K Number
K180127

Validate with FDA (Live)

Date Cleared
2019-03-15

(422 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The THINK Surgical TCAT® THA Instrument Tray Set is intended to protect, organize, and deliver to the surgical field TCAT tools, instruments and accessories. The trays allow sterilization of tools, instruments and accessories, and maintain sterility of the enclosed devices until used. The trays are wrapped with an FDA-cleared sterilization wrap during the pre-vacuum autoclave sterilization process.

Device Description

The TCAT THAT Instrument Tray Set is intended only for use with TSolution One® Surgical System instruments, tools and accessories. These trays are used to enclose and hold the instruments, tools, and accessories in an organized manner during the sterilization process and subsequent storage and transportation to and from the surgical suite. The trays are designed to fit any standard autoclave and are constructed primarily of anodized aluminum meeting biocompatibility requirements and compatible with repeated steam sterilizations. The trays have perforations to facilitate sterilant penetration, evacuation, and drying. Since the trays are perforated, an FDA cleared sterilization wrap must be used to maintain sterility of the contents. The trays have the same size and same basic configuration: a rectangular base with latchable lid. The trays have perforations with an evenly distributed hole pattern on the lid, bottom, and sides to allow sterilant penetration. The bottom surface and interior shelves of the trays contain stanchions designed to separately hold each individual instrument, tool, and accessory for effective sterilant exposure, evacuation and drying during the entire duration of the sterilization process, as well as ease of locating each instrument when placing in or removing from the tray.

AI/ML Overview

The presented document describes the non-clinical testing performed on the TCAT® THA Instrument Tray Set to demonstrate its safety and effectiveness.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria / TestReported Device Performance
Sterilization Efficacy (ISO 17665-1: 2006, Sterilization of health care products-Moist Heat Part 1 Requirements for the development, validation and routine control of sterilization process for medical devices.) Purpose: Demonstrate sterilization efficacy of the TCAT THA Instrument Tray Sets containing the maximum inoculated load.The test resulted in no growth and a six-log reduction at the half-cycle for both devices (TCAT® Base Instrument Tray and TCAT® THA Instrument Tray) of the indicator organism.
Packaging (ASTM D4169-16 Standard Practice for Performance Testing in Shipping Containers and Systems) Purpose: Demonstrate that the THINK Surgical TCAT® THA Instrument Tray Set was evaluated with maximum load, wrapped according to use instructions and in the shipping configurations according to a shipping validation test.The results for both trays indicate packaging endpoints were met.
Cleaning (FDA. 2015. Guidance for Industry and FDA Staff-Processing/Reprocessing Medical Devices in Health Care Settings: Validation and Labeling) Purpose: Demonstrate that the cleaning instructions for the TCAT THA Instrument Tray Set are valid.The results for both trays indicated cleaning endpoints were met.
Biocompatibility (ISO 10993-5:2009 Cytotoxicity Test) Purpose: Demonstrate that the TCAT THA Instrument Tray Set is biocompatible.Under the conditions of the test, the device extract was not cytotoxic.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not explicitly state the numerical sample size for each test (e.g., number of trays tested for sterilization efficacy). However, it consistently refers to "both devices" or "both trays" in the results, implying that at least one of each of the two different tray types (TCAT® Base Instrument Tray and TCAT® THA Instrument Tray) were tested for each criterion where applicable. For "maximum inoculated load" and "maximum load," it implies that the trays were loaded to their full capacity when tested.
  • Data Provenance: The studies are described as "non-clinical performance testing," indicating they were conducted in a controlled, laboratory environment. The provenance is internal to the manufacturer or a contracted testing facility, following established international (ISO, ASTM) and FDA guidance standards. This is prospective testing to validate the device's design and performance. There is no mention of country of origin for the data beyond the standards applied.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • This information is not applicable as the document describes non-clinical performance testing (e.g., sterilization efficacy, packaging, cleaning, biocompatibility). These tests do not involve human interpretation of medical images or diagnoses, and, therefore, do not require medical experts to establish a "ground truth" in that context. The "ground truth" for these tests comes from objective laboratory measurements and adherence to specified scientific standards (e.g., absence of microbial growth, meeting packaging integrity metrics, non-cytotoxicity).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • This information is not applicable for the same reasons as point 3. Adjudication methods like "2+1" or "3+1" are used in studies where human reviewers provide initial assessments, and a consensus or tie-breaking mechanism is needed to establish a definitive ground truth, typically in image interpretation or clinical trial settings. The non-clinical tests described here rely on objective, quantifiable results from laboratory procedures.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This information is not applicable. The device under review is an instrument tray set for sterilization, not an AI-powered diagnostic or assistive technology for human readers. Therefore, an MRMC comparative effectiveness study involving human readers or AI assistance would not be relevant to its evaluation.

