K Number
K173571

Validate with FDA (Live)

Date Cleared
2018-03-21

(121 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Element system is intended to be used exclusively for myoelectric exoprosthetic fittings of the upper limb.

Device Description

The Element system is a surface electromyography (EMG) electrode system that is to be used with upper limb prosthetic devices. Element outputs standard envelop EMG signals detected from EMG electrodes (IBT Electrodes) placed on the user's skin. These signals are used as inputs by connected prosthetic devices, such as hands, wrists or elbows. The Element system is an alternative to standard suction socket myoelectrodes, with the advantages of a lower profile, digital signal processing, and wireless gain adjustment. Element does not replace or modify any functionality of connected prosthetic components.

Element is compatible with most hands, wrists, and elbows that accept standard analog EMG electrode inputs. The Element system is typically sold with three-port kidney-style output connectors, however alternative connectors may be used to ensure compatibility with other terminal devices. The Element system accepts power from IBT's FlexCell Battery system and outputs control EMG signals to hands, wrists, or other prosthetic components. The Element system is installed in the prosthesis by a trained prosthetist and connected to prosthetic components selected to meet the needs of the individual user.

System components:

  • IBT Electrodes (up to 2)
  • Signal Processing Box
  • Element Desktop Software
AI/ML Overview

The provided document is a 510(k) summary for the Infinite Biomedical Technologies (IBT) Element System with IBT Electrodes. It outlines the device's characteristics, comparison to a predicate device, and the testing conducted to demonstrate substantial equivalence.

Here's an analysis of the acceptance criteria and study information, addressing your specific questions:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly present a table of acceptance criteria with corresponding pass/fail results for each criterion to demonstrate "device performance" in terms of clinical outcomes or specific quantitative metrics for the Element System's functionality as a myoelectric signal processor. Instead, it focuses on demonstrating safety and effectiveness through adherence to recognized standards and successful completion of various verification and validation tests.

However, we can infer acceptance criteria from the "Performance Data" section. The acceptance criterion for each test is generally "Pass," indicating that the device met the requirements of the specified standard or test protocol.

Here's a table based on the provided "Performance Data":

CategoryTest/StandardAcceptance Criteria (Implied)Reported Device Performance
SafetyIEC/EN 60601-1: Medical electrical equipment Part 1: General requirements for basic safety and essential performanceCompliance with standardEquivalent (to predicate)
IEC 60601-1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environmentCompliance with standardPredicate was not tested to this standard, implying the Element system does comply.
Electromagnetic CompatibilityIEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and testsCompliance with standardEquivalent
BiocompatibilityISO 10993-1: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management processCompliance with standardEquivalent
ISO 10993-5: Biological evaluation of medical devices Part 5: Test for in vitro cytotoxicityCompliance with standardEquivalent
ISO 10993-10: Biological evaluation of medical devices Part 10: Test for irritation and skin sensitizationCompliance with standardEquivalent
Design V&V/Usability (Element System)Compatibility with IBT ElectrodesDevice is compatiblePass
Simulated installation of ElementSuccessful simulated installationPass
Simulated Use of Element Software and Electrode PlacementSuccessful simulated usePass
Simulated Use with Commonly used Prosthetic ComponentsSuccessful simulated usePass
Lifetime and Reliability TestingDevice meets lifetime/reliability requirementsPass
Testing Report SummaryReport deemed satisfactoryPass
Signal Performance during Simulated UseAcceptable signal performancePass
Simulated battery life with ElementAcceptable battery lifePass
Simulated use with Region Specific NoiseAcceptable performance in noisePass
Packaging Drop TestPackaging integrity maintainedPass
Simulated Installation of IBT ElectrodesSuccessful simulated installationPass
Simulated SealAcceptable seal performancePass
Simulated CleaningDevice withstands cleaningPass
FlexCell Battery TestingIEC 60601-1: Medical electrical equipment Part 1: General requirements for basic safety and essential performanceCompliance with standardComplies
IEC 62133: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applicationsCompliance with standardComplies
UN38.3: Transportation Testing for Lithium BatteriesCompliance with standardComplies
FlexCell Major Component TestMajor components perform as specifiedPass
FlexCell V&V Test Specification PlansPlans are executed successfullyPass
FlexCell Charger Update V&V Test PlanPlan is executed successfullyPass
FlexCell Charger IC Update V&V Test PlanPlan is executed successfullyPass

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify the sample sizes used for the various verification and validation tests listed. It also does not provide details on data provenance (e.g., country of origin of the data, retrospective or prospective) for these tests. The tests mentioned are primarily laboratory-based simulations and regulatory standard adherence, not clinical trials with patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not provided in the document. The tests described are engineering and regulatory compliance validations rather than diagnostic or interpretative studies requiring expert ground truth establishment. For example, "Signal Performance during Simulated Use" would likely involve objective measurements against predefined specifications, not expert consensus on an interpretation.

