(28 days)
The Crystalsert Delivery System is intented to fold and deliver the Crystalens accommodating intraocular lens and other intraocular lenses identifying the Crystalsert Delivery System in their approved labeling.
The Crystalsert Delivery System is a device used for folding and delivering the Crystalens accommodating intraocular lens and other IOLs indicating use of the Crystalsert Delivery System in their approved labeling into the eye. The Crystalsert Delivery System consists of a syringe shaped body and tip with a plunger and drawer. The Crystalsert Delivery System is a sterile, disposable plastic device, with a small tubular pathway in which the lens can be placed into the eye with one continuous motion, designed for single use only.
The provided document is a 510(k) summary for the Bausch & Lomb Crystalsert Lens Delivery System (Model CI-26). It primarily focuses on demonstrating substantial equivalence to a predicate device (Model CI-28) through nonclinical bench and laboratory tests.
Here's an analysis of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly "Pass" for all tests, indicating that the device met the established benchmarks for each test. The document emphasizes that the proposed device demonstrated equivalence to the predicate device.
| Item | Test | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|---|
| Function Performance | (Before and after insertion) IOL surface & bulk homogeneity | Pass | Pass |
| (Before and after insertion) IOL lens power | Pass | Pass | |
| (Before and after insertion) IOL image quality | Pass | Pass | |
| (Before and after insertion) IOL dimensions | Pass | Pass | |
| IOL delivery outcome | Pass | Pass | |
| Damage to insertion device | Pass | Pass | |
| Coating transfer study | Pass | Pass | |
| Particle counting study | Pass | Pass | |
| Sterilization Verification | Evaluation and rationale for inclusion in "family grouping" of similar devices per ISO11135:2014 and AAMI TIR 28:2009 | Pass | Pass |
| Packaging Verification | Evaluation and rationale for inclusion in "family grouping" of similar devices. | Pass | Pass |
| Product Shelf Life | Evaluation and rationale for inclusion in "family grouping" of similar devices for a one-year shelf life. | Pass | Pass |
| Biocompatibility | Evaluation and rationale was performed per ISO 10993-1:2009 and ISO 10993-18:2005. Components are either the same part or considered to be materially equivalent to parts in the predicate device. As such, no additional biocompatibility testing was warranted or conducted. | Pass | Pass |
| EO/ECH Residuals Transfer Test | Lab testing per ISO 10993-7:2008 with protocols and acceptance criteria previously reviewed by FDA. | Pass | Pass |
| Bacterial Endotoxin Validation | Lab testing per EN ISO 13485:2012 | Pass | Pass |
2. Sample Size Used for the Test Set and Data Provenance
The document states: "Bench tests, laboratory tests, and evaluations were completed on the proposed B+L Crystalsert Lens Delivery System (Model Cl-26). No animal or clinical testing was required for this submission."
- Sample Size: The exact sample sizes for each nonclinical test are not specified in the provided text.
- Data Provenance: The data provenance is from nonclinical bench and laboratory testing. The country of origin of the data is not explicitly stated, but the submitter (Bausch + Lomb) is based in St. Louis, Missouri, USA, implying the testing likely occurred within the US or by US-affiliated labs. The nature of the testing is prospective as it was conducted for the purpose of this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Since this submission relies solely on nonclinical bench and laboratory testing, the concept of "ground truth" established by human experts in the way it applies to diagnostic or prognostic devices is not directly applicable. The "ground truth" for these tests would be defined by the technical specifications, standards (e.g., ISO, AAMI), and predefined acceptance criteria for each test. The individuals performing and evaluating these tests would be laboratory technicians and engineers, whose qualifications are not detailed but are assumed to be appropriate for conducting such tests.
