K Number
K173330

Validate with FDA (Live)

Device Name
PanNAT STEC Test
Manufacturer
Date Cleared
2018-06-01

(221 days)

Product Code
Regulation Number
866.3990
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Micronics PanNAT STEC Test is a qualitative, in vitro nucleic acid amplification-based tor the simultaneous detection and identification of the stx1 and stx2 Shiga toxin genes and the O-antigen gene cluster of E.coli O157 (fc/). Testing is performed in a unitized cartridge on the PanNAT System on soft to diarrheal unpreserved or Cary-Blair preserved stool specimens from individuals with signs and symptoms of gastrointestinal infection. The PanNAT STEC Test is intended for use, in conjunction with clinical presentation, laboratory findings and epidemiological risk factors, as an aid in detection of specific Shiga-toxin expressing strains of E. coli ("STEC") from patients with diarrhea.

The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Positive PanNAT STEC Test results do not rule out the potential for coinfection with other pathogens that are not detected by this device and may not be indicative of the sole or definitive cause of patient illness. Negative PanNAT STEC Test results in the context of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Device Description

The PanNAT System consists of the instrument with onboard software and 9-inch touchscreen used to process PanNAT STEC Test cartridges. The instrument automates all steps of the assay after sample addition to the test cartridge and insertion into the instrument; including DNA purification, nucleic acid amplification and detection of the target nucleic acid sequences using qualitative real-time PCR.

The system is a portable device that is powered by an external mains supply with a voltage range of 100-240VAC and a frequency range of 50-60 Hz. An onboard battery is included to provide power to the instrument for up to one hour in the event that mains power supply is interrupted. The instrument includes a pneumatic subsystem, a thermal control unit, fluorescent optical detectors, and software needed to process a test cartridge,

The PanNAT STEC Test cartridge is a unique, single use, disposable device in which all test reagents and controls are incorporated and all steps of the assay are performed. The PanNAT STEC Test cartridge uses a PanNAT Sample Transfer Pack accessory that contains a flocked swab, a prefilled Sample Buffer Tube and an Adaptor Cap.

There are three integrated controls for the PanNAT STEC Test: an endogenous human DNA internal process control that is coextracted and coamplified with the target nucleic acids, a negative control, and a positive control. These controls are performed automatically and do not require any action from the operator. External quality controls for the PanNAT STEC Test are also commercially available. Alternatively, laboratories may prepare their own controls.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study that proves the device meets them:

The document describes the PanNAT STEC Test, a qualitative, in vitro nucleic acid amplification-based test for detecting stx1, stx2 Shiga toxin genes, and the E.coli O157 O-antigen gene cluster.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a numerical or percentage format next to performance metrics. However, the results presented in the analytical and clinical performance studies implicitly serve as the demonstration of the device's ability to meet expected performance standards for diagnostic assays. I will summarize the key performance metrics from the studies provided.

Performance MetricAcceptance Criteria (Implied / Demonstrated Performance)Reported Device Performance
Reproducibility (Overall Agreement %)High agreement expected across sites, lots, and operators for both positive and negative samples.Overall: 97.8%. Site A: 98.8%. Site B: 99.2%. Site C: 95.3%. (Lower for "All Low" panel member at Site C: 76.7%)
Limit of Detection (LoD)Consistent and reliable detection at specified concentrations.Unpreserved stool: 1.00E+06 CFU/mL for EDL933 and 3007-85 (stx1, stx2, O157). Cary-Blair preserved stool: 1.75E+06 CFU/mL for EDL933 and 3007-85. (Confirmed with ≥95% positive replicates.)
Analytical Reactivity (Sensitivity)Detection of targeted Shiga toxin genes (stx1, stx2) and E. coli O157 in various strains, including subtypes. A high percentage of expected results.stx1: 92% (24/26 strains, with 2 false negatives for stx1d variants due to weak detection). stx2: 96% (25/26 strains, with 1 false negative for an stx2b-like variant due to extensive mismatches). O157 O-antigen (fcl): 100% (26/26 strains).
Analytical Specificity (Cross-Reactivity)No cross-reactivity with common microbial flora or other potential pathogens.Bacteria: 100% agreement with expected negative results for 37 strains (except Shigella dysenteriae which is expected to be detected due to sequence homology with stxA gene). Viruses: 100% agreement for 8 viruses. Yeast/Parasites: 100% agreement for 4 organisms.
Analytical Specificity (Microbial Interference)No inhibition of target detection in the presence of other organisms.100% agreement with expected positive results for stx1, stx2, and O157 targets in the presence of 49 non-target organisms (bacteria, viruses, yeast, parasites). No competitive inhibition observed.
Analytical Specificity (Interfering Substances)No significant interference from common stool contaminants or substances.Minimal interference: Gaviscon® and Imodium AD® caused one false negative for all three targets in one of three replicates, and stearic acid caused one false negative for all three targets in one of three replicates, when spiked into a positive sample. Otherwise, no interference with 25 substances.
Carry Over / Cross ContaminationNo false positives due to run-to-run contamination.No carryover or cross-contamination observed across three instruments over five rounds of alternating high positive and negative samples. (1 invalid result in 15 negative samples, but no false positive).
Clinical Performance (PPA - Prospective)High positive percent agreement for detected analytes.stx1 (preserved): 100.0% (4/4). stx2 (preserved): 100.0% (5/5). stx2 (unpreserved): 100.0% (2/2). O157 (preserved): 100.0% (3/3). O157 (unpreserved): 100.0% (2/2).
Clinical Performance (NPA - Prospective)High negative percent agreement for detected analytes.stx1 (preserved): 99.3% (1172/1180). stx1 (unpreserved): 100.0% (96/96). stx2 (preserved): 99.4% (1172/1179). stx2 (unpreserved): 98.9% (93/94). O157 (preserved): 99.8% (1179/1181). O157 (unpreserved): 100.0% (94/94).
Clinical Performance (PPA - Retrospective)High positive percent agreement, especially for samples with low prevalence in prospective studies.stx1 (unpreserved): 88.5% (23/26). stx2 (unpreserved): 90.0% (27/30). O157 (unpreserved): 89.7% (26/29).
Clinical Performance (NPA - Retrospective)High negative percent agreement.stx1 (unpreserved): 97.4% (75/77). stx2 (unpreserved): 98.6% (72/73). O157 (unpreserved): 98.6% (72/73).

2. Sample Size Used for the Test Set and Data Provenance

Test Set for Reproducibility:

  • Sample Size: 773 observations overall from 90 panel members (10 for each of 9 panel members evaluated in triplicate over 6+ days at each of 3 sites). Specific counts per panel member vary from 82 to 88.
  • Data Provenance: The study was conducted at three sites (2 external and 1 internal). This suggests a mix of internal testing and external validation, likely within the US given the FDA submission. The data appears to be prospective in the sense of being generated specifically for the study under controlled conditions (testing coded panel members).

