K Number
K173452

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-05-09

(184 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EmboTrap® II Revascularization Device is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Device Description

The EmboTrap® II Revascularization Device is composed of a retrievable, self-expanding, Nitinol shaped section at the distal end of a tapered Nitinol shaft. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. The EmboTrap® II Revascularization Device is supplied sterile, and is intended for single-use only by physicians trained in neuro-interventional catheterization and the treatment of ischemic stroke. It is indicated for patients who are ineligible for intravenous tissue plasogen activator (IV t-PA) or who fail IV t-PA therapy.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:

Acceptance Criteria and Device Performance

Criteria CategoryAcceptance CriteriaReported Device Performance and Conclusion
BiocompatibilityAll biocompatibility tests (MTT Cytotoxicity, ISO Guinea Pig Maximization Sensitization, ISO Intracutaneous, ISO Systemic Toxicity, USP Rabbit Pyrogen, ASTM Hemolysis, Complement Activation Assay) must meet pre-assigned acceptance criteria.All biocompatibility tests completed met the pre-assigned acceptance criteria. The device was found to be non-cytotoxic, not a sensitizer, not an irritant, non-toxic, non-pyrogenic, non-hemolytic, and had acceptable levels of C3a and SC5b-9.
SterilizationDevice must be EO-sterilized to achieve a minimum SAL of 10-6.The sterilization process was successfully validated and process monitoring controls assure a minimum SAL of 10-6.
Shelf LifeShelf life studies must establish that the product and packaging remain functional and sterile for a period of 3 years.Shelf life studies established that the product and packaging remain functional and sterile for a 3-year shelf life.
In Vitro (Bench) TestingDevice must fulfill all pre-determined product performance specification requirements for: Dimensional Testing (Radial Force, Outer Cage Recovery, Device Durability/Integrity, Joint Strength), Flexibility & Kink Resistance Testing (Torque Durability, Corrosion Resistance, Radiopacity, Tip Flexibility, Re-sheathing Force), Simulated Use/Performance Testing (Coating Integrity, Particulate Evaluation). Performance must be comparable to predicate devices.The device fulfilled all pre-determined product performance specifications. Side-by-side bench testing showed the EmboTrap® II to be comparable or superior to its predicate devices in terms of performance and functionality for all tested characteristics (System Dimensions, Tip Flexibility, Re-sheathing force, Delivery and re-sheathing force during simulated use, Durability Testing, System Kink Resistance, Kink Resistance - Deployed Shaped Section, Distal Coil Tensile Testing, System Tensile Testing, Torque Testing in a Microcatheter, Torque Strength Testing, Radial Force Testing, Radiopacity, Clot retrieval and performance, and Physician usability testing).
In Vivo (Animal) StudiesUsability, effectiveness, and safety must be assessed in a swine model compared to predicate devices. Local and end organ tissue response must be comparable to predicate devices.Acute performance on Day 0 showed equivalent usability and performance to the predicate device. Histological evaluation after 3 and 28 days demonstrated comparable local and end organ tissue response between the EmboTrap® II and predicate devices.
Clinical Efficacy (Primary Endpoint)Revascularization rate (mTICI 2b or greater in the target vessel following 3 or less passes of the EmboTrap® device without rescue therapy) must be statistically significantly greater than a performance goal (NL) of 0.56 (56%). Null Hypothesis (H0: PEmboTrap ≤ 0.56) should be rejected in favor of the Alternative Hypothesis (H1: PEmboTrap > 0.56) at a one-sided Exact test for a binomial proportion at the 5% significance level.The successful revascularization rate was 72% (137/191) with a 95% confidence interval of [65%, 78%]. The p-value was <0.0001, demonstrating that the revascularization rate was statistically significantly greater than the performance goal of 56%.
Clinical Safety (Primary Endpoint)The occurrence of Symptomatic Intracerebral Hemorrhage (sICH) within 24 hours post-procedure, together with any other Serious Adverse Device Effects (SADE), excluding those already counted in sICH, must meet acceptable safety standards.The occurrence of sICH within 24h and/or SADE was 5% (9/191) with an unadjusted 95% Confidence Interval of [2%, 9%].

