K Number
K180144

Validate with FDA (Live)

Date Cleared
2018-09-21

(246 days)

Product Code
Regulation Number
878.3610
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Agile Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.

Device Description

The Agile™ Esophageal Stent System is intended for the endoscopic placement of the selfexpanding nitinol stent for maintaining luminal patency in esophageal strictures. The flexible Agile Stent and through-the scope Delivery System allow for stent placement through direct visualization. The Agile™ Esophageal Stent System is supplied sterile.

The Agile™ Stent is a flexible. MR conditional, self-expanding, braided nitinol stent. The proximal and distal ends of the stent have a flare which is larger in diameter than the stent body. The stent wires at each end of the flare are looped. A suture is threaded within the looped ends of both the proximal and distal stent flares to aid in repositioning during initial placement. The proposed Agile Stent is both fully and partially covered in a silicone covering.

The Agile™ Stent comes preloaded and constrained in the delivery system. The delivery system consists of two coaxial shafts compatible with a minimum 3.7mm working channel of a therapeutic gastroscope. The exterior tube is used to constrain the stent before deployment and to reconstrain the stent after partial deployment. The delivery system has radiopaque and visual markers to aid the user in accurate stent placement. There are three visual markers on the delivery system handle to aid in the placement of the stent under endoscopic visualization and four radiopaque markers on the interior of the delivery system to aid in placement of the stent under fluoroscopic visualization.

AI/ML Overview

The provided document describes the Agile Esophageal Stent System, a medical device intended for maintaining esophageal luminal patency in esophageal strictures. The information does not detail diagnostic or AI/ML-based performance. Instead, it focuses on the engineering and biocompatibility validation of the stent system itself. Therefore, many of the requested fields related to AI/ML model performance, such as MRMC studies, ground truth establishment for training/test sets, and sample sizes for training sets, are not applicable.

Here's an analysis based on the available information:

Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

The acceptance criteria for the Agile Esophageal Stent System are implicitly defined by the successful completion of a series of biocompatibility and bench tests, along with demonstrating substantial equivalence to predicate devices. The study proving the device meets these criteria involved performing these tests and comparing the results to pre-defined specifications or to predicate device performance.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific TestReported Device Performance
Material BiocompatibilityMEM Elution CytotoxicityPass
Sub-acute (Subchronic)/Intravenous ToxicityPass
Guinea Pig Maximization SensitizationPass
Intracutaneous ReactivityPass
Sub-acute (Subchronic)/Intraperitoneal ToxicityPass
Materials Mediated Rabbit PyrogenPass
Ames MutagenicityPass
Acute Systemic InjectionPass
Mouse LymphomaPass
Analytical Testing (NVR; GCMS; LC-MS; ICPMS; HS-GC-MS)Pass
Implantation (Intramuscular, 4 weeks)Pass
Implantation (Intramuscular, 13 weeks)Pass
Functional Performance (Bench)Guidewire PassagePass
Trackability/PushabilityPass
Deployment and ReconstrainmentPass
Delivery System RemovalPass
Immediate Stent OpeningPass
Delivery System WithdrawalPass
Stent Hoop Expansion and Compression ForcePass
Delivery System Working LengthPass
Delivery System Outer DiameterPass
Visual Transition Zone LengthPass
Stent ForeshorteningPass
Final Unconstrained Stent LengthPass
Stent Flare LengthPass
Non-Covered Length (Partially Covered Codes Only)Pass
Stent Covering Length (Partially Covered Codes Only)Pass
Stent DiameterPass
Flare DiameterPass
Tip to Inner Member Bond TensilePass
Inner Jacket to Handle Assembly TensilePass
Reconstrainment Band to Inner Jacket TensilePass
Exterior Tube to Handle Bond TensilePass
Inner Jacket to Inner Member Bond TensilePass
Exterior Tube Transition Zone TensilePass
Suture Pull Strength TestPass
Corrosion TestingPass
Fatigue TestingPass
Deployment AccuracyPass
Magnetic Resonance TestingPass
Comparative Performance (Bench)Stent Hoop Expansion & Compression ForceFalls within range of predicate devices

