K Number
K172883

Validate with FDA (Live)

Date Cleared
2018-02-12

(144 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted hair from small areas such as, but not limited to, underarm and facial hair below the chin line and large areas such as legs, in patients with Fitzpatrick Skin Phototypes I-V.

Device Description

The Home Use Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using IPL technology. And it consists of IPL main device and AC adapter, and a detachable lamp cap located in the IPL main device which is the source of optical radiation, namely a Xenon flashlamp.

The Home Use Hair Removal Device includes three models, T1, T2 and T3. All three models are exactly the same with the exception of the LCD display contents (mainly as the symbol of treatment level). T1 and T3 both adopt shape identification to display the treatment levels, which T1 adopts fan-shaped display and T3 adopts rectangle-shaped display. And T2 direct adopts digital symbol to display treatment levels.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device (Home Use Hair Removal Device). It describes the device, its intended use, and compares it to predicate devices to establish substantial equivalence for regulatory purposes.

However, it does not contain any information about a study proving the device meets specific acceptance criteria related to its performance in hair removal. Instead, it focuses on the engineering and safety performance of the device to demonstrate that it is as safe and effective as a previously approved device.

Therefore, many of the requested sections below cannot be populated with information from the provided text.

Here is the information that can be extracted or derived from the text, with an explanation of why other sections cannot be completed:


1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria in terms of hair reduction efficacy or clinical outcomes, as it primarily focuses on the technical and safety equivalence to predicate devices. It reports on various engineering and safety tests.

Acceptance Criteria CategoryReported Device Performance (from text)
BiocompatibilityPassed ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Skin Sensitization)
Electrical SafetyPassed IEC 60601-1-2, IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57
Eye SafetyPassed IEC 62471 (Photobiological safety)
Software ValidationAll software requirement specifications met and hazards mitigated.

2. Sample size used for the test set and the data provenance

The document does not describe a clinical study or a "test set" in the context of hair removal efficacy. The "performance data" mentioned refers to engineering and safety testing. Therefore, there is no sample size for a test set or data provenance related to clinical performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The document does not describe a clinical study with a test set requiring expert ground truth for hair removal efficacy.

4. Adjudication method for the test set

Not applicable. No clinical test set described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The device is a physical hair removal device, not an AI-assisted diagnostic or imaging tool. There are no "human readers" or AI assistance involved in its direct function.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. The device is a physical hair removal device, not an algorithm.

7. The type of ground truth used

For the engineering and safety performance data, the "ground truth" is defined by the requirements of the international standards listed (e.g., ISO 10993, IEC 60601, IEC 62471). Deviations from these standards would indicate failure.

8. The sample size for the training set

Not applicable. This device is not an AI/ML algorithm that requires a "training set".

9. How the ground truth for the training set was established

Not applicable. This device is not an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 12, 2018

Shenzhen Mareal Tech Co., Ltd % Rain Yip Official Correspondent Feiying Drug & Medical Consulting Technical Service Group Rm. 3005, Area B, Bldg.1, Southward Ruifeng Business Center Guimiao Road Shenzhen, 518000 Cn

Re: K172883

Trade/Device Name: Home Use Hair Removal Device/T1, T2, T3 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: January 12, 2018 Received: January 16, 2018

Dear Rain Yip:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172883

Device Name Home Use Hair Removal Device

Indications for Use (Describe)

The Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted hair from small areas such as, but not limited to, underarm and facial hair below the chin line and large areas such as legs, in patients with Fitzpatrick Skin Phototypes I-V.

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

"510(k) Summary" as required by 21 CFR Part 807.92.

Date: 2017.09.16

I. Submitter

Shenzhen Mareal Tech Co., Ltd 403A Block, Xinlong Technology Park, No.50 Fengtang Road, Xintian Community, Fuyong Town, Bao'an District, Shenzhen, China

Tel.: +86 755 2372 9558 Tax: +86 755 2998 0779

Libin Liu Tel: +86 188 2288 1173 Email: 1935068684@qq.com

II. Device

Name of Device: Home Use Hair Removal Device Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Description: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

III. Predicate Device

  • 1 ) The predicate devices are listed as below:
ManufacturerPredicate Device510(k) NumberApproval Date
Shen Zhen CosBeautyCo., LtdPerfectsmoothK161428March 23, 2017
CyDen LimitediPulse SmoothSkin Gold HairRemoval SystemK160968April 04, 2016

2 ) No reference devices were used in this submission.

