AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NexSite™ HD, Hemodialysis Symmetric Tip Catheter for long term use is indicated for use in attaining long term vascular access for chronic hemodialysis and apheresis. It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein which should only be used when no other upper extremity or chest wall options are available. Catheters greater than 40cm are indicated for femoral vein insertion.

Device Description

The NexSite™ HD, Hemodialysis Symmetric Tip Catheter for long term use (19 cm, 23 cm, 37 cm, 31 cm, 35 cm and 55 cm (tip to cuff lengths) is a dual lumen, symmetric tip radiopaque polyurethane catheter which contains a Dacron biomaterial cuff and two female luer connectors. The DISC (Dermal Ingrowth Support Collar) assists with the direction of the catheter and consists of a biomaterial tissue ingrowth scaffold. The biomaterial scaffolds on the catheter shaft and DISC are aligned and facilitate tissue ingrowth, which is considered important in reducing the source of extraluminal infection in patients requiring long term catheterisation. The Catheter, DISC and the following ancillary components required for the procedure (0.038" guidewire, Stainless Steel Tunneler and Sleeve, 16Fr Introducer / Dilator and Luer Caps) are provided in a tray, sealed with a Tyvek lid and placed in a second sterile pouch so that the tray can be delivered to the sterile field.

AI/ML Overview

The provided text is a 510(k) summary for the Marvao NexSite™ HD, Hemodialysis Symmetric Tip Catheter for long term use. This document describes the device, its intended use, and argues for its substantial equivalence to predicate devices, which is a regulatory pathway for medical devices in the US.

However, the information requested in your prompt (acceptance criteria, details of a study proving device meets criteria, sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set information) is typically found in detailed clinical trial reports or performance study reports, which are not present in this 510(k) summary.

A 510(k) summary focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a de novo study with acceptance criteria for a novel device. The document states that clinical studies were not deemed necessary because in vivo and in vitro testing were sufficient to demonstrate safety and effectiveness by comparison to a legally marketed predicate device.

Therefore,Based on the provided text, I cannot extract the specific information requested about acceptance criteria and a study proving the device meets those criteria.

Here's why and what can be inferred:

  • No specific acceptance criteria or study proving they are met are explicitly detailed. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices, not on proving a novel device meets specific acceptance criteria through a full-scale clinical trial.
  • The document states: "In vitro testing was performed on the NexSite HD, Hemodialysis Symmetric Tip Catheter to assure reliable design and performance in accordance with ISO 10555-1:2013." and "The non-clinical tests performed include visual and dimensional of the catheter, DISC and stylet, stylet joint strength, priming volume, catheter leakage, catheter pressure, catheter flow rate testing, catheter recirculation testing and mechanical hemolysis. The test results demonstrate that the NexSite™ HD, Hemodialysis Symmetric Tip Catheter meets the requirements in the applicable standards and specifications, and is substantially equivalent to legally marketed predicate devices."
  • It also mentions: "In vivo implantation studies were also performed to demonstrate that the device would perform as intended."
  • Crucially, it states: "Clinical studies were not deemed necessary since in vivo and in vitro testing were sufficient to demonstrate safety and effectiveness by way of comparison to a legally marketed predicate device."

This means that while performance testing and in vivo implantation studies were conducted, they were done primarily to establish substantial equivalence to existing devices, usually by showing the new device performs at least as well as the predicate device on the relevant metrics, rather than proving it meets pre-defined acceptance criteria for a novel functionality.

Therefore, I cannot populate the table or answer most of your questions as the document does not contain this level of detail about a dedicated study proving specific acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 8, 2018

Marvao Medical Devices, Ltd. % Mary P. LeGraw Principal Regulatory Consultant Boston Biomedical Associates 100 Crowley Drive, Suite 216 Marlborough, MA 01752

Re: K173805

Trade/Device Name: NexSite™ HD, Hemodialysis Symmetric Tip Catheter for long term use Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood Access Device and Accessories Regulatory Class: II Product Code: MSD Dated: December 18, 2017 Received: December 19, 2017

Dear Mary P. LeGraw:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the tray have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Charles Viviano -S

For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173805

Device Name

NexSite™ HD, Hemodialysis Symmetric Tip Catheter for long term use

Indications for Use (Describe)

The NexSite™ HD, Hemodialysis Symmetric Tip Catheter for long term use is indicated for use in attaining long term vascular access for chronic hemodialysis and apheresis. It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient. Alternate include the subclavian vein which should only be used when no other upper extremity or chest wall options are available. Catheters greater than 40cm are indicated for femoral vein insertion.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1 510(K) SUMMARY

This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92. Summary preparation date is December 11, 2017 [21 CFR 807.92(a)(1)].