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample size for each bench test or biocompatibility test. It reports "Pass" for each test, indicating that the tested samples met the pre-defined criteria. The tests are bench tests performed on the device components or complete systems, and biocompatibility tests conducted in vitro and in vivo (e.g., animal models). The provenance of this data is from the manufacturer's (Boston Scientific Corporation) internal testing. These are prospective tests conducted on the newly designed device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not Applicable. This request is relevant for AI/ML validation where expert consensus establishes ground truth for diagnostic accuracy. Given this is a physical medical device (stent), the "ground truth" is established by engineering specifications, material science standards, and biocompatibility guidelines, rather than expert interpretation of medical images or data.

4. Adjudication Method for the Test Set

Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are used to resolve disagreements among human readers or experts in diagnostic studies to establish a consensus ground truth. This is not relevant for the bench and biocompatibility testing of a physical device.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

Not Applicable. This is a physical medical device (stent), not an AI-based diagnostic tool that assists human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not Applicable. This is a physical medical device (stent), not a standalone algorithm.

7. The type of ground truth used

The "ground truth" for this device's performance is based on:

  • Engineering specifications and standards: For mechanical properties, dimensions, deployment accuracy, etc.
  • Biocompatibility standards (e.g., ISO 10993): For material safety tests.
  • Comparison to predicate devices: For comparative bench testing, the performance of the predicate devices serves as a benchmark for what is considered acceptable.

8. The sample size for the training set

Not Applicable. This document describes a physical medical device, not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established

Not Applicable. As there is no AI/ML model, there is no training set or associated ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 21, 2018

Boston Scientific Corporation Kayla Mackey Regulatory Affairs Specialist II 100 Boston Scientific Way Marlborough, Massachusetts 01752

Re: K180144

Trade/Device Name: Agile Esophageal Stent System Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal Prosthesis Regulatory Class: Class II Product Code: ESW Dated: August 22, 2018 Received: August 23, 2018

Dear Kayla Mackey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/7 description: The image shows the FDA logo on the left side. On the right side of the image, there is text that says "Jeffrey W" followed by "2018.09" and then "-04'00'". The FDA logo is a stylized version of the letters FDA.

Jeffrey W. Cooper -S 2018.09.21 15:47:15

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known)Unknown K180144
Device NameAgile Esophageal Stent System
Indications for Use (Describe)The Agile Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

19

{3}------------------------------------------------

SECTION 5: 510(K) SUMMARY

1. Submitter:

Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752

Kayla Mackey Regulatory Affairs Specialist II Telephone: 508-683-4534 Fax: 508-683-5939

Date Prepared: January 16, 2018

2. Device:

Trade Name:Agile™ Esophageal Stent System
Classification Name:Esophageal Prosthesis
Regulation Number:21 CFR 878.3610
Product Code:ESW
Classification:Class II

3. Predicate Device(s):

Trade Name:Wallflex Esophageal Fully &Partially Covered Stent System
510(k) Number:K091510/ K073266
Classification Name:Esophageal Prosthesis
Regulation Number:21 CFR 878.3610
Product Code:ESW
Classification:Class II
Trade Name:Taewoong Niti-S Esophageal TTSStent
510(k) Number:K123205
Classification Name:Esophageal Prosthesis
Regulation Number:21 CFR 878.3610
Product Code:ESW
Classification:Class II
Trade Name:Merit Alimaxx-ES EsophagealStent
510(k) Number:K080838

{4}------------------------------------------------

Classification Name: Regulation Number: Product Code: Classification:

Esophageal Prosthesis 21 CFR 878.3610 ESW Class II

Reference Device:

Trade Name: 510(k) Number: Classification Name: Regulation Number: Product Code: Classification:

Ultraflex Esophageal NG Stent System K091816 Esophageal Prosthesis 21 CFR 878.3610 ESW Class II

4. Device Description:

The Agile™ Esophageal Stent System is intended for the endoscopic placement of the selfexpanding nitinol stent for maintaining luminal patency in esophageal strictures. The flexible Agile Stent and through-the scope Delivery System allow for stent placement through direct visualization. The Agile™ Esophageal Stent System is supplied sterile.