6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

  • This information is not applicable. The device is a physical medical device (sterilization tray), not an algorithm or software. Its performance is evaluated through physical and biological tests described, not through standalone algorithm performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth used for these non-clinical tests is based on objective scientific and engineering standards and measurements:
    • Sterilization Efficacy: Absence of microbial growth and a 6-log reduction of the indicator organism, as per ISO 17665-1.
    • Packaging: Meeting predefined "packaging endpoints" which are likely quantitative metrics for package integrity after simulated shipping (e.g., no damage, seals intact), as per ASTM D4169-16.
    • Cleaning: Meeting predefined "cleaning endpoints" which are likely quantitative metrics for residual soil (e.g., protein, hemoglobin levels below specified limits), as per FDA guidance.
    • Biocompatibility: Absence of cytotoxicity (i.e., no adverse cellular response to device extracts), as per ISO 10993-5:2009.

8. The sample size for the training set:

  • This information is not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI that would be relevant to these non-clinical validation tests. The testing performed is to validate the manufactured product against performance criteria.

9. How the ground truth for the training set was established:

  • This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 15, 2019

THINK Surgical Inc. % George Prendergast Manager of Regulatory Affairs THINK Surgical, Incorporated 47201 Lakeview Blvd. Fremont, California 94538

Re: K180127

Trade/Device Name: TCAT® THA Instrument Tray Set Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: February 13, 2019 Received: February 14, 2019

Dear George Prendergast:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

51O(k) Number (if known) K180127

Device Name TCAT® THA Instrument Tray Set

Indications for Use (Describe)

The THINK Surgical TCAT® THA Instrument Tray Set is intended to protect, organize, and deliver the surgical field TCAT tools, instruments and accessories. The trays allow sterilization of tools, instruments and accessories, and maintain sterility of the enclosed devices until used. The trays are wrapped with an FDAcleared sterilization wrap during the pre-vacuum autoclave sterilization process.

Table 1 lists the THINK Surgical compatible devices:

Table 1: THINK Surgical Compatible devices

Intended Instrument Tray Set ContentsDescription
TCAT General Instruments, Tools and AccessoriesInstruments, tools and accessories intended forgeneral use with the TSolution One® SurgicalSystem (TCAT)
TCAT THA Instruments, Tools, and AccessoriesInstruments, tools and accessories intended for THAuse with the TSolution One® Surgical System(TCAT)

Table 2 lists the sterilization parameters under which the TCAT® THA Instrument Tray Set was validated:

Table 2: Sterilization Parameters

CycleCycle TemperatureExposure TimeDry Time
Pre-vacuum132°C (270°F)4 minutes45 minutes

Table 3 provides a description of the TCAT® THA Instrument Tray Set (107985):

Table 3: Description of the TCAT THA Instrument Tray Set

Tray NameIntended InstrumentTray Set ContentsDimensions, L x W x H(inches)Weight
TCAT® BaseInstrument TrayPN 107735TCAT general useinstruments, tool andaccessories20 x 9.8 x 4.5Unloaded: 14.21lbs.Loaded: 17.33lbs.
TCAT® THAInstrument TrayPN107736Instruments, tools andaccessories for THAonly20 x 9.8 x 4.5Unloaded: 14.39lbs.Loaded: 19.50lbs.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "THINK Surgical". The word "THINK" is in large, gray, block letters, with a blue square above the "I". Below "THINK" is the word "SURGICAL" in smaller, blue letters. The logo is simple and modern, with a focus on the company name.