4. Adjudication Method for the Test Set

This information is not provided and is generally not applicable to the types of engineering and regulatory compliance tests described. Adjudication methods (like 2+1, 3+1) are common in clinical studies where multiple human readers interpret data, and an agreed-upon truth is needed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done and is not applicable to this device. The Element System is a myoelectric signal processing unit for prosthetic control, not an AI-assisted diagnostic or interpretive tool that human readers would use. The document describes the system as an alternative to standard suction socket myoelectrodes, with advantages like a lower profile, digital signal processing, and wireless gain adjustment, but it doesn't involve "AI assistance" in the sense of improving human reader performance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

The Element System itself is a standalone algorithm/device for processing myoelectric signals. The performance tests ("Signal Performance during Simulated Use", "Simulated battery life with Element", "Simulated use with Region Specific Noise") would effectively be standalone performance evaluations of the device's electrical and signal processing capabilities, without human user interaction as part of the performance measurement. However, its intended use is with a human user providing the EMG signals and benefiting from its processing for prosthetic control.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the specific tests mentioned, the "ground truth" would be established by:

  • Regulatory Standards: For tests like IEC 60601-1, ISO 10993, and UN38.3, the ground truth is adherence to the requirements and specifications defined by these international standards.
  • Engineering Specifications: For tests like "Compatibility with IBT Electrodes," "Signal Performance during Simulated Use," and "Simulated Seal," the ground truth would be predefined engineering specifications, performance tolerances, or established benchmarks for electrical signals and mechanical functions.
  • Simulated Conditions: For tests involving "simulated use" or "simulated battery life," the ground truth is the controlled and measurable conditions established within the simulation.

There is no mention of expert consensus, pathology, or outcomes data being used as ground truth in this submission, as it's not a diagnostic or treatment outcome-focused device.

8. The Sample Size for the Training Set

The document does not mention any training set size. As the Element System with IBT Electrodes is a device that processes myoelectric signals based on predefined algorithms and hardware, it is unlikely to involve "training" in the machine learning sense that would require a distinct training set. The "digital signal processing" and "signal smoothing" functionalities are described as inherent design features or adjustable parameters, rather than learned models.

9. How the Ground Truth for the Training Set Was Established

Since no training set is mentioned or implied, the question of how its ground truth was established is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 21, 2018

Infinite Biomedical Technologies, LLC Rahul Kaliki CEO 1101 E 33rd Street, Suite E305 Baltimore, Maryland 21218

Re: K173571

Trade/Device Name: Element System with IBT Electrodes Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY, IQZ Dated: January 19, 2018 Received: January 23, 2018

Dear Rahul Kaliki:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vivek J. Pinto -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173571

Device Name Element System with IBT Electrodes

Indications for Use (Describe)

The Element system is intended to be used exclusively for myoelectric exoprosthetic fittings of the upper limb.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a red infinity symbol against a black background. The symbol is centered in the frame and is the only object in the image. The red color of the symbol contrasts sharply with the black background, making it stand out.

510(k) Summarv Element System with IBT Electrodes

SUBMITTER'S INFORMATION 1.

Name/ManufacturerInfinite Biomedical Technologies, LLC
Address1101 E. 33rd Street, Suite E305, Baltimore, MD 21218
Phone Number(443) 451-7177
Fax Number(443) 451-7179
Contact PersonRahul Kaliki, PhD, Chief Executive Officer
Date PreparedMarch 21, 2018

DEVICE INFORMATION 2.

Trade NameElement System with IBT Electrodes
Common NamePowered, External Upper Limb Prosthetic System
ClassificationCutaneous Electrode (21 CFR § 882.1320)
Product CodeGXY (Electrode, Cutaneous)
Subsequent Product CodeIQZ (Hand, External Limb Component, Powered)

PREDICATE DEVICE INFORMATION 3.