Test Set for Clinical Performance:

  • Prospective Specimens:
    • Sample Size: 1331 prospective specimens enrolled, 1301 evaluated after initial exclusions, and 1280 (1184 Cary-Blair preserved and 96 unpreserved) qualified for agreement analysis.
    • Data Provenance: Collected at six clinical sites. Specimens were from pediatric and adult patients as part of routine patient care. This is prospective regional data.
  • Retrospective Specimens:
    • Sample Size: 114 specimens initially enrolled, leading to 103 evaluable results.
    • Data Provenance: Frozen, retrospective samples. Likely collected from a broader range of past cases. This is retrospective data.
  • Contrived Specimens:
    • Sample Size: 16 contrived specimens.
    • Data Provenance: Unpreserved, spiked with enumerated stocks of specific E. coli strains and tested at clinical sites. These are contrived samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the "number of experts" or their specific qualifications (e.g., "radiologist with 10 years of experience") for establishing the ground truth.

  • For Reproducibility: The ground truth was established by the composition of the coded 9-member panel, which consisted of E. coli strains with known analyte presence (stx1, stx2, O157) at low or moderate concentrations, or negative for all. This is a defined ground truth based on strain characteristics.
  • For Analytical Reactivity, Specificity, Interference: The ground truth for these studies was based on the known characteristics of the microbial strains, viruses, yeast, parasites, and interfering substances used in the experiments.
  • For Clinical Performance:
    • Prospective study: The ground truth was established using "standard of care testing (comparator method)" which included "an immunoassay for detection of Shiga toxins 1 and 2 following enrichment culture, and a SMAC plate for presumptive detection of O157, confirmed by latex agglutination and biochemical analysis."
    • Discordant analysis: Performed by an "independent molecular reference laboratory" using "alternative PCR and bidirectional (BDS) sequencing methods." This implies expert molecular biologists or laboratory staff established the final ground truth for discrepant cases.
    • Retrospective study: The ground truth was established by "alternative PCR (nested)/bidirectional sequencing." This also implies expert molecular biologists or laboratory staff.
    • The document doesn't specify the exact number of experts or their years of experience for these reference methods.

4. Adjudication Method for the Test Set

  • For Reproducibility and Analytical Studies: No explicit "adjudication method" beyond the inherent known characteristics of the panel members and strains. For the quantitative results in reproducibility, only valid results were used for mean/SD/CV calculation.
  • For Clinical Performance:
    • Prospective study: A discordant analysis was performed on discrepant samples by an independent molecular reference laboratory. This serves as an adjudication method, where the reference method's result is used to resolve discrepancies between the device and the initial standard of care comparator.
    • Retrospective study: The comparator method itself was an "alternative PCR (nested)/bidirectional sequencing," which acted as the reference standard, therefore, no separate adjudication process is described for this stage beyond the initial testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done.

This device is a diagnostic test (nucleic acid amplification-based assay) for pathogen detection, not an imaging device or one that relies on human interpretation for its primary output. Therefore, measuring "how much human readers improve with AI vs without AI assistance" is not applicable to this type of device. The study evaluates the device's performance against comparator methods and predefined ground truths, not assistance to human interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance study was done.

The entire performance evaluation described focuses on the PanNAT STEC Test system (instrument and cartridges) operating autonomously after sample loading. The "PanNAT System automates all steps of the assay after sample addition to the test cartridge and insertion into the instrument." The results are displayed on the screen and stored in the instrument database. There is no human interpretation or human-in-the-loop component in the detection and reporting of results for this assay. The human interaction is limited to preparing and loading the sample.

7. The Type of Ground Truth Used

  • Analytical Studies (Reproducibility, LoD, Reactivity, Specificity, Interference, Carryover):
    • Defined Ground Truth: Known concentrations of specific bacterial strains, or known presence/absence of target analytes in well-characterized cultures, viruses, yeast, parasites, and chemical substances.
  • Clinical Performance Studies (Prospective and Retrospective):
    • Prospective: Initial ground truth established by "standard of care testing" (immunoassay + SMAC/latex agglutination/biochemical). Discrepant results were then adjudicated by an "independent molecular reference laboratory" using "alternative PCR and bidirectional (BDS) sequencing methods."
    • Retrospective: Ground truth established by "alternative PCR (nested)/bidirectional sequencing."
    • Contrived: Defined ground truth based on spiking known concentrations of specific strains into negative matrix.

8. The Sample Size for the Training Set

The document describes no separate "training set" in the context of an AI/ML algorithm that would undergo a distinct training phase. This device is a real-time PCR assay, not a machine learning model that is "trained" on data in the traditional sense. The data presented are for performance validation of the established PCR assay.

9. How the Ground Truth for the Training Set Was Established

As noted above, there is no explicit "training set" for an AI/ML algorithm. The ground truth for the analytical studies and clinical studies (which act as validation data) was established as described in section 7.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 1, 2018

Micronics, Inc Karen Hedine President 8464 154th Ave NE Redmond, Washington 98052

Re: K173330

Trade/Device Name: PanNAT STEC Test Regulation Number: 21 CFR 866.3990 Regulation Name: Gastrointestinal microorganism multiplex nucleic acid-based assay Regulatory Class: Class II Product Code: PCI, PCH, OOI Dated: April 27, 2018 Received: May 1, 2018

Dear Karen Hedine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173330

Device Name PanNAT STEC Test

Indications for Use (Describe)

The Micronics PanNAT STEC Test is a qualitative, in vitro nucleic acid amplification-based tor the simultaneous detection and identification of the stx1 and stx2 Shiga toxin genes and the O-antigen gene cluster of E.coli O157 (fc/). Testing is performed in a unitized cartridge on the PanNAT System on soft to diarrheal unpreserved or Cary-Blair preserved stool specimens from individuals with signs and symptoms of gastrointestinal infection. The PanNAT STEC Test is intended for use, in conjunction with clinical presentation, laboratory findings and epidemiological risk factors, as an aid in detection of specific Shiga-toxin expressing strains of E. coli ("STEC") from patients with diarrhea.

