(55 days)
The Anterior Cervical Plate System is indicated for anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the following indications; degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion.
The Anterior Cervical Plate System consists of self-tapping screws and plates. Screws are available in a variety of diameter-length combination. Plates are available in a variety of lengths.
This document is a 510(k) premarket notification for a medical device called the Anterior Cervical Plate System. It is focused on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria in a clinical study. Therefore, much of the requested information regarding AI/algorithm performance, detailed study design, and expert review is not applicable.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (from predicate device standards) | Reported Device Performance (Subject Device) |
|---|---|
| Mechanical performance according to ASTM F1717 for Static compression bending, torsion, and dynamic compression bending. | The mechanical test results demonstrated that the Subject device is substantially equivalent to the predicate devices when tested for static compression bending, torsion, and dynamic compression bending of the worst-case construct per ASTM F1717. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not applicable. This was a bench testing study, not a clinical study with patient data. The "sample size" would refer to the number of physical constructs tested, but this specific number is not provided, only that the "worst case" construct was tested.
- Data Provenance: Not applicable. This refers to bench test results in a lab environment.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This was a mechanical bench test against a standard, not a study requiring expert clinical ground truth.
4. Adjudication method for the test set
- Not applicable. No clinical test set to adjudicate.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI/algorithm-driven device. It is a physical implant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an AI/algorithm-driven device.
7. The type of ground truth used
- Mechanical Performance Standards: The ground truth for performance was established by recognized mechanical testing standards (ASTM F1717) and comparison to the predicate device's established performance.
8. The sample size for the training set
- Not applicable. This is not an AI/algorithm-driven device. There is no "training set."
9. How the ground truth for the training set was established
- Not applicable.
Summary of the Study:
The study proving the device meets the acceptance criteria was a bench testing study. The purpose of this study was to demonstrate substantial equivalence of the Anterior Cervical Plate System to a legally marketed predicate device (Synthes Anterior CSLP System, K030866).
- Acceptance Criteria for Substantial Equivalence: The primary acceptance criterion was that the mechanical performance of the subject device, when tested according to ASTM F1717 (covering static compression bending, torsion, and dynamic compression bending), would be comparable or equivalent to the predicate device.
- Device Performance: The study reported that "The mechanical test results demonstrated that the Subject device is substantially equivalent to the predicate devices."
- Test Conduct: The testing was performed on the "worst case Anterior Cervical Plate construct."
- Ground Truth: The "ground truth" for the mechanical performance comparison was the established performance characteristics of the predicate device and the requirements of the ASTM F1717 standard.
- Purpose: The overall conclusion was that the Anterior Cervical Plate System has the same intended use, identical indications for use, similar technological characteristics and materials as its predicate(s), and minor differences do not raise new safety or efficacy concerns after bench testing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 21, 2017 Evolution Spine, LLC Mr. Douglas Davis Chief Operating Officer 2300 N. Haskell Dallas, Texas 75204
Re: K173375
Trade/Device Name: Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: September 15, 2017 Received: October 27, 2017
Dear Mr. Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vincent J. Devlin -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K173375
Device Name Anterior Cervical Plate System
Indications for Use (Describe)
The Anterior Cervical Plate System is indicated for anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the following indications; degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion.
Type of Use (Select one or both, as applicable)
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Anterior Cervical Plate SystemTraditional 510(k) | K173375 - page 1 of 2 | |
|---|---|---|
| Anterior Cervical Plate System | ||
| Premarket Notification 510(k) Summary | ||
| DATE PREPARED | September 1st, 2017 | |
| MANUFACTURER | Evolution Spine | |
| 2300 North Haskell Avenue | ||
| Dallas, TX 75204 | ||
| CONTACT PERSON | Douglas Davis | |
| Vice President of Product Development | ||
| 214.228-6252 | ||
| ddavis@evolutionspine.com | ||
| PANEL CODE | Orthopaedics/87 | |
| CLASSIFICATION NAME | KWQ 888.3060 – Spinal Intervertebral Body Fixation Orthosis, Cervical | |
| CLASS | Class II | |
| COMMON NAME | Anterior Cervical Plate System (KWQ) | |
| TRADE NAME | Anterior Cervical Plate System | |
| PREDICATE DEVICES | Synthes Anterior CSLP System manufactured by Synthes Spine(K030866) identified as the primary predicate device | |
| Anodyne Anterior Cervical Plate System manufactured by Corelink(K121514) identified as an additional predicate device |
DEVICE DESCRIPTION
The Anterior Cervical Plate System consists of self-tapping screws and plates. Screws are available in a variety of diameter-length combination. Plates are available in a variety of lengths.
INDICATIONS FOR USE
The Anterior Cervical Plate System is indicated for anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the treatment of the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e.,
{4}------------------------------------------------
fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion.
MATERIALS
The Anterior Cervical Plate system is manufactured from titanium alloy (Ti-6Al-4V) as described by ASTM F136.
TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
Documentation was provided to demonstrate that the Subject Anterior Cervical Plate System is substantially equivalent to the predicate Synthes Anterior CSLP System (K030866). The Subject device is substantially equivalent to the predicate device in intended use, indications for use, materials, technological characteristics, and labelling.
PERFORMANCE DATA
Static compression bending and torsion, and dynamic compression bending of the worst case Anterior Cervical Plate construct was performed according to ASTM F1717. The mechanical test results demonstrated that the Subject device is substantially equivalent to the predicate devices.
CONCLUSION
The Anterior Cervical Plate System and its predicate(s) have the same intended use, to provide mechanical stability in the cervical disc space to facilitate biologic fusion. The indications for use of the Subject device are identical to those of the predicate device. Minor differences between the Subject and predicate devices do not raise any new questions of safety or efficacy. Bench testing demonstrated that the differences do not adversely impact device mechanical performance. Based on the intended use, indications for use, technological characteristics, materials, and comparison to the predicate devices, the Subject Anterior Cervical Plate System has been shown to be substantially equivalent to legally marketed predicate devices.
N/A