K Number
K173658

Validate with FDA (Live)

Date Cleared
2018-08-21

(265 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Confidence Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip. The Confidence Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The Confidence Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.

The Confidence Blood Glucose Monitoring System should not be used for the diagnosis of or diabetes or for neonatal use. The Confidence Blood Glucose Test Strips are for use with the Confidence Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip.

Device Description

The Confidence™ BGMS consists of test strips stored in a vial with desiccant lining, a hand held blood qlucose meter and a control solution. To perform blood glucose test a test strip is removed from the vial and inserted into Confidence™ meter. On inserting the strip meter gets activated and displays a code, user must select a code corresponding to their strip vial using meter's interface, after selecting correct code, meter will be ready to apply sample by showing a droplet on the LCD display. On applying blood on the end of the test strip a countdown starts and meter LCD displays the plasma equivalent value in the blood sample within 10 seconds in mg/dL unit. The user can validate the operation of the system by using provided control solution. The system uses an electrochemical method based on the GDH-FAD enzyme and a known algorithm, where generated current is directly proportional to glucose concentration.

AI/ML Overview

Here's an analysis of the provided text to fulfill your request:

The document describes the Confidence Blood Glucose Monitoring System and its performance studies for FDA clearance.


1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for glucose monitoring systems are typically defined by regulatory bodies like the FDA, often referring to standards such as ISO 15197. While the specific standard is not explicitly stated in the provided text as the "acceptance criteria" for the tables below, the data presented aligns with common metrics used in these standards: percentage of results falling within certain accuracy ranges compared to a reference method (YSI 2300).

System Accuracy Results (Trained Person)

Glucose Concentrations RangeWithin ± 10 mg/dL or ±10%Within ±10 mg/dL or ±15%Within ±10 mg/dL or ± 20 %
60-500 mg/dL74.8% (202/270)99.6% (269/270)100% (270/270)

User Performance Results (Lay Users)

Glucose Concentrations RangeWithin ±10 mg/dL or ± 10 %Within ±15 mg/dL or ± 15 %
60-480 mg/dL90.4% (169/187)100% (187/187)

2. Sample Size and Data Provenance for Test Set

  • Sample Size (Trained Person Study): 265 subjects
  • Sample Size (Lay User Study): 181 users (results reported for N=187)
  • Data Provenance: Not explicitly stated, but clinical studies for FDA submissions are typically conducted in the US and are prospective in nature. The term "subjects" and "users" implies prospective data collection for these specific studies.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not applicable in this context. The ground truth for glucose measurements is established by a reference laboratory instrument, not by human experts.

4. Adjudication Method for Test Set

  • Adjudication Method: Not applicable. The ground truth is established by a quantitative reference instrument, eliminating the need for expert adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This type of study is more common for diagnostic imaging or subjective interpretations rather than quantitative measurement devices like a blood glucose meter. The studies focused on comparing device performance against a reference standard and user performance.

6. Standalone Performance Study

  • Yes, a standalone (algorithm only) performance study was done. The "System Accuracy Results" table (Page 6) and the corresponding linear regression data (Page 7) directly report the performance of the Confidence™ BGMS itself when operated by trained personnel, comparing its readings to the YSI reference standard. The "User Performance Results" also show standalone performance, but in the hands of lay users.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth used was quantitative reference measurements obtained from an FDA cleared reference method (YSI 2300). The YSI 2300 is a laboratory-grade glucose analyzer often used as the gold standard for glucose measurement in clinical studies.

8. Sample Size for the Training Set

  • Sample Size for Training Set: The document does not specify the sample size used for the training set. The descriptions focus on the validation/test set studies for substantial equivalence. For medical devices, particularly those involving electrochemical sensors, "training" might refer to the development and calibration process rather than a machine learning model's training set in the typical sense. Since it's a device with an algorithm for glucose calculation (based on current measurement), the development process would involve extensive characterization and calibration using various glucose concentrations and interfering substances, but a distinct "training set" in the context of predictive algorithms is not detailed.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: The document does not specify how the ground truth for any potential "training set" was established. Assuming "training set" refers to the data used during the development and calibration of the device's measurement algorithm, the ground truth would typically be established using a highly accurate laboratory reference method (like the YSI 2300) with precisely prepared glucose solutions and spiked blood samples across the relevant concentration range, including potentially interfering substances. However, this is inferred based on standard practices for such devices and not explicitly stated in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

August 21, 2018

Applied Biomedical LLC % Jigar Shah mdi Consultants, Inc. 55 Northern Blvd, Suite 200 Great Neck, NY 11021

Re: K173658

Trade/Device Name: Confidence Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: July 19, 2018 Received: July 20, 2018

Dear Jigar Shah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind vou, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173658

Device Name Confidence Blood Glucose Monitoring System

Indications for Use (Describe)

The Confidence Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip. The Confidence Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The Confidence Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.