The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Positive PanNAT STEC Test results do not rule out the potential for coinfection with other pathogens that are not detected by this device and may not be indicative of the sole or definitive cause of patient illness. Negative PanNAT STEC Test results in the context of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5. 510(k) Summary

5.1 Applicant

Micronics, Inc. 8463 154th Avenue NE Redmond. WA 98052 Phone: (425) 895-9197 Fax: (425) 895-1183

Contact Person:

Karen Hedine President Phone: (425) 896-3026 Fax: (425) 895-1183 Email: khedine@micronics.net

Consultant:

Gail E. Radcliffe, PhD 2311 Fairbanks St. West Boylston, MA 01583 Phone: (508) 835-1688 Email: gradcliffe@RMDCI.com

Date of Preparation: May 22, 2018

5.2 Device

Trade Name: PanNAT® STEC Test Common name: Gastrointestinal Bacterial Panel Multiplex Nucleic Acid-Based Assay System Regulation number: 866.3990, Gastrointestinal microorganism multiplex nucleic acidbased assay Product Code: PCI, PCH, OOI Class: 2

5.3 Predicate Device

BD Max Enteric Bacterial Panel (K140111)

5.4 Intended Use

The Micronics PanNAT STEC Test is a qualitative, in vitro nucleic acid amplificationbased test intended for the simultaneous detection and identification of the stx2 Shiga toxin genes and the O-antigen gene cluster of E.coli O157 (fcl). Testing is performed in a unitized cartridge on the PanNAT System on soft to diarrheal unpreserved or Cary-Blair preserved stool specimens from individuals with signs and symptoms of

{4}------------------------------------------------

gastrointestinal infection. The PanNAT STEC Test is intended for use, in conjunction with clinical presentation, laboratory findings and epidemiological risk factors, as an aid in detection of specific Shiga-toxin expressing strains of E. coli ("STEC") from patients with diarrhea.

The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Positive PanNAT STEC Test results do not rule out the potential for coinfection with other pathogens that are not detected by this device and may not be indicative of the sole or definitive cause of patient illness. Negative PanNAT STEC Test results in the context of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

5.5 Device Description

The PanNAT System consists of the instrument with onboard software and 9-inch touchscreen used to process PanNAT STEC Test cartridges. The instrument automates all steps of the assay after sample addition to the test cartridge and insertion into the instrument; including DNA purification, nucleic acid amplification and detection of the target nucleic acid sequences using qualitative real-time PCR.

The system is a portable device that is powered by an external mains supply with a voltage range of 100-240VAC and a frequency range of 50-60 Hz. An onboard battery is included to provide power to the instrument for up to one hour in the event that mains power supply is interrupted. The instrument includes a pneumatic subsystem, a thermal control unit, fluorescent optical detectors, and software needed to process a test cartridge,

The PanNAT STEC Test cartridge is a unique, single use, disposable device in which all test reagents and controls are incorporated and all steps of the assay are performed. The PanNAT STEC Test cartridge uses a PanNAT Sample Transfer Pack accessory that contains a flocked swab, a prefilled Sample Buffer Tube and an Adaptor Cap.

There are three integrated controls for the PanNAT STEC Test: an endogenous human DNA internal process control that is coextracted and coamplified with the target nucleic acids, a negative control, and a positive control. These controls are performed automatically and do not require any action from the operator. External quality controls for the PanNAT STEC Test are also commercially available. Alternatively, laboratories may prepare their own controls.

5.6 Principle of Operation

Using the swab from the Sample Transfer Pack, the operator transfers a specimen (either soft to diarrheal unpreserved stool specimen or Cary-Blair preserved stool specimen) to the Sample Buffer Tube, which is then sealed using the Adaptor Cap. The Sample Buffer Tube is then attached to the sample inlet port of the cartridge by means of the attached Adaptor Cap; fully containing all fluids within the cartridge-tube assembly. Once the test cartridge with attached Sample Buffer Tube is inserted into the instrument, the cartridge barcode is scanned and the test starts automatically.

Utilizing proprietary microfluidic design, and under control of the instrument, sample and reagents are moved through the cartridge to complete all of the steps of the procedure.

{5}------------------------------------------------

DNA is extracted from lysed cells, captured on silica membranes, and eluted to accomplish purification. DNA sequences specific to stx1, stx2 and O157, together with the internal process control, are amplified and then detected by fluorescent probe hybridization for qualitative real-time PCR detection.

When the test is completed, sample results are displayed on the screen and stored in the instrument database. Test results can be exported to a USB flash drive or printer.

5.7 Substantial Equivalence

A comparison of similarities and differences between the PanNAT STEC Test and the predicate device is provided in Table 5-1.

FeaturesPanNAT STEC TestPredicate: BD Max™ Enteric Bacterial Panel (K140111)
Similarities
Intended UseThe Micronics PanNAT STEC Test is a qualitative, in vitro nucleic acid amplification-based test intended for the simultaneous detection and identification of the stx1 and stx2 Shiga toxin genes and the O-antigen gene cluster of E.coli O157 ( fcl ). Testing is performed in a unitized cartridge on the PanNAT System on soft to diarrheal unpreserved or Cary-Blair preserved stool specimens from individuals with signs and symptoms of gastrointestinal infection. The PanNAT STEC Test is intended for use in conjunction with clinical presentation, laboratory findings and epidemiological risk factors, as an aid in detection of specific Shiga-toxin expressing strains of E. coli (“STEC”) from patients with diarrhea.The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Positive PanNAT STEC Test results do not rule out the potential for coinfection with other pathogens that are not detected by this device and may not be indicative of the sole or definitive cause of patient illness Negative PanNAT STEC Test results in the context of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerativeThe BD MAX™ Enteric Bacterial Panel performed on the BD MAX™ System is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric bacterial pathogens. The BD MAX™ Enteric Bacterial Panel detects nucleic acids from:Salmonella spp.Campylobacter ssp. (jejuni and coli)Shigella spp./Enteroinvasive E. coli (EIEC)Shiga Toxin 1 ( stx1 )/Shiga toxin 2 ( stx2 ) genes (found in Siga toxin-producing E. coli as well as Shigella dysenteriae , which can possess a Shiga toxin gene ( stx ) that is identical to the stx1 gene of STEC.Testing is performed on unpreserved soft to diarrheal unpreserved stool specimens or Cary Blair preserved stool symptomatic patients with suspected acute gastroenteritis, enteritis or colitis. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of SpaO , a Campylobacter specific tuf gene sequence, ipaH and stx1/stx2 . The test utilizes fluorogenic sequence-specific hybridization probes for detection of the amplified DNA.This test is intended for use, in conjunction with clinical presentation, laboratory findings and epidemiological

Table 5-1. Substantial Equivalence Table Comparing PanNAT STEC Test to Predicate

{6}------------------------------------------------

colitis, irritable bowel syndrome, orCrohn's disease.information, as an aid in the differential diagnosis of Salmonella, Shigella/EIEC, Campylobacter and Shiga toxin-producing E. coli (STEC) infections.Results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
Specimen TypeUnpreserved and Cary-Blair preserved stoolSame
Assay FormatReal-time PCR amplification:Fluorogenic target-specific hybridizationSame
Detection of stx1/stx2Presence of stx1 and stx2 genes specific to Shiga toxin-producing organismsSame
Sample PreparationAutomated by the PanNAT SystemAutomated by the BD MAXTM System
Detection probesFluorescent ProbesMGB PleiadesTM ProbeFluorescent ProbesTaqMan® Probe
Assay controlsInternal Process Control and Integrated positive and negative controlsSample Processing Control (SPC)
Differences
Detection of E. colifcl gene for E. coli O157ipaH gene for E coli (EIEC)
Analysis platformPanNAT® SystemBD MAXTM System

5.8 Analytical Performance Studies

Precision/ Reproducibility

Reproducibility of the PanNAT STEC Test was evaluated at three sites (2 external and 1 internal) utilizing 3 test cartridge lots, 6 operators (2 per site), and 13 PanNAT instruments. A total of 90 samples, 10 for each of 9 panel members were evaluated in triplicate over a minimum of 6 days at each site.

