K Number
K180080

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-04-06

(86 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FIBERGRAFT® BG Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine.

Device Description

FIBERGRAFT® BG Matrix product is composed of 45S5 bioactive glass components (M-45 granules, MS-45 microspheres) and bovine type I collagen. The BG Matrix after hydration with saline or blood can be applied to the defect site or can be molded into the desired shape and gently packed into the defect site as a non-setting putty. In the posterolateral spine fusion applications, the product is intended to be hydrated with bone marrow aspirate (BMA) and mixed with autograft in a 1:1 ratio.

The FIBERGRAFT® BG Matrix product was previously cleared in K171284. There has not been any change to the device since the last clearance.

AI/ML Overview

The provided text describes the FIBERGRAFT® BG Matrix Bone Graft Substitute but does not contain a discrete table of acceptance criteria and reported device performance. It also lacks specific details on sample sizes, ground truth establishment, expert involvement, or comparative effectiveness studies in the format requested.

However, based on the information provided, here's an attempt to structure the answer, highlighting what is available and noting the missing information:

Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria for FIBERGRAFT® BG Matrix Bone Graft Substitute

The performance of the FIBERGRAFT® BG Matrix was established through physical and chemical property evaluation studies, functional performance animal studies, and biocompatibility tests. The device's substantial equivalence to predicate devices for its intended use as a synthetic bone void filler was the primary goal, demonstrating that it functions as intended without raising new safety or effectiveness concerns.

1. Table of Acceptance Criteria and the Reported Device Performance

Acceptance Criteria CategoryReported Device Performance
In Vitro Functionality & BioactivityConfirmed by physical and chemical property studies. (Note: "The in vitro bioactivity test results have not been correlated to clinical performance.")
BiocompatibilityDemonstrated by ISO 10993 testing. Long history of clinical use of the bioactive glass material for the same intended use. Composed of the same bioactive glass material and the same type and duration of patient contact as the predicates.
Packaging & Shelf LifeEvaluations and real-time aging testing performed with passing results.
Bacterial Endotoxin LimitsTesting performed using the limulus amebocyte lysate (LAL) method, showing the device meets established endotoxin limits.
In Vivo Performance (Posterolateral Spine)Demonstrated performance at 26-week follow-up in a rabbit posterolateral spine fusion study. Evaluated using radiographic, histological, histomorphometric, and biomechanical data. Compared favorably to a predicate device and controls.
In Vivo Performance (Extremities & Pelvis)Demonstrated substantial equivalence to the predicate device and positive control at 26-week follow-up in a critical size distal femur defects study in rabbits. Evaluated using radiographic, histomorphometric, and biomechanical data. "Minor technological differences between the device groups do not raise new types of safety or effectiveness concerns."
Overall Safety and Effectiveness ConcernsPerformance testing demonstrated that the device functions as intended and meets the requirements of Class II bone void fillers as compared to the predicate device. Minor technological differences do not raise any new issues of safety or effectiveness, supporting substantial equivalence.

2. Sample size used for the test set and the data provenance

  • Rabbit Posterolateral Spine Fusion Study:
    • Sample Size: 53 skeletally mature rabbits.
    • Data Provenance: Animal study (specifically, rabbit model). The text implies prospective data collection within the study design. Country of origin not specified.
  • Rabbit Critical Size Distal Femur Defects Study:
    • Sample Size: 42 skeletally mature rabbits.
    • Data Provenance: Animal study (specifically, rabbit model). The text implies prospective data collection within the study design. Country of origin not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The provided text describes animal studies with data evaluated using radiographic, histological, histomorphometric, and biomechanical data. It does not mention human experts establishing ground truth for the test set in the context of clinical images or diagnostic performance. Instead, ground truth is inferred from the scientific measurements and observations made in the animal models.

4. Adjudication method for the test set

Not applicable, as the clinical adjudication of diagnostic assessments by human readers is not the focus of these animal performance studies. The evaluation methods (radiographic, histological, histomorphometric, biomechanical data) are objective measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The studies described are animal studies evaluating the performance of a bone graft substitute, not a diagnostic AI device requiring human reader assessment.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a bone graft substitute, not an algorithm.

7. The type of ground truth used

The ground truth in these animal studies was established through:

  • Histology: Microscopic examination of tissue.
  • Histomorphometry: Quantitative analysis of tissue structure.
  • Radiography: X-ray imaging for bone formation/healing.
  • Biomechanical data: Measurements of mechanical properties of the bone.
    These are direct scientific measurements of material integration and functional outcomes in a living organism.

8. The sample size for the training set

Not applicable. As this is a medical device (bone graft substitute) and not an AI/algorithm, there is no concept of a "training set" in the context of machine learning. The studies described are performance evaluations of the physical product.

