K Number
K172899

Validate with FDA (Live)

Device Name
Vygon PICCs
Manufacturer
Date Cleared
2018-02-26

(157 days)

Product Code
Regulation Number
880.5970
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and allow(s) for central venous pressure monitoring, infusion or therapy, use 4 French or larger catheter. The maximum recommended infusion rate is 5mL/section of contrast media. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Device Description

The Vygon PICCs are sterile, single-use, polyurethane PICCs with the same intended use, indications for use, similar design, similar sizes, similar or the same materials, and similar properties as the predicate device.

AI/ML Overview

The provided document details the Vygon PICCs device, its intended use, and substantial equivalence to a predicate device, focusing on various performance and safety tests. However, the document does not describe an algorithm or AI-driven device, nor does it present acceptance criteria and performance data in the context of such a device.

Therefore, many of the requested items (e.g., sample size for AI test sets, number of experts for ground truth, MRMC study, standalone performance, training set details) are not applicable to this submission, as it pertains to a physical medical device (Vygon PICCs) and not a software or AI-based diagnostic/screening tool.

The document primarily focuses on bench testing, sterilization testing, and biocompatibility testing to demonstrate that the Vygon PICCs are substantially equivalent to the predicate device (5 Fr DL PowerPICC® Catheter).

Here's an attempt to extract the relevant information based on the provided document, addressing the prompt's categories where applicable:

1. Table of Acceptance Criteria and Reported Device Performance

For a traditional medical device like a PICC, the "acceptance criteria" are typically the satisfactory completion of standardized tests and meeting the requirements outlined in those standards. The "reported device performance" is whether the device "Passed" each test.

Test Standard / RequirementAcceptance Criteria (Implied: "Pass" per standard)Reported Device Performance
Bench Testing
ISO 9626:2016 (Dimensional)Meets dimensional specificationsPassed
ISO 10555-1:2014 (Conditioning Saline)Maintains integrity after conditioningPassed
ISO 10555-1:2014 (Leak Test)No leaks detectedPassed
Internal Requirement (Priming Volume)Meets specified priming volumePassed
ISO 10555-1:2014 (Gravity Flow)Achieves specified flow ratePassed
ISO 10555-1:2014 (Pump Flow)Achieves specified flow ratePassed
ISO 10555-1:2014 (Tensile)Withstands specified tensile forcePassed
ISO 10555-1:2014 (Burst)Withstands specified burst pressurePassed
ISO 10555-1:2014 (Power Injection Conditioning)Maintains integrity after power injection conditioningPassed
Internal Requirement (Printing Font Size)Meets specified font sizePassed
ASTM F1842-15 (Ink Permanence)Ink remains permanentPassed
ISO 594-1: 1986 Via / ISO 80369-7:2016 (Luer Taper)Complies with Luer taper standardsPassed
ASTM F1842-15 (Ink Integrity)Ink integrity maintainedPassed
Vygon Internal (Aspiration Flow)Achieves specified aspiration flowPassed
Vygon Internal (Luer Color/Orientation)Correct color and orientationPassed
Vygon Internal (Clamp Closure Maintenance)Clamp maintains closurePassed
Vygon Internal (Clamp and ID Tag Fit)Clamp and ID tag fit correctlyPassed
Vygon Internal (Clamp Actuation)Clamp actuates correctlyPassed
Aged Performance Testing
ISO 10555-1:2014 (Power Injection Conditioning)Maintains integrity after conditioning (aged)Passed
ISO 10555-1:2014 (Burst)Withstands specified burst pressure (aged)Passed
Sterilization Testing
ASTM F88 / F88M-15 (Seal Strength)Packaging seal strength maintainedPassed
ASTM F2096-11 (Bubble Leak)No leaks in packagingPassed
ANSI/AAMI ST72:2011 (LAL)Meets endotoxin limitsPassed
ISO 10993-7:2008 (EO Residual)Meets EO residual limitsPassed
Biocompatibility Testing
ISO 10993-5:2009 (Cytotoxicity)Non-cytotoxicNon-cytotoxic
AAMI/ANSI/ISO 10993-10 (Sensitization)Non-sensitizingNon-sensitizer
AAMI/ANSI/ISO 10993-10 (Irritation)Non-irritatingNon-irritant
ISO 10993-11:2006 (Acute Systemic)Non-toxicNon-toxic
USP37 (Material Mediated Pyrogen)Non-pyrogenicNon-pyrogenic
ISO 10993-4 (Hemolysis, extract)Non-hemolyticNon-hemolytic
ISO 10993-4 (Hemolysis, direct)Non-hemolyticNon-hemolytic
ISO 10993-4 (Complement Activation)Similar to predicateSimilar when compared to predicate
ISO 10993-4 (Partial Thromboplastin Time)Minimal activator (same as predicate)Minimal activator (same as predicate)
ISO 10993-4 (Dog Thrombogenicity)Similar to predicateEquivocal similar
AAMI/ANSI/ISO 10993-6 (Implantation)Non-irritatingNon-irritant
ISO 10993-18 (Extractable/Leachable)Analysis summarizedSummarized in project #MJ16357-BIO01
ISO 10993-17 (Toxicity)Evaluated satisfactorilyEvaluated in Toxicological Risk Assessment

