K Number
K172893

Validate with FDA (Live)

Date Cleared
2017-12-21

(90 days)

Product Code
Regulation Number
870.1390
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SafeSept® is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. The SafeSept® Transseptal Guidewire is intended for single use only.

Device Description

The SafeSept® Needle Free Transseptal Guidewire (SafeSept NF) is 0.031" in diameter and 180cm long and consists of a flexible memory wire material called nitinol. The components of the SafeSept NF include the 0.031" transseptal guidewire, dispenser and straightener, and sterile packaging and labeling. There is one model number for the SafeSept NF, SSNF. The SafeSept NF is sterilized by 100% ethylene oxide cvcle and is for single-use only. The SafeSept is used in a healthcare facility/hospital.

The SafeSept NF is used in transseptal procedures to gain access to the left atrium through the right side of the heart. The device is used by inserting the device through a transseptal dilator and sheath located in the femoral vein. The SafeSept NF is then advanced to the fossa ovalis with the support of the dilator and sheath. The SafeSept NF's sharp distal tip is able to perforate through the fossa ovalis, a thin wall between the right and left atrium, when it is supported by a transseptal dilator and sheath. When the wire is no longer supported, it is atraumatic and operates as a typical 0.031" diameter guidewire. A radiopaque coil located along the shaft allows for fluoroscopic visualization of the wire within the left atrium and proximal marker bands help determine the location of the quidewire tip in relation to the needle. The guidewire duration in the body is less than 24 hours.

AI/ML Overview

This document is an FDA 510(k) clearance letter and summary for a medical device, the SafeSept Needle Free Transseptal Guidewire. It does not describe a study involving an algorithm or AI, nor does it present acceptance criteria and performance data for such a device. Instead, it focuses on demonstrating substantial equivalence to a legally marketed predicate device through physical and biological performance testing.

Therefore, many of the requested categories are not applicable to the provided text.

Here's a breakdown of what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

The document lists performance tests but does not explicitly state quantitative acceptance criteria or raw performance data (e.g., specific tensile strength values, fracture thresholds, etc.). It only states that the device "performed equivalently to the predicate device" and "met the requirements set forth in ISO-10993."

Acceptance CriteriaReported Device Performance
Performance Testing:
Visual and dimensional inspectionPerformed equivalently to the predicate device
Particulate testing (USP <788>, Light Obscuration Method)Performed equivalently to the predicate device
Surface finish assessmentPerformed equivalently to the predicate device
Corrosion resistancePerformed equivalently to the predicate device
Guidewire tensile strengthPerformed equivalently to the predicate device
Fracture and Flex testingPerformed equivalently to the predicate device
Curve Integrity testingPerformed equivalently to the predicate device
Torque TestingPerformed equivalently to the predicate device
Packaging testingPerformed equivalently to the predicate device
Radiopacity assessmentPerformed equivalently to the predicate device
Sterilization (ANSI/AAMI/ISO 11135-1 to SAL of 10-6)Validated to SAL of 10-6
Biocompatibility Data (ISO 10993 standards):
ISO 10993-5 CytotoxicityMet the requirements set forth in ISO-10993
ISO 10993-10 SensitizationMet the requirements set forth in ISO-10993
ISO 10993-10 Irritation/Intracutaneous ReactivityMet the requirements set forth in ISO-10993
ISO 10993-11 Acute Systemic ToxicityMet the requirements set forth in ISO-10993
ISO 10993-11 PyrogenicityMet the requirements set forth in ISO-10993
ISO 10993-3 GenotoxicityMet the requirements set forth in ISO-10993
ISO 10993-4 Hemocompatibility ASTM Hemolysis CompleteMet the requirements set forth in ISO-10993
ISO 10993-4 Hemocompatibility In-vivo Dog ThromboresistanceMet the requirements set forth in ISO-10993
ISO 10993-4 Hemocompatibility Complement Activation Complete with C3a & SC5b-9Met the requirements set forth in ISO-10993

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The testing described is primarily laboratory/bench testing and animal testing for biocompatibility, not clinical data sets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is not an AI/algorithm study involving expert review for ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document does not describe an AI device or an MRMC study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/algorithm device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For the performance testing, the "ground truth" would be the established engineering and ISO standards/requirements for the mechanical and biological properties of the device, and the performance of the predicate device.

8. The sample size for the training set

Not applicable. This is not an AI/algorithm device that requires a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2017

Pressure Products Medical Device Manufacturing LLC Andrew Armour Managing Director 1 School Street Morton, Pennsylvania 19070

Re: K172893

Trade/Device Name: SafeSept Needle Free Transseptal Guidewire Regulation Number: 21 CFR 870.1390 Regulation Name: Trocar Regulatory Class: Class II Product Code: DRC Dated: September 21, 2017 Received: September 22, 2017

Dear Mr. Armour:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Nicole G. Ibrahim -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172893

Device Name

SafeSept Needle Free Transseptal Guidewire

Indications for Use (Describe)

The SafeSept® is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. The SafeSept® Transseptal Guidewire is intended for single use only.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter

Pressure Products Medical Device Manufacturing LLC. 1 School Street Morton, PA 19070 Phone: 610-285-9858 Fax: 610-285-9859