Device NameAxon-Bus Prosthetic System
510(k) NumberK123795

4. INDICATIONS FOR USE STATEMENT

The Element system is intended to be used exclusively for myoelectric exoprosthetic fittings of the upper limb.

DEVICE DESCRIPTION 5.

The Element system is a surface electromyography (EMG) electrode system that is to be used with upper limb prosthetic devices. Element outputs standard envelop EMG signals detected from EMG electrodes (IBT Electrodes) placed on the user's skin. These signals are used as inputs by connected prosthetic devices, such as hands, wrists or elbows. The Element system is an alternative to standard suction socket myoelectrodes, with the advantages of a lower profile, digital signal processing, and wireless gain adjustment. Element does not replace or modify any functionality of connected prosthetic components.

Element is compatible with most hands, wrists, and elbows that accept standard analog EMG electrode inputs. The Element system is typically sold with three-port kidney-style output connectors, however alternative connectors may be used to ensure compatibility with other terminal devices. The Element system accepts power from IBT's FlexCell Battery system and outputs control EMG signals to hands, wrists, or other prosthetic components. The Element system is installed in the prosthesis by a trained prosthetist and connected to prosthetic components selected to meet the needs of the individual user.

System components:

  • IBT Electrodes (up to 2) ●
  • Signal Processing Box ●
  • . Element Desktop Software

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a red infinity symbol against a black background. The symbol is a figure eight on its side, with smooth, rounded curves. The red color of the symbol contrasts sharply with the black background, making it stand out. The image is simple and minimalist, with the focus entirely on the infinity symbol.

6. TECHNOLOGICAL CHARACTERISTICS

Table 1 provides a summary of technological characteristics of the product in comparison to the predicate device.

ManufacturerInfinite BiomedicalTechnologies, LLCOtto Bock HealthCare Product GmbHDeviceComparison
Trade NameElement System withIBT ElectrodesAxon bus ProstheticSystem
System
Power Source included?NoYesDiffers
Terminal device (Hand,Wrist or Elbow) included?NoYesDiffers
Wireless communicationBluetooth®Bluetooth®Same
Clinical Software ToolYesElement ApplicationYesAxonSoftSimilar
Software/Firmware/Microprocessor Control?YesYesSimilar
Input Voltage5 to 10 VDC11.1 VDCDiffers
Output Signal0-5 V analog0-5 V analogSame
Processing Unit
Processing Unit(L x W x H)Signal ProcessingBox38 x 23 x 8 mmAxonMaster53 x 28 x 9 mmSimilar
Control optionsNoneMultipleDiffers
Signal SmoothingYesNoDiffers
Electrode
Electrode(L x W x H)IBT Electrode29 x 17 x 7Electrode27 x 18 x 9.5 mmSimilar
Temperature range (use)-10°C to 50°C-15°C to 60°CSimilar
Housing MaterialPlastics (ABS/PCBlend)Plastics (ASA)Similar
Contact AreaTitanium (Grade 1)Titanium (Grade 1)Same
Bonding AgentCyanoacrylateCyanoacrylateSame
Signal processingDigitalAnalogDiffers
Frequency Bandwidth90 - 500 Hz90 - 450 HzSimilar
AdjustmentDigital gain 1-7Potentiometer 1-7Differs
Installationsuspension arms /suction socketsuspension arms /suction socketSame

Table 1: Technological Summary

7. Substantial Equivalence Discussion

IBT believes that its Element system with IBT Electrodes is substantially equivalent to specific parts of the Ottobock Axon-Bus Prosthetic System (K123795). A comparison of intended use, indications for use, technology, and performance is provided herein to support this determination. For instances where technological differences are identified, additional discussion describes how these differences do not raise

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a red infinity symbol against a black background. The symbol is centered in the frame and is the only object in the image. The red color of the symbol contrasts with the black background, making it stand out.

new or different questions of safety and effectiveness. When appropriate, performance data is cited to provide evidence that the subject device is as safe and as effective as the legally marketed predicate device.

Intended Use and Indications for Use

Both the subject device and the predicate device aim to detect, process, and transmit physiological signals (detected via surface electrodes), providing them to a downstream prosthetic device. Both products are prescription use devices, intended to be installed by a prosthetist or trained clinician.