The coded 9-member panel consisted of E. coli strains containing a single analyte at low or moderate concentrations of stxl, stx2 or 0157, a strain positive for all analytes at low or moderate concentrations and a strain negative for all 3 analytes.

Agreement results by panel member by site are shown in Table 5-2. Each site used a different lot of cartridges; therefore lot-to-lot variability is demonstrated by site. In Table 5-3, agreement for each analyte is shown by concentration and by site/lot.

{7}------------------------------------------------

AnalytePanelPanel 1All NegPanel 2Low stx1Panel 3Med stx1Panel 4Low stx2Panel 5Med stx2Panel 6Low O157Panel 7Med O157Panel 8All LowPanel 9All MedTotal
Overall
%Agmt100.00%97.60%98.8%97.70%100.00%98.90%100.00%88.60%98.90%97.8%
(95% CI)(95.6,100)(91.8,99.4)(93.4,99.8)(91.9, 99.4)(95.8, 100)(93.8, 99.8)(95.8, 100)(80.3, 93.7)(93.8, 99.8)
Total (n)838582868788878887773
Agmt (n)838381848787877886756
Site A
% Agmt100.0%100.0%100.0%100.0%100.0%100.0%100.0%93.3%96.3%98.8%
(95% CI)(87.9,100)(87.9,100)(87.5,100)(87.5,100)(87.9,100)(88.6,100)(88.6,100)(78.7,98.2)(81.7,99.3)
Total (n)282827272830303027255
Agmt (n)282827272830302826252
Site B
% Agmt100.0%100.0%96.7%100.0%100.0%100.0%100.0%96.4%100.0%99.2%
(95% CI)(88.3,100)(88.3,100)(83.3,99.4)(88.3,100)(88.3,100)(88.3,100)(87.9,100)(82.3,99.4)(88.6,100)
Total (n)292930292929282830261
Agmt (n)292929292929282730259
Site C
% Agmt100.0%92.90%100.0%93.30%100.0%96.60%100.0%76.70%*100.0%95.3%
(95% CI)(87.1, 100)(77.3, 98.0)(87.1, 100)(78.7, 98.2)(88.6, 100)(82.8, 99.4)(88.3, 100)(59.1, 88.2)(88.6, 100)
Total (n)262825303029293030257
Agmt (n)262625283028292330245

Table 5-2 Agreement hy Panel Member by Site

*83/90 possible results for Panel Member #8 were concordant (92% [80.0, 100]). The individual test results for this panel member at Site C were: stx 1: 93.3% (80, 100), 28/30 concordant results; stx 2: 86.6% (70, 90), 26/30 concordant results; fcl: 96.7% (80, 100), 29/30 concordant results. For all of the seven Panel Member #8 test results that did not demonstrate with the expected result, only one target was missed at each test point. None of these missed 2 targets and none missed 3 targets. Further, all three target concentrations in Panel Member #8 were at the LoD of the assay (1.00E+06 CFUmL) and as such are not expected to be detected 100% of the time. When looking at the individual targets, the detection rate ranged from 86.6%-96.7%.

Table 5-3. Agreement by Concentration, Analyte and Site/Lot for Positive Panel Members

Analytestx1stx2O157
NegLowMedNegLowMedNegLowMed
Total^%Agmt100.0%97.1%98.8%100.0%96.0%100.0%100.0%97.2%99.4%
(95% CI)(98.9,100)(93.4,98.8)(95.8,99.7)(98.9,100)(91.9,98.0)(97.8,100)(98.9,100)(93.5,98.8)(96.8,99.9)
Total (n)348173169342174174340176174
Agmt (n)348168167342167174340171173
Site A%Agmt100.0%100.0%98.1%100.0%100.0%100.0%100.0%96.7%98.2%
Lot 1(95% CI)(96.8,100)(93.8,100)(90.2,99.7)(96.8,100)(93.7,100)(93.5,100)(96.6,100)(88.6,99.1)(90.7,99.7)
Total (n)115585411557551106057
Agmt (n)115585311557551105856
Site B%Agmt100.0%98.2%98.3%100.0%98.2%100.0%100.0%98.2%100.0%
Lot 2(95% CI)(96.8,100)(90.7,99.7)(91.1,99.7)(96.8,100)(90.7,99.7)(93.9,100)(96.8,100)(90.7,99.7)(93.8,100)
Total (n)115576011657591175758
Agmt (n)115565911656591175658
Site C%Agmt100.0%93.1%100.0%100.0%90.0%100.0%100.0%96.6%100.0%
Lot 3(95% CI)(96.9,100)(83.6,97.3)(93.5,100)(96.7,100)(79.8,95.3)(94.0,100)(96.7,100)(88.5,99.1)(93.9,100)
Total (n)118585511160601135959
Agmt (n)118545511154601135759

Total^: total combined across sites. stx1 neg: panel members 4, 5, stx1 10w: panel 2 ,8, stx1 medium: panel members 3, 9, st:2 neg: panels 2, 3, 6, 7, st:2 low: panel members 4, 8, stx2 medium: panel members 5,9, 0157 neg: panels 2, 3, 4, 5. 0157 low: panels 6, 8, 0157 medium: panel members 7, 9 *95% Cl: 95% confidence intervals calculated utilizing Wilson's Method.

For evaluation of the quantitative results, all negative samples in panel member 1 reported "0" and were therefore in 100% agreement. For all positive samples with a valid result, the mean, standard deviation and coefficient of variation (%CV) are reported.

To further evaluate performance by Site/Lot the mean, standard deviation, and coefficient of variation of cycle number were calculated. These parameters are calculated excluding samples that were expected to yield a positive result but where the PanNAT STEC Test cycle number was 0 (Table 5-4).

{8}------------------------------------------------

SiteAnalyte:stx 1stx 2O157Process Control 1
LowMedLowMedLowMed
Site ALot 1n58535755585684
Mean (std)36.2 (0.8)34.8 (0.9)35.5 (1.0)33.6 (1.1)35.4 (0.8)34.0 (1.0)28.6 (1.0)
CV2.2%2.6%2.7%3.2%2.1%3.0%3.6%
Site BLot 2n56595659565887
Mean (std)36.0 (0.8)34.9 (0.7)35.4 (1.0)33.9 (0.8)35.6 (0.8)34.2 (0.8)28.7 (1.0)
CV2.2%1.9%2.7%2.3%2.1%2.4%3.4%
Site CLot 3n54555460575978
Mean (std)35.3 (0.7)34.0 (0.7)35.1 (0.9)33.5 (0.8)35.4 (1.0)33.9 (0.7)28.9 (0.9)
CV2.0%2.2%2.7%2.5%2.7%2.0%3.0%

Table 5-4. Reproducibility Cycle Numbers by Site, Analyte and Concentration

In negative panel members

n=number of observations, mean = mean cycle number, std = standard deviation for mean cycle number, CV=percent coefficient of variation.