510(k) Summary of K180127

Submitter Name:THINK Surgical, Inc.
Submitter Address:47201 Lakeview Blvd., Fremont, CA 94538
Contact Person:George J. Prendergast
Phone Number:(510) 249-2323
Fax Number:(510) 249-2396
Date Prepared:March8, 2019
Device Trade Name:TCAT ®THA Instrument Tray Set
Device Common Name:Sterilization Cassettes, Instrument Tray, Sterilization Tray.Instrument Delivery System
Classification Name:Sterilization wrap,
Product CodeKCT
Regulation Number:21 CFR 880.6850
Predicate Device:Intuitive Surgical® Instrument and Accessory SterilizationTrays (K072211)

Device Description:

The TCAT THAT Instrument Tray Set is intended only for use with TSolution One® Surgical System instruments, tools and accessories. These trays are used to enclose and hold the instruments, tools, and accessories in an organized manner during the sterilization process and subsequent storage and transportation to and from the surgical suite. The trays are designed to fit any standard autoclave and are constructed primarily of anodized aluminum meeting biocompatibility requirements and compatible with repeated steam sterilizations. The trays have perforations to facilitate sterilant penetration, evacuation, and drying. Since the trays are perforated, an FDA cleared sterilization wrap must be used to maintain sterility of the contents.

The trays have the same size and same basic configuration: a rectangular base with latchable lid. The trays have perforations with an evenly distributed hole pattern on the lid, bottom, and sides to allow sterilant penetration. The bottom surface and interior shelves of the trays contain stanchions designed to separately hold each individual instrument, tool, and accessory for effective sterilant exposure, evacuation and drying during the entire duration of the sterilization process, as well as ease of locating each instrument when placing in or removing from the tray.

The sterilization trays have been tested for use only with THINK Surgical instruments. tools and accessories for a sterilization cycle of pre-vacuum steam sterilization at 132 °C,

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Think Surgical. Below the logo, the text "4 min cycle with 45-min dry time" is present. The logo is in gray and blue.

The type and maximum number of instruments, tool and accessories for each of the trays listed above are as follows:

TCAT® Base Tray (Tray 1) Top
Part NumberDescriptionQuantity
103426Cutter Motor1
106298-04Digitizer Ball Probe1
106382Digitizer Probe Hub1
103593Recovery Marker Installation Tool1
102940Cutter Verification Gauge Set1
103636Cutter Wrench Set1
105716Wrench, Hex, 2.5mm1
103279Wrench, Hex 8mm1
101884Wrench, T-handle, Square, 8mm1
104948BMM Probe Assembly1
Hospital SuppliedFixation Clamp, Ø4-5mm pin/Ø8mm rod2
TCAT® Base Tray (Tray 1) Bottom
Part NumberDescriptionQuantity
103205Fixation Arm, Straight1
106219Fixation Adapter Rod, Right-Angle, 4 inch1
103414Swivel Block1
100086Wrench, Open End, 1-1/8 inch1
103384Cutter Motor Cable1
TCAT® THA Application Tray (Tray 2) Top
Part NumberDescriptionQuantity
105653Cutter Bearing Sleeve, 244mm1
102613Recovery Marker, Tack Ø1.5 x 11mm1
102709Cutter Ball Probe, 276mm1
102862-NSCutter, Flat, Ø8.25x268mm1
106298-03Digitizer Percutaneous Probe1
107720Recovery Marker, Groove, Ø4x125mm2
105734Cuter, Ball, Ø6x268mm1
107719Recovery Marker, Divot, Ø4x70mm1
HospitalSuppliedFixation Screw, Self-Drilling, Ø5x150mm3

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Think Surgical. The word "THINK" is in large, gray, block letters. Above the "I" is a small, blue square. Below the word "THINK" is the word "SURGICAL" in smaller, blue letters.