Study Details for Clinical Performance (ARISE II Study)

1. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: 191 patients (FDA Intention-to-Treat (ITT) Cohort).
  • Data Provenance: The study was a multi-center clinical trial. Explicit countries are not mentioned for the entire cohort, but results are broken down by "US cohort (N=94)" and "EU cohort (N=97)", indicating the data came from both the United States and European Union countries. The study was prospective in nature, as it was a clinical trial assessing the device's efficacy and safety before marketing approval.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The text states: "Angiographic data were obtained at baseline (prior to patient treatment), and again at the conclusion of each revascularization attempt, and after treatment." and "Successful achievement of the endpoint is defined as achieving an mTICI score of 2b or greater in the target vessel following 3 or less passes of the EmboTrap® device without the use of rescue therapy at any point during the procedure." The evaluation of revascularization (mTICI scores) was done by a Core Laboratory adjudicated data. While the exact number and qualifications of the experts in the Core Laboratory are not specified, a core lab typically consists of highly qualified, independent experts (e.g., neuro-interventional radiologists) blinded to treatment arms, who provide standardized, unbiased assessments of imaging.
  • "The Primary Safety Endpoint was adjudicated by the Clinical Events Committee (CEC)." Again, the exact number and specific qualifications are not provided, but a CEC typically comprises independent medical experts relevant to the study's focus (e.g., neurologists, intensivists).

3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • The text explicitly mentions "Core Laboratory adjudicated data" for the primary efficacy endpoint (mTICI scores) and "Clinical Events Committee (CEC)" for the primary safety endpoint. This implies a formal adjudication process was used. However, the exact "X+Y" method (e.g., two readers, one tie-breaker) is not specified in the provided document.

4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, this was not an MRMC comparative effectiveness study involving AI-assistance. This was a clinical trial comparing the performance of a medical device (EmboTrap® II Revascularization Device) against a pre-specified performance goal derived from other clinical trials (SWIFT and TREVO 2) of similar, predicate devices. The study did not involve AI or assess human readers' improvement with or without AI assistance.

5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No, this was not a standalone algorithm study. The study evaluated the clinical performance of a physical medical device used by human operators (physicians trained in neuro-interventional catheterization). There is no mention of an algorithm being tested in a standalone capacity.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For Efficacy: The ground truth for the primary efficacy endpoint (successful revascularization) was based on expert adjudication of angiographic data by a Core Laboratory, using modified Thrombolysis in Cerebrovascular Infarction (mTICI) scores.
  • For Safety: The ground truth for the primary safety endpoint (sICH and SADE) was established by expert adjudication from a Clinical Events Committee (CEC).
  • For Secondary Outcomes: Good clinical outcome (mRS score), mortality, and neurological deterioration were based on patient outcomes data (mRS, NIHSS scores) collected by healthcare professionals during follow-up.

7. The sample size for the training set:

  • This was a clinical trial and does not involve a "training set" in the context of an AI/algorithm study. The study evaluates the device's performance in a patient cohort (test set) based on pre-defined endpoints.

8. How the ground truth for the training set was established:

  • As stated above, this study does not involve an AI "training set." The clinical trial design assessed real-world patient outcomes with the device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 9, 2018

Neuravi Ltd. Mairsíl Claffey VP Clinical, Quality and Regulatory Affairs Block 3, Ballybrit Business Park Galway, Ireland

Re: K173452

Trade/Device Name: EmboTrap® II Revascularization Device Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: April 6, 2018 Received: April 9, 2018

Dear Mairsíl Claffey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carlos L. Pena -S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173452

Device Name

EmboTrap® II Revascularization Device

Indications for Use (Describe)

The EmboTrap® II Revascularization Device is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K173452

l. SUBMITTER:

510(k) Owner: Neuravi Ltd. Block 3, Ballybrit Business Park, Galway H91 K5YD, Ireland Contact Person: Mairsíl Claffey VP Clinical, Quality and Regulatory Affairs Tel.: +353-91-394121 Fax: +353-91-394025 E-mail: mclaffey@neuravi.com Date Prepared: May 09, 2018

II. DEVICE

Trade Name of Device: EmboTrap® II Revascularization Device Common Name of Device: Catheter, Thrombus Retriever Classification Name: 21 CFR 870.1250 - Class II Product Code: NRY