2. Sample sized used for the test set and the data provenance:
For a physical device, "sample size" refers to the number of units tested. The document does not explicitly state the sample sizes for each bench, sterilization, or biocompatibility test. The data provenance is internal testing performed by Vygon USA (the submitter). The context implies these are prospective tests conducted on the final device. No country of origin for the data is specified beyond the submitter's location (Lansdale, PA).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This device is subjected to standardized physical and chemical tests, not an assessment requiring expert "ground truth" derived from clinical cases or images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. Adjudication methods are relevant for subjective assessments, typically in clinical studies or image interpretations. The tests performed are objective, pass/fail criteria based on adherence to international standards and internal requirements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI-driven device or study.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. This is not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
Not applicable. The "ground truth" for this device is compliance with established engineering and biological standards and specifications.

8. The sample size for the training set:
Not applicable. This is not an AI/algorithm-based device, so there is no training set.

9. How the ground truth for the training set was established:
Not applicable. There is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

February 26, 2018

Vygon USA Jillian Mikovich Regulatory Affairs Manager 2750 Morris Rd. Suite A200 Lansdale, Pennsylvania 19460

Re: K172899

Trade/Device Name: Vygon PICCs Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: January 26, 2018 Received: January 29, 2018

Dear Jillian Mikovich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina
Kiang-S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172899

Device Name

Vygon PICCs

Indications for Use (Describe)

The catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and allow(s) for central venous pressure monitoring, infusion or therapy, use 4 French or larger catheter. The maximum recommended infusion rate is 5mL/section of contrast media. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Vygon logo. The logo consists of the word "VYGON" in white letters on a green background, with a red triangle above the "V". Below the main logo, the words "Value Life" are written in a smaller, gray font.

K172899 510(k) Summary (21 CFR 807.92(c))

I. Submitter Information

Submitter Name:Vygon
Address:2750 Morris Road, Suite A200Lansdale, PA 19446
Telephone:215-390-2002
Fax:215-672-6740
Contact Person:Jillian Mikovich
Email of Contact:jmikovich@vygonus.com
Date Prepared:February 23, 2018

II. Device

Device Name:Vygon PICCs
Common Name:Vygon PICCs
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Percutaneous, Implanted, Long-Term Intravascular
Catheter
Classification Panel:General Hospital
Regulation Device Name:Catheter, Intravascular, Therapeutic, Long-Term Greater
Than 30 Days
Regulation Number:21 CFR 880.5970
Regulatory Class:2
Product Code:LIS

III. Predicate Device

Predicate Name:5 Fr DL PowerPICC® Catheter
Trade Name:5 Fr DL PowerPICC® Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:LJS - Percutaneous, Implanted, Long-Term IntravascularCatheter
Premarket Notification:K051672
Manufacturer:Bard Access Systems, Inc.

Per the FDA Medical Device Recalls Database, this predicate has not been subject to a design related recall.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Vygon logo. The logo consists of the word "VYGON" in white letters inside a green rounded rectangle with a red triangle above the "V". Below the rectangle, the words "Value Life" are written in a smaller, gray font.

IV. Indications for Use

The catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and allow(s) for central venous pressure monitoring. For blood sampling, infusion or therapy, use 4 French or larger catheter. The maximum recommended infusion rate is 5mL/sec for power injection of contrast media. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

The Indications for Use statement for the Vygon PICCs is not identical to the predicate device, 5 Fr DL PowerPICC® Catheter, Indications for Use; however, the differences do not alter the intended therapeutic use of the device. Both the subject and predicate devices have the same intended use for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.

V. Technological Characteristics

All characteristics of the Vygon PICCs are substantially equivalent to the predicate device, the Bard Access Systems 5 Fr DL PowerPICC® catheter. Both the subject and the predicate catheters are sterile, single-use, polyurethane PICCs with the same intended use, indications for use, similar design, similar sizes, similar or the same materials, and similar properties. See comparison table below.

