K Number
K173511

Validate with FDA (Live)

Date Cleared
2018-07-12

(241 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ActiveCare TD-4121 Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood taken from the finger. It is intended to be used by people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. The ActiveCare TD-4121 Blood Glucose Monitoring System is comprised of the ActiveCare TD-4121 glucose meter and the ActiveCare TD-4121 blood glucose test strip. The ActiveCare TD-4121 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. The ActiveCare TD-4121 Blood Glucose Monitoring System is not intended for the diagnosis of, or screening of diabetes. It is not intended for use on neonates.

Device Description

The ActiveCare TD-4121 Blood Glucose Monitoring System is comprised of the ActiveCare TD-4121 glucose meter and the ActiveCare TD-4121 blood glucose test strip. They have been designed, tested, and proven to work together as a system to produce accurate blood glucose test results. Use only ActiveCare TD-4121 blood glucose test strip with ActiveCare TD-4121 Blood Glucose Monitoring System.

AI/ML Overview

The document describes the ActiveCare TD-4121 Blood Glucose Monitoring System, an over-the-counter device intended for quantitative measurement of glucose in fresh capillary whole blood from the finger. It is used by people with diabetes mellitus at home to monitor their diabetes control program but is not for diagnosis or screening of diabetes, nor for use on neonates.

The device's acceptance criteria are based on its performance compared to a predicate device (TD-4277 Blood Glucose Monitoring System) and its conformance to established accuracy standards for blood glucose meters.

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document implicitly references ISO 15197 for accuracy standards, as is common for blood glucose monitoring systems. Although a specific table of acceptance criteria with numerical targets is not explicitly provided in the excerpt, the typical acceptance criteria for accuracy of blood glucose meters, as per ISO 15197:2013, are:

Acceptance Criteria (typically ISO 15197:2013)Reported Device Performance (from the context of SE determination)
Accuracy relative to laboratory reference method:Demonstrated to be accurate.
- For glucose concentrations < 100 mg/dL:≥ 95% of results within ± 15 mg/dL
- For glucose concentrations ≥ 100 mg/dL:≥ 95% of results within ± 15%
- Overall, across the measuring range:≥ 99% of results within specific error boundaries (not detailed)
Repeatability/Reproducibility:Not explicitly detailed in the provided text.
User Performance (Lay User Study):Implied by "over-the-counter use" and "at home" indication.

The document states under "Similarities" that the "Accuracy" of the ActiveCare TD-4121 Blood Glucose Monitoring System is similar to the predicate device (TD-4277 Blood Glucose Monitoring System). This implies that the device met the accuracy standards expected for the predicate, which would have been established during its 510(k) clearance (K100322). The "Conclusion" also states that the ActiveCare TD-4121 and TD-4277 systems are "substantially equivalent" based on the information provided, which includes meeting performance criteria.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not explicitly stated. The document refers to "clinical and non-clinical studies" but does not give specific numbers of participants or samples.
  • Data Provenance: Not explicitly stated. However, the submitter is TaiDoc Technology Corporation, based in New Taipei City, Taiwan. It is reasonable to infer that the studies were conducted in Taiwan or by the manufacturer. The document does not specify if the studies were retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

  • Number of Experts: Not applicable. For blood glucose monitoring systems, ground truth is typically established by comparing device results to a highly accurate laboratory reference method, not by expert consensus on interpretations.
  • Qualifications of Experts: Not applicable in the traditional sense of clinical experts.

4. Adjudication Method for the Test Set:

  • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used when there's subjective interpretation by human readers (e.g., in diagnostic imaging). For quantitative measurements like blood glucose, the "ground truth" is determined by a reference device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was not done. This type of study (comparing human readers with and without AI assistance) is not relevant for a standalone blood glucose monitoring system, which does not involve human interpretation or AI assistance in this context.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Yes, a standalone performance study was implicitly done. The "clinical and non-clinical studies" mentioned involved comparing the device's measurements directly against a reference laboratory method (YSI-2300 Glucose Analyzer). This represents the performance of the device (meter and strip) as a standalone system, as it provides a quantitative measurement without human intervention for result interpretation beyond reading the display.

7. Type of Ground Truth Used:

  • Laboratory Reference Method: The ground truth was established using the YSI-2300 Glucose Analyzer. The document explicitly states: "This system is compared to the YSI-2300 Glucose Analyzer in the clinical and non-clinical studies."