III. PREDICATE DEVICE(S)

Trevo ProVue Retriever (K122478) Solitaire™ FR Revascularization Device (K113455)

IV. DEVICE DESCRIPTION

The EmboTrap® II Revascularization Device is composed of a retrievable, self-expanding, Nitinol shaped section at the distal end of a tapered Nitinol shaft. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. The EmboTrap® II Revascularization Device is supplied sterile, and is intended for single-use only by physicians trained in neuro-interventional catheterization and the treatment of ischemic stroke. It is indicated for patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.

{4}------------------------------------------------

INDICATIONS FOR USE V.

The EmboTrap® II Revascularization Device is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

A summary of the technological characteristics of the EmboTrap® II Revascularization Device in comparison to those of the predicate device is presented below.

CharacteristicsPredicate Devices Referenced in this SubmissionSubject Device
Trevo ProVueSolitaire™ FREmboTrap® II Device
ManufacturerStrykerCovidien/MedtronicNeuravi
510(k) NumberK122478K113455K173452
ClassificationClass II (21CFR 870.1250)Same
Device ClassificationNameCatheter, Thrombus RetrieverSame
ClassificationProduct CodeNRYSame
Intended UseRestoration of blood flow by removal of thrombus from theneurovasculature (large intracranial vessel) in patientsexperiencing ischemic stroke within 8 hours of symptomonsetSame
Target PopulationPatients with symptoms of an ischemic stroke within 8 hoursof symptom onset, who are ineligible for intravenous tissueplasminogen activator (IV t-PA) or who fail IV t-PA therapyare candidates for treatmentSame
Design/TechnologicalPrinciplesSelf-expanding Nitinol shaped sectionNitinol guide-wire like shaftSame
Principal Device Materials
Distal ShapedSectionNitinolNitinolSame
Core Wire (Shaft)NitinolNitinolSame
Distal Marker/CoilPlatinum/TungstenPlatinum/IridiumPlatinum/Tungsten
ProximalMarker/Coil304 Stainless SteelPlatinum/IridiumPlatinum/Tungsten
Design Characteristics & Technology
Size(s) Offered(Retriever Diameter× Length)4×20 mm4×15 mm, 4×20 mm,6×20 mm, 6×30 mm5×21 mm, 5×33 mm
MinimumMicrocatheter ID0.021"0.021" (4 mm diameter size)0.027" (6 mm diameter size)0.021"
Additional Characteristics
How SuppliedSterile/Single UseSterile/Single UseSame
Sterilization MethodEthylene OxideEthylene OxideSame

{5}------------------------------------------------

VII. PERFORMANCE DATA Biocompatibility Testing:

The biocompatibility evaluation for the EmboTrap® II Revascularization Device was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process" as recognized by FDA (Recognition Number 2-156) and FDA Biocompatibility Guidance (Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", June 16, 2016).

Per ISO 10993-1, the EmboTrap® II device is categorized as an external communicating device with limited exposure, i.e. whose contact with circulating blood is less than 24 hours.

TestResultsConclusions
MTT Cytotoxicity StudyBased on the percentage viabilityvalues for the test article extractdilutions, the device is non-cytotoxic.Device is non-cytotoxic per theMTT Cytotoxicity Study
ISO Guinea Pig MaximizationSensitization TestTest article extracts showed noevidence of causing delayeddermal contact sensitization in theguinea pig.Device is not considered asensitizer per the Guinea PigMaximization Test
ISO Intracutaneous Study inRabbitsThe difference between the testextract overall mean score and thecorresponding control overallmean score was 1.0 or less.Device is not an irritant wheninjected intracutaneously per theISO Intracutaneous Study inRabbits
ISO Systemic Toxicity Study inMiceThere was no mortality or evidenceof systemic toxicity from theextracts injected into mice.Device is non-toxic per the ISOSystemic Toxicity Study in Mice
USP Rabbit Pyrogen Study,Material MediatedNo individual rabbit showed a risein temperature of ≥0.5°C above itsbaseline temperature and the totalmaximum temperature rise of allthree animals was withinacceptable USP limits.Device is non-pyrogenic per theMaterial Mediated Rabbit PyrogenStudy
ASTM HemolysisBoth the test article in directcontact with blood and the testarticle extract were non-hemolytic.Device is non-hemolytic per theASTM Hemolysis Test
Complement Activation AssayStudiesSC5b-9 Complement ActivationAssay StudyThe C3a and SC5b-9concentrations of the test articlesamples were acceptable.Levels of C3a and SC5b-9 wereacceptable.