From a clinical perspective, the Element system with IBT Electrodes is intended to be used with upper limb exoprosthetic fittings, exclusively. This clinical usage profile is identical to the predicate device.

CharacteristicSubject DevicePredicate Device (K123795)
Element System with IBT ElectrodesAxon-Bus Prosthesis System
Intended UseDetect, process, and transmitphysiological signals for use with aprosthesisSame as subject device
Indications for UseThe Element system is intended to beused exclusively for myoelectricexoprosthetic fittings of the upper limb.The Axon-Bus Prosthetic System is to beused exclusively for exoprosthetic fittingsof the upper limbs.

Table 2. Intended Use and Indications for Use

Minor differences in the written text do not affect the meaning (in other words, they are semantics).

Technology

The Element system contains a subset of components within the Axon-Bus Prosthetic System (K123795), as delineated in Table 3.

ComponentSubject DevicePredicate Device (K123795)
Element SystemAxon-Bus Prosthesis System
Terminal deviceN/AMichelangelo Hand
Passive wrist flexion deviceN/AAxonFlexion Adapter
Passive wrist rotation deviceN/AAxonRotation Adapter
Passive elbow jointN/AAxonArm
BatteryN/AAxonEnergy Integral
ChargerN/AAxonCharge
Processing unitSignal Processing BoxAxonMaster
EMG electrodeIBT ElectrodeElectrode
Adjustment softwareSoftware ApplicationAxonSoft
Prosthetic gloveN/AAxonSkin

Table 3: System Components

IBT is only claiming equivalence to the predicate's components that are comparable to or part of the Element system. Based upon a technical review of both systems, IBT has identified the following technical differences:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a red infinity symbol against a black background. The symbol is centered in the frame and appears to be illuminated, as the red color is vibrant. The infinity symbol is a mathematical symbol representing the concept of limitless or endless.

  • Digital Signal Processing ●
  • Digital Gain Adjustment Software
  • Signal Smoothing
  • Control Options
  • Terminal Options
  • . Battery Characteristics
  • Electrode Materials
  • Electrode Size
  • . Software

These areas are described and discussed herein.

Digital Signal Processing

Description & Comparison: The IBT Electrodes digitize the signals at the electrode site using an analog front end (AFE) chip. The AFE samples the raw EMG signals with a high-resolution ADC. By digitizing the signal near the site of the electrode contacts, this system architecture reduces the potential for external noise sources to corrupt the EMG signals. The predicate Axon-Bus Prosthetic System Electrodes only use analog components in its design.

Discussion: Digital signal processing was incorporated to reduce noise corruption and allow for further signal processing downstream. These low-power components also are protected throughout the circuit layout with appropriate ESD protective components. The deviation does not result in any additional risk to the user. Both subject and predicate devices need meet electrical safety and electromagnetic compatibility to requirements.

Digital Gain Adjustment Software

Description & Comparison:The Element system allows for wireless adjustments of gain and furthersignal processing at the downstream Signal Processing Box. Specifically,the Signal Processing Box includes a microcontroller that communicateswith the electrodes, processes the EMG signals, and outputs envelopeEMG signals to connected prosthetic components. The Signal ProcessingBox contains a Bluetooth module, which allows the controller tocommunicate with a software application developed in Java called theElement Application. The user (clinician) can visualize EMG signals andadjust electrode gains through this mechanism. The predicate Axon-BusProsthetic System Electrode has a gain that can be adjusted through apotentiometer on the back of the electrode.
Discussion:Access to the gain adjustment can be difficult to reach for prosthetistsand, often times, they must create holes in the exterior shell and socketof the prosthesis to enable access to the gain adjustment, creating highstress areas in the material as well as making the prosthesis vulnerable todebris and liquids. The Element system does not have a physicalmanipulator to adjust the gains. Instead, the adjustments can all be made

through the Element Application software. Verification and Validation tests were conducted to ensure that the Element System's gain

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a red infinity symbol against a black background. The symbol is centered in the image and is the only object present. The red color of the symbol contrasts sharply with the black background, making it stand out. The infinity symbol is a mathematical symbol representing the concept of infinity.

adjustment method resulted in an equivalent signal output to the predicate device.