IV. Device Description

The Home Use Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using IPL

{4}------------------------------------------------

technology. And it consists of IPL main device and AC adapter, and a detachable lamp cap located in the IPL main device which is the source of optical radiation, namely a Xenon flashlamp.

The Home Use Hair Removal Device includes three models, T1, T2 and T3. All three models are exactly the same with the exception of the LCD display contents (mainly as the symbol of treatment level). T1 and T3 both adopt shape identification to display the treatment levels, which T1 adopts fan-shaped display and T3 adopts rectangle-shaped display. And T2 direct adopts digital symbol to display treatment levels.

V. Indications for Use

The Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted hair from small areas such as, but not limited to, underarm and facial hair below the chin line and large areas such as legs, in patients with Fitzpatrick Skin Phototypes I-V.

VI. Comparison of Technological Characteristics With the Predicate Device

The Home Use Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the Home Use Hair Removal Device may be found substantially equivalent to its predicate device.

Home Use Hair Removal Device is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

  1. K161428, "PerfectSmooth", manufactured by "Shen CosBeauty Co., Ltd" in Guangdong, China

  2. K160968, "iPulse SmoothSkin Gold Hair Removal System ", manufactured by "CyDen Limited" in Wales, UK

The following table shows similarities and differences of use, design, material, safety and effectiveness between the subject device and predicate devices.

Information for predicate device was obtained from publicly available sources, including the 510(k) Summary and device instruction manual. A technical comparison to the predicate is provided below:

ComparisonElementsSubject DevicePredicate Device 1Predicate Device 2
K NumberPendingK161428K160968
Common nameLight Based Over-The-Counter HairRemovalLight Based Over-The-Counter HairRemovalLight Based Over-The-Counter HairRemoval
ComparisonElementsSubject DevicePredicate Device 1Predicate Device 2
Trade nameHome Use HairRemoval DevicePerfectSmoothiPulse SmoothSkinGold Hair RemovalSystem
Energy mediumXenon Arc FlashlampXenon Arc FlashlampXenon Arc Flashlamp
Pulse duration11-13 milliseconds11-13 milliseconds2-10 milliseconds
Energy density4.8 J/cm24.7J/cm23~6 J/cm2
Spot size3.9 cm24.5 cm23 (3cm by 1cm)
Delivery deviceDirect illumination totissueDirect illumination totissueDirect illumination totissue
Pulsing controlFinger switchFinger switchFinger switch
Indicationforuse/Intended useThe Home Use HairRemoval Device is anover-the-counterdevice intended forremoval of unwantedhair such as but notlimited to small areassuch as underarm andfacial hair below thechin line and largeareas such as legs,with Fitzpatrick SkinPhototypes I-V.The PerfectSmooth isan over-the-counterdevice intended forremoval of unwantedhair such as but notlimited to small areassuch as underarm andfacial hair below thechin line and largeareas such as legs.The iPulseSmoothSkin GoldHair Removal Systemis indicated for theremoval of unwantedhair. The iPulseSmoothskin Gold isalso indicated for thepermanent reductionin hair regrowth,defined as the long-term, stable reductionin the number of hairsregrowing whenmeasured at 6, 9 and12 months after thecompletion of atreatment regime.
Location for useOTCOTCOTC

{5}------------------------------------------------

VII.Performance Data

The following performance data were provided in support of the substantial equivalence determination.

{6}------------------------------------------------

1) Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the Home Use Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on June 16, 2016", as recognized by FDA. The battery of testing was performed to, and passed, including:

  • ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices –Par t 5: Tests for In Vitro Cytotoxicity

  • ISO 10993-10:2010, Biological Evaluation of Medical Devices -Par t 10: Tests for Irritation and Skin Sensitization

2) Electrical Safety and Eve Safety

Electrical safety and Eye safety testing was performed to, and passed, the following standards:

  • IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility

  • IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance

  • IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment

  • IEC 60601-2-57 Medical electrical equipment -Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

3) Eye Safety

  • IEC 62471 Photobiological safety of lamps and lamp systems

4) Software Verification and Validation

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

Summary

Based on the above performance as documented in this application, Home Use Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device.

VIII. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the Home Use Hair Removal Device is to be concluded substantial equivalent to its predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.