The biocompatibility evaluation included the following tests:

All biocompatibility tests completed met the pre-assigned acceptance criteria as specified in the test protocol and in accordance with the requirements of the applicable standards.

{6}------------------------------------------------

Sterilization and Shelf Life:

The EmboTrap® II Revascularization Device is labeled as a single-use, sterile device, with a shelf life of 3 years. The sterilization process for the EmboTrap® II device has been successfully validated and process monitoring controls are in place to assure that the device is EO-sterilized to achieve a minimum SAL of 10-6.

Shelf life studies have been conducted for the EmboTrap® II device and establish that the product and packaging remain functional and sterile for the shelf life period of 3 years.

In Vitro (Bench) Testing:

The bench testing performed to evaluate the EmboTrap® II device included the following tests:

  • Dimensional Testing -
  • -Radial Force Testing
  • -Outer Cage Recovery Testing
  • -Device Durability/Integrity Testing
  • -Joint Strength Testing
  • Flexibility & Kink Resistance Testing -
  • -Torque Durability Testing
  • -Corrosion Resistance Testing
  • -Radiopacity Testing
  • -Tip Flexibility Testing
  • -Re-sheathing Force Testing
  • -Simulated Use/Performance Testing
  • -Coating Integrity Testing
  • -Particulate Evaluation Testing

The results of design verification testing demonstrate that the device fulfils all pre-determined product performance specification requirements. In addition, the results of the side-by-side bench testing completed show that the EmboTrap® II device is comparable to its predicate devices in terms of performance and functionality. A summary of testing performed to evaluate key descriptive characteristics between the proposed and predicate devices is presented in the table below.

{7}------------------------------------------------

Characteristic/TestMethodConclusions
System DimensionsDevice attributes (overall devicelength, length and OD of theshaped section, andforeshortening) were measuredcompared with one or morepredicate devices.The subject device dimensions arewithin the range of existingpredicate dimensions for thisdevice type. The longer overalllength of the subject device doesnot affect performance, safety oreffectiveness.
Tip FlexibilityTip flexibility was evaluatedcompared with a predicate device.The subject device demonstratedgreater flexibility than thepredicate device tested at a 90°approach angle, and equivalent tipflexibility at a 70° approach angle.
Re-sheathing forceThe force required to re-sheaththe device was evaluatedcompared with one or morepredicate devices.The force required to re-sheaththe subject device is within thesame range as that measured forthe predicate devices.
Delivery and re-sheathing forceduring simulated useThe delivery and re-sheathingforces measured during simulateduse of the subject device werecompared with the forcesmeasured for one or morepredicate devices.The delivery and re-sheathingforces of the subject device arecomparable to that of thepredicate device tested duringsimulated use.
Durability TestingDamage was evaluated afterdelivery and withdrawal of thedevice beyond the recommendednumber of passes and re-sheathings recommended in theinstructions for use compared withone or more predicate devices.Devices tested demonstrated nodamage after delivery andwithdrawal testing. Durability ofthe subject device is equivalent tothat of the predicate devicetested.
System Kink ResistanceKink resistance of the entire device(shaft and shaped section) wasevaluated when wrapped around aseries of mandrels of decreasingradii until permanent deformationwas observed or until the smallestradius was used.Kink resistance of the shapedsection was comparable to that ofthe predicate device tested.
Kink Resistance - DeployedShaped SectionKink resistance of the deployedshaped section was evaluated in aseries of bend radii within a rangeof vessel lumen diameters.Kink resistance of the deployedshaped section was equivalent toor superior than that of thepredicate device tested.
Distal Coil (Markercoil) TensileTestingThe tensile strength of the distalcoil bond was evaluated andcompared with one or morepredicate devices.The tensile strength of the distalmarker coil met acceptancecriteria and comparable to that ofthe predicate device.
System Tensile TestingThe peak tensile strength to failureof the fully assembled device wasevaluated post-simulated use andcompared with that measured forone or more predicate devices.The system tensile strength of thesubject device is equivalent to thatof the predicate device tested.
Characteristic/TestMethodConclusions
Torque Testing in a MicrocatheterDevices were tracked through a microcatheter in a tortuous path anatomical model and evaluated for damage following a number of rotations with the distal end constrained.Torque durability performance of the subject device is equivalent to that of the predicate device tested.
Torque Strength TestingThe number of rotations to separate the subject device when torqued in a representative tortuous model was compared with that measured for one or more predicate devices.Torque durability performance of the subject device is better than that of the predicate device tested.
Radial Force TestingThe radial force of the subject device was measured within a range of lumen diameters applicable to the intended vasculature and compared with the radial forces measured for one or more predicate devices.The radial force of the subject device, when tested in representative lumen sizes, is comparable to that of the predicate devices tested.
RadiopacityRadiopacity of the subject device was evaluated fluoroscopically and compared with one or more predicate devices.Radiopacity of the subject device is equivalent to that of the predicate devices tested.
Clot retrieval and performancePerformance of the subject device and its effectiveness at retrieval of firm and soft clots of various sizes in a tortuous anatomical model was evaluated and compared with one or more predicate device.The subject device effectively retrieved clot and restored flow in the test model. Device performance was rated equivalent to or better than the predicate devices tested for loading, delivery, deployment and retrieval in an in vitro tortuous path anatomical model. In vitro clot retrieval performance of the subject device is comparable to or better than that of the predicate devices tested.
Physician usability testingThe user's ability to reliably deploy and use the subject device in a tortuous benchtop model was evaluated and compared with one or more predicate devices.The subject device usability performance was rated higher than or equivalent to the predicate device tested for deliverability, deployment, retrieval, and re-sheathing.