4. Adjudication Method for the Test Set
As the evaluation is based on objective nonclinical tests against predetermined acceptance criteria, there is no human adjudication method in the sense of a 2+1 or 3+1 design. The results are binary (Pass/Fail) based on direct measurement and comparison to specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No animal or clinical testing was required for this submission." This type of study is relevant for diagnostic imaging or AI devices involving human interpretation, which is not the case here.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, this is not an algorithm or AI device. It's a medical device for delivering intraocular lenses. Therefore, a standalone algorithm performance study is not applicable and was not done.
7. The Type of Ground Truth Used
The "ground truth" for the nonclinical tests is based on engineering specifications, established industry standards (ISO, AAMI), and predefined acceptance criteria for each test (e.g., IOL surface homogeneity, lens power, dimensions, sterility, biocompatibility). It is objective and measurable, rather than expert consensus, pathology, or outcomes data, which are typically associated with diagnostic or therapeutic efficacy studies in clinical settings.
8. The Sample Size for the Training Set
Not Applicable. This device is not an AI or machine learning algorithm, so there is no "training set" in the conventional sense. The device itself is manufactured, and its performance is evaluated against engineering and safety standards.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. (See point 8).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 11, 2017
Bausch & Lomb, Inc Gary Rauvola Regulatory Affairs Manager 3365 Tree Ct. Industrial Blvd. St. Louis, Missouri 63122
Re: K173480
Trade/Device Name: Crystalsert Lens Delivery System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: MSS Dated: November 2, 2017 Received: November 13, 2017
Dear Gary Rauvola:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
Crystalsert Lens Delivery System CI-26
Indications for Use (Describe)
The Crystalsert Delivery System is intented to fold and deliver the Crystalens accommodating intraocular lens and other intraocular lenses identifying the Crystalsert Delivery System in their approved labeling.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This summary document has been prepared in accordance with section 21 CFR 807.92(c).
The submitter of the 510(k) is:
Gary Rauvola Regulatory Affairs Manager Bausch + Lomb 3365 Tree Court Industrial Blvd St. Louis, MO. 63122 Phone: 636-226-3402
Date Summary Prepared: November 02, 2017
1. Subject Device:
| Trade name: | Crystalsert Lens Delivery System |
|---|---|
| Common Name: | Intraocular lens Guide |
| Classification Name: | 21 CFR 886.4300 |
2. Predicate Device:
| Primary Predicate Device | K132593 | Crystalsert Lens Delivery System | CI-28 |
|---|---|---|---|
| -------------------------- | --------- | ---------------------------------- | ------- |
3. Device Description:
The Crystalsert Delivery System is a device used for folding and delivering the Crystalens accommodating intraocular lens and other IOLs indicating use of the Crystalsert Delivery System in their approved labeling into the eye. The Crystalsert Delivery System consists of a syringe shaped body and tip with a plunger and drawer. The Crystalsert Delivery System is a sterile, disposable plastic device, with a small tubular pathway in which the lens can be placed into the eye with one continuous motion, designed for single use only.
4. Indications for Use:
The Crystalsert Delivery System is intended to be used to fold and deliver the Crystalens accommodating intraocular lens and other intraocular lenses identifying the Crystalsert Delivery System in their approved labeling.
The indications for use of the proposed device are identical to the primary predicate device.
{4}------------------------------------------------
5. Brief Summary of Nonclinical Tests and Results:
Bench tests, laboratory tests, and evaluations were completed on the proposed B+L Crystalsert Lens Delivery System (Model Cl-26). No animal or clinical testing was required for this submission. Please refer to the summary table shown below.