The Confidence Blood Glucose Monitoring System should not be used for the diagnosis of or diabetes or for neonatal use. The Confidence Blood Glucose Test Strips are for use with the Confidence Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

646-3233 • Fax. (714) 646-3238

510(K) SUMMARY

The assigned 510(K) Number: K 173658

1. Submitter's Identification:

APPLIED BIOMEDICAL LLC 1981 E Miraloma Ave, Unit B Placentia, CA 92870 TEL: 714 646 3233 FAX: 714 646 3238 Date Summary Revised: August 16, 2018 Official Correspondent's name: Dr. Zakir Murtaza Ph.D.

2. Name of the Device:

Confidence™ Blood Glucose Monitoring System

3. Common or Usual Name:

Glucose Test System, Over-the-Counter

    1. Predicate Device Information: K150214. OneTouch Verio Flex Blood Glucose Monitoring System

5. Classification:

Confidence™ Blood Glucose Meter and Confidence™ Test Strips are Class II devices (US 21 CFR § 862.1345), Product Code NBW.

6. Device Description:

The Confidence™ BGMS consists of test strips stored in a vial with desiccant lining, a hand held blood qlucose meter and a control solution. To perform blood glucose test a test strip is removed from the vial and inserted into Confidence™ meter. On inserting the strip meter gets activated and displays a code, user must select a code corresponding to their strip vial using meter's interface, after selecting correct code, meter will be ready to apply sample by showing a droplet on the LCD display. On applying blood on the end of the test strip a countdown starts and meter LCD displays the plasma equivalent value in the blood sample within 10 seconds in mg/dL unit. The user can validate the operation of the system by using provided known algorithm, where generated current is directly proportional to glucose concentration.

{4}------------------------------------------------

7.

lntended Use:
The Confidence™ Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip.

The Confidence™Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Confidence™ Blood Glucose Monitoring is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.

The Confidence™Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. The Confidence™ Blood Glucose Test for use with the Confidence™ Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples freshly drawn from the fingertip.

8. Comparison to Predicate Devices:

Predicate Device: a.

OneTouch Verio Flex Blood Glucose Monitoring System,

  • b. Predicate Device 510K number and date of cleared K150214 (July 31 2015)
  • Predicate Device Comparison Table C.
ItemCandidate DeviceConfidenceTMBlood GlucoseMonitoring SystemOneTouch Verio Flex Blood GlucoseMonitoring System
--------------------------------------------------------------------------------------------------------------------------------------

Similarities

510K Issued on510 (k) numberCurrent SubmissionJuly 31, 2015
K173658K150214
Intended/Indications forUseSameFor self-testing outside the body (In vitrodiagnostic use) by people with diabetes athome as an aid to monitor the effectiveness ofdiabetes control. Not to be used for diagnosisor screening diabetes or for neonatal use
CalibrationSamePlasma Equivalent
Measurement technologySameAmperometry
Sample typeSameCapillary whole blood
Sample sitesSameFinger Tip
EnzymeSameGDH-FAD
Maximum AltitudeSameUp to 10,000 feet
ButtonsSame3 button user input
WeightSameApproximately 1.7 ounces
Auditory IndicatorSameYes

{5}------------------------------------------------

Differences

Operating Temperature41-104 °F (5-40 °C)43°F - 111°F (6°C - 44°C)
Operating humidity10%-85%10% - 90%
Test range50-600 mg/dL20-600 mg/dL
Sample volume1.2 µL0.4 µL
Hematocrit30 – 55 %20%-60%
Sample Test Time10 seconds5 seconds
Calibration CodingManual code inputwith date & time (14 & 30 days average)Non-coding
Memory capacity1500 with date &time (14 & 30 daysaverage)Non (data transfer to remote station)
Size (L x W x H)0.88 x 2.28 x 3.15inches0.63 x 2.05 x 3.39 inches
Power SourceOne 3 Volts, 0.2A(CR2032) batteryTwo 1.5 V (AAA) battery
Microprocessor16 Bit with 16KmemoryMultichip Module
DisplaySegmentedmonochrome liquidcrystal displayDot matrix thin film transistor color display
Compatible off-metersoftware accessoriesPrimosAppOne Touch Diabetes Management Software
Alert AlarmOption to set alertsfor testing timeNo Alarm
Remote data storageNoYes
Controls solution(s)Single Control(either Low, Mediumor High)Two Controls (Medium and High)
Target Range IndicatorNoYes
Data DownloadYes, USB onlyYes (via USB or Blue tooth)
Test StripPlastic molded stripmade of palladiumplated solid copperelectrodes covered byinsert & over moldingwith plasticMetal foil electrodes trapped between layers ofplastic

Candidate device is similar or very close to the predicate device in its all major claims. The following are main differences between predicate and candidate device.