The Agile™ Stent is a flexible. MR conditional, self-expanding, braided nitinol stent. The proximal and distal ends of the stent have a flare which is larger in diameter than the stent body. The stent wires at each end of the flare are looped. A suture is threaded within the looped ends of both the proximal and distal stent flares to aid in repositioning during initial placement. The proposed Agile Stent is both fully and partially covered in a silicone covering.

The Agile™ Stent comes preloaded and constrained in the delivery system. The delivery system consists of two coaxial shafts compatible with a minimum 3.7mm working channel of a therapeutic gastroscope. The exterior tube is used to constrain the stent before deployment and to reconstrain the stent after partial deployment. The delivery system has radiopaque and visual markers to aid the user in accurate stent placement. There are three visual markers on the delivery system handle to aid in the placement of the stent under endoscopic visualization and four radiopaque markers on the interior of the delivery system to aid in placement of the stent under fluoroscopic visualization.

Table 5-1: Agile Partially Covered Stent Sizes
UPNStent Size
Stent DiameterStent LengthFlare Diameter
M0051720014mm62 mm19mm
M0051721014mm102 mm19mm
M0051722014mm119 mm19mm
M0051723014mm148 mm19mm
M0051724018mm59 mm23mm
M0051725018mm97 mm23mm
M0051726018mm119 mm23mm
M0051727018mm149 mm23mm

Proposed Model Configurations:

{5}------------------------------------------------

Table 5-1: Agile Partially Covered Stent Sizes
UPNStent Size
Stent DiameterStent LengthFlare Diameter
M0051728023mm62 mm28mm
M0051729023mm101 mm28mm
M0051730023mm120 mm28mm
M0051731023mm150 mm28mm
Table 5-2: Agile Fully Covered Stent Sizes
UPNStent Size
Stent DiameterStent LengthFlare Diameter
M0051740014mm62 mm19mm
M0051741014mm102 mm19mm
M0051742014mm119 mm19mm
M0051743014mm148 mm19mm
M0051744018mm59 mm23mm
M0051745018mm97 mm23mm
M0051746018mm119 mm23mm
M0051747018mm149 mm23mm
M0051748023mm62 mm28mm
M0051749023mm101 mm28mm
M0051750023mm120mm28mm
M0051751023mm150mm28mm

5. Indications for Use:

The Agile™ Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.

6. Technological Characteristics

The intended use of the proposed Agile™ Esophageal Stent System is identical to the predicate Wallflex Esophageal Fully and Partially Covered Stent System. Taewoong Niti-S TTS Esophageal Stent, and Merit Alimaxx-ES Esophageal Stent.

The proposed Agile™ Esophageal Stent Delivery system is similiar to the predicate Taewoong Niti-S TTS Esophageal delivery system in overall length and working length, outer diameter, method of placement, and guidewire compatibility.

The proposed Agile™ Esophageal Stent is similar to the predicate Wallflex Esophageal Fully and Partially Covered Stents, Taewoong Niti-S TTS Esophageal Stents, and Merit Alimaxx-ES Esophageal Stents in terms of performance, stent diameter, and stent length. The

{6}------------------------------------------------

stent materials are similar between the proposed Agile Esophageal Stent System and the predicate Wallflex Esophageal Fully and Partially Covered Stents.

7. Biocompatibility Analysis

The following biocompatibility tests have been completed for the proposed Agile Esophageal Fully Covered Stent System.