{8}------------------------------------------------

In Vivo (Animal) Studies:

Acute and chronic animal studies have been performed to assess the usability, effectiveness and safety of the EmboTrap® II Revascularization Device compared to the predicate devices in the swine model. Acute performance evaluated on Day 0 showed that the usability and performance of the EmboTrap® II device was equivalent to that of the predicate device tested. Histological evaluation performed on treated vessels after 3 and 28 days demonstrated that the local and end organ tissue response was comparable between the EmboTrap® II device and the predicate devices tested.

{9}------------------------------------------------

Clinical Studies:

Substantial equivalence of the EmboTrap® II Revascularization Device has been demonstrated through the results of the ARISE II (Analysis of Revascularization in Ischemic Stroke with EmboTrap) Study.

The ARISE II (Analysis of Revascularization in Ischemic Stroke with EmboTrap®) clinical trial assessed the efficacy and safety of the EmboTrap® Revascularization Device against a performance goal derived from the SWIFT and TREVO 2 clinical trials. Key inclusion criteria were: Patients with a large-vessel occlusion who could be treated within 8 hours of stroke symptom onset; age 18-85; 8<NIHSS≤25; Angiographic confirmation of an ICA, MCA, MCA, M1, or M2, vertebral or basilar arteries, and IV-tPA, if used, was initiated within 3 hrs of stroke onset. Key exclusion criteria were: Stenosis, or any occlusion, in a proximal vessel that required treatment or would prevent access to the site of occlusion.

Two hundred and forty-four (244) patients signed informed consent, but 16 were identified by the treating physician as not meeting the angiographic enrolment criteria. Of the 228 identified as suitable for treatment one became too agitated for treatment and was not treated with any device. Two hundred and twenty-seven (227) patients were treated with the EmboTrap® device and included in the study. Of these, 191 patients met all inclusion and exclusion criteria as requested by FDA for Intention-to-Treat (ITT) Cohort. FDA further requested that within this FDA ITT cohort, use of rescue therapy at any point in the procedure be imputed as a failure to achieve the revascularization and good clinical outcome endpoints. Among the 36 patients excluded to meet the FDA defined ITT criteria, some had more than one deviation from the criteria. Those removed include 12 Patients with IV-tPA initiated between 3-4.5hr, 5 missing pregnancy tests, 1 ≥86 yrs, 4 with treated carotid stenosis, 1 dissection, and 1 with evidence of multiple occlusions, as well as 12 who did not meet the criteria around laboratory ranges for GFR, Glucose, Platelets and/or INR.

