the previous and proposed CALLISTO eye with a focus on the interaction with RESCAN 700.

CALLISTO eye FunctionsSWVersion3.2SWVersion3.6
as cleared in premarket notificationK141844
Network connectivityXX
User managementXX
UpdatedPatient managementXX
Surgery documentation (video recording and stillimages)XX
Ophthalmic surgical microscope remote control(OPMI LUMERA 700)XX
Live videoXX
Eye trackingX
NewExport of OCT data to connected PACS system(FORUM)*X

Table 5-1: CALLISTO eye Functions by Software Version (Note: The functionality marked with an asterisk (*) is not the subject of this 510(k) Premar

{6}------------------------------------------------

CALLISTO eye FunctionsSWVersion3.2SWVersion3.6
as cleared in premarket notificationK141844
UpdatedControl of the RESCAN 700 (OCT Camera)connected to the OPMI LUMERA 700XX
Visualization of optical coherence tomography(OCT) dataXX
Capture and review of OCT data(images and B-scans)XX
Support of 1-Line, 2-Line (OCT cube) and5-Line OCT dataXX
NewXY-tracking support for Scan Location Marker(for OCT views based on vessels)X
NewSupport for 2.9 mm and 5.8 mm scan depthsX
NewDefault combination of scan length and scandepth.X
UpdatedCombined full screen mode for microscope livevideo and OCT live viewX
NewIcon-based ocular overlays to visualize theprovided information (“Cockpits”) in the datainjection system of the connected microscopeX
Import / Export of patient treatment data to aUSB storage deviceXX
NewUSB recording: Surgery documentation can berecorded in parallel to a USB hard driveX

RISK MANAGEMENT

Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by design means, protection measures and user instructions. To confirm that the measures are effective and that the product meets its intended uses, verification of requirements and standards, and validation of the clinical workflow was performed. Carl Zeiss Meditec adheres to recognized and established industry practice and relevant international standards where indicated.

{7}------------------------------------------------

PERFORMANCE DATA & SUMMARY OF VERIFICATION AND VALIDATION ACTIVITY (21 CFR §807.92(B))

Performance testing was conducted on the subject device to address modifications on RESCAN 700. Existing performance testing remains valid. The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate devices to which it has been compared.

RESCAN 700 was designed and tested to the applicable standards for electrical and optical safety and verified to established standards. Testing of the device with RESCAN 700, OPMI Lumera 700 and CALLISTO eye Panel PC was conducted to demonstrate conformance to the third edition of 60601-1 in addition to the IEC 60601-1-2 standard for electromagnetic compatibility.

RESCAN 700 and CALLISTO eye conform to the applicable FDA recognized and international IEC and ISO standards with regards to performance and safety (see Table 5-2).

FDA Recognized Standards
IdentificationDescription
ISO 14971: 2007Medical Devices – Application of RiskManagement to Medical Devices
ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012,C1:2009/(R)2012 andA2:2010/(R)2012Electrical equipment -- Part 1: Generalrequirements for basic safety and essentialperformance
IEC 60601-1-2:2014Medical electrical equipment - Part 1-2:General requirements for basic safety andessential performance - Collateral standard:Electromagnetic compatibility - Requirementsand tests
IEC 60601-1-6:2013Medical electrical equipment - Part 1-6:General requirements for basic safety andessential performance - Collateral standard:Usability
IEC 62366-1:2015Medical devices – Application of usabilityengineering to medical devices
ISO 10936-2:2010Optics and photonics - Operation microscopes- Part 2: Light hazard from operationmicroscopes used in ocular surgery
ISO 15004-2:2007Ophthalmic Instruments - Fundamentalrequirements and test methods Part 2: Lighthazard protection

Table 5-2: FDA Recognized Standards

{8}------------------------------------------------

FDA Recognized Standards
IdentificationDescription
IEC 62471:2006Photobiological safety of lamps and lampsystems
IEC 62304:2015Medical device software - Software life cycleprocesses
NEMA PS 3.1-3.20Digital Imaging and Communications inMedicine (DICOM)

In addition to systems testing, software verification activities completed were divided into three phases:

  • Tests accompanying development (including code inspections)
  • Integration test phase - stabilization phase
  • System verification ●

Verification and validation activities were successfully completed and prove that RESCAN 700 and CALLISTO eye meet their requirements and perform together as intended.

Clinical performance testing

A clinical study was conducted to compare visualization quality of relevant posterior- and anterior-segment ocular structures during ocular surgery from OCT images between two RESCAN 700 systems using different software versions at the three available scan depths (2.0 mm [predicate device], 2.9 mm, 5.8 mm). Obtained images were anonymized and graded for clinical utility in a randomized order by three independent, masked graders using a pre-specified grading scale. A total of 22 participants were enrolled (11 with corneal conditions, seven with glaucoma, four with retina conditions) and 84 images were obtained. All images were successfully assessed by all graders, with fair to good concordance between graders. Image quality was similar between the 2.0 mm and 2.9 mm scan depths in the corneal and glaucoma surgery sub-groups. Scans at the 2.9 mm scan depth were rated slightly higher than those at the 2.0 mm scan depth in the retinal surgery sub-group. Scans at 5.8 mm were rated lower than those at 2.0- and 2.9-mm scan depths in the corneal and glaucoma surgery sub-groups. In summary, the results demonstrate that OCT image quality at the new scan depth of 2.9 mm is acceptable. Images at the 5.8 mm scan depth may allow for a view of ocular structures with additional spatial context at the expense of finer detail.

TECHNOLOGICAL CHARACTERISTICS (21 CFR §807.92(a)(6)) AND SUBSTANTIAL EQUIVALENCE (21 CFR §807.92(a)(3))

It is the opinion of Carl Zeiss Meditec AG that the proposed device, RESCAN 700 is substantially equivalent to RESCAN 700 (K141844) and CIRRUS photo (K112184) with regards to indications for use. RESCAN 700 and CIRRUS photo are both intended to provide OCT images of the anterior and posterior segment of the eye. Both devices employ a noninvasive, non-contact low-coherence interferometry technique to generate cross-sectional images of internal ocular tissue microstructures by measuring optical reflections from tissue.

{9}------------------------------------------------

Both devices obtain high-resolution images using spectral domain optical coherence tomography.

Differences with the proposed device include modifications to improve visualization constituting increased OCT power that remains below the level of CIRRUS photo and increased scan depth.

The proposed product, RESCAN 700, will be used with CALLISTO eye, Software version 3.6. The predicate device, RESCAN 700, was used with CALLISTO eye Software version 3.2 (K141844). Both versions of CALLISTO eye provide non-diagnostic video documentation and image capture for ophthalmic surgeries and allow remote control of the surgical microscope.

Both products inject information such as light, zoom and recording into the right ocular; the proposed device uses "Cockpits", icon-based ocular overlays to visualize the provided information in the data injection system of the connected microscope.

SUMMARY (21 CFR §807.92(b)(3))

RESCAN 700 has the same intended use and technological characteristics as the predicate devices, RESCAN 700 (K141844) and the same intended use as CIRRUS photo (K112184). RESCAN 700 was and continues to be used with CALLISTO eye.

The modifications to the RESCAN 700 device were implemented to improve OCT visualization while remaining compliant with the applicable standards for optical light hazards. These modifications do not raise new issues of safety or effectiveness. Therefore, ZEISS believes that the subject device, RESCAN 700 is substantially equivalent to the predicate devices, RESCAN 700 (K141844) and CIRRUS photo (K112184).

§ 886.1570 Ophthalmoscope.

(a)
Identification. An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.(b)
Classification. Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.