  • The strip of candidate device has palladium plated solid copper electrodes which l. are covered by insert and over molding of plastic, whereas the predicate device has metal foil electrodes trapped between plastic layers.
  • II. In the candidate device data can be transfer using a cable whereas predicate device transfer data wirelessly.

{6}------------------------------------------------

  • III. The candidate device has monochrome display with coding whereas predicate has color displav with no coding.

9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

The new Applied Biomedical LLC Confidence, a blood glucose monitoring device was verified against users need and intended uses as to fulfill through Design Input requirements.

The subject device demonstrated substantial equivalent performance to the predicate device and YSI. The following standards were used for non-clinical test:

  • EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline i) Second Edition (2005).
  • ii) EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (2003).
  • iii) ISO 14971 Medical devices —Application of risk management to medical devices (2007).
  • iv) IEC/EN 60601-1-2 (EMC Testing)
  • v) IEC 61010-1-04 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1: General Requirements.
  • vi) IEC 61010-1 (2nd Edition) Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1: General Requirements.
  • vii) IEC 61010-2-101:04-Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory use - Part 2-101: Particular Requirements for In-Vitro Diagnostic (IVD) Medical Equipment

10. Discussion of Clinical Tests Performed:

A system accuracy evaluation was performed by comparing glucose value obtained by trained person using Confidence™ BGMS from finger stick blood of the 265 subiects with the scaled glucose value obtained by FDA cleared reference method (YSI 2300). The study was performed at two sites using three strip lots and eleven meters. The results demonstrated that Confidence™ BGMS is substantially equivalent in its performance with YSI.

(Glucose Concentrations Range 60-500 mg/dL)
Percent (and Number ) of meter results that match the YSI reference (N=270)
Within ± 10 mg/dL or±10%Within ± 10 mg/dL or±15%Within ±10 mg/dL or± 20 %
74.8% (202/270)99.6% (269/270)00% (270/270)

System Accuracy Results Relative to YSI Values

The Confidence™ BGMS liner regression data relative to YSI reference value was compared with liner regression data of Touch® VerioFlex™ BGMS (Predicate Device) relative to YSI reference value in following table. All values

{7}------------------------------------------------

are showing excellent linearity, the slope and R2 are comparable to predicate device, indicating that the Confidence™ BGMS is substantially equivalent to the VerioFlex™ BGMS as per predicate device labeling.

Liner Regression data for Accuracy of Confidence™ BGMS and Predicate Device Relative to YSI Values

BGMSParticipantsSlopeInterceptR2
Touch® VerioFlex™300*0.992.610.99
Confidence™2700.993.830.99
*From Product labeling

A user study was performed to demonstrate that lay users could use the Confidence Glucose Monitoring Systems and obtain accurate results. The study was performed by 181 users; they never have used the candidate device before. The results obtained by using candidate device were compared with the scaled glucose values obtained by YSI. The results demonstrated the Confidence™ BGMS is substantially equivalent in lay users' hand to its performance with YSI.

Users Performance Results Relative to YSI Values (Glucose Concentrations Range 60-480 mg/dL)

Percent (and number) of meter results that match YSI reference (N=187)
Within ±10 mg/dL or ± 10 %Within ±15 mg/dL or ± 15 %
90.4% (169/187)100% (187/187)

The linear regression data obtained by user with the Confidence™ BGMS relative to YSI reference value was compared with liner regression data of values obtained by user using Touch® VerioFlex™ BGMS (Predicate Device) relative to YSI are showed in following table. All values are showing excellent linearity, the slope and R2 are comparable with predicate device, indicating that the Confidence™ BGMS is substantially equivalent to the VerioFlex™ BGMS as per predicate device labeling.

Liner Regression data of Confidence™ BGMS and Predicate Device Relative to YSI

BGMSParticipantsSlopeInterceptR2
Touch® VerioFlex™172*0.992.980.98
Confidence™1871.010.290.99
*From Product's labeling

A survey was also conducted using 181 lay users about the instruction to use and steps to perform the test, cleaning and error codes description. The participants had no difficulty to perform the test and understood the instructions about cleaning and error codes.

11. Conclusions:

The Confidence™ Glucose Monitoring Systems is substantially equivalent in its intended use, performance, safety, effectiveness and underlying scientific and operating principal used, to the predicate device, OneTouch Verio Flex Blood Glucose Monitoring System.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.