Although the methodology is different, both systems contain a mechanism to adjust the signal gain.

Signal Smoothing
Description & Comparison:The Element system includes functionality to change the smoothness of the EMG signal. This feature is adjusted through the Element Application . The smoothing function increases the stability of the EMG signal at the cost of responsiveness. The higher smoothness allows for users to control their signal amplitude more steadily if they have trouble maintaining extended signals. A lower smoothness allows the user to achieve triggers for degree-of-freedom switching more easily. The smoothness function does have limits to ensure that the reduced responsiveness does not significantly impede the performance of the electrodes when used with a prosthetic device (see draft instructions for use). Default settings can be used to mimic similar signal qualities of the predicate device. The predicate device does not include a smoothing feature in the design, however other electrodes on the market include this functionality. 1
Discussion:The smoothness function is a feature to aid prosthetists in making adjustments unique to the individual, similar to how the predicate device provides various control options depending on the type of signals generated by the patient/end user. The risks involved in this customization step are no different between subject and predicate devices.
Control Options
Description & Comparison:Complete prosthetic systems typically contain a set of "control options" that allow modification of the way a user would operate the prosthetic device. This may include "first-over" versus "differential" activation of a degree of freedom or an option to reverse input directions. It may also specify if inputs besides EMG are used in the prosthesis. The Axon-Bus Prosthetic System (predicate under K123795) contains this functionality, whereas the subject Element system does not.
Discussion:Element does not provide control mechanisms for the downstream prosthetic system; the output of the Element system is similar to that of the predicate device EMG Electrode, which provides filtered signals to the prosthesis based on user inputs. Therefore, the device has inherently simpler operation than the predicate. This deviation does not raise any additional concerns with respect to risk to the user.
Terminal Options
Description & Comparison:The Element system does not include a terminal device, passive wrist flexion/extension devices, or an elbow joint. These options are available on the predicate device. Instead, the Element system outputs its signal

1 For example, the Triad Electrodes from Motion Control, Inc. have a smoothness functionality from Motion Control is part of a Class I, 510(k) exempt component.

{8}------------------------------------------------

8

infinite biomedical technologies

through industry-standard 3-pin kidney shaped connectors, with an output range exactly the same as other similar electrode devices on the market (including the predicate system's electrode).

  • Discussion: Both subject and predicate devices need to ensure that the output signal is transferrable to a downstream component. The subject device does this through industry-standard connections. If this specification is met, then the subject device does not introduce new or different questions of safety or effectiveness.
  • Battery Characteristics Description & Comparison: The Element system does not include a battery and has an input voltage range of 5-10 V. The predicate device uses a rechargeable 11.1 V Li-lon Battery Pack. This difference is thought to be due to voltage requirements in the Axon-Bus Prosthetic System's Michelangelo Hand, which requires higher voltage than other terminal devices on the market.
  • Discussion: The input voltage range of the Element system is similar to most electrodes currently sold on the market today. The Element system is only intended to be compatible with a separately sold battery, FlexCell, which is rated at a nominal voltage of 7.4V. This battery system is compatible with most upper limb prosthetic terminal devices on the market. This difference in voltage requirements has no impact on the safety or effectiveness profile.
  • Electrode Materials Description & Comparison: The housing of the IBT Electrode is manufactured with a PC-ABS blend, whereas the predicate device housing uses ASA. Both products utilize titanium grade 1 electrode contact material. The type of contact, duration of contact, and cleaning profiles are identical.
  • Discussion: Both subject and predicate devices have patient contacting materials that experience the same use scenarios. The risk of a tissue response to the material used is equivalent, and both products are tested to the applicable parts of ISO 10993.
  • Electrode Size Description & Comparison: A specific requirement for certain prosthetic applications is to include electrodes capable of maintaining suction on the patient's limb when the prosthesis is donned. The Element system with IBT Electrodes can maintain a sealed socket with a small form factor.
  • The smaller electrode size could impact installation (maintaining suction) Discussion: and performance (signal detection). Neither aspect is unique or representative of a new/different question of safety or effectiveness. Testing demonstrated equivalent performance.
  • Software Description & Comparison: The Element system includes embedded system software within the microprocessor of the Signal Processing Unit and a separate software application. The predicate device also has a software application (AxonSoft) and embedded software within the Axon Master and

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a red infinity symbol against a black background. The symbol is centered in the image and is the only object present. The red color of the symbol contrasts with the black background, making it stand out.

infinite biomedical technologies

Michaelangelo hand. The Element's software application is used for viewing signals, gain adjustment, and adjusting smoothness of the EMG signal, whereas the predicate device's software application is used to change control options, adjust the parameters of the control options, and view signals. The Element's embedded software is used to process the digital signals (filtering, gain adjustment, smoothness), whereas the embedded software in the predicate device is responsible for both lowlevel and high-level control functions for the prosthetic hand.