8. Sample Size for the Training Set:

  • Not explicitly stated. The document describes a 510(k) submission for a device, not an AI/ML model, so the concept of a "training set" for an algorithm in the machine learning sense is not directly applicable here. The device's calibration and development would have involved internal testing and validation, but these are not typically referred to as "training sets" in the context of conventional medical devices.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable in the machine learning "training set" sense. For the device itself, the accuracy and calibration would have been established against a reference method. The document mentions that "The YSI is calibrated with NIST (SRM) 917A reference material." This indicates that the primary reference method (YSI-2300) used for evaluating the device's accuracy is traceable to a recognized national standard, ensuring the reliability of the ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 12, 2018

TaiDoc Technology Corporation Anne Kuo Scientific Reviewer B1-7F, No. 127, Wugong 2nd Rd. Wugu District New Taipei City 24888. Taiwan

Re: K173511

Trade/Device Name: ActiveCare TD-4121 Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: June 9, 2018 Received: June 13, 2018

Dear Anne Kuo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173511

Device Name

ActiveCare TD-4121 Blood Glucose Monitoring System

Indications for Use (Describe)

The ActiveCare TD-4121 Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood taken from the finger. It is intended to be used by people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program.

The ActiveCare TD-4121 Blood Glucose Monitoring System is comprised of the ActiveCare TD-4121 glucose meter and the ActiveCare TD-4121 blood glucose test strip.

The ActiveCare TD-4121 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared.

The ActiveCare TD-4121 Blood Glucose Monitoring System is not intended for the diagnosis of, or screening of diabetes. It is not intended for use on neonates.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray letters above three red hearts. Below the hearts is a gray bar with the website address "www.taidoc.com" written in white.

新北市24888万股區五工二路127號6樓 6F., No.127. Wugong 2nd Rd., Wugu Dist.

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

The assigned 510(k) number is: K173511

1. Submitter information

Company nameTaiDoc Technology Corporation
AddressB1-7F, No. 127, Wugong 2nd Rd., Wugu District,New Taipei City, 24888, Taiwan
Phone+886-2-6625-8188 #1195
Fax+886-2-6625-0288
Contact personAnne Kuo
TitleRegulatory Affairs Senior Specialist
E-mailanne.kuo@taidoc.com.twra.cert@taidoc.com.tw
Prepared dateJune 09, 2018

2. Device name

Proprietary nameActiveCare TD-4121 Blood Glucose Monitoring System
Common nameBlood Glucose Monitoring System
Product codeNBW, Blood Glucose Test System, Over-the-Counter
Classification panelClinical chemistry
ClassificationClass II
Regulation citation21 CFR §862.1345, Glucose test system

3. Predicate Device

Proprietary Name:TD-4277 Blood Glucose Monitoring System
Common Name:Blood Glucose Monitoring System
510(k) Number:K100322

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray letters above three red hearts. Below the hearts is the website address "www.taidoc.com" in white letters on a gray background.

科技股份有限公司

Intended use 4.

The ActiveCare TD-4121 Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood taken from the finger.

It is intended to be used by people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program.

The ActiveCare TD-4121 Blood Glucose Monitoring System is comprised of the ActiveCare TD-4121 glucose meter and the ActiveCare TD-4121 blood glucose test strip.

The ActiveCare TD-4121 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared.

The ActiveCare TD-4121 Blood Glucose Monitoring System is not intended for the diagnosis of, or screening of diabetes. It is not intended for use on neonates.

5. Device Description

The ActiveCare TD-4121 Blood Glucose Monitoring System is comprised of the ActiveCare TD-4121 glucose meter and the ActiveCare TD-4121 blood glucose test strip. They have been designed, tested, and proven to work together as a system to produce accurate blood glucose test results. Use only ActiveCare TD-4121 blood glucose test strip with ActiveCare TD-4121 Blood Glucose Monitoring System.

Comparison to the Predicate 6.

The modified ActiveCare TD-4121 Blood Glucose Monitoring System has the following similarities and modifications to the predicate device:

SimilaritiesModifications
1. Operating principle1. Changed meter appearance and size
2. Detection method and scientific technology2. Change battery type from 2 pieces of 3V AAA batteries to 1 Li-polymer battery.
3. Employs the same test strip3. PCB layout change
4. Operation and storage conditions4. Changed memory type from EEprom to FLASH
5. Accuracy5. Revise the contents of user manuals and labeling
6. Component materials
7. Packaging materials
8. Shelf life

TaiDoc Technology Corporation

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a document from TaiDoc Technology Corp. It lists several modifications made to a user interface, including changing from buttons to a touchscreen. The Gen/AC/PC function was modified into 8 tags, and error messages are different from predicate. Also, the manufacturing process and transmission function were removed.

7. Test principle

The blood glucose is based on the measurement of electrical current generated by the reaction of glucose with the reagent of the strip. The meter utilizes the current signal to calculate the blood glucose level.

Traceability 8.

This system is compared to the YSI-2300 Glucose Analyzer in the clinical and non-clinical studies. The YSI is calibrated with NIST (SRM) 917A reference material.

9. Conclusion

Based on the information provided in this submission, the ActiveCare TD-4121 Blood Glucose Monitoring System and TD-4277 Blood Glucose Monitoring System are substantially equivalent.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.