Contact Person: Andrew Armour Prepared: October 23, 2017

Identification of the Device

Proprietary-Trade Name: SafeSept® Needle Free Transseptal Guidewire Device Class: Class II Classification Name: Trocar (CFR 870.1390) Common/Usual Name: Needle Free Transseptal Guidewire Product Code: DRC

Equivalent Legally Marketed Devices

Oscor, SafeSept® Needle Free Transseptal Trocar Guidewire, K130843

Description of the Devices

The SafeSept® Needle Free Transseptal Guidewire (SafeSept NF) is 0.031" in diameter and 180cm long and consists of a flexible memory wire material called nitinol. The components of the SafeSept NF include the 0.031" transseptal guidewire, dispenser and straightener, and sterile packaging and labeling. There is one model number for the SafeSept NF, SSNF. The SafeSept NF is sterilized by 100% ethylene oxide cvcle and is for single-use only. The SafeSept is used in a healthcare facility/hospital.

The SafeSept NF is used in transseptal procedures to gain access to the left atrium through the right side of the heart. The device is used by inserting the device through a transseptal dilator and sheath located in the femoral vein. The SafeSept NF is then advanced to the fossa ovalis with the support of the dilator and sheath. The SafeSept NF's sharp distal tip is able to perforate through the fossa ovalis, a thin wall between the right and left atrium, when it is supported by a transseptal dilator and sheath. When the wire is no longer supported, it is atraumatic and operates as a typical 0.031" diameter guidewire. A radiopaque coil located along the shaft allows for fluoroscopic visualization of the wire within the left atrium and proximal marker bands help determine the location of the quidewire tip in relation to the needle. The guidewire duration in the body is less than 24 hours.

Indications for Use

The SafeSept is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. The SafeSept®Transseptal Guidewire is intended for single use only.

The Indications for Use statement for the SafeSept Transseptal Guidewire device is identical to the predicate device. Both devices have the same intended use and are used in conjunction with a transseptal dilator and sheath to create the primary puncture in the interatrial septum. The devices quide the dilator and introducer through the septum from the right side of the heart to the left side.

Comparison of Technological Characteristics with the Predicate Device

The technological characteristics of this device are similar to the predicate device. The main difference between the subject device and the predicate device is the guidewire. The predicate SafeSept NF is 150cm in length and the proposed device is 180cm. The subject and predicate device are based on the follow technological elements:

  • Nitinol wire that is super-elastic so that it can have a 'J' curve with a sharp tip .

{4}------------------------------------------------

  • Radiopaque coil used by the physician to locate SafeSept and to guide supporting devices like . the transseptal dilator and sheath across the fossa
  • Device inserted through the transseptal dilator so that it can be used in the transseptal procedure
  • Creates the primary puncture of the fossa with the sharp tip
  • Taper is used to dilate the fossa further so that the dilator can then be inserted through the fossa . rather than using a transseptal needle
  • Printed markers are along the body for the physician to use as a guide to understand how far the ● SafeSept is in relation to the supporting devices

Performance Testing

The following performance tests were performed in support of the substantial equivalence to the predicate:

  • Visual and dimensional inspection
  • Particulate testing (USP <788>, Light Obscuration Method) ●
  • Surface finish assessment ●
  • Corrosion resistance ●
  • Guidewire tensile strength ●
  • Fracture and Flex testing
  • . Curve Integrity testing
  • Torque Testing
  • Packaging testing ●
  • . Radiopacity assessment

In addition, the sterilization conditions have been validated in accordance with ANSI/AAMI/ISO 11135- 1. Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of sterilization process for medical device is sterilized to a SAL of 10-6. The results of the testing demonstrated the subject device performed equivalently to the predicate device.

Biocompatibility Data

The biocompatibility evaluation for the SafeSept NF was conducted in accordance with FDA 510(k) Memorandum- #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" June 16, 2016, and International Standard ISO 10993-"Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by the FDA. The following tests were performed:

  • ISO 10993-5 Cytotoxicity ●
  • . ISO 10993-10 Sensitization
  • ISO 10993-10 Irritation/Intracutaneous Reactivity ●
  • ISO 10993-11 Acute Systemic Toxicity
  • ISO 10993-11 Pyrogenicity ●
  • ISO 10993-3 Genotoxicity
  • ISO 10993-4 Hemocompatibility ASTM Hemolysis Complete ●
  • ISO 10993-4 Hemocompatibility In-vivo Dog Thromboresistance .
  • ISO 10993-4 Hemocompatibility Complement Activation Complete with C3a & SC5b-9 ●

The SafeSept NF is considered an external communicating device with circulating blood contact and limited exposure (less than 24 hours). The SafeSept NF met the requirements set forth in ISO-10993.

Conclusion

When compared to the predicate device, the SafeSept Needle Free Transseptal Guidewire is substantially equivalent in design, technological characteristics, materials and performance testing.

§ 870.1390 Trocar.

(a)
Identification. A trocar is a sharp-pointed instrument used with a cannula for piercing a vessel or chamber to facilitate insertion of the cannula.(b)
Classification. Class II (special controls). Except for trocars that are reprocessed for multiple use, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.