Discussion:

The differing features of the Element system's software application and embedded software do not introduce new/different safety or efficacy concerns.

8. Performance Data

The Element system with IBT Electrodes was tested to ensure its safety and effectiveness. The following Performance Standards were used for performance testing of the Element system:

CategoryNoTitleVersionComparison to PredicateDevice Axon-Bus(K123795)
SafetyIEC/EN 60601-1Medical electrical equipmentPart 1: General requirementsfor basic safety and essentialperformance2012(IEC)/2006(EN)Equivalent
IEC 60601-1-11General requirements for basicsafety and essentialperformance - CollateralStandard: Requirements formedical electrical equipmentand medical electrical systemsused in the home healthcareenvironment2010Predicate was not tested tothis standard

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a red infinity symbol against a black background. The symbol is horizontally oriented and centered in the frame. The red color of the symbol contrasts sharply with the black background, making it stand out. The infinity symbol is smooth and symmetrical.

infinite biomedical technologies

CategoryNoTitleVersionComparison to PredicateDevice Axon-Bus(K123795)
ElectromagneticCompatibilityIEC 60601-1-2Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests2007Equivalent
BiocompatibilityISO 10993-1Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process2009Equivalent
BiocompatibilityISO 10993-5Biological evaluation of medical devices Part 5: Test for in vitro cytotoxicity2009Equivalent
BiocompatibilityISO 10993-10Biological evaluation of medical devices Part 10: Test for irritation and skin sensitization2009Equivalent

The Element system with IBT Electrodes also underwent design verification and validation, software verification and validation, and usability testing to demonstrate its ability to achieve its intended use safely and effectively. The following validation testing was performed on the device:

Test NameResult
Compatibility with IBT ElectrodesPass
Simulated installation of ElementPass
Simulated Use of Element Software andElectrode PlacementPass
Simulated Use with Commonly used ProstheticComponentsPass
Lifetime and Reliability TestingPass
Testing Report SummaryPass
Signal Performance during Simulated UsePass
Simulated battery life with ElementPass
Simulated use with Region Specific NoisePass
Packaging Drop TestPass
Simulated Installation of IBT ElectrodesPass
Simulated SealPass
Simulated CleaningPass

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows a red infinity symbol against a black background. The infinity symbol is a mathematical symbol representing the concept of infinity. It is shaped like a figure eight on its side. The red color of the symbol contrasts with the black background, making it stand out.

The Flexcell battery was also tested to ensure its safety and effectiveness. The following Performance Standards were used for performance testing of the FlexCell battery:

CategoryNoTitleVersion
SafetyIEC 60601-1Medical electrical equipment Part 1: Generalrequirements for basic safety and essentialperformance2012 (IEC)
IEC 62133Secondary cells and batteries containing alkalineor other non-acid electrolytes - Safetyrequirements for portable sealed secondary cells,and for batteries made from them, for use inportable applications2002 (1st edition)
TransportationUN38.3Transportation Testing for Lithium BatteriesST/SG/AC.10/11/Rev.5/Amend.1&ST/SG/AC.10/11/Rev.5/Amend.2

The FlexCell battery also underwent design verification and usability testing to demonstrate its ability to function safely and effectively. The following testing was performed on the device:

Test NameResult
FlexCell Major Component TestPass
FlexCell V&V Test Specification PlansPass
FlexCell Charger Update V&V Test PlanPass
FlexCell Charger IC Update V&V Test PlanPass

9. Conclusions

Based upon the discussion provided herein and the supporting data, IBT believes its Element system with IBT Electrodes are as safe and as effective as the predicate device (Axon-Bus K123795) for its intended use, making it substantially equivalent to a legally marketed predicate device.

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).