AttributePredicate:Predicate DeviceDLSubject Device 1SLSubject Device 2DLSubject Device 3TL
Device Name5 Fr DLPowerPICC®CatheterVygon PICC 4 FrSLVygon PICC 5 FrDLVygon PICC 6 FrTL
510(k) NumberK051672K172899K172899K172899
510(k)SubmitterBard AccessSystems (BAS)Vygon USAVygon USAVygon USA
Trade Name5 Fr DLPowerPICC®CatheterVygon PICC 4 FrSLVygon PICC 5 FrDLVygon PICC 6 FrTL
CommonNamePeripherallyInserted CentralCatheter (PICC)Same as PredicateSame as PredicateSame as Predicate
AttributePredicate:Predicate DeviceDLSubject Device 1SLSubject Device 2DLSubject Device 3TL
Device Name5 Fr DLPowerPICC®CatheterVygon PICC 4 FrSLVygon PICC 5 FrDLVygon PICC 6 FrTL
ClassificationNameCatheter,Intravascular,Therapeutic, Long-Term IntravascularCatheterSame as PredicateSame as PredicateSame as Predicate
RegulatoryClass2Same as PredicateSame as PredicateSame as Predicate
ClassificationRegulation21 CFR 880.5970Same as PredicateSame as PredicateSame as Predicate
Product CodeLJSSame as PredicateSame as PredicateSame as Predicate
Review Panel /DivisionGeneral HospitalSame as PredicateSame as PredicateSame as Predicate
Indications forUseThe PowerPICC®catheter isindicated for shortor long termThe catheter isindicated for shortor long termperipheral access tothe central venoussystem forintravenoustherapy, powerinjection of contrastmedia, and allow(s)for central venouspressuremonitoring. Forblood sampling,infusion or therapy,use 4 French orlarger catheter. Themaximumrecommendedinfusion rate is5mL/sec for powerinjection of contrastmedia. For centralvenous pressuremonitoring, it isThe catheter isindicated for shortor long termperipheral access tothe central venoussystem forintravenoustherapy, powerinjection of contrastmedia, and allow(s)for central venouspressuremonitoring. Forblood sampling,infusion or therapy,use 4 French orlarger catheter. Themaximumrecommendedinfusion rate is5mL/sec for powerinjection of contrastmedia. For centralvenous pressuremonitoring, it isThe catheter isindicated for shortor long termperipheral access tothe central venoussystem forintravenoustherapy, powerinjection of contrastmedia, and allow(s)for central venouspressuremonitoring. Forblood sampling,infusion or therapy,use 4 French orlarger catheter. Themaximumrecommendedinfusion rate is5mL/sec for powerinjection of contrastmedia. For centralvenous pressuremonitoring, it is
AttributePredicate:Predicate DeviceDLSubject Device 1SLSubject Device 2DLSubject Device 3TL
Device Name5 Fr DLPowerPICC®CatheterVygon PICC 4 FrSLVygon PICC 5 FrDLVygon PICC 6 FrTL
catheter may notexceed 300 psi.catheter lumen of20 gauge or largerbe used.catheter lumen of20 gauge or largerbe used.catheter lumen of20 gauge or largerbe used.
Intended UseThe catheters areintended for shortor long termperipheral accessto the centralvenous system forintravenoustherapy and bloodsampling.Same as PredicateSame as PredicateSame as Predicate

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Vygon logo. The logo consists of the word "VYGON" in white letters inside a green rounded rectangle with a red triangle above the "V". Below the rectangle, the words "Value Life" are written in gray.

2750 Morris Road, Suite A200 • Lansdale, PA 19446 • Telephone: 215-390-2002 • Fax: 215-672-6740 Email: email@vygonus.com • Website: www.vygonusa.com

{6}------------------------------------------------

Table 1: Substantial Equivalence Comparison Table

VI. Performance Testing

Vygon PICCs were tested per Table 2 and Table 3 below.

Special controls for the Product Code LJS include the guidance document "Guidance on Premarket Notification [510(K)] Submissions for Short-Term and Long-Term Intravascular Catheters" and the consensus standards:

  • ISO 10555-1 Second edition 2013-06-15 Intravascular catheters -- Sterile and single-● use intravascular catheters -- Part 1: General requirements
  • ISO 10555-3 Second edition 2013-06-15 Intravascular catheters -- Sterile and singleuse catheters -- Part 3: Central venous catheters
  • . ISO 10555-1 and ISO 10555-3 requirements have been addressed and met.

Other requirements listed in the guidance document "Guidance on Premarket Notification [510(K)] Submissions for Short-Term and Long-Term Intravascular Catheters" have been met in the development of the device and are described below.

Bench Testing

Bench tests are listed below.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for Vygon. The logo consists of the word "VYGON" in white letters on a green rounded rectangle. There is a red triangle above the "V" in Vygon. Below the green rectangle, the words "Value Life" are written in gray.