Table 5-3 Summary of Agile Esophageal Fully Covered Stent Biocompatibility Testing
TestResults
MEM Elution Cytotoxicity/Part 5Pass
Sub-acute (Subchronic)/Part 11/ Intravenous ToxicityPass
Guinea Pig Maximization Sensitization/Part 10Pass
Intracutaneous Reactivity/Part 10Pass
Sub-acute (Subchronic)/Part 11/ Intraperitoneal ToxicityPass
Materials Mediated Rabbit Pyrogen/Part 11Pass
Ames Mutagenicity/ Part 3Pass
Acute Systemic Injection/Part 11Pass
Mouse Lymphoma / Part 3Pass
Analytical Testing/ Part 18 (NVR; GCMS; LC-MS; ICPMS;HS-GC-MS)Pass
Implantation (Intramuscular)/ Part 6(4weeks)Pass
Implantation (Intramuscular)/ Part 6(13 weeks)Pass
TestResults
MEM Elution Cytotoxicity/Part 5Pass
Intracutaneous Reactivity/Part 10Pass
Guinea Pig Maximization Sensitization/Part 10Pass
Analytical testingPass

{7}------------------------------------------------

Ultraflex Esophageal NG Stent System (K091816) was used as a reference device to support the biocompatibility of the Agile Esophageal Partially Covered Stent. The following biocompatibility tests have been leveraged from Ultraflex Esophgeal NG Stent System (K091816).

Table 5-5 Summary of Ultraflex Esophageal NG Stent System Biocompatibility Testing
TestResults
CytotoxicityPass
SensitizationPass
Intracutaneous ReactivityPass
Systemic Toxicity- Acute Systemic ToxicityPass
Subacute Toxicity- IntravenousPass
Subacute Toxicity- IntraperitonealPass
Genotoxicity- Ames AssayPass
Mouse LymphomaPass
Mouse Lymphoma / Part 3Pass
Implantation (12 weeks)Pass
USP PhysiochemicalPass

8. Performance Data

The proposed Agile™ Esophageal Stent System successfully passed all pre-defined product specifications for the tests performed. Below is a summary of the tests performed on the Agile Esophageal Stent System:

Table 5-6 Summary of Bench Tests and Results
No.TestResults(Pass/Fail)
1Guidewire PassagePass
2Trackability/PushabilityPass
3Deployment and ReconstrainmentPass
4Delivery System RemovalPass
5Immediate Stent OpeningPass
6Delivery System WithdrawalPass
7Stent Hoop Expansion and Compression ForcePass
8Delivery System Working LengthPass
9Delivery System Outer DiameterPass
10Visual Transition Zone LengthPass
11Stent ForeshorteningPass
12Final Unconstrained Stent LengthPass
13Stent Flare LengthPass

{8}------------------------------------------------

14Non-Covered Length (Partially Covered Codes Only)Pass
15Stent Covering Length (Partially Covered Codes Only)Pass
16Stent DiameterPass
17Flare DiameterPass
18Tip to Inner Member Bond TensilePass
19Inner Jacket to Handle Assembly TensilePass
20Reconstrainment Band to Inner Jacket TensilePass
21Exterior Tube to Handle Bond TensilePass
22Inner Jacket to Inner Member Bond TensilePass
23Exterior Tube Transition Zone TensilePass
24Suture Pull Strength TestPass
25Corrosion TestingPass
26Fatigue TestingPass
27Deployment AccuracyPass
28Magnetic Resonance TestingPass
Table 5-7 Summary of the Comparative Bench Tests
No.Test
29Stent Hoop Expansion & Compression Force

9. Comparative Bench Testing

Comparative bench testing was conducted on the Agile Esophageal Stent System and its substantially equivalent predicate device. The comparative testing demonstrated that the Stent Hoop Expansion and Compression Force of the Agile Esophageal Stents falls within the range of the currently marketed predicates for the propose stent diameters.

Conclusion:

The information Boston Scientific provided in this submission demonstrates that the proposed Agile Esophageal Stent System is substantially equivalent to the currently cleared predicate devices thru the design and performance characteristics and indication for use.

§ 878.3610 Esophageal prosthesis.

(a)
Identification. An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”