ARISE II Primary Endpoints:

Primary Efficacy Endpoint

The primary effectiveness endpoint was revascularization measured using modified Thrombolysis in Cerebrovascular Infarction (mTICI inclusive of the 2c rating).

Successful achievement of the endpoint is defined as achieving an mTICI score of 2b or greater in the target vessel following 3 or less passes of the EmboTrap® device without the use of rescue therapy at any point during the procedure.

Primary Safety Endpoint

The primary safety endpoint was the occurrence of Symptomatic Intracerebral hemorrhage (sICH) within 24 hours (-8/+12 hrs) post-procedure, together with any other Serious Adverse Device Effects (excluding those already counted in sICH).

{10}------------------------------------------------

ARISE II Statistical Analysis:

The Primary Efficacy endpoint was designed to evaluate performance of the EmboTrap® device compared to predicate devices and was based on a one-sided Exact test for a binomial proportion of the hypothesis:

HO: PEmboTrap ≤ NL versus H1: PEmboTrap > NL, where NL = 0.56

ARISE II Study Procedures:

Angiographic data were obtained at baseline (prior to patient treatment), and again at the conclusion of each revascularization attempt, and after treatment. Twenty-four hour postprocedure follow-up included CT or MR imaging and NIHSS examination. Follow-up at Day 7 (±12 hours) or discharge, if earlier, and at Day 90 (±14 days) included NIHSS exam and modified Rankin Scale (mRS) evaluation. Adverse events were recorded throughout the hospitalization period and at each follow-up.

ARISE II Inclusion Criteria:

  • The patient or the patient's legally authorized representative has signed and dated an Informed Consent Form.
  • Aged between 18 years and 85 years (inclusive). ●
  • A new focal disabling neurologic deficit consistent with acute cerebral ischemia.
  • NIHSS score ≥8 and ≤25.
  • Pre-ictal mRS score of 0 or 1.
  • The interventionalist estimates that at least one deployment of the EmboTrap® device can be completed within 8 hours from the onset of symptoms.
  • Patients for whom IV-tPA is indicated and who are available for treatment, are treated with IVtPA.
  • . IV-tPA, if used, was initiated within 3 hrs of stroke onset (onsettime is the last time when the patient was witnessed to beat baseline), with investigator verification that the subject has received/is receiving the correct IV t-PA dose for the estimated weight.
  • Angiographic confirmation of an occlusion of an ICA (including T or L occlusions), M1 or M2 MCA, VA, or BA with mTICI flow of 0 – 1.
  • . For strokes in the anterior circulation, the following imaging criteria should also be met:
    • MRI criterion: volume of diffusion restriction visually assessed ≤50 mL; or
    • । CT criterion: ASPECTS 6 to 10 on baseline CT or CTA-source images, or, volume of significantly lowered CBV ≤50 mL.
  • . The patient is indicated for neurothrombectomy treatment by the interventionalist and it is confirmed by diagnostic angiography that the device will be able to reach the target lesion proximally.

{11}------------------------------------------------

ARISE II Exclusion Criteria:

  • . Life expectancy likely less than 6 months.
  • Females who are pregnant or breastfeeding. ●
  • History of severe allergy to contrast medium. ●
  • . Known nickel allergy at time of treatment.
  • Known current use of cocaine at time of treatment.
  • Patient has suffered a stroke in the past 3 months. .
  • The patient presents with an NIHSS score <8 or >25 or is physician assessed as being in a clinically relevant uninterrupted coma.
  • Subject participating in another study involving an investigational device or drug.
  • Use of warfarin anticoagulation or any Novel Anticoagulant with International Normalized Ratio (INR) >3.0.
  • Platelet count <50,000/μL.
  • . Glucose <50 mg/dL.
  • Any known hemorrhagic or coagulation deficiency.
  • Unstable renal failure with serum creatinine >3.0 or Glomerular Filtration Rate (GFR) <30.
  • . Patients who have received a direct thrombin inhibitor within the last 48 hours; must have a partial thromboplastin time (PTT) less than 1.5 times the normal to be eligible.
  • . All patients with severe hypertension on presentation (SBP > 220 mmHg and/or DBP > 120 mm Hg). All patients, in whom intravenous therapy with blood pressure medications is indicated, with hypertension that remains severe and sustained despite intravenous antihypertensive therapy (SBP >185 mmHg and/ or DBP >110 mmHg).
  • Known cerebral vasculitis.
  • Rapidly improving neurological status.
  • Clinical symptoms suggestive of bilateral stroke in multiple territories. ●
  • Ongoing seizure due to stroke.
  • Evidence of active systemic infection.
  • Known cancer with metastases.
  • Computed tomography (CT) or Magnetic Resonance Imaging (MRI) evidence of recent/ fresh hemorrhage on presentation.
  • Baseline computed tomography (CT) or MRI showing mass effect or intracranial tumor (except small meningioma).
  • Suspicion of aortic dissection, presumed septic embolus, or suspicion of bacterial endocarditis.
  • Stenosis, or any occlusion, in a proximal vessel that requires treatment or prevents access to the site of occlusion.
  • Evidence of dissection in the extra or intracranial cerebral arteries.
  • . Occlusions in multiple vascular territories (e.g., bilateral anterior circulation, or anterior/posterior circulation).

{12}------------------------------------------------

ARISE II Results:

The ARISE II (Analysis of Revascularization in Ischemic Stroke with EmboTrap®) clinical trial assessed the efficacy and safety of the EmboTrap® Revascularization Device against a performance goal derived from the SWIFT and TREVO 2 clinical trials. Patients with a large-vessel occlusion who could be treated within 8 hours of stroke symptom onset were eligible for enrolment. Of the 227 patients treated with the EmboTrap® device, 191 patients met all inclusion and exclusion criteria (FDA Intention-to-Treat (ITT) Cohort).

The Primary Efficacy Endpoint was successful revascularization defined as an mTICI score of at least 2b in the target vessel, following three or less passes of the EmboTrap® device without rescue, using Core Laboratory adjudicated data. For the FDA ITT cohort, successful revascularization was modified to count any case in whom rescue therapy was used, after the EmboTrap® device, as a failure to meet the primary efficacy endpoint.

The Primary Safety Endpoint was the occurrence of symptomatic intracerebral hemorrhage (sICH) within 24 hours of the procedure, together with any serious adverse device effects (SADE), excluding those already counted as sICH. The Primary Safety Endpoint was adjudicated by the Clinical Events Committee (CEC).

ARISE II Study ResultsFDA ITT Cohort (N=191)
Primary Efficacy Endpoint
Successful Revascularization172% (137b/191)
[95% Conf. Interval]2[65%, 78%]
p-value<0.0001a
Primary Safety Endpoint
Occurrence of sICH within 24h and/or SADE5% (9c/191)
[Unadjusted 95% Conf. Interval*][2%, 9%]
Key Secondary Endpoints
Good clinical outcome353% (102b,d/191)
[Unadjusted 95% Conf. Interval*][46%, 61%]
Procedure-related mortality at 7 days post-procedure0% (0/191)
[Unadjusted 95% Conf. Interval*][0%, 2%]
All-cause mortality at 90 days post-procedure11% (21d/191)
[Unadjusted 95% Conf. Interval*][7%, 16%]
Neurological deterioration49% (18d/191)
[Unadjusted 95% Conf. Interval*][6%, 14%]
Procedure Related Serious Adverse Events (PRSAE)5% (9/191)
[Unadjusted 95% Conf. Interval*][2%, 9%]
  1. Successful revascularization was defined as achieving an mTICI rating of 2b or greater, in 3 or less passes of the study device. 2. All confidence intervals reported for binary variables are Clopper-Pearson Exact binomial confidence intervals, and those for continuous variables are based on a normal distribution assumption for the parameter of interest; 3. Modified Rankin Scale score of ≤2 at 90 days post-procedure (missing patients imputed as failure); 4. Increase of ≥4 points on the NIHSS score at 24 hours post-procedure.

