K Number
K173561

Validate with FDA (Live)

Date Cleared
2018-02-12

(87 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vinyl Patient Examination Glove (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

The proposed device is Powder Free Vinyl Patient Examination Gloves. The proposed device is yellow. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The proposed device is non-sterile.

AI/ML Overview

The acceptance criteria and study proving the device meets these criteria are detailed below, based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

ITEMAcceptance Criteria (Proposed Device K173561)Reported Device Performance (Proposed Device K173561)
Product CodeLYZLYZ
Regulation No.21 CFR 880.625021 CFR 880.6250
ClassII
Intended UseDisposable device for medical purposes, worn on examiner's hands to prevent contamination between patient and examiner.Meets this description.
Powdered or Powered freePowder-freePowder-free
Design FeatureAmbidextrousAmbidextrous
Labeling InformationSingle-use indication, powder free, device color, device name, glove size and quantity, Vinyl Examination Gloves, Non-SterileMeets these requirements.
Dimensions (Length, Width)Complies with ASTM D 5250 specifications (specific values depend on size: e.g., Length min 230mm, Width ±5mm of specific values)All proposed device meets the specifications of ASTM D 5250.
Thickness (Finger, Palm)Finger: 0.05 min, Palm: 0.08 minFinger: 0.05 min, Palm: 0.08 min
ColorantYellowYellow
Physical PropertiesBefore Aging: Tensile Strength: 12 MPa, min; Ultimate Elongation: 300 % minAchieved 12 MPa, min (Tensile Strength); 300 % min (Ultimate Elongation)
After Aging: Tensile Strength: 12 MPa, min; Ultimate Elongation: 300 % minAchieved 12 MPa, min (Tensile Strength); 300 % min (Ultimate Elongation)
Freedom from HolesBe free from holes when tested in accordance with ASTM D5151 Level=G-I, AQL=1.5Be free from holes when tested in accordance with ASTM D5151 AQL=1.5
Powder ContentMeets the requirements of ASTM 52501.7 mg (within ASTM 5250 requirements for powder-free, which is < 2mg)
MaterialVinylVinyl
BiocompatibilityIrritation: not an irritant (under study conditions)Met (evaluated per ISO 10993-10)
Sensitization: not a sensitizer (under study conditions)Met (evaluated per ISO 10993-10)
Cytotoxicity: did not show potential toxicity to L-929 cells (under study conditions)Met (evaluated per ISO 10993-5)
Label and LabelingMeets FDA's Requirements of 21 CFR 801Meets FDA's Requirements of 21 CFR 801

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes for each non-clinical test (e.g., number of gloves for tensile strength, number of samples for biocompatibility). However, it indicates compliance with recognized standards that prescribe sampling methods.

  • ASTM D5151-06 (Reapproved 2015), Standard Test Method for Detection of Holes in Medical Gloves: This standard outlines sampling procedures to determine freedom from holes.
  • ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection: This standard was used for sampling procedures.

The data provenance is from non-clinical laboratory testing performed by or for Yurun Glove Co., Ltd. in China. The study is prospective in the sense that tests were conducted specifically for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. The ground truth for this device (a medical glove) is established through adherence to recognized international and national consensus standards (e.g., ASTM, ISO) for physical properties, barrier integrity, and biocompatibility, rather than expert consensus on interpretive data like medical images. The "experts" would be the certified laboratories and personnel performing the standardized tests.

4. Adjudication Method for the Test Set

Not applicable. As noted above, performance is measured against objective, standardized laboratory tests, not subjective interpretations requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This type of study is typically performed for AI/CAD devices involving interpretation of medical images or data by human readers. This device is a medical glove, and its performance is assessed via non-clinical laboratory tests.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No. This concept is not applicable to a physical medical device like a glove. The device itself is not an algorithm.

7. Type of Ground Truth Used

The ground truth is based on:

  • Established consensus standards: Such as ASTM D5250 for Poly(vinyl chloride) Gloves for Medical Application, ASTM D6124 for Residual Powder, ASTM D5151 for Detection of Holes, ISO 10993-10 for Irritation and Sensitization, and ISO 10993-5 for Cytotoxicity. These standards define the acceptable performance parameters.
  • Objective laboratory measurements: The device's physical properties (e.g., tensile strength, elongation, thickness, freedom from holes, powder content) and biocompatibility are measured against the specifications outlined in these standards.

8. Sample Size for the Training Set

Not applicable. This device is a physical product and does not involve AI/machine learning requiring a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is blue.