9. How the ground truth for the training set was established

Not applicable, as there is no "training set" for this type of medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized depiction of a human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

April 6, 2018

Prosidyan, Inc. % Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103

Re: K180080

Trade/Device Name: FIBERGRAFT® BG Matrix Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: January 10, 2018 Received: January 10, 2018

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K180080 Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known)

K180080

Device Name

FIBERGRAFT® BG Matrix Bone Graft Substitute

Indications for Use (Describe)

FIBERGRAFT® BG Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)

□ Over-The-Counter Use (21 CFR 801

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

510(k) SUMMARY

Prosidyan, Inc.'s FIBERGRAFT® BG Matrix - Bone Graft Substitute

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Prosidvan. Inc. 30 Technology Drive Warren, NJ 07059 Phone: (610)-945-5640 Facsimile: (908) 396-1151 Contact Person: Charanpreet S. Bagga

Date Prepared: March 7, 2018

Name of Device and Name

FIBERGRAFT® BG Matrix Bone Graft Substitute

Common or Usual Name

Bone Void Filler

Classification Name/CFR Requlation/Product Code

Resorbable Calcium Salt Bone Void Filler, 21 CFR 888.3045, product code MQV

Predicate Devices

  • . Novabone Products LLC, Novabone MacroFORM (K140946) (Primary predicate device)
  • . Prosidyan Inc, FIBERGRAFT BG Matrix (K171284) (Reference device)
  • . Prosidyan Inc, FIBERGRAFT BG Putty (K143533, K170306) (Reference device)

Intended Use / Indications for Use

FIBERGRAFT® BG Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine.

Device Description

FIBERGRAFT® BG Matrix product is composed of 45S5 bioactive glass components (M-45 granules, MS-45 microspheres) and bovine type I collagen. The BG Matrix after hydration with saline or blood can be applied to the defect site or can be molded into the desired shape and gently packed into the defect site as a non-setting putty. In the posterolateral spine fusion

{4}------------------------------------------------

applications, the product is intended to be hydrated with bone marrow aspirate (BMA) and mixed with autograft in a 1:1 ratio.

The FIBERGRAFT® BG Matrix product was previously cleared in K171284. There has not been any change to the device since the last clearance.

Technological Characteristics

FIBERGRAFT® BG Matrix is a bioactive osteoconductive, resorbable, biocompatible bone graft substitute. The product is composed of 45S5 bioactive glass components (M-45 granules, MS-45 microspheres) and bovine type I collagen. The BG Matrix hydrated with saline or blood can be applied to the defect site or can be used as a moldable and malleable material that can be placed at the defect site. In addition, for posterolateral spine fusion applications, the product can be mixed with autograft.

The addition of collagen creates a single piece bone graft in a shape and size that is clinically relevant for ease of use during implantation.

Performance Data

The performance of the FIBERGRAFT® BG Matrix has been established by undertaking physical and chemical property evaluation studies, functional performance animal studies and biocompatibility tests. The physical and chemical property studies confirmed the in vitro functionality and bioactivity of the BG Matrix. The in vitro bioactivity test results have not been correlated to clinical performance. The biocompatibility of the FIBERGRAFT® BG Matrix is demonstrated by ISO 10993 testing and the long history of clinical use of the bioactive dlass material for the same intended use. In addition, the BG Matrix is composed of the same bioactive glass material and the same type and duration of patient contact as the predicates. Packaging evaluations, shelf life testing and real time aging testing were performed with passing results. Bacterial endotoxin testing was performed using the limulus amebocyte lysate (LAL) method and showed that the device meets the endotoxin limits of established guidelines .

The FIBERGRAFT® BG Matrix product was evaluated in a rabbit posterolateral spine fusion study to further support device performance for its indications for use in posterolateral spine. The FIBERGRAFT® BG Matrix product was compared to its predicate device as well as controls. The animal study evaluated device performance in critical sized cancellous bone in the posterolateral spine of 53 skeletally mature rabbits. The performance was evaluated using radiographic, histological, histomorphometric, and biomechanical data. Testing of the FIBERGRAFT® BG Matrix in the rabbit model is representative of the indications for use and range of anatomical sites proposed for the subject device. The results of the study through 26 week follow up demonstrated the performance of the FIBERGRAFT® BG Matrix device.

The FIBERGRAFT® BG Matrix product was also evaluated in a critical size distal femur defects in skeletally mature rabbits to further support device performance for its indications for use in extremities and pelvis applications. The FIBERGRAFT® BG Matrix product was compared to its predicate device as well as controls. The animal study evaluated device performance in critical sized defects in the distal femur of 42 skeletally mature rabbits. The performance was evaluated using radiographic, histomorphometric, and biomechanical data, Testing of the FIBERGRAFT® BG Matrix in the rabbit model is representative of the indications for use and range of anatomical sites proposed for the subject device. The results of the study through 26 week follow up demonstrated that the FIBERGRAFT® BG Matrix device performs substantially equivalently to the predicate device and positive control, and any minor technological differences between the device groups do not raise new types of safety or effectiveness concerns.

{5}------------------------------------------------

Therefore, performance testing demonstrated that the FIBERGRAFT® BG Matrix device functions as intended and meets the requirements of class II bone void fillers as compared to the predicate device.

Substantial Equivalence

As demonstrated in performance testing, the FIBERGRAFT® BG Matrix has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between BG Matrix and its predicate devices do not raise any new issues of safety or effectiveness. The data demonstrate that BG Matrix is substantially equivalent to the predicate devices.

Conclusion

FIBERGRAFT® BG Matrix is an osteoconductive, resorbable, biocompatible bone graft substitute composed of bioactive glass, mixed with Type I collagen. The FIBERGRAFT® BG Matrix is substantially equivalent to its predicate device for its intended use as a synthetic bone void filler. Performance testing, including in vivo data, demonstrated that the device functions as intended without raising new safety or effectiveness questions.

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.