*The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical inferences.

a. One-sided Exact test at the two-sided 5% significance level for the revascularization rate against a performance goal of 56%; b. Use of rescue therapy at any point in the procedure was imputed as a failure to achieve the endpoint; c. No occurrence of SADE; d. Unknown/Missing data were imputed as worst-case (this affected 4 subjects for Good Clinical Outcome, 8 subjects for Neurological Deterioration, and 5 subjects for All-Cause Mortality at 90 Days).

{13}------------------------------------------------

Primary Efficacy Endpoint by Region:

Population: FDA ITT Cohort (N=191)Primary Efficacy EndpointUS cohort(N=94)EU cohort (N=97)
Successful Revascularization1,267% (63/94)76% (74/97)
[95% Conf. Interval]3[57%, 76%][67%, 84%]
  1. Successful revascularization was defined as achieving an mTICI rating of 2b or greater, in 3 or less passes of the study device. 2. Use of rescue therapy at any point in the procedure was imputed as a failure the endpoint; 3. All confidence intervals reported for binary variables are Clopper-Pearson Exact binomial confidence intervals, and those for continuous variables are based on a normal distribution assumption for the parameter of interest.

ARISE II Summary of Serious Adverse Events

One hundred and seventeen (117) SAE occurred within the FDA ITT Cohort. The most common (incidence of ≥1%) SAE are presented below.

Serious Adverse Events occurring at an incidence of 1% or more in any population
MedDRA Preferred Term% pts w/ SAE [Number of SAE]FDA ITT Cohort (N=191)
Cardiac disorders
Myocardial infarction1.57% [3]
Atrial fibrillation1.05% [2]
Cardiac failure1.05% [2]
Gastrointestinal disorders
Gastrointestinal haemorrhage1.05% [2]
Infections and infestations
Septic shock2.09% [4]
Urinary tract infection1.57% [3]
Pneumonia1.57% [3]
Sepsis1.05% [2]
Nervous system disorders
Ischaemic stroke6.81% [13]
Haemorrhagic transformation stroke6.28% [12]
Brain oedema3.66% [7]
Neurological decompensation2.09% [4]
Seizure1.57% [3]
Subarachnoid haemorrhage1.57% [3]
Carotid artery stenosis1.05% [2]
Respiratory, thoracic and mediastinal disorders
Pneumonia aspiration1.57% [3]
Respiratory distress1.05% [2]
Bronchospasm1.05% [2]
Respiratory failure1.05% [2]
Surgical and medical procedures
Coronary arterial stent insertion1.05% [2]
Vascular disorders
Vessel perforation1.05% [2]
Deep vein thrombosis1.05% [2]

{14}------------------------------------------------

Reasons for exclusion from treatment:

Two-hundred and forty-four (244) patients were consented for enrolment in the ARISE II Study, of which, 17 were excluded from treatment for the reasons outlined below.

Angiographic reasons for exclusion from treatmentNumber of Subjects(N=16)*
No angiographic confirmation of an occlusion of an ICA (including T or L occlusions), M1 orM2 MCA, VA, or BA with mTICI flow of 0 – 1.7
The patient is indicated for neurothrombectomy treatment by the interventionalist and itis confirmed by diagnostic angiography that the device will not be able to reach the targetlesion proximally.7
Stenosis, or any occlusion, in a proximal vessel that requires treatment or prevents accessto the site of occlusion.4
Evidence of dissection in the extra or intracranial cerebral arteries.2
Occlusions in multiple vascular territories (e.g., bilateral anterior circulation, oranterior/posterior circulation).2
Other reasons for exclusion from treatmentNumber of Subjects(N=1)
Severe agitation1

*7 patients failed 2 angiographic criteria.

In addition to the pre-specified performance goal evaluation of the ARISE II study, FDA also considered the ARISE II revascularization rates (mTICI scores) relative to more recent studies of stent retriever devices in acute ischemic stroke patients as additional supportive information in the substantial equivalence assessment.

CONCLUSIONS

Non-clinical and clinical studies demonstrate that the EmboTrap® II Revascularization Device is substantially equivalent to the predicate devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).