TCAT® THA Application Tray (Tray 2) Bottom
Part NumberDescriptionQuantity
107645Cutter Drive Assembly1
105751Impactor Coupler Assembly,1
Magnetic
107667Impactor, Acetabular Cup, 0101
107666Reamer Driver, Acetabular Cup, 0012
107642Arm Tool Coupler1

Indications for Use:

The THINK Surgical TCAT® THA Instrument Tray Set is intended to protect, organize, and deliver to the surgical field TCAT tools, instruments and accessories. The trays allow sterilization of tools, instruments and accessories, and maintain sterility of the enclosed devices until used. The trays are wrapped with an FDA-cleared sterilization wrap during the pre-vacuum autoclave sterilization process.

Table 1 lists the THINK Surgical compatible devices:

Intended Instrument Tray SetContentsDescription
TCAT General Instruments, Toolsand AccessoriesInstruments, tools andaccessories intended for generaluse with the TSolution One®Surgical System (TCAT)
TCAT THA Instruments, Tools,and AccessoriesInstruments, tools andaccessories intended for THA usewith the TSolution One®Surgical System (TCAT)

Table 1: THINK Surgical Compatible Devices

Table 2 lists the sterilization parameters under which the TCAT® THA Instrument Tray Set was validated:

Table 2: Sterilization Parameters

Cycle TypeCycleTemperatureExposure TimeDry Time
Pre-vacuum132°C (270°F)4 minutes45 minutes

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Think Surgical. The logo consists of the word "THINK" in gray block letters, with a small blue square above the "I". Below the word "THINK" is the word "SURGICAL" in a smaller, blue font. The logo is simple and modern, and the use of color is effective in creating a visual hierarchy.

Table 3 provides a description of the TCAT® THA Instrument Tray Set (107985):

Tray NameIntendedInstrument TraySet ContentsDimensions,L x W x H(inches)Weight
TCAT® BaseInstrument TrayPN 107735TCAT generaluse instruments,tool andaccessories20x9.8x4.5Unloaded:14.21lbs.Loaded:17.33lbs.
TCAT® THAInstrument TrayPN107736Instruments,tools andaccessories forTHA only20x9.8x4.5Unloaded:14.39lbs.Loaded:19.50lbs.

Table 3: Description of the TCAT® THA Instrument Tray Set

Predicate Device:

The TCAT® THA Instrument Tray Sets are similar to the Intuitive Surgical® Instrument Accessory Trays, K072211.

Technological Characteristics Comparison:

Table 4 provides a comparison of technological characteristics and principles of operation between the TCAT THA Instrument Tray Set System and its predicate devices.

FeatureTCAT® THAInstrument Tray Set(Subject Device)Intuitive Surgical®Instrument AccessoryTrays, K072211Comparison
Intended UseA sterilization tray is adevice intended to beused to enclose anothermedical device that is tobe sterilized by a healthcare provider. It isintended to allowsterilization of theenclosed medical device.A sterilization tray is adevice intended to be used toenclose another medicaldevice that is to be sterilizedby a health care provider. Itis intended to allowsterilization of the enclosedmedical device.Same
LoadContentsTCAT GeneralInstrument, Tools andAccessories:Instruments, tools andaccessories intended forgeneral use with theTCAT Surgical SystemTCAT THA,Instruments Toolsand Accessories:Instruments, tools andaccessories intended forTHA use with theTCAT Surgical SystemIntuitive SurgicalEndoWrist Instruments:DaVinci and DaVinci Sfamilies of 8mm and45mm Endo WristInstrumentsIntuitive SurgicalReusable Accessories:DaVinci and Davinci Sfamilies of 8mm and 5mmaccessories includingcanulae, obturators,cannula mounts, sterileadapters, etc.
ContentsMaximumLoadMedicaldevices/instrumentsweighing a maximumof 17.33lbs for TCATbase instrument trayandweighing a maximumof 19.50lbs for theTHA instrument tray.Medicaldevices/instrumentsweighing a maximum of1lbs for Endo Wristinstrument trayandweighing a maximum of7lbs for DaVinci andDavinci S trays.Similar
Design Characteristics
DeviceCompositionBase, lift out tray, lidBase, lidSimilar
InsertsYesYesSame
HandlesYesYesSame
LatchesYesYesSame
ReusableYesYesSame
Materials
Lid/base/liftout tray, notincludinginsertsAluminumAluminumSame
InsertsSilicone, aluminum,stainless steel, nylon--
LatchStainless steel--
AssembledDimensions20" x 9.8" x 4.5"24.1" x 9.7" x 5.5"Similar
Max loadweight17lbs for TCAT basetray19.5lbs for THAinstrument tray21lbs.Similar
Sterilization
Percent of surface perforations forboth subject devices:Percent of surface perforations for thepredicate device:
-lid21%-lid, base-
-base19%--
-lift out tray22%--
SterilizationMethodPre-vacuum (Steam)Pre-vacuum (Steam)Same
CycleTemperature132°C (270°F)132°C (270°F)Same
Cycle Time4 minutes4 minutesSame
Drying time45 minutes30 minutesSimilar