Controls

The PanNAT STEC Test has an internal process control, a negative control, and a positive control, integrated within the cartridge. External controls are not provided, but are commercially available. Alternatively, a strain of E. coli that is positive for stx2 and fcl, such as strain EDL933 (ATCC® #43895™) may be used as an external positive control. A known negative sample such as a suspension of human A549 cells (ATCC® CCL-185TM) may be used as an external negative control.

Specimen Stability

The recommended storage time and temperature for Cary-Blair preserved and unpreserved stool specimens is:

  • . Refrigerated storage (2-8℃) for up to 7 days
    To assess the stability of PanNAT STEC Test targets, testing was performed on both Cary-Blair preserved and unpreserved specimens at 3X LoD. The study tested the 0157 STEC reference strain (ATCC #43895), which contains all three PanNAT STEC Test targets. Specimens were stored at 2-8℃ and tested at intervals over the course of the study. Five replicates per specimen type per time point were tested. Results demonstrated that both specimen types were stable for 7 days under these conditions for all PanNAT STEC Test analytes.

Analytical Sensitivity/ Limit of Detection (LoD)

An estimation of the LoD for seven E. coli strains with different combinations of the stx1, stx2 and 0157 was performed by testing dilutions of each strain with the PanNAT STEC Test. Unpreserved and preserved specimens were tested at different concentrations. The estimated LoD was the lowest concentration at which all replicates were positive. The estimated LoD was then confirmed with strains EDL933 (stx1+, stx2+, 0157+) and 3007-85 (stx1+, stx2+, 0157-) tested in replicates of 28 at the estimated LoD in both preserved and unpreserved specimen types. To be considered confirmed, ≥95% of replicates had to produce positive results at the LoD target level.

Results:

{9}------------------------------------------------

  • In unpreserved stool specimens, the LoD for EDL933 and 3007-85 (stxl, stx2, and 0157) is 1.00E+06 CFU/mL.
  • In stool specimens preserved in Cary-Blair media, the LoD for EDL933 and 3007-85 (stx1, stx2 and O157) is 1.75E+06 CFU/mL.

Analytical Reactivity

A collection of well-characterized O157 and non-O157 strains was tested using the PanNAT STEC Test. The presence of the two Shiga toxin genes, stx1 and stx2, was assessed and measured in seven (7) 0157 strains, twelve (12) non-0157 strains, six (6) un-typeable (0) strains, and in a single O-antigen negative laboratory strain as a control (K12).

A summary of results is listed below:

  • The PanNAT STEC Test detected stx1 and stx2 in O157, non-O157 serotype strains . and un-typeable O strains.
  • The PanNAT STEC Test detected at least 2 subtypes of each toxin gene.
  • . For stxl, the PanNAT STEC Test generated the expected results for 24/26 strains (92%), including strains with subtypes stx1a and stx1c. The 2 strains that gave false negative results (TW07740, TW07754) were determined to be stx1d variants. The PanNAT STEC Test only weakly detects stxld due to nucleotide mismatches in the probe-binding region.
  • For stx2, the PanNAT STEC Test generated the expected results for 25/26 strains (96%), including strains with subtypes stx2a and stx2c/d. The strain that gave false negative results (TW07731) was determined to be an stx2 variant that closely resembles stx2b. The PanNAT STEC Test cannot detect this variant due to extensive nucleotide mismatches in the primer and probe binding regions.
  • . For the 0157 O-antigen (fcl) gene cluster, the PanNAT STEC Test generated the expected results in 26/26 strains (100%).

Analytical Specificity - Cross Reactivity

Thirty-seven (37) unique bacterial strains, eight (8) viruses, one (1) yeast organism, and three (3) parasites were spiked into a STEC negative pool matrix and evaluated for cross reactivity (Table 5-5).

