Intended Use

Relieves the symptoms of dry mouth, while moisturizing and lubricating oral dryness.

Device Description

Moisyn is a specifically formulated artificial saliva substitute which contains moisturizers, humectants, and biopolymers that are designed to relieve dry mouth symptoms, soothe irritated oral surfaces, refresh, clean, and lubricate. The rinse is supplied as an 80z PET bottle or as a 20z PET spray bottle. Both bottles contain the same solution without any changes in formulation or intended use. Both the rinse and the Mist are formulated with and without spearmint flavor to provide the customer with a choice in flavoring.

AI/ML Overview

The provided text is a 510(k) summary for the device "Moisyn Dry Mouth Oral Rinse/Mist". It describes the device, its intended use, comparison with predicate devices, and performance data. However, the information requested regarding acceptance criteria and a study proving the device meets these criteria (especially related to AI/algorithm performance, expert review, MRMC studies, etc.) is not present in this document.

This document is a regulatory submission for a medical device that relieves symptoms of dry mouth. It focuses on demonstrating substantial equivalence to existing predicate devices based on intended use, technological characteristics, and safety aspects (biocompatibility, preservation effectiveness, shelf-life, viral inactivation).

Here's a breakdown of why the requested information cannot be extracted:

  • No AI/Algorithm Component: The device described is a physical oral rinse/mist, not an AI or algorithm-driven diagnostic or medical imaging device. Therefore, concepts like acceptance criteria for AI performance (sensitivity, specificity, AUC), test sets, training sets, ground truth establishment by experts, MRMC studies, or standalone algorithm performance are not applicable.
  • Performance Data Focus: The "Performance data" section refers to standard biocompatibility testing, preservative effectiveness, and shelf-life, which are relevant for a chemical/physical product, not an AI.
  • "Clinical Study" Context: The mention of a "Clinical Study" refers to a human study evaluating the effectiveness of the oral rinse in relieving dry mouth symptoms, not the performance of an AI or diagnostic algorithm. The study cited is about symptom improvement and saliva production, not diagnostic accuracy.

Therefore, I cannot provide a table of acceptance criteria for AI performance or details about an AI-related study because the device in question is not an AI/algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 19, 2018

Syendgen, Inc. Christopher Ryan, Ph.D. Director of Manufacturing 1420 N. Claremont Blvd. Suite 105D Claremont, California 91711

Re: K173237

Trade/Device Name: Moisyn Dry Mouth Oral Rinse. Moisyn Dry Mouth Oral Mist. Moisyn Free Dry Mouth Oral Rinse, Moisyn Free Dry Mouth Oral Mist

Regulatory Class: Unclassified Product Code: LFD Dated: February 15, 2018 Received: February 16, 2018

Dear Christopher Ryan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173237

Device Name

Moisyn Dry Mouth Oral Rinse, Moisyn Dry Mouth Oral Mist, Moisyn Free Dry Mouth Oral Rinse, Moisyn Free Dry Mouth Oral Mist

Indications for Use (Describe)

Relieves the symptoms of dry mouth, while moisturizing and lubricating oral dryness.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K173237

Synedgen, Inc.

Moisyn Dry Mouth Oral Rinse, Moisyn Dry Mouth Oral Mist, Moisyn Free Dry Mouth Oral Rinse, Moisyn Free Dry Mouth Oral Mist

Submitter:

Synedgen, Inc. 1420 North Claremont Blvd, Suite 105D Claremont, CA 91711 Phone: 909-447-6858 909-447-6801 Fax: Contact Person: Shenda Baker Correspondent: Christopher Ryan Date Prepared: March 19, 2018

Device:

Name of Device: Moisyn Dry Mouth Oral Rinse, Moisyn Dry Mouth Oral Mist, Moisyn Free Dry Mouth Oral Rinse, Moisyn Free Dry Mouth Oral Mist Dry Mouth Spray Common or Usual name: Artificial Saliva

Classification: Unclassified Product Code: LFD

Primary Predicate Device:

Dr. Fresh Dry Mouth Mouthwash K111250 (LFD) This predicate has not been subject to a design-related recall.

Reference Predicate Devices:

Oasis Dry Mouth(K041563). Moi-Stir(K810157). SynePure (K143444). Oral 7 Moisturizing Mouth Spray(K142549), Mouth Kote Oral Moisturizer (K062653)

Device Description:

Moisyn is a specifically formulated artificial saliva substitute which contains moisturizers, humectants, and biopolymers that are designed to relieve dry mouth symptoms, soothe irritated oral surfaces, refresh, clean, and lubricate.

The rinse is supplied as an 80z PET bottle or as a 20z PET spray bottle. Both bottles contain the same solution without any changes in formulation or intended use.

Both the rinse and the Mist are formulated with and without spearmint flavor to provide the customer with a choice in flavoring.

Indications for use:

Relieves the symptoms of dry mouth, while moisturizing and lubricating oral dryness.

{4}------------------------------------------------

Comparison with predicate:

The candidate device. Moisyn, has the same intended use and the substantially equivalent technological characteristics as the predicate device: Dr. Fresh Dry Mouth Mouthwash. Both the predicate and the proposed device are aqueous based solutions that use neutral sugars (glycerol and sorbitol) as well as polymers to assist in the moisturizing dry mouth. The proposed device and predicate use sorbitol to balance the solution and to provide a denser solution. The proposed device and predicate are preserved to control microbial content. The proposed device and predicate are delivered with a pH that is not significantly acidic or basic. Like the predicate device, Moisyn acts as an artificial saliva that utilizes moistening properties to aid in the moisturizing and lubricating of oral dryness.

New Device (Moisyn)Predicate Device (Dr. Fresh)
Intended UseSymptomatic treatment of Xerostomia(Dry Mouth)Symptomatic treatment of Xerostomia
Indications for UseRelieves the symptoms of dry mouth, while moisturizing and lubricating oral dryness.Relieves the symptoms of dry mouth, while moisturizing and lubricating oral dryness.
Sterile or preservativeNon-sterile, preservedNon-sterile, preserved
CompositionPurified waterSorbitolXylitolGlycerolChitosan derivativesBetaineFlavorSodium Hydroxide for pH balancePurified waterSorbitolXylitolPropylene GlycolPoloxamer 407PolyvinylpyrrodlidoneSodium BenzoateBenzoic AcidFlavorMentholCalcium lactateZinc GluconateAloe VeraSodium PhosphateBactase Oral
DeliveryOral Cavity via PET bottle withremovable cap or PET bottlewith nozzle sprayOral Cavity viaPET bottles with flip caps
Mode of actionMoisturizing and lubricating oral drynessMoisturizing and lubricating oral dryness
Type of ProductLiquid SolutionLiquid Solution
pH5.5 - 8.05.3

Comparison of New and Predicate Devices

The proposed device has substantially equivalent technological characteristics as a non-sterile, preserved, non-toxic, polymer-based mouthwash and is similar in design and configuration to the

{5}------------------------------------------------

predicate device. The proposed device is designed and assembled with components found in the predicate and reference devices.

Performance data:

Biocompatibility Testing

Moisyn has been tested in accordance with ISO 10993 and was shown to meet the requirements of biocompatibility testing for Cytotoxicity, Maximization Test for Delayed-Type Hypersensitivity, Dermal Irritation, Oral Mucosal Irritation, and Acute Systemic Toxicity.

TestDescriptionResult
Cytotoxicity Direct ContactCytotoxicity was evaluatedusing ISO-10993-5,Non-toxic
Biological evaluation of
Medical Devices-Part 5: Tests
for In Vitro Cytotoxicity.
Maximization Test forThe ability of Moisyn OralNo Sensitization reaction was
Delayed-TypeRinse/Mist to cause delayed-observed in any of the test
Hypersensitivitytype hypersensitivity wasanimals
evaluated according to ISO
10993-10:2010, Biological
evaluation of medical devices
– Part 10: Tests for Irritation
and skin sensitization.
Dermal IrritationTo assess dermal irritationNo erythema or edema (score
due to exposure of skin toof () was observed on test
Moisyn Oral Rinse and Mist,sites at the 1, 24, 48, and 72-
the test articles was evaluatedhour scoring. The Primary
according to ISO 10993-Irritation Index for the test
10:2010, Biologicalarticle was 0.
evaluation of medical devices
– Part 10: Tests for irritation
and skin sensitization.
Oral Mucosal IrritationTo assess oral mucosalThe Oral Irritation Index for
irritation due to exposure tothe test article was 0.0.
Moisyn Oral Mist, the test
article was evaluated
according to ISO 10993-
10:2010, Biological
evaluation of medical devices
– Part 10: Tests for irritation
and skin sensitization.To assess the acute systemicThe test article met the
Acute Systemic Toxicitytoxicity of the Moisyn Oralrequirements for the Acute
Mist, which was testedSystemic Toxicity Test and
according to ISO 10993-11,the requirements of ISO
Biological Evaluation of10993-11
Medical Devices – Tests for
Systemic Toxicity.

{6}------------------------------------------------

Preservative Effectiveness

The Moisyn Dry Mouth Oral Rinse/Mist was tested for its preservative effectiveness. The data indicate that Moisyn Dry Mouth Oral Rinse/Mist meet the requirements of USP 51.

Shelf-Life

The stability data indicate the product meets specifications after two years of storage and is scheduled to test up to 3 years.

Viral Inactivation

Viral inactivation was assessed.

Clinical Study

An initial study of 57 patients with Xerostomia was conducted to treatment with Moisyn. Substantial equivalence is based in part on this study. Statistically significant improvements in dry mouth symptoms, reduction in oral pain, and improvement in taste/diet were demonstrated. Whole unstimulated/resting saliva and whole stimulated saliva improved.

Epstein, Joel B, Dana C Villines, Mabi Singh, and Athena Papas. 2017. "Management of dry mouth: assessment of oral symptoms after use of a polysaccharide-based oral rinse." Oral Medicine 76-83

Conclusions:

Based on the Indications for Use and the data presented in this submission, Moisyn is substantially equivalent to the predicate device Dr. Fresh Dry Mouthwash (K111250).