February 12, 2018

Yurun Glove Co., Ltd. % Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. 5-402, Building #27, No. 56, LiangXiang East Rd., FangShan District Beijing, 102401 Cn

Re: K173561

Trade/Device Name: Vinyl Patient Examination Glove (Yellow) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I Product Code: LYZ Dated: November 15, 2017 Received: November 17, 2017

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K173561

Device Name Vinyl Patient Examination Glove (Yellow)

Indications for Use (Describe)

The Vinyl Patient Examination Glove (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K173561

    1. Date of Preparation:2018/01/26
    1. Sponsor Identification

YURUN GLOVE CO., LTD. No.10 ZHAOCAI STREET, LUANNAN COUNTY, SHITANGSHAN CITY, HEBEI,CHINA

Contact Person: Chao kevin Position: General Manager Tel: +86- 315-4168700 Fax: +86- 315-4168700 Email: zhplylyp@outlook.com

    1. Designated Submission Correspondent
      Mr. Ray Wang

Beijing Believe-Med Technology Service Co., Ltd.

Tel: +86-18910677558, Fax: +86-21-68093116 Email: Ray.Wang@believe-med.com

{4}------------------------------------------------

    1. Identification of Proposed Device
      Trade Name: Vinyl Patient Examination Glove (Yellow) Common Name: Vinyl Patient Examination Gloves (Powder Free) Model(s): S M L XL

Regulatory Information Classification Name: Vinyl Patient Examination Gloves (Powder Free) Classification: I Product Code: LYZ Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

Indication for use Statement:

The Vinyl Patient Examination Glove (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    1. Device Description
      The proposed device is Powder Free Vinyl Patient Examination Gloves. The proposed device is yellow. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The proposed device is non-sterile.
    1. Identification of Predicate Device(s)
      Predicate Device K163168 Vinyl Examination Gloves (White, Blue or Yellow) Hebei Hongtai Plastic Products Company Limited

{5}------------------------------------------------

ITEMProposed Device (K173561)Predicate Device (K163168)Remark
Product CodeLYZLYZSAME
Regulation No.21 CFR 880.625021 CFR 880.6250SAME
ClassIISAME
Intended UseThe Vinyl Patient Examination Glove(Yellow) is a disposable device intendedfor medical purposes that is worn on theexaminer's hands to prevent contaminationbetween patient and examiner.The Vinyl Examination Glove (White,Blue, or Yellow) is a disposable deviceintended for medical purposes that is wornon the examiner's hands to preventcontamination between patient andexaminer.SAME
Powdered orPowered freePowdered freePowdered freeSAME
Design FeatureambidextrousambidextrousSAME
LabelingInformationSingle-use indication, powder free, devicecolor, device name, glove size and quantity,Vinyl Examination Gloves, Non-SterileSingle-use indication, powder free, devicecolor, device name, glove size and quantity,Vinyl Examination Gloves, Non-SterileSAME

Technological Comparison Table 7.

Table 1 General Comparison

{6}------------------------------------------------

Size
DesignationXSSMLXLTolerance
Length, mm230230235245245min
Predicate Device (K163168)Width, mm808595105115±5
Thickness, mm:
Finger0.05min
Palm0.08min
Size
DesignationSMLXLTolerance
Length, mm230230230230min
Proposed Device (K173561)Width, mm8595105115±5
Thickness, mm:
Finger0.05min
Palm0.08min
RemarkAnalysis 1

Table 2 Device Dimensions Comparison

Analysis 1:

The proposed device has different size specification to the predicate device, but all proposed device meets the specifications of ASTM D 5250.

Table 3 Performance Comparison
ITEMProposed Device (K173561)Predicate Device (K163168)Remark
ColorantYellowWhite, Blue, YellowSAME
PhysicalPropertiesBeforeAgingTensileStrength12 MPa, min15 MPa, minAnalysis 2
BeforeAgingUltimateElongation300 % min380 % min
AfterAgingTensileStrength12 MPa, min15 MPa, min
AfterAgingUltimateElongation300 % min380 % min
Freedom from HolesBe free from holes when tested inaccordance with ASTM D5151Level=G-I, AQL=1.5Be free from holes when testedin accordance with ASTMD5151AQL=1.5SAME
Powder Content1.7 mgMeet the requirements of ASTM5250SAME

{7}------------------------------------------------

ITEMProposed Device (K173561)Predicate Device (K163168)Remark
MaterialVinylVinylSAME
BiocompatibilitySAME
IrritationUnder the conditions of the study, not an irritantComply with ISO 10993-10
SensitizationUnder conditions of the study, not a sensitizer.N/A
CytotoxicityUnder conditions of the study, did not show potential toxicity to L-929 cells.
Label and LabelingMeet FDA's Requirements of 21CFR 801Meet FDA's Requirements of 21CFR 801SAME
Table 4 Safety Comparison

Analysis 2:

The proposed device has different Ultimate Elongation after aging specification to the predicate device, but all proposed device meets the specification requirements of ASTM D 5250.

8. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all specifications. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

  • ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

  • ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves

  • ASTM D 5151-06 (Reapproved 2015), Standard Test Method for Detection of Holes in Medical Gloves.

  • ASTM D5250-06 (Reapproved 2015), Standard Specification for Poly(vinyl chloride) Gloves for Medical Application.

  • ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.

    1. Clinical Test Conclusion

No clinical study is included in this submission.

{8}------------------------------------------------

10. Comparison Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.