{10}------------------------------------------------

OrganismStrain ID/Catalog #Input Tested% Agreementwith ExpectedResult(stx1-/stx2-/0157-)
Bacteria
Acinetobacter baumanniiZeptoMetrix 307-0294 08015971.00E+06 CFU/mL3/3 (100%)
Aeromonas hydrophilaZeptoMetrix Z161 0804098 and ATCC 356541.00E+06 CFU/mL3/3 (100%)
Bacillus cereusZeptoMetrix ZO91 08018231.00E+06 CFU/mL3/3 (100%)
Bacteroides fragilisZeptoMetrix Z029 08015831.00E+06 CFU/mL3/3 (100%)
Campylobacter jejuniZeptoMetrix Z086 08016501.00E+06 CFU/mL3/3 (100%)
Citrobacter koseriZeptoMetrix Z039 08017451.00E+06 CFU/mL3/3 (100%)
Clostridium difficileZeptoMetrix NAP1 08016191.00E+06 CFU/mL3/3 (100%)
Clostridium parabotulinumATCC 178631.00E+06 CFU/mL3/3 (100%)
Clostridium perfringensZeptoMetrix TypeA 08015851.00E+06 CFU/mL3/3 (100%)
Enterobacter cloacaeZeptoMetrix Z101 08018301.00E+06 CFU/mL3/3 (100%)
Enterococcus faecalisZeptoMetrix VRE 08016931.00E+06 CFU/mL3/3 (100%)
Enterococcus faeciumZeptoMetrix VRE, vanA 08018921.00E+06 CFU/mL3/3 (100%)
Escherichia coli K12ATCC 294251.00E+06 CFU/mL3/3 (100%)
Escherichia coli O111a (EAEC)ATCC 295521.00E+06 CFU/mL3/3 (100%)
Escherichia coli O29:NM (EIEC)STEC Center O29:NM TW010951.00E+06 CFU/mL3/3 (100%)
Escherichia coli O111a:2 (EPEC)STEC Center O111a:2 TW078861.00E+06 CFU/mL3/3 (100%)
Escherichia coli 08:H19 (ETEC)STEC Center O8:H19 TW090681.00E+06 CFU/mL3/3 (100%)
Escherichia hermanniiATCC 336501.00E+06 CFU/mL3/3 (100%)
Hafnia alveiATCC 133371.00E+06 CFU/mL3/3 (100%)
Klebsiella pneumoniaeZeptoMetrix Z026 08015061.00E+06 CFU/mL2/2^ (100%)
Listeria monocytogenesZeptoMetrix serotype 1/2b 08015341.00E+06 CFU/mL3/3 (100%)
Morganella morganiiATCC 258301.00E+06 CFU/mL3/3 (100%)
Plesiomonas shigelloidesZeptoMetrix Z130 08018991.00E+06 CFU/mL3/3 (100%)
Proteus mirabilisZeptoMetrix Z050 08015441.00E+06 CFU/mL3/3 (100%)
Pseudomonas aeruginosaZeptoMetrix clinical isolate 08015191.00E+06 CFU/mL3/3 (100%)
Salmonella entericaATCC 356641.00E+06 CFU/mL3/3 (100%)
Serratia marcescensZeptoMetrix Z053 08017231.00E+06 CFU/mL3/3 (100%)
Shigella dysenteriaeATCC 133131.00E+06 CFU/mL3/3^ (100%)
Shigella flexneriZeptoMetrix Z046 08017571.00E+06 CFU/mL3/3 (100%)
Shigella sonneiZeptoMetrix clinical isolate 08016271.00E+06 CFU/mL3/3 (100%)
Staphylococcus aureusZeptoMetrix Z020 08015311.00E+06 CFU/mL2/2^ (100%)
Staphylococcus epidermidisMSSE:HER1292 08016891.00E+06 CFU/mL3/3 (100%)
Streptococcus agalactiaeZeptoMetrix Z019 08015451.00E+06 CFU/mL3/3 (100%)
Streptococcus dysgalactiaeZeptoMetrix Z068 08015161.00E+06 CFU/mL3/3 (100%)
Vibrio sppATCC 119851.00E+06 CFU/mL3/3 (100%)
Yersinia enterocoliticaZeptoMetrix Z036 08017341.00E+06 CFU/mL3/3 (100%)
Yersinia pseudotuberculosisATCC 42841.00E+06 CFU/mL3/3 (100%)
Viruses
Adenovirus 18ZeptoMetrix D.C.(VR-19) VPL-0301.00E+05 PFU/mL3/3 (100%)
Adenovirus 40ZeptoMetrix Dugan HEK293A 0810084CF1.00E+05 PFU/mL3/3 (100%)
Coxsackie B4ZeptoMetrix LLC-mk2 0810075CF1.00E+05 PFU/mL3/3 (100%)
CytomegalovirusZeptoMetrix AD-169 MRC-5 0810003CF1.00E+05 PFU/mL3/3 (100%)
Echovirus 11ZeptoMetrix LLC-MK2 0810023CF1.00E+05 PFU/mL3/3 (100%)
Enterovirus 68ZeptoMetrix FO2-3607 corn (VR-1197) HelaVPL-0301.00E+05 PFU/mL3/3 (100%)
Norovirus (recombinant)ZeptoMetrix Vero Group 10810086CF1.00E+05 PFU/mL3/3 (100%)
RotavirusZeptoMetrix WA MA-104 MRC-5 0810041CF1.00E+05 PFU/mL3/3 (100%)
Yeast and Parasites
Candida albicansZeptoMetrix Z006 08015041.00E+06 CFU/mL3/3 (100%)
Cryptosporidium parvumWaterborne Iowa Isolate1.00E+06 Oocysts/mL3/3 (100%)
Entamoeba histolytica (lysate)ZeptoMetrix DS4-868 BacT-035R051.00E+06 Cysts/mL3/3 (100%)
Giardia lambliaWaterborne H3 isolate1.00E+06 Cysts/mL3/3 (100%)

Table 5-5. Summary of Cross Reactivity Data

{11}------------------------------------------------

Indicates that one of the replicates was "INCOMPLETE" and was removed from analysis.

  • S. dysenteriae detection is expected.

Analytical Specificity – Microbial Interference

For microbial interference, the same 49 non-target organisms were spiked into Shiga toxin negative pooled, unpreserved matrix containing 1X LoD of the reference strain E. coli O157:H7 (EDL933 ATCC #43895), which is positive for target analytes stx1, stx2 and O157 (Table 5-6).

Table 5-6. Summary of Microbial Interference Data

OrganismStrain ID/Catalog #Input TestedCFU/mL% Agreementwith ExpectedResult(stx1+ / stx2+ /0157+)
Bacteria
Acinetobacter baumanniiZeptoMetrix 307-0294 08015971.00E+06 CFU/mL3/3 (100%)
Aeromonas hydrophiliaZeptoMetrix Z161 0804098 and ATCC 356541.00E+06 CFU/mL3/3 (100%)
Bacillus cereusZeptoMetrix ZO91 08018231.00E+06 CFU/mL3/3 (100%)
Bacteroides fragilisZeptoMetrix Z029 08015831.00E+06 CFU/mL3/3 (100%)
Campylobacter jejuniZeptoMetrix Z086 08016501.00E+06 CFU/mL3/3 (100%)
Citrobacter koceriZeptoMetrix Z039 08017451.00E+06 CFU/mL3/3 (100%)
Clostridium difficileZeptoMetrix NAP1 08016191.00E+06 CFU/mL3/3 (100%)
Clostridium parabotulinumATCC 178631.00E+06 CFU/mL3/3 (100%)
Clostridium perfringensZeptoMetrix TypeA 08015851.00E+06 CFU/mL3/3 (100%)
Enterobacter cloacaeZeptoMetrix Z101 08018301.00E+06 CFU/mL3/3 (100%)
Enterococcus faecalisZeptoMetrix VRE 08016931.00E+06 CFU/mL3/3 (100%)
Enterococcus faeciumZeptoMetrix VRE, vanA 08018921.00E+06 CFU/mL3/3 (100%)
Escherichia coli K12ATCC 294251.00E+06 CFU/mL3/3 (100%)
Escherichia coli O111a (EAEC)ATCC 295521.00E+06 CFU/mL3/3 (100%)
Escherichia coli O29:NM (EIEC)STEC Center O29:NM TW010951.00E+06 CFU/mL3/3 (100%)
Escherichia coli O111a:2 (EPEC)STEC Center O111a:2 TW078861.00E+06 CFU/mL3/3 (100%)
Escherichia coli 08:H19 (ETEC)STEC Center O8:H19 TW090681.00E+06 CFU/mL3/3 (100%)
Escherichia hermanniiATCC 336501.00E+06 CFU/mL3/3 (100%)
Hafnia alveiATCC 133371.00E+06 CFU/mL3/3 (100%)
Klebsiella pneumoniaeZeptoMetrix Z026 08015061.00E+06 CFU/mL3/3 (100%)
Listeria monocytogenesZeptoMetrix serotype 1/2b 08015341.00E+06 CFU/mL3/3 (100%)
Morganella morganiiATCC 258301.00E+06 CFU/mL3/3 (100%)
Plesiomonas shigelloidesZeptoMetrix Z130 08018991.00E+06 CFU/mL3/3 (100%)
Proteus mirabilisZeptoMetrix Z050 08015441.00E+06 CFU/mL3/3 (100%)
Pseudomonas aeruginosaZeptoMetrix clinical isolate 08015191.00E+06 CFU/mL3/3 (100%)
Salmonella entericaATCC 356641.00E+06 CFU/mL3/3 (100%)
Serratia marcescensZeptoMetrix Z053 08017231.00E+06 CFU/mL3/3 (100%)
Shigella dysenteriaeATCC 133131.00E+06 CFU/mL3/3 (100%)
Shigella flexneriZeptoMetrix Z046 08017571.00E+06 CFU/mL3/3 (100%)
Shigella sonneiZeptoMetrix clinical isolate 08016271.00E+06 CFU/mL3/3 (100%)
Staphylococcus aureusZeptoMetrix Z020 08015311.00E+06 CFU/mL3/3 (100%)
Staphylococcus epidermidisZeptoMetrix MSSE:HER1292 08016891.00E+06 CFU/mL3/3 (100%)
Streptococcus agalactiaeZeptoMetrix Z019 08015451.00E+06 CFU/mL3/3 (100%)
Streptococcus dysgalactiaeZeptoMetrix Z068 08015161.00E+06 CFU/mL3/3 (100%)
Vibrio sppATCC 119851.00E+06 CFU/mL3/3 (100%)