| Item | Test | Results |
|---|---|---|
| FunctionPerformance | (Before and after insertion) IOL surface & bulkhomogeneity | Pass |
| (Before and after insertion) IOL lens power | Pass | |
| (Before and after insertion) IOL image quality | Pass | |
| (Before and after insertion) IOL dimensions | Pass | |
| IOL delivery outcome | Pass | |
| Damage to insertion device | Pass | |
| Coating transfer study | Pass | |
| Particle counting study | Pass | |
| Sterilizationverification | Evaluation and rationale for inclusion in "family grouping" ofsimilar devices per ISO11135:2014 and AAMI TIR 28:2009 | Pass |
| Packaging verification | Evaluation and rationale for inclusion in "family grouping" ofsimilar devices. | Pass |
| Product shelf life | Evaluation and rationale for inclusion in "family grouping" ofsimilar devices for a one-year shelf life. | Pass |
| Biocompatibility | Evaluation and rationale was performed per ISO 10993-1:2009 and ISO 10993-18:2005. Components are either thesame part or considered to be materially equivalent to partsin the predicate device. As such, no additionalbiocompatibility testing was warranted or conducted. | Pass |
| EO/ ECH residualstransfer test | Lab testing per ISO 10993-7:2008 with protocols andacceptance criteria previously reviewed by FDA. | Pass |
| Bacterial EndotoxinValidation pre- andpost-sterilization | Lab testing per EN ISO 13485:2012 | Pass |
Device functional performance testing all passed and demonstrates equivalence to predicate device. EO/ ECH residual transfer tests all passed. Endotoxin validation testing passed. Sterilization evaluation, packaging verification, shelf-life evaluation, biocompatibility evaluation all met requirements and demonstrated equivalence to the predicate device. All results support the conclusion that the proposed device (Model Cl-26) is as safe and as effective, and therefore equivalent to the predicate device (Model CI-28).
{5}------------------------------------------------
6. Comparative Analysis
A table comparing the proposed device to the primary predicate device is provided below.
| Table 6-1: Comparison of Crystalsert Lens Delivery System Predicate Device | ||
|---|---|---|
| (Model CI-28) to the Proposed Minor Modification (Model CI-26) | ||
| Characteristic | Primary Predicate DeviceCI-28 (K132593) | Proposed DeviceCI-26 |
| Indications for use | The Crystalsert Delivery System isintended to be used to fold and deliverthe Crystalens accommodatingintraocular lens and other intraocularlenses identifying the CrystalsertDelivery System in their approvedlabeling. | The Crystalsert Delivery System isintended to be used to fold and deliverthe Crystalens accommodatingintraocular lens and other intraocularlenses identifying the CrystalsertDelivery System in their approvedlabeling. |
| Contraindications | None | None |
| Anatomical site | Eye | Eye |
| Injector configuration | Body, drawer, plunger, bearing | Body, cartridge, drawer, plunger,bearing, haptic guide |
| Materials | Body, drawer, plunger, bearing:Polypropylene | Body: drawer, plunger, bearing:PolypropyleneCartridge: Polyimide |
| How is the deviceused | An IOL is placed into the loading areaand the drawer is closed. Thiscompresses the IOL. The plunger isadvanced until it stops at a detentposition. The distal end is filled withviscoelastic or balanced salt solutionand placed through an incision into theeye. Once the tip is in the eye, theplunger is advanced until the lens isfully expressed into the capsular bag. | An IOL is placed into the loading areaand the drawer is closed. Thiscompresses the IOL. The plunger isadvanced until it stops at a detentposition. The haptic guide is removedand distal end is filled with viscoelasticor balanced salt solution and placedthrough an incision into the eye. Oncethe tip is in the eye, the plunger isadvanced until the lens is fullyexpressed into the capsular bag. |
| Single use? | Yes | Yes |
| Is the product sterile? | Sterile | Sterile |
| How sterilized/SAL | Ethylene oxide/ 10-6 | Ethylene oxide/ 10-6 |
| Shelf life | 12 months | 12 months |
{6}------------------------------------------------
7. Conclusion
The proposed Crystalsert Lens Delivery System CI-26 is substantially equivalent to the CI-28 predicate device.
§ 886.4300 Intraocular lens guide.
(a)
Identification. An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.(b)
Classification. Class I (general controls). Except when used as folders or injectors for soft or foldable intraocular lenses, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 886.9.