Submitter 1.1

Name:Marvao Medical Devices, Ltd.
Address:GMIT Innovation in Business Centre, Dublin Road
Galway, Ireland
Phone:+353 (0)91 759 301
Fax:+353 (0)91 762 055
Contact Person:Mary P. LeGraw,
Boston Biomedical Associates
100 Crowley Drive Suite 216,
Marlborough, MA 01752
mlegraw@boston-biomedical.com
Phone: (508) 691-7038
Fax: (508) 351-8637
1.2 Device Name
Device/Trade Name:NexSite™ HD, Hemodialysis Symmetric Tip
Catheter for long term use
Date of Preparation:December 14, 2017.
Common Name:Catheter, hemodialysis, implanted
Classification Name:Blood access devices and accessories
ClassificationNumber:876.5540
Product Code/Classification Panel:MSD/Gastroenterology /Urology
1.3 Predicate Device(s)
PrimaryK161026Marvao NexSite HD HemodialysisCatheter for long term use, Manufacturedby Marvao Medical Devices, Ltd.
SecondaryK171571Marvao NexSite HD HemodialysisCatheter for long term use (60cm)
Manufactured by Marvao MedicalDevices, Ltd.
ReferenceK123196Palindrome™ Precision Symmetric TipDual Lumen Catheter, Manufactured byCovidien

{4}------------------------------------------------

Traditional 510(l) Premarket Notification Submission: Marvao Medical Devices Ltd. NexSite™ HD Hemodialysis Symmetric Tip Catheter for long term use

1.4 Device Description

The NexSite™ HD, Hemodialysis Symmetric Tip Catheter for long term use (19 cm, 23 cm, 37 cm, 31 cm, 35 cm and 55 cm (tip to cuff lengths) is a dual lumen, symmetric tip radiopaque polyurethane catheter which contains a Dacron biomaterial cuff and two female luer connectors. The DISC (Dermal Ingrowth Support Collar) assists with the direction of the catheter and consists of a biomaterial tissue ingrowth scaffold. The biomaterial scaffolds on the catheter shaft and DISC are aligned and facilitate tissue ingrowth, which is considered important in reducing the source of extraluminal infection in patients requiring long term catheterisation. The Catheter, DISC and the following ancillary components required for the procedure (0.038" guidewire, Stainless Steel Tunneler and Sleeve, 16Fr Introducer / Dilator and Luer Caps) are provided in a tray, sealed with a Tyvek lid and placed in a second sterile pouch so that the tray can be delivered to the sterile field.

1.5 Indications for Use

The NexSite™ HD, Hemodialysis Symmetric Tip Catheter for long term use is indicated for use in attaining long term vascular access for chronic hemodialysis and apheresis. It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein which should only be used when no other upper extremity or chest wall options are available. Catheters greater than 40cm are indicated for femoral vein insertion.

1.6 Comparison Of Technological Characteristics With The Predicate Devices

The descriptive characteristics of the proposed device and the identified predicates were compared and determined to be substantially equivalent in terms of intended use, dimensions, materials, and basic design. Comparative performance testing using finished NexSite™ HD Symmetric Tip devices and identified predicate devices was completed. The test results support the substantial equivalence of the NexSite™ HD Symmetric Tip device to the predicate devices.

1.7 Performance Data

In vitro testing was performed on the NexSite HD, Hemodialysis Symmetric Tip Catheter to assure reliable design and performance in accordance with ISO 10555-1:2013. The non

{5}------------------------------------------------

clinical tests performed include visual and dimensional of the catheter, DISC and stylet, stylet joint strength, priming volume, catheter leakage, catheter pressure, catheter flow rate testing, catheter recirculation testing and mechanical hemolysis. The test results demonstrate that the NexSite™ HD, Hemodialysis Symmetric Tip Catheter meets the requirements in the applicable standards and specifications, and is substantially equivalent to legally marketed predicate devices.

In vivo implantation studies were also performed to demonstrate that the device would perform as intended. Clinical studies were not deemed necessary since in vivo and in vitro testing were sufficient to demonstrate safety and effectiveness by way of comparison to a legally marketed predicate device.

Guidance 1.8

The FDA Guidance on Premarket Notification [(510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, dated 3/16/95, was utilized in order to meet the FDA requirements for content and organization of this submission, as well as the Guidance Industry and Food and Drug Administration Staff Implanted Blood Access Devices for Hemodialysis, January 21, 2016.

1.9 Conclusions

Marvao Medical believes the proposed NexSite HD Hemodialysis Symmetric Tip Catheter for long term use is substantially equivalent to legally marketed predicate devices. The indications for use, methods of operation, design and materials used are either identical or substantially equivalent to existing legally marketed predicate products. In addition, performance testing supports substantial equivalence of the proposed and predicate devices.

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.