{12}------------------------------------------------

Yersinia enterocoliticaZeptoMetrix Z036 08017341.00E+06 CFU/mL3/3 (100%)
Yersinia pseudotuberculosisATCC 42841.00E+06 CFU/mL3/3 (100%)
Viruses
Adenovirus 18ZeptoMetrix D.C.(VR-19) VPL-0301.00E+05 PFU/mL3/3 (100%)
Adenovirus 40ZeptoMetrix Dugan HEK293A 0810084CF1.00E+05 PFU/mL3/3 (100%)
Coxsackie B4ZeptoMetrix LLC-mk2 0810075CF1.00E+05 PFU/mL3/3 (100%)
CytomegalovirusZeptoMetrix AD-169 MRC-5 0810003CF1.00E+05 PFU/mL3/3 (100%)
Echovirus 11ZeptoMetrix LLC-MK2 0810023CF1.00E+05 PFU/mL3/3 (100%)
Enterovirus 68ZeptoMetrix FO2-3607 corn (VR-1197) HelaVPL-0301.00E+05 PFU/mL3/3 (100%)
Norovirus (recombinant)ZeptoMetrix Vero Group 10810086CF1.00E+05 PFU/mL3/3 (100%)
RotavirusZeptoMetrix WA MA-104 MRC-5 0810041CF1.00E+05 PFU/mL3/3 (100%)
Yeast and Parasites
Candida albicansZeptoMetrix Z006 08015041.00E+06 CFU/mL2/2^ (100%)
Cryptosporidium parvumWaterborne lowa Isolate1.00E+06 Oocysts/mL3/3 (100%)
Entamoeba histolytica (lysate)ZeptoMetrix DS4-868 BacT-035R051.00E+06 Cysts/mL3/3 (100%)
Giardia lambliaWaterborne H3 isolate1.00E+06 Cysts/mL3/3 (100%)

^ Indicates that one of the replicates was "INCOMPLETE" and was removed from analysis.

The analytical reactivity studies demonstrated no cross reactivity for any analytes tested except for Shigella dysenteriae, which is expected due to sequence homology between the S. dysenteriae stxA gene and the stxl region targeted by the PanNAT STEC Test. No competitive inhibition of PanNAT STEC Test analytes was demonstrated in any of the samples tested.

Analytical Specificity - Interfering Substances

The PanNAT STEC Test was evaluated for potential interference from 25 endogenous and exogenous substances that are common stool contaminants, or likely to be present in patients with diarrhea (Table 5-7). All substances were tested in unpreserved stool specimens in the presence and absence of two stx1+/stx2+/0157+ strains of E. coli, with three replicates per condition. In the presence of an stx 1+/stx2+/0157+ strain, Gaviscon® and Imodium AD® each gave a stx1+/stx2+/0157result on one of three replicates, while stearic acid gave an stxl-/stx2-/0157- result on one of three replicates. No interference was demonstrated for any of the remaining substances tested.

InterferingSubstanceConcentration tested
Whole Blood (EDTA)40% v/v
Mucin3.5% w/v (35mg/mL)
Cholesterol4.80% w/v (48mg/mL)
Human genomic DNA10 ug/mL
Bile25% w/v (250mg/mL)
Kaopectate®3.5% v/v
Pepto-Bismol®5% v/v
Imodium AD®5% v/v

Table 5-7. List of Interferents Tested and Concentrations

{13}------------------------------------------------

Petroleum Jelly (Vaseline®)50% w/v (500mg/mL)
Baby Wipes50% v/v
Hydrocortisone Cream50% w/v (500mg/mL)
Diaper Rash Cream50% w/v (500mg/mL)
Stearic Acid4.80% w/v (48mg/mL)
Palmitic Acid4.80% w/v (48mg/mL)
Gaviscon®10% w/v (100mg/mL)
Milk of Magnesia®10% w/v (100mg/mL)
Aleve®0.5 mg/mL
Preparation H®9.5% w/v (95mg/mL)
Mineral oil50% w/v (500mg/mL)
Amoxicillin5% w/v (50mg/mL)
Nystatin0.43% w/v (4.28mg/mL)
Metronidazole5% w/v (50mg/mL)
Ciprofloxacin1.25% w/v (12.5mg/mL)
Sennica laxatives1% w/v (10mg/mL)
Diaper Eluent25% v/v

Carry Over / Cross Contamination

The potential for false positive results with the PanNAT STEC Test due to run-to-run contamination was evaluated by testing an alternating series of contrived high positive stool samples and clinically negative stool samples in direct succession five (5) times on three (3) PanNAT instruments.

All positive samples reported stx1+/stx2+/0157+ results, and 14 of the 15 negative samples reported stx1-/stx2-/0157- results. One of the negative samples yielded an invalid result (negative sample, instrument 3, round 5). No carryover or cross contamination was observed.

Assay Cut-Off

The PanNAT STEC Test detects Shiga toxin 1 (stxr}, Shiga toxin 2 (stx2), and E. coli 0157. For each target, the valid cycle number range is 16 to 39. Cycle numbers are automatically calculated by the PanNAT system software. Each of the target analytes (stx1, stx2, 0157), as well as the Process Control, must have a cycle number within the valid range in order to be reported as "Positive," otherwise it is reported as "Negative." Cycle number cutoffs were verified on a panel of seven (7) STEC strains of known genotypes, and subsequently validated in the multi-site clinical study.

5.9 Clinical Performance Studies

Prospective and Retrospective Specimen Testing

Performance characteristics of the PanNAT STEC Test were determined on stool samples prospectively collected at six (6) clinical sites and on frozen, retrospective samples. Prospective data were collected from pediatric and adult patients as part of routine patient

{14}------------------------------------------------

care. Results for standard of care testing (comparator method) were obtained using multiple methods, including a test for Shiga toxins 1 and 2 and a test for E.coli O157. A presumptive result for 0157 on specimens that tested positive with sorbitol on MacConkey agar was confirmed by latex agglutination and biochemical analysis.