StandardSterile Device TestResult
ISO 9626:2016Dimensional InspectionsPassed
ISO 10555-1:2014Catheter Conditioning (Saline)Passed
ISO 10555-1:2014Leak TestPassed
Internal RequirementPriming Volume TestPassed
ISO 10555-1:2014Gravity Flow TestPassed
ISO 10555-1:2014Pump Flow TestPassed
ISO 10555-1:2014Tensile TestPassed
ISO 10555-1:2014Burst TestPassed
ISO 10555-1:2014Power Injection ConditioningPassed
Internal RequirementPrinting Font SizePassed
ASTM F1842-15Ink PermanencePassed
ISO 594-1: 1986 ViaLuer Taper TestPassed
ISO 80369-7:2016
Functional Testing
ASTM F1842-15Ink Integrity TestPassed
Vygon InternalRequirementAspiration Flow TestPassed
Vygon InternalRequirementLuer Color/Orientation TestPassed
Vygon InternalRequirementClamp Closure Maintenance TestPassed
Vygon InternalRequirementClamp and ID Tag Fit TestPassed
Vygon InternalRequirementClamp Actuation TestPassed

Table 2: Vygon PICCs Performance Testing

StandardSterile, Aged (Accelerated 12 Months)Device TestResult
ISO 10555-1:2014Power Injection ConditioningPassed
ISO 10555-1:2014Burst TestPassed

Table 3: Vygon PICCs Aged Performance Testing

Sterilization Testing

Sterilization tests are listed in Table 4 and Table 5 below.

StandardSterile, Aged (Accelerated 13 Months)Packaging TestResult
ASTM F88 / F88M-15Seal StrengthPassed

2750 Morris Road, Suite A200 • Lansdale, PA 19446 • Telephone: 215-390-2002 • Fax: 215-672-6740 Email: email@vygonus.com • Website: www.vygonusa.com

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Vygon logo. The word "VYGON" is in white letters on a green background. Above the "V" is a red triangle pointing downwards. Below the logo is the text "Value Life" in a smaller, gray font.

ASTM F2096-11Bubble LeakPassed
Table 4: Vygon PICCs Aged Packaging Testing
StandardSterile Device TestResult
ANSI/AAMI ST72:2011LAL TestingPassed
ISO 10993-7:2008EO Residual TestingPassed

Table5: Vygon PICCs Sterile Testing

In addition, performance testing for catheter stiffness and catheter flexural fatigue tolerance was completed on the final finished sterile devices with acceptable results.

Clinical Testing

No clinical testing was performed.

Biocompatibility Testing

Biocompatibility tests are listed below in Table 6.

StandardTest NameResult
ISO 10993-5:2009/(R)2014CytotoxicityNon-cytotoxic
AAMI/ANSI/ISO 10993-10:2010/(R)2014SensitizationNon-sensitizer
AAMI/ANSI/ISO 10993-10:2010/(R)2014IrritationNon-irritant
ISO 10993-11:2006Acute Systemic InjectionNon-toxic
USP37Material Mediated PyrogenNon-pyrogenic
ISO 10993-4: 2002/(R) 2013& A1: 2006/ (R) 2013Hemolysis (extract)Non-hemolytic
ISO 10993-4: 2002/(R) 2013& A1: 2006/ (R) 2013Hemolysis (direct)Non-hemolytic
ISO 10993-4: 2002/(R) 2013& A1: 2006/ (R) 2013Complement ActivationPredicate: PowerPICC®CatheterSimilar when compared topredicate
ISO 10993-4: 2002/(R) 2013& A1: 2006/ (R) 2013Partial Thromboplastin Time(PTT)Predicate: PowerPICC®CatheterMinimal activator (same aspredicate)
ISO 10993-4: 2002/(R) 2013& A1: 2006/ (R) 2013Dog ThrombogenicityEquivocal similar
AAMI/ANSI/ISO 10993-6:2007/(R)2014ImplantationNon-irritant

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the Vygon logo. The logo features the word "VYGON" in white letters on a green, rounded rectangle. Above the "V" is a small, red triangle. Below the logo is the text "Value Life" in a smaller, gray font.

StandardTest NameResult
ISO 10993-18 (Nelson LabsReport does not specifyyear.)Extractable/Leachable AnalysisSummarized in project#MJ16357-BIO01
ISO 10993-17:2008Subacute/Subchronic Toxicity,Genotoxicity, Chronic Toxicityand CarcinogencityEvaluated in ToxicologicalRisk Assessment

Table 6: Vygon PICCs Biocompatibility Testing

VII. Conclusion

Through performance bench testing the subject devices have demonstrated that they are substantially equivalent to the predicate device, 5 Fr DL PowerPICC® Catheter, K051672.

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”