K Number
K173225

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-06-15

(255 days)

Product Code
Regulation Number
884.5470
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Interlude and private label Unscented 3-piece applicator tampons are inserted into the vagina and used to absorb menstrual or other vaginal discharge.

Device Description

Interlude and private label Unscented 3-piece compact applicator tampons are menstrual tampons used to absorb menstrual fluid. These tampons are provided in 2 absorbencies, regular and super. These tampons are made from rayon fibers and polymeric overwrap. The withdrawal cord is made of cotton and polyester. The applicator tubes are made of polyethylene (PE) and include a cylindrical barrel with a finger grip and a two-piece plunger. The assembled tampon with applicator is wrapped in a printed PE wrapper.

AI/ML Overview

This document is a marketing authorization (510(k) clearance) for a medical device: "Interlude and private label Unscented 3-piece compact applicator tampon". It does not describe acceptance criteria for an AI/ML powered device, nor does it detail a study proving such a device meets acceptance criteria.

The document clarifies that the device is a menstrual tampon, and the data provided relates to its physical properties, material composition, biocompatibility, and microbial testing, all conducted to demonstrate substantial equivalence to a predicate device (Rostam Fragranced and Unfragranced Compact Applicator Tampons).

Therefore, it is not possible to extract the requested information based on the provided text, as it concerns a completely different type of medical device and regulatory process.

To directly answer your prompt based on the provided text:

1. A table of acceptance criteria and the reported device performance:
There is no table of acceptance criteria for an AI/ML powered device. The document provides performance characteristics for a menstrual tampon.

Performance CharacteristicReported Device Performance (Interlude tampon)Acceptance Criteria (from FDA guidance)
Regular size length114.4 mmn/a (compared to predicate)
Regular size diameter14 mmn/a (compared to predicate)
Regular size weight2.25 gn/a (compared to predicate)
Super size length114.4 mmn/a (compared to predicate)
Super size diameter16 mmn/a (compared to predicate)
Super size weight2.4 gn/a (compared to predicate)
Regular size absorbency6-9 gramsn/a (compared to predicate)
Super size absorbency9-12 gramsn/a (compared to predicate)
BiocompatibilityNon-cytotoxic, non-irritating, non-sensitizing, non-systemically toxicMeets ISO 10993-1, 10993-5, 10993-10, 10993-11 standards
Microbiological effectsDoes not enhance Staphylococcus aureus growth, does not increase TSST-1 production, does not alter normal vaginal microflora growth.Meets recommendations of 2005 FDA guidance document.
Other performance tests (Dimensional, Absorbency, Chemical residues, Withdrawal cord strength, Fiber shedding, Tampon integrity)Assessed in accordance with 2005 FDA guidance. Details not specified, but implied to meet requirements.Meets recommendations of 2005 FDA guidance document.

2. Sample sizes used for the test set and the data provenance:
Not applicable for an AI/ML device. For the tampon, details of sample sizes for the performance tests (biocompatibility, microbial) are not specified in this summary. The data provenance is derived from laboratory testing of the physical product, not patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. Ground truth for a tampon is established through objective laboratory testing, not expert interpretation of images or data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
For the menstrual tampon, the "ground truth" for performance characteristics was based on established laboratory testing methods and FDA guidance standards for physical properties, material safety, and microbiological impact.

8. The sample size for the training set:
Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established:
Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 15, 2018

ALBAAD fem % Robert Staab Official Correspondent Regulatory Technical Associates 30 Neck Road Old Lyme, Connecticut 06371

Re: K173225

Trade/Device Name: Interlude and private label Unscented 3-piece compact applicator tampon Regulation Number: 21 CFR 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: Class II Product Code: HEB Dated: May 14, 2018 Received: May 14, 2018

Dear Robert Staab:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael T. Bailey -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173225

Device Name

Interlude and private label Unscented 3-piece compact applicator tampon

Indications for Use (Describe)

Interlude and private label Unscented 3-piece applicator tampons are inserted into the vagina and used to absorb menstrual or other vaginal discharge.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K173225 510(k) SUMMARY

ALBAAD fem1.0Submitter:
---------------------------------

Name: Shlomo Helvits Address: 1, Alon Hatavor St. Caesarea Ind. Park 3800900 Israel

FDA Est Reg#: 9613295 Phone No: +972-4-6223860 Fax No: +972-4-6277022

Date 510(k) Summary was prepared: 06/14/2018

  • 2.0 Submission Correspondent:
Name:Robert J. Staab
Address:30 Neck RoadOld Lyme, CT 06371
Phone No:+1 (860) 434-5872
Fax No:+1 (860) 434-5892
Email:rta1ali1@aol.com

3.0 Name of the device:

Trade Name:Interlude and private label Unscented 3-piece compact applicator tampon
Device Name:Unscented 3-piece compact applicator tampon
Common Name:Unscented menstrual tampon
Device Class:Class II
Regulation Name:Unscented menstrual tampon
Regulation Number:21 CFR 884.5470
Product Code:HEB (tampon, menstrual, unscented)

4.0 Predicate Device Information:

The predicate device is K090071, Rostam Fragranced and Unfragranced Compact Applicator Tampons, Rostam Limited

The predicate device has not been subject to a design-related recall

{4}------------------------------------------------

  • 5.0 Description of the Device:
    • A. Device Description and Dimensions

Interlude and private label Unscented 3-piece compact applicator tampons are menstrual tampons used to absorb menstrual fluid. These tampons are provided in 2 absorbencies, regular and super.