Table 4:Technological Characteristics Comparison Table

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Think Surgical. The word "THINK" is in large, gray, block letters, with a blue square above the "T". Below "THINK" is the word "SURGICAL" in smaller, blue letters.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Think Surgical. The word "THINK" is in large, gray, block letters, with a blue square above the "I". Below "THINK" is the word "SURGICAL" in a smaller, blue font.

The TCAT® THA Instrument Tray Sets were found to be similar to the predicate. Three fundamental similarities are:

  1. Basic design: Both the TCAT THA Instrument Trays and the predicate have a basic lid/base design with latches, handles, perforations, and contoured inserts or sections for containing items for sterilization, storage, and transport. General size, shape, weight, and materials are not identical but are similar.

  2. Fundamental Technology: The TCAT THA Instrument Tray Sets and the predicate allow the sterilant (steam) to penetrate and render its contents sterile by relying on surface perforations.

Primary differences are related to size, shape, intended contents, surface perforations, and appearance.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Think Surgical. The logo consists of the word "THINK" in large, bold, gray letters, with a blue square above the "I". Below "THINK" is the word "SURGICAL" in smaller, blue letters. The logo is simple and modern, and it conveys a sense of innovation and precision.

Summary of Non-Clinical Testing:

Table 5 summarizes the non-clinical performance testing of the TCAT® THA Instrument Tray Sets:

  • Sterilization Efficacy ●
  • . Packaging
  • Cleaning
  • Biocompatibility ●
Table 5: Summary of Non-Clinical Performance Testing
Non-ClinicalPerformanceTestingPurposeResults
Sterilization Efficacy(ISO 17665-1: 2006Sterilization of healthcare products-Moist HeatPart 1 Requirements forthe development,validation and routinecontrol of sterilizationprocess for medicaldevices.)Demonstrate sterilization efficacy ofthe TCAT THA Instrument TraySets containing the maximuminoculated load.The testresulted in nogrowth andsix-logreduction atthe half-cyclefor bothdevices

Table 5: Summary of Non-Clinical Performance Testing

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Think Surgical. The logo is composed of the word "THINK" in gray, block letters, with a small blue square above the "I". Below the word "THINK" is the word "SURGICAL" in a smaller, blue font. The logo is simple and modern, and the colors are clean and professional.

of the indicator organism.
Packaging(ASTM D4169-16Standard Practice forPerformance Testing inShipping Containers andSystems)Demonstrate that the THINKSurgical TCAT® THA InstrumentTray Set was evaluated withmaximum load, wrapped accordingto use instructions and in theshipping configurations accordingto a shipping validation test.The resultsfor bothtraysindicatepackagingendpointswere met.
Cleaning(FDA. 2015.Guidance forIndustry and FDA Staff-Processing/ReprocessingMedical Devices in HealthCare Settings: Validationand Labeling)Demonstrate that that the cleaninginstructions for the TCAT THAInstrument Tray Set are valid.The resultsfor bothtrayindicatedcleaningendpointswere met.
Biocompatibility(ISO 10993-5:2009Cytotoxicity Test)Demonstrate that the TCATTHA Instrument Tray Set isbiocompatible.Under theconditionsof the test,the deviceextract wasnotcytotoxic.

Conclusions

The conclusions drawn from the nonclinical that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device Intuitive Surgical® Instrument and Accessory Sterilization Trays (K072211).

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).