K Number
K173729

Validate with FDA (Live)

Device Name
REVIAN RED
Manufacturer
Date Cleared
2018-02-28

(84 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

REVIAN RED is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

Device Description

REVIAN RED is a non-invasive, hands-free device indicated to treat androgenetic alopecia in both men and women. It is operated via a mobile application to promote hair growth using Modulated Light Therapy (MLT) for a recommended 10-minute daily treatment. REVIAN RED is a system comprised of a wearable soft textile Cap using driver electronics, a rechargeable battery, and integrated light emitting diode (LED) flexible printed circuit board. REVIAN RED is designed as an over-the-counter (OTC) home-based device to treat the entire scalp area of the user and is controlled via a wireless Bluetooth connection to a mobile application operating on a wireless device.

AI/ML Overview

The provided document is a 510(k) summary for the Revian Red device. It states that no clinical trial data for the REVIAN RED System was submitted for this 510(k). Therefore, the device's acceptance criteria and studies demonstrating its performance in a clinical setting against those criteria are not detailed in this document.

The document focuses on demonstrating substantial equivalence to predicate devices through non-clinical performance testing.

Here's a breakdown of the information that is available in the document, framed by your request for acceptance criteria and studies:

1. A table of acceptance criteria and the reported device performance

Since no clinical trial data was submitted, there are no specific performance metrics like sensitivity, specificity, or accuracy that would constitute clinical acceptance criteria in this document. Instead, the "performance" here refers to meeting safety and engineering standards:

Acceptance Criteria CategoryReported Device Performance (as stated in the 510(k) Summary)
Electrical SafetyConforms to IEC 60601-1
Electromagnetic Compatibility (EMC)Conforms to IEC 60601-1-2
Specific Safety for Medical Electrical EquipmentConforms to ANSI/AAMI HA60601-1-11:2015
Safety of Lamps & Lamp SystemsConforms to IEC 60601-2-57 (Edition 1.0)
Photobiological SafetyConforms to IEC 62471 (Edition 1.0)
BiocompatibilityConforms to ISO 10993-1, no effects of applied parts noted for cytotoxicity, skin irritation, or skin sensitization.
Performance Verification & Validation"Demonstrates REVIAN RED meets user needs and design inputs and met requirements for its intended use." (General statement, no specific metrics provided.)
Human Factors Engineering & Usability"Evaluated representative users of the device under expected use conditions during simulated and actual use of the device" following ANSI/AAMI/IEC 62366-1 and 60601-1-6. (No specific performance metrics provided.)

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • No clinical trial test set was used or described for clinical performance. The testing mentioned in the 510(k) is non-clinical (e.g., electrical safety, EMC, biocompatibility, human factors).
  • For human factors evaluation, the study protocol "was designed to evaluate representative users of the device." The specific sample size is not stated, nor is the country of origin or whether it was retrospective/prospective, though formative and summative user testing implies a prospective approach.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • This is not applicable as no clinical test set requiring expert ground truth was described in the submitted 510(k).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This is not applicable as no clinical test set requiring adjudication was described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This is not applicable. The Revian Red is a light therapy device, not an AI-powered diagnostic tool for human readers. No MRMC study was mentioned or required for this type of device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This is not applicable. The Revian Red is a device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the non-clinical tests mentioned, the "ground truth" would be the standards themselves (e.g., IEC 60601-1, ISO 10993-1) and the demonstrated adherence to their requirements.
  • For the biocompatibility testing, the "ground truth" was likely established through standardized laboratory procedures and observation of biological responses.
  • For human factors, the "ground truth" relates to usability and safety observed during user interactions, evaluated against usability principles and safety critical tasks.

8. The sample size for the training set

  • This is not applicable as no machine learning algorithm development (which would require a training set) is described for this device.

9. How the ground truth for the training set was established

  • This is not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 28, 2018

PhotonMD. Inc Mr. William Knape Senior Director, Regulatory Quality and Clinical Affairs 627 Davis Drive, Suite 400 Morrisville, North Carolina 27560

Re: K173729

Trade/Device Name: Revian Red Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP Dated: December 4, 2017 Received: December 6, 2017

Dear Mr. Knape:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

K173729

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173729

Device Name REVIAN RED

Indications for Use (Describe)

REVIAN RED is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PhotonMD

510(K) Summary 1

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 890.5500.

1.1 Submitter Information:

Sponsor:PhotonMD, Inc.627 Davis Drive, Suite 400Morrisville, NC 27560
Contact Person:William A. KnapeSenior Director of Regulatory, Quality and Clinical AffairsPhone: (919) 757-2033Email: bknape@photonmd.com
Submission Date:December 4, 2017

1.2 Subject Device Information

Trade/Proprietary Name:REVIAN RED (Model 101)
Common/Usual Name:Lamp for Promotion of Hair Growth
Regulation Name:Infrared Lamp per 21 CFR 890.5500
Review Panel:General & Plastic Surgery
Product Code:OAP (laser, comb, hair)
Regulation Number:21 CFR 890.5500
Regulatory Class:Class 2

1.3 Predicate Device Information

Predicate Devices:Device510(k)NumberManufacturer
iHelmet Hair Growth SystemK162782Slinph Technologies
HairMDK152019Trophy Skin
HairMax LaserComb 82K142573Lexington International
LH80 PROK122950Theradome

1.4 Device Description

REVIAN RED is a non-invasive, hands-free device indicated to treat androgenetic alopecia in both men and women. It is operated via a mobile application to promote hair growth using Modulated Light Therapy (MLT) for a recommended 10-minute daily treatment. REVIAN RED is a system comprised of a wearable soft textile Cap using driver electronics, a rechargeable battery, and integrated light emitting diode (LED) flexible printed circuit board. REVIAN RED is designed as an over-the-counter (OTC) home-based device to treat

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for PhotonMD. The word "Photon" is in gray, with a colorful atom symbol in the middle of the word. The letters "MD" are in green.

February 26, 2018 Traditional 510(k) Premarket Notification

the entire scalp area of the user and is controlled via a wireless Bluetooth connection to a mobile application operating on a wireless device.

1.5 Indications for Use

The REVIAN RED device is indicated to treat Androgenetic Alopecia and to Promote Hair Growth in males who have Norwood-Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and Promote Hair Growth in females who have Ludwig-Savin Scale I-1 to I-4, II -1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

1.6 Performance Test Summary

No clinical trial data for the REVIAN RED System was submitted for this 510(k). Nonclinical performance testing to demonstrate substantial equivalence to the predicate device included evaluation to IEC 60601-1 and 60601-1-2 to confirm the device's electrical safety and electromagnetic compatibility and conformance with ANSI/AAMI HA60601-1-11:2015, IEC 60601-2-57 (Edition 1.0), and IEC 62471 (Edition 1.0). Biocompatibility testing of the Cap per ISO 10993-1 confirmed no effects of the applied parts were noted for cytotoxicity, skin irritation or skin sensitization. Performance verification and validation testing demonstrates REVIAN RED meets user needs and design inputs and met requirements for its intended use.

Studies Related to Human Factors Evaluation

Human factors evaluation of REVIAN RED was tested following ANSVAAMI/IEC 62366-1 and 60601-1-6 standards. The study protocol was designed to evaluate representative users of the device under expected use conditions during simulated and actual use of the device. The Human Factors Engineering and Usability Engineering process involved several methodologies, including observational user research, formative and summative user testing.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for PhotonMD. The word "Photon" is in gray, with a colorful graphic in the middle of the word. The letters "MD" are in green. The logo is simple and modern.

February 26, 2018 Traditional 510(k) Premarket Notification

1.7 Comparison to Predicate Devices

REVIAN RED has similar technological characteristics as the iHelmet Hair Growth System (K162782) including wavelength, light energy delivery method, mobile app interface with interactive audible timers and intended use. REVIAN RED is also similar technologically to the other helmet designed devices; with HairMD (K152019) and LH80 PRO (K122959) having fewer diodes (51 laser diodes/LEDS and 80 laser diodes, respectively) and longer treatment regimens of 20 minutes.

510(k) #K173729K162782K152019K142573K122950
DeviceREVIAN REDiHelmet HairGrowth System(LTD200S)HairMDHairMaxLaserComb 82(HMLC 82)LH80 PRO
CompanyPhotonMD, Inc.SlinphTechnologiesTrophy SkinLexingtonInternationalTheradome
Product CodeOAPOAPOAPOAPOAP
Rx/OTCOTCOTCOTCOTCOTC
Intended Use/Indications forUseTreatAndrogeneticAlopecia andpromote hairgrowth in maleswho haveNorwood-HamiltonClassification IIato V patterns ofhair loss and totreat AndrogeneticAlopecia andpromote hairgrowth in femaleswho haveLudwig-Savinscale I-1 to I-4,II-1, II-2 or frontalpatterns of hairloss; both withFitzpatrick SkinTypes I to IVPromote hairgrowth in femaleswithAndrogeneticAlopecia whohave Ludwig-SavinClassification ofI-II, in males withAndrogeneticAlopecia whohave NorwoodHamiltonClassifications ofIIa-V and forboth, FitzpatrickClassification ofSkin Phototypes I- IV.Promote hairgrowth in femaleswithAndrogeneticAlopecia whohave Ludwig-SavinClassification ofI-II, in males withAndrogeneticAlopecia whohave NorwoodHamiltonClassifications ofIIa-V and forboth, FitzpatrickClassification ofSkin Phototypes I- IV.TreatAndrogeneticAlopecia andpromote hairgrowth in maleswho haveNorwoodHamiltonClassificationIIa to V patternsof hair loss andtreatAndrogeneticAlopecia and,promote hairgrowth in femaleswho have Ludwig(Savin) scale I-4,II-1, II-2 orfrontal, both withFitzpatrick skintypes I to IV.TreatAndrogeneticAlopecia, topromote hairgrowth infemales withfemale patternhair loss (FPHL)on the Ludwigand Savin HairLoss Scale I-II,Fitzpatrick Skin-Types I to IV.
WearableMountingTextile CapHelmetHelmetHead BandHelmet
Wave-Length(s)620 - 660 nm650 nm ±10nm650 nm655 nm ±10nm678 nm
Visible Lightsource119 RedLEDs200 Red LaserDiodes21 Red LaserDiodes and30 Red LEDs82 Red LaserDiodes80 Red LaserDiodes
BatteryLithiumPolymerLithiumPolymerPowered by 120VWall PlugLithiumPolymerLithiumPolymer

Table 1 – Summary of Predicate Devices Compared to REVIAN RED

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for PhotonMD. The word "Photon" is in gray, with a graphic of colorful people surrounding a white sphere in the middle of the word. The letters "MD" are in green. The logo is simple and modern.

February 26, 2018 Traditional 510(k) Premarket Notification

Table 1 - Summary of Predicate Devices Compared to REVIAN RED (cont'd)

510(k) #K173729K162782K152019K142573K122950
Treatment TimeEvery day10 minsEvery other day20-35 minsEvery other day20 mins3 times / week90 seconds2 times / week20 mins
Digital MobileControllerYesYesNoNoNo
Restrictions onDaily IrradianceYes, daily limitfor treatmentsNoNoNoNo
Other DesignCharacteristicsCap pausestherapy if subject'shead movesoutside zone ofradiation; aftercompletion oftreatment, unitpowers down,limits dailytreatment to 10mins, andprovides treatmentreminders andmessages viamobile appcontrollerHelmet pausestherapy ifsubject's headmoves outsidezone of radiation;after completionof time, unitpowers down andaudible beep.Helmet pausestherapy ifsubject's headmoves outsidezone of radiation;after completionof time, unitpowers down andaudible beep.Vibrate so usercan move to thenext section, 3sections @ 30seconds each,proximity sensorto shut off lasersif not in contact,7.5 cm scalpsection, automaticturn offHelmet pausestherapy ifsubject's headmoves outsidezone ofradiation; aftercompletion oftime, unitpowers downand audiblebeep.

1.8 Conclusion

This notification contains all information required by 21 CFR 807.87. Performance testing including electrical, electromagnetic compatibility, biocompatibility, software, and human factors demonstrates REVIAN RED meets the requirements for its intended use and does not raise any new types of safety or effectiveness questions when compared to the predicate device.

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.