For prospective specimens (both unpreserved and Cary-Blair preserved), the leftover, deidentified specimens were tested by two comparator methods: 1) an immunoassay for detection of Shiga toxins 1 and 2 following enrichment culture, and 2) a SMAC plate for presumptive detection of 0157. Discordant analyses were performed by an independent molecular reference laboratory on discrepant samples using alternative PCR and bidirectional (BDS) sequencing methods.

A total of 1331 prospective specimens (1231 Cary-Blair preserved and 100 unpreserved) from a range of age groups (Table 5-8) were enrolled in the clinical evaluation. Of these, 30 were removed from the analysis due to pre- or post-analytical errors. Of the remaining 1301 specimens, 7.6% (99/1301) were initially invalid. Specimens with initial invalid results were retested. On retest, 78 were valid resulting in a repeat invalidity rate of 1.6% (21/1301). The number of prospective specimens that qualified for the agreement analysis was 1280 (1184 Cary-Blair preserved and 96 unpreserved) (Table 5-9).

Total ResultsValid ResultsInvalid Results
Age Groupnn%n%
<=515415097.4%42.6%
6-2119619197.4%52.6%
22-5950149598.8%61.2%
60+45044498.7%61.3%
Overall1301128098.4%211.6%

Table 5-8. Samples with Valid and Invalid Results by Age Group

Total ResultsValid ResultsInvalid Results
Typenn%n%
Preserved1201118498.6%171.4%
Unpreserved1009696.0%44.0%
Overall1301128098.4%211.6%

In the prospective clinical study with preserved specimens, although the number of positive specimens was small, positive percent agreement (PPA) for each analyte (stxl, stx2 and 0157) was 100% while negative percent agreement (NPA) ranged from 99.3% to 99.8% (Table 5-10). With unpreserved specimens, no stxl positive specimens were obtained, although PPA for stx2 and 0157 was 100% and NPA ranged from 98.6 to 100%.

Due to the low prevalence of positive specimens in the prospective clinical study, additional testing was also performed using unpreserved, retrospective frozen specimens. The comparator method for the retrospective study was alternative PCR (nested)/bi

{15}------------------------------------------------

directional sequencing. Of the 114 specimens initially enrolled in the study, eight (8) were excluded from the analysis of performance because they did not yield sufficient DNA for comparator testing (7) or previous enrollment (1). On initial testing 8/106 (7.5%) retrospective specimens produced invalid results, all of which were retested and three (3) of which remained invalid for a final invalid rate of 2.8% (3/106). The retrospective study data set therefore contained 103 evaluable results (Table 5-10). PPA for retrospective specimens ranged from 88.5% for stxl to 90.0% for stx2 and NPA ranged from 97.4% to 98.6%.

To supplement the analysis of prospective and retrospective clinical specimens, additional testing was also performed with 16 contrived unpreserved specimens, each of which was spiked with an enumerated stock of a different strain of Shiga-toxin producing E. coli or E. coli 0157:H7 to a final concentration of 5 x 106 CFU/mL (Table 5-10). The contrived specimens were frozen and tested at the clinical sites interspersed with the retrospective specimens.

AnalyteStudySample TypePPA (95% CI)NPA (95% CI)
stx1ProspectivePreserved100.0% (51.0,100.0)(4/4)99.3% (98.7,99.7)(1172/1180)
Unpreserved100.0% (96.2,100.0)(96/96)
Retrospective^*Unpreserved88.5% (71, 96.0)(23/26)97.4% (91.0, 99.3)(75/77)
Contrived^Unpreserved100.0% (75.8, 100.0)(12/12)100.0% (51.0, 100.0)(4/4)
stx2ProspectivePreserved100.0% (56.6,100.0)(5/5)99.4% (98.8,99.7)(1172/1179)
Unpreserved100.0% (34.2,100.0)(2/2)98.9% (94.2,99.8)(93/94)
Retrospective^ΦUnpreserved90.0% (74.4, 96.5)(27/30)98.6% (92.6, 99.8)(72/73)
Contrived^Unpreserved100.0% (67.6, 100.0)(8/8)100.0% (67.6, 100.0)(8/8)
0157ProspectivePreserved100.0% (43.9,100.0)(3/3)99.8% (99.4,100.0)(1179/1181)
Unpreserved100.0% (34.2,100.0)(2/2)100.0% (96.1,100.0)(94/94)
Retrospective^ # δUnpreserved89.7% (73.6, 96.4)(26/29)98.6% (92.6, 99.8)(72/73)
Contrived^Unpreserved100.0% (34.2, 100.0)(2/2)100.0% (78.5, 100.0)(14/14)

Table 5-10. Prospective and Retrospective study results

PPA=positive percent agreement, NPA=negative percent agreement. 95% CI = 95% confidence interval using Wilson's method

^The alternative PCR/BDS comparator assay detects all known variants of 0157, although it may demonstrate reduced sensitivity and/or inability to detect stx1d, variants of stx2b, stx2e and stx2f.

One specimen produced insufficient PCR product for sequencing of this target only.

{16}------------------------------------------------

  • 3/3 specimens that were negative by the PanNAT STEC Test but positive by PCR/BDS were negative by standard of care culture

� 2/3 specimens that were negative by the PanNAT STEC Test but positive by PCR/BDS were negative by standard of care culture

o 1/3 specimens that was negative by the PanNAT STEC Test but positive by PCR/BDS was negative by standard of care culture

A summary of PanNAT STEC Test performance stratified by analyte is shown in Table 5-11.

Table 5-11. Summary of PanNAT STEC Test performance by study and analyte/analyte combination

Analyte(s)1ProspectiveRetrospectiveContrived
PreservedUnpreservedUnpreservedUnpreserved
stx1 Only100%(4/4)Not applicable25.0%(1/4)100%(7/7)
stx2 Only100%(2/2)Not applicable0.0%(0/2)100%(3/3)
O157 OnlyNot applicableNot applicableNot applicable100%(1/1)
stx1 + stx2Not applicableNot applicableNot applicable100%(4/4)
stx1 + O157Not applicableNot applicable0.0%(0/1)Not applicable
stx2 + O157100%(3/3)100%(2/2)85.7%(6/7)Not applicable
stx1 + stx2 +O157Not applicableNot applicable95.2%(20/21)100%(1/1)
Negative98.7%(1160/1175)298.9%(93/94)398.5%(67/68)4Not applicable

1 Analyte(s) present as determined by the applicable comparator method for the study 2 Of the 15 specimens with false positive results by the PanNAT STEC Test, 8 were positive for str.1, 5 for stx2 and 2 for both st:2

and 0157

3 1/94 negative specimens was positive by the PanNAT STEC Test for stx2

4 1/68 negative specimens was positive by the PanNAT STEC Test for stx1, stx2 and 0157

5.10 Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3990 Gastrointestinal microorganism multiplex nucleic acid-based assay.

(a)
Identification. A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitativein vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.(b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).