K Number
K173662

Validate with FDA (Live)

Date Cleared
2017-12-20

(21 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents.

Device Description

The speX Support Catheter is a device intended to provide additional support to a steerable guidewire when accessing discrete regions of the peripheral and/or coronary vasculature. The device consists of a support catheter body with a luer end. The through-lumen of the device can serve as a conduit for the delivery of diagnostic and therapeutic agents.

AI/ML Overview

The provided document is a 510(k) summary for the speX Support Catheter, a medical device. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving that an AI-driven device meets specific acceptance criteria.

As such, the information requested in the prompt, which is typically relevant for AI/ML medical devices, is largely absent from this document. The document describes a physical medical device (a support catheter), and the "performance data" refers to physical and mechanical tests of the catheter, not the performance of an algorithm.

Therefore, I cannot provide most of the requested information based on the provided text.

Here is what I can extract:

1. A table of acceptance criteria and the reported device performance:

Based on the document, specific numerical acceptance criteria and reported device performance are not explicitly detailed in a table format. The document states:

  • "The modified speX Support Catheter met all specified criteria and did not raise new safety or performance questions."
  • "Based on the performance testing the modified speX Support Catheter was found to be substantially equivalent to the predicate device."

The types of tests performed were:

  • Simulated Use Testing
  • Component/Assembly Integrity Testing
  • Corrosion Resistance Testing
  • Lubricity Testing

2. Sample sized used for the test set and the data provenance:

  • This information is not available in the provided text, as the study involves physical device testing, not AI data analysis.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not applicable/not available, as the study pertains to physical device performance and not diagnostic or interpretive tasks requiring expert ground truth in the context of AI.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • This information is not applicable/not available for this type of physical device testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This information is not applicable. This document describes a physical medical device, not an AI system that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • This information is not applicable. This document describes a physical medical device, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

  • This information is not applicable. For physical device testing, "ground truth" would typically refer to established engineering specifications, mechanical properties, and functional requirements, as opposed to medical diagnoses.

8. The sample size for the training set:

  • This information is not applicable/not available as this is a physical medical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established:

  • This information is not applicable/not available as this is a physical medical device, not an AI model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is a stylized graphic of an eagle, while the FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are placed side-by-side, with the HHS logo on the left and the FDA logo on the right. The logos are both in color and are of high resolution.

December 20, 2017

ReFlow Medical, Inc. Ms. Rebecca Pine Vice President, Regulatory, Quality and Clinical Affairs 1003 Calle Sombra San Clemente, California 92673

Re: K173662

Trade/Device Name: speX Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: November 28, 2017 Received: November 29, 2017

Dear Ms. Pine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173662

Device Name speX Support Catheter

Indications for Use (Describe)

The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K173662 510(k) Summarv

Page 1 of 2

510(k) Summary

I. SUBMITTER

ReFlow Medical, Inc. 1003 Calle Sombra San Clemente, CA 92673

Contact person: Rebecca K Pine Phone: (760) 809-5178 Fax: (760) 290.3216 Date prepared: March 31, 2017

II. DEVICE

Name of the device: speX Support Catheter Common of usual name: Support Catheter Classification name: Percutaneous Catheter Regulatory Class: 2 Product Code: DQY

III. PREDICATE DEVICE

speX Support Catheter (K141649) This predicate has not been subject to a design-related recall The Wingman 14 Crossing Catheter (K132420) and Wingman 18 Crossing Catheter (K151880) were used as reference predicates in this submission

IV. DEVICE DESCRIPTION

The speX Support Catheter is a device intended to provide additional support to a steerable guidewire when accessing discrete regions of the peripheral and/or coronary vasculature.

The device consists of a support catheter body with a luer end. The through-lumen of the device can serve as a conduit for the delivery of diagnostic and therapeutic agents.

V. INDICATIONS FOR USE

The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents.

{4}------------------------------------------------

Page 2 of 2

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The technological characteristics of the modified speX Support Catheter are highly analogous to the technological characteristics of the speX Support Catheter previously cleared (K141649) version of the device.

At a high level, the subject and predicate devices are based on the following same technological elements:

  • all delivered to the target site using an over-the-wire percutaneous technique
  • I all have a through lumen to allow passage and exchange of guidewires
  • all have a smooth inner lumen to provide reduced friction for guidewire movement
  • all have a polymer catheter shaft with specific geometry to control I the torque and push movements associated with lesion crossing

The following technological differences exist between the subject and predicate devices:

  • The speX Support Catheter incorporates a gold plating within the shapeable region at the distal end of the catheter for increased radiovisibility
  • Additional product diameters have been incorporated into the product family to facilitate use with .014" and .018" guidewires.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence.

  • l Simulated Use Testing
  • l Component/Assembly Integrity Testing
  • l Corrosion Resistance Testing
  • Lubricity Testing

The modified speX Support Catheter met all specified criteria and did not raise new safety or performance questions. Based on the performance testing the modified speX Support Catheter was found to be substantially equivalent to the predicate device.

VIII. CONCLUSIONS

The design testing performed for the speX Support Catheter demonstrated that the performance of the device is substantially equivalent to the legally marketed predicate devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).