These tampons are made from rayon fibers and polymeric overwrap. The withdrawal cord is made of cotton and polyester. The applicator tubes are made of polyethylene (PE) and include a cylindrical barrel with a finger grip and a two-piece plunger. The assembled tampon with applicator is wrapped in a printed PE wrapper.

  • B. Physical Properties and Weight
    • a. Dimensions and Weight

Regular size length = 114.4 mm Regular size diameter = 14 mm Regular size weight = 2.25 g

Super size length = 114.4 mm Super size diameter = 16 mm Super size weight = 2.4 g

  • b. Absorbency
    Regular size absorbency = 6-9 grams Super size absorbency = 9-12 grams

  • C. Device Materials

    • The pledget absorbing fibers are made of rayon. The finishing a. agent for the rayon fibers is a polyglycol ester and fatty alcohol mixture. Leomin or glycerol.
    • b. The non-woven over-wrap is a bi-component thermoplastic nonwoven.
    • The withdrawal cord is made from polyester and cotton, and c. includes an antiwicking agent consisting of a paraffin wax emulsion.
    • d. The applicator is made from high density polyethylene (HDPE), low density polyethylene (LDPE) and polypropylene

{5}------------------------------------------------

  • 6.0 Indications for Use
    Interlude and private label Unscented 3-piece applicator tampons are inserted into the vagina and used to absorb menstrual or other vaginal discharge.

9.0 Predicate Device Comparison

The table below shows similarities and differences between the predicate device and the subject device.

CharacteristicsPredicate DeviceSubject Device
Device NameRostam Scented andUnscented plasticCOMPACT applicatorTamponsInterlude and private labelUnscented 3-piece compactapplicator tampon
ManufacturerRostam LimitedALBAAD fem
510(k)K090071K173225
Indications for UseThe Rostam Fragranced andUnfragranced CompactApplicator Tamponsare inserted into the vaginaand used to absorbmenstrual or other vaginaldischargeInterlude and private labelUnscented 3-pieceapplicator tampons areinserted into the vagina andused to absorb menstrual orother vaginal discharge.
Device DesignCompact applicator withcylindrical barrel and one-piece plunger. W foldedtampon with roundmushroom tipCompact style applicatorwith cylindrical barrel withfinger grip and two-pieceplunger. Tampon withround tip and straightgrooves.
OverwrapcompositionPolyethylene/PolypropyleneNonwovenPolyethylene/PolyethyleneTerephthalate Nonwoven
Withdrawal cordcompositionCottonPolyester and Cotton
Pledget compositionViscose RayonViscose Rayon
Plastic ApplicatorPolyethyleneHDPE, LDPE,polypropylene
Primary Packaging(i.e. wrapper)Printed PolypropylenePrinted Polyethylene

The Interlude and private label Unscented 3-piece compact applicator tampon is a modification of the Rostam Scented and Unscented plastic COMPACT applicator Tampons. The modifications include material and physical design changes, including differences in the overwrap composition, applicator composition, packaging, and minor dimensional differences in the applicator and pledget.

{6}------------------------------------------------

These differences do not raise different questions of safety and effectiveness. The intended use of the subject and predicate device are identical, to absorb menstrual or other vaginal discharge

10.0 Summary of Performance Testing:

Biocompatibility

Biocompatibility studies were performed in accordance with the 2016 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993-1:2009 as follows:

  • Cytotoxicity (ISO 10993-5:2009) ●
  • Sensitization (ISO 10993-10:2010) ●
  • Irritation (ISO 10993-10:2010)
  • Acute Systemic Toxicity (ISO 10993-11:2006) .

All tests were performed on the tampon and applicator separately. The results of this testing demonstrated that the subject device is non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic.

Performance Characteristics

The following performance characteristics were assessed in accordance with the 2005 FDA guidance document Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s) – Guidance for Industry and FDA Staff:

  • Dimensional information ●
  • . Absorbency range
  • Chemical residues ●
  • Withdrawal cord strength
  • Fiber shedding ●
  • Tampon integrity ●

Microbial Testing

Per the 2005 FDA guidance document Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s) – Guidance for Industry and FDA Staff, tampon manufacturers should demonstrate that the subject device in its final, finished form does not:

  • enhance the growth of Staphylococcus aureus
  • increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1)
  • alter the growth of normal vaginal microflora ●

{7}------------------------------------------------

To assess these parameters, the following tests were performed:

  • Zone of inhibition assay assessing the effects of the subject device on the . representative microorganisms considered as normal microflora and/or microflora from the infected genitourinary tract
  • Evaluation of the tampon's effect on Staphylococcus aureus growth and . Toxic Shock Syndrome Toxin-1 production using the tampon-sac/devicesac method

The tests performed met the recommendations of the 2005 FDA guidance document.

11.0 Conclusion:

The results of the performance testing described above demonstrate that the Interlude and private label Unscented 3-piece compact applicator tampon is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5470 Unscented menstrual tampon.

(a)
Identification. An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.(b